SCAVENGING SYSTEM
INTRODUCTION :
Trace level - conc of anesthetic gas that is far below the one
that is needed for clinical anesthesia.
Varies greatly – depending on FGF, ventilation system , the
length of time of anesthesia, measurement site and other
variables
Degree of pollution is more during paediatric surgeries
Workplace exposures :
- Hospital-based and stand-alone operating rooms,
- Recovery rooms,
- Dental operatories,
- Veterinary facilities .
Two different classes of
chemicals:
nitrous oxide and halogenated
agents
In 1977 National Institute for
Occupational Safety and Health
(NIOSH) –
recommended exposure limits
(RELs) for both nitrous oxide and
halogenated agent
Methods of studies
Animal Investigations
Human Volunteer studies
Epidemiological studies of exposed humans
Mortality studies
Problems
Spontaneous abortions :
Epidemiological studies - Higher rates in OR personnel
than in women in other settings.
study by Viera et al. (1980), spontaneous abortion was
observed in rats at 1000 ppm or more of N2O
In summary with possible exception of N2O the conc of
halogenated agents should be some 100 times more than
what is present usually
Infertility : Studies found higher than expected rates of
involuntary infertility among exposed
Birth Defects : Studies show that conc of inhalation agents
are well above those found in even unscavenged OR
Impaired Skilled performance : many studies failed to
establish , but one study showed that neuropsychological
symptoms and tiredness were reported more by individuals
in OR that are less scavenged
Carcinogenicity : A large study found higher risk of cancer in
females than males who are exposed, but data has been
questioned.
Liver disease : recurrent hepatitis – halothane reported in
few individuals
Renal Disease
Hematological : N2O Inactivates Vit B12 – impaired DNA
sysnthesis
Cardiac Disease : higher Freq of HTN and dysrthythmias
Scavenging and Waste Anesthetic
Gases (WAGs)
SCAVENGING
Definition
 Scavenging is the collection and
removal of vented anesthetic gases
from the OR.
 Since the amount of anesthetic gas
supplied usually far exceeds the
amount necessary for the patient, OR
pollution is decreased by scavenging.
Scavenger and operating room
ventilation efficiency are the two
most important factors in reduction of
waste anesthetic gases (WAGs).
The use of scavenging devices with anesthesia delivery systems is the
most effective way to decrease waste anesthetic gases.
 An efficient scavenging system is capable of reducing ambient
concentrations of waste gases by up to 90%.
 Anesthesia machines and breathing systems delivering
halogenated hydrocarbon anesthetics and/or nitrous
oxide should not be operated unless they are equipped
with a functional scavenging system.
To be effective, a scavenging system must not leak and must control
the concentration of trace anesthetics in ambient air.
 Waste gases should not be discharged into the outside air in an area
where reentry into the building is likely.
Types
Scavenging may be active (suction applied) or passive (waste gases proceed
passively down corrugated tubing through the room ventilation exhaust
grill of the OR).
 Active systems require a means to protect the patient's airway from the
application of suction, or buildup of positive pressure.
 Passive systems require that the patient be protected from positive
pressure buildup only.
The 2 major causes of waste gas contamination in the O.R :
1. The anesthetic technique used and
2. The equipment issues
Regarding anesthesia technique the following will increase anesthetic gas
pollution when inhalant anesthetics are used:
Administering inhalant anesthetics by open drop (eg, periodically
dripping liquid volatile anesthetic onto a gauze sponge) or insufflation
(eg, delivery of a relatively high flow of anesthetic in oxygen into the
trachea or pharynx through a catheter) techniques.
Turning on flowmeters and vaporizers before attaching the breathing
system to the patient.
Allowing flowmeters and vaporizers to remain on after the patient is
disconnected from the breathing system.
Using uncuffed endotracheal tubes that do not create a completely
sealed airway or using cuffed tubes without inflating the cuff.
Disconnecting a patient from a breathing system without eliminating as
much of the residual gases as reasonably possible through the scavenging
system .
Spilling liquid anesthetic during the filling of vaporizers, especially
during an anesthetic procedure.
Poorly fitting masks.
Flushing the circuit.
Use of breathing circuits that are difficult to scavenge, such as Jackson-
Rees.
NIOSH recommendations for trace gas levels
anesthetic gas Max. TWA conc.[ppm]
Halogenated agent alone 2
Nitrous oxide 25
Halogenated gas +nitrous oxide 0.5 + 25
Dental facilities[nitrous oxide alone] 50
Components of the scavenger system:
Gas collection assembly.
Transfer tubing.
Scavenging interface .
Gas disposal tubing .
Gas disposal assembly.
1.Gas collecting assembly
Captures excess anesthetic gases
and delivers it to the transfer
tubing.
WAG are vented from anesthesia
system through either adjustable
pressure limiting valve or
ventilator relief valve.
Gas passing through these valves
accumulates in the gas collecting
assembly and is directed to the
transfer means.
2.Transfer tubing
The transfer tubes carries excess gas from gas collecting assembly to
scavenging interface.
The tubes must be either 19 or 30 mm, sometimes yellow color-coded .
The tubes should be sufficiently rigid to prevent kinking and as short as
possible to minimize the chance of occlusion.
Separate tubes from the APL valve and ventilator relief valve merge into a
single hose before they enter scavenging interface.
If the transfer means is occluded, baseline breathing circuit pressure will
increase and barotrauma can occur.
3.Scavenging interface
The scavenger interface is the most important component because it
protects the breathing circuit or ventilator from excess positive or
negative pressure.
Positive-pressure relief is mandatory, irrespective of the type of
disposable system used, to vent excess gas in case of occlusion distal to
interface.
scavenger interfaces may be open
closed
B&C closed interfaces
A&B are active systems
C is passive system
OPEN INTERFACE:
It contains no valves and is open to
the atmosphere, allowing positive and
negative pressure relief.
Open interfaces should be used only
with active disposable systems that
have a central evacuation system.
 open interfaces require a reservoir
because waste gases are intermittently
discharged in surges whereas flow
from the evacuation system is
continuous.
The efficiency of it depends on several factors:
A.The vacuum flow rate per min must equal or exceed the
minute volume of excess gases to prevent spillage.
B.Spillage will occur if the volume of a single exhaled
breath exceeds the capacity of reservoir.
Open interfaces
are safer for the patients.
CLOSED INTERFACE:
It communicates with the atmosphere through valves.
Two types of closed interfaces are commercially available:
positive pressure relief only
positive and negative pressure relief
POSITIVE PRESSURE RELIEF ONLY:
 It has a single positive pressure relief valve and is designated to be used
only with passive disposable systems.
Transfer of the waste gas from the interface to the disposable system
relies on weight or pressure of the waste gas itself because a negative
pressure evacuation system is not used.
In this system reservoir bag is not required.
POSITIVE AND NEGATIVE PRESSURE RELIEF:
 It has positive and negative pressure relief valve in addition to a reservoir
bag.
It is used with active disposable systems.
The effectiveness of a closed system in preventing spillage depends on:
the rate of waste gas inflow
the evacuation flow rate
the size of the reservoir
Leakage occurs only when the reservoir bag becomes fully inflated and
pressure increases sufficiently to open the positive pressure relief valve .
In contrast, the effectiveness of a open system in preventing spillage
depends not only on the volume of the reservoir but also on the flow
characteristics within the interface.
4.Gas disposal tubing
The gas disposable tubing conducts waste gas from the Scavenging
interface to the gas disposable assembly.
It should be collapse proof and should run overhead, if possible to
minimize the chance of accidental occlusion.
5.Gas disposal assembly
It ultimately eliminates excess waste gas.
It is of two types: Active
Passive
Active assembly:
most commonly used. It uses central evacuation system.
A vacuum pump serves as mechanical flow inducing device that removes
the waste gases.
An interface with a negative pressure relief valve is mandatory because
the pressure within the system is negative.
A reservoir is very desirable and the larger the reservoir, the lower the
suction flow rate needed.
- Piped Vacuum
- Active duct system - employs flow inducing devices (fans ,
pumps , blowers etc..) to move even at low pressures
Advantages
 Convenient in large hospitals, where many machines are in
use in different locations
Disadvantages
  Vacuum system and pipe work is a major expense 
 Needle valve may need continual adjustment
Passive disposable system:
 does not use a mechanical flow inducing device.
The weight or pressure from the heavier-than-air anesthetic gases
produces flow through the system.
Positive pressure relief is mandatory, but a negative pressure relief and
reservoir are unnecessary.
Passive – simpler and less expensive but not effective
- Room ventilation system
- Nonrecirculating or – Recirculating
- Piping direct to atmosphere
- Adsorption devices
- Catalyst decomposition
Charcoal canisters :
activated charcoal
connected to the outlet of the breathing
system
removes halogenated anaesthetics by
filtration
Catalytic Decomposition :
• Catalytic decomposition can be used
to convert nitrous oxide to nitrogen
and oxygen, reducing its contribution
to the greenhouse effect
Advantages
  No set-up cost
 Mobile - moves with the machine
 
Disadvantages
  Continuing cost of replacement.
 Continuos replacement is messy
  Does not remove nitrous oxide
  Heating the canister causes the release of the
inhalational agents
Evaluation of Anesthetic Equipment
Each piece of equipment involved in the delivery of inhalant anesthetics
should be evaluated regularly to assure its function and integrity.
Procedures for checkout of anesthesia equipment, depending on
the equipment to be used, should include the following:
Status of the high-pressure system, including the oxygen supply and nitrous oxide
supply - The nitrous oxide supply should not leak when the cylinder valve
is on and the nitrous oxide flowmeter is off.
Status of the low-pressure system (flowmeter function) - A negative-pressure
leak test should be performed at the common gas outlet or the outlet of
the vaporizer immediately upstream from the breathing system.
Status of the breathing system - An appropriate leak test for a circle
system and Noncircle systems can usually be done. The quantity of
leakage can be measured by determining the flow rate of oxygen
necessary to maintain a constant pressure in the system, and the
leak rate should be less than 300 ml/min at 30 cm of h2O.
Status of the scavenging system - The scavenging system should be
properly attached at all connectors, and the appropriate vacuum
should be assured for active systems. If charcoal canisters are
employed for scavenging, they should be changed at appropriate
intervals.
Status of mechanical ventilators - Ventilators should be connected
properly to the anesthesia machine, and an absence of leaks should
be assured.
Monitoring of the Effectiveness of
Antipollution Techniques
Monitoring trace-gas concentrations in the workplace provides a
quantitative assessment of the effectiveness of a waste-gas control
program.
Measuring the concentration of anesthetic in the breathing zone of the
most heavily exposed workers is the usual procedure.
An air-monitoring program is most appropriately started after anesthesia
delivery systems have been equipped with scavenging systems and after
other techniques for minimizing waste gas pollution are in place.
 An ideal approach would include frequent air monitoring, preferably at least
semiannual evaluations.
Effectiveness
Unscavenged operating rooms show 10-70 ppm halothane, and 400-3000
ppm N2O.
 scavenging brings these levels down to 1 and 60 ppm respectively.
 Adding careful attention to leaks and technique can yield levels as low as
0.005 and 1 ppm.
HAZARDS
Scavenging system functionally extends the anesthesia circuit all the way
from the anesthesia machine to the disposable site.
Obstruction of scavenging pathway can cause excessive positive pressure in
the breathing circuit and barotrauma can occur.
 excessive vacuum applied can result in undesirable negative pressures
within breathing system.
Loss of Monitoring Input – it may mask the strong odor of a volatile
anesthetic agent, delaying recognition of an overdose
Alarm failure – neg. pressure from the scavenging system interface prevent
the bellows from collapsing when breathing system is disconnected . Low
airway pressure alarm not activated.
Avoiding waste gas exposure
Good mask fit.
Avoid unscavengeable techniques if possible (insufflation).
Prevent flow from breathing system into room air (only turn on agent and
nitrous oxide after mask is on face, turn them off before suctioning).
Washout anesthetics (into the breathing circuit) at the end of the
anesthetia.
Don’t spill liquid agent.
Use low flows .
Use cuffed tracheal tubes when possible.
Check the machine regularly for leaks.
Disconnect nitrous oxide pipeline connection at wall at the day’s end.
Total intravenous anesthesia.
THANK U….THANK U….

Contenu connexe

Tendances

Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
Ubaidur Rahaman
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Dhritiman Chakrabarti
 

Tendances (20)

Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machine
 
Anaesthesia workstation
Anaesthesia workstationAnaesthesia workstation
Anaesthesia workstation
 
High pressure system- Anaesthesia Machine
High pressure system- Anaesthesia MachineHigh pressure system- Anaesthesia Machine
High pressure system- Anaesthesia Machine
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway device
 
Patient positioning and anaesthetic consideration
Patient positioning and anaesthetic considerationPatient positioning and anaesthetic consideration
Patient positioning and anaesthetic consideration
 
Low flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringLow flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas Monitoring
 
Safety features of anaesthesia machine
Safety features of anaesthesia machineSafety features of anaesthesia machine
Safety features of anaesthesia machine
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
 
One lung ventilation
One lung ventilationOne lung ventilation
One lung ventilation
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supply
 
Breathing System
Breathing SystemBreathing System
Breathing System
 
caudal anesthesia.pdf
caudal anesthesia.pdfcaudal anesthesia.pdf
caudal anesthesia.pdf
 
Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features
 
Laryngeal mask-airway
Laryngeal mask-airwayLaryngeal mask-airway
Laryngeal mask-airway
 
Mapleson circuits
Mapleson circuitsMapleson circuits
Mapleson circuits
 
Monitoring in anaesthesia ro
Monitoring in anaesthesia roMonitoring in anaesthesia ro
Monitoring in anaesthesia ro
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
 
Capnography
CapnographyCapnography
Capnography
 

En vedette (7)

Muscle relaxant
Muscle relaxantMuscle relaxant
Muscle relaxant
 
Anesthesia Ventilator
Anesthesia VentilatorAnesthesia Ventilator
Anesthesia Ventilator
 
13236530 Anesthesia Pharmacology
13236530 Anesthesia Pharmacology13236530 Anesthesia Pharmacology
13236530 Anesthesia Pharmacology
 
Face masks, laryngeal tube, airways yuvaraj
Face masks, laryngeal tube, airways  yuvarajFace masks, laryngeal tube, airways  yuvaraj
Face masks, laryngeal tube, airways yuvaraj
 
Breathing systems (2)
Breathing systems (2)Breathing systems (2)
Breathing systems (2)
 
hypertension anesthesia, general management. antihypertensive pharmacology
hypertension anesthesia, general management. antihypertensive pharmacologyhypertension anesthesia, general management. antihypertensive pharmacology
hypertension anesthesia, general management. antihypertensive pharmacology
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 

Similaire à Scavenging system in operating room

Vets 238 Anesthetic Equipment Final
Vets 238   Anesthetic Equipment FinalVets 238   Anesthetic Equipment Final
Vets 238 Anesthetic Equipment Final
meckelbt
 
Trace anesthetic exposure
Trace anesthetic exposureTrace anesthetic exposure
Trace anesthetic exposure
Dr Sandeep
 
Brochure MedAir AHU & HVAC Services
Brochure MedAir AHU & HVAC ServicesBrochure MedAir AHU & HVAC Services
Brochure MedAir AHU & HVAC Services
Frank Crook
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth Testing
Eric Ferguson
 

Similaire à Scavenging system in operating room (20)

Scavanging system
Scavanging systemScavanging system
Scavanging system
 
scavenging system and hazards
scavenging system and hazardsscavenging system and hazards
scavenging system and hazards
 
Breathing systems - Mapleson Classification
Breathing systems - Mapleson ClassificationBreathing systems - Mapleson Classification
Breathing systems - Mapleson Classification
 
scavenging system
scavenging systemscavenging system
scavenging system
 
Vets 238 Anesthetic Equipment Final
Vets 238   Anesthetic Equipment FinalVets 238   Anesthetic Equipment Final
Vets 238 Anesthetic Equipment Final
 
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
Anesthesia machine basics_V1.0_EN讲 - Copy.pdfAnesthesia machine basics_V1.0_EN讲 - Copy.pdf
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
 
Máquina de Anestesia
Máquina de AnestesiaMáquina de Anestesia
Máquina de Anestesia
 
Anesthestic Breathing Systems by Dr. Mohammad abdeljawad
Anesthestic Breathing Systems by Dr. Mohammad abdeljawad Anesthestic Breathing Systems by Dr. Mohammad abdeljawad
Anesthestic Breathing Systems by Dr. Mohammad abdeljawad
 
Trace anesthetic exposure
Trace anesthetic exposureTrace anesthetic exposure
Trace anesthetic exposure
 
Anesthesia machine
Anesthesia machine Anesthesia machine
Anesthesia machine
 
Central Medical Gas Distribution System
Central Medical Gas Distribution SystemCentral Medical Gas Distribution System
Central Medical Gas Distribution System
 
Good containment practices
Good containment practicesGood containment practices
Good containment practices
 
Anesthesia Machine.pptx
Anesthesia Machine.pptxAnesthesia Machine.pptx
Anesthesia Machine.pptx
 
Brochure MedAir AHU & HVAC Services
Brochure MedAir AHU & HVAC ServicesBrochure MedAir AHU & HVAC Services
Brochure MedAir AHU & HVAC Services
 
Trouble Shooting of Equipment.pptx
Trouble Shooting of Equipment.pptxTrouble Shooting of Equipment.pptx
Trouble Shooting of Equipment.pptx
 
PIPELINE.pptx
PIPELINE.pptxPIPELINE.pptx
PIPELINE.pptx
 
Leak Tests.pptx
Leak Tests.pptxLeak Tests.pptx
Leak Tests.pptx
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth Testing
 
Nasopulmonary dds
Nasopulmonary ddsNasopulmonary dds
Nasopulmonary dds
 
Aerosol
AerosolAerosol
Aerosol
 

Plus de Dr Kumar

Hypertension , crf post renal transplant patient for surgery
Hypertension , crf post renal transplant patient for surgeryHypertension , crf post renal transplant patient for surgery
Hypertension , crf post renal transplant patient for surgery
Dr Kumar
 

Plus de Dr Kumar (20)

Anaesthesia for robotic surgery
Anaesthesia for robotic surgeryAnaesthesia for robotic surgery
Anaesthesia for robotic surgery
 
Telemedicine in Anaesthesia
Telemedicine in AnaesthesiaTelemedicine in Anaesthesia
Telemedicine in Anaesthesia
 
TOURNIQUET AND ITS EFFECTS
TOURNIQUET AND ITS EFFECTS TOURNIQUET AND ITS EFFECTS
TOURNIQUET AND ITS EFFECTS
 
Supra glottic airway children
Supra glottic airway childrenSupra glottic airway children
Supra glottic airway children
 
shock
shock shock
shock
 
Renal replacement therapy
Renal replacement therapyRenal replacement therapy
Renal replacement therapy
 
Pheochromocytoma and its anaesthetic management
Pheochromocytoma and its anaesthetic managementPheochromocytoma and its anaesthetic management
Pheochromocytoma and its anaesthetic management
 
Neuromuscular weakness related to critical illness
Neuromuscular weakness related to critical illnessNeuromuscular weakness related to critical illness
Neuromuscular weakness related to critical illness
 
Mechanisms of cerebral injury and cerebral protection
Mechanisms of cerebral injury and cerebral protectionMechanisms of cerebral injury and cerebral protection
Mechanisms of cerebral injury and cerebral protection
 
Latex allergy and its management
Latex allergy and its managementLatex allergy and its management
Latex allergy and its management
 
Intra operative hypoxia and hypercarbia
Intra operative hypoxia and hypercarbiaIntra operative hypoxia and hypercarbia
Intra operative hypoxia and hypercarbia
 
Hypotensive anesthesia
Hypotensive anesthesiaHypotensive anesthesia
Hypotensive anesthesia
 
Hypertension , crf post renal transplant patient for surgery
Hypertension , crf post renal transplant patient for surgeryHypertension , crf post renal transplant patient for surgery
Hypertension , crf post renal transplant patient for surgery
 
Extubation problems and its management
Extubation problems and its managementExtubation problems and its management
Extubation problems and its management
 
Effects of anaesthesia on immune system
Effects of anaesthesia on immune systemEffects of anaesthesia on immune system
Effects of anaesthesia on immune system
 
Buffers in the body
Buffers in the bodyBuffers in the body
Buffers in the body
 
Anaesthetic management of conjoined twins’
Anaesthetic management of conjoined twins’Anaesthetic management of conjoined twins’
Anaesthetic management of conjoined twins’
 
Anaesthesia for interventional neuroradiology
Anaesthesia for interventional neuroradiologyAnaesthesia for interventional neuroradiology
Anaesthesia for interventional neuroradiology
 
Ambulatory anaesthesia
Ambulatory anaesthesiaAmbulatory anaesthesia
Ambulatory anaesthesia
 
Acute pain and its management
Acute pain and its managementAcute pain and its management
Acute pain and its management
 

Dernier

Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
adilkhan87451
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
Sheetaleventcompany
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
mahaiklolahd
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
adilkhan87451
 

Dernier (20)

Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls  * UPA...
Call Girl in Indore 8827247818 {LowPrice} ❤️ (ahana) Indore Call Girls * UPA...
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near MeTop Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Top Rated Pune Call Girls (DIPAL) ⟟ 8250077686 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls (DIPAL) ⟟ 8250077686 ⟟ Call Me For Genuine Sex Serv...Top Rated Pune Call Girls (DIPAL) ⟟ 8250077686 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls (DIPAL) ⟟ 8250077686 ⟟ Call Me For Genuine Sex Serv...
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
 

Scavenging system in operating room

  • 1. Presentor : Dr.T.Kumar Moderator : Dr.Venkatesh (Asst. professor) SCAVENGING SYSTEM
  • 2. INTRODUCTION : Trace level - conc of anesthetic gas that is far below the one that is needed for clinical anesthesia. Varies greatly – depending on FGF, ventilation system , the length of time of anesthesia, measurement site and other variables Degree of pollution is more during paediatric surgeries
  • 3. Workplace exposures : - Hospital-based and stand-alone operating rooms, - Recovery rooms, - Dental operatories, - Veterinary facilities .
  • 4. Two different classes of chemicals: nitrous oxide and halogenated agents In 1977 National Institute for Occupational Safety and Health (NIOSH) – recommended exposure limits (RELs) for both nitrous oxide and halogenated agent
  • 5. Methods of studies Animal Investigations Human Volunteer studies Epidemiological studies of exposed humans Mortality studies
  • 6. Problems Spontaneous abortions : Epidemiological studies - Higher rates in OR personnel than in women in other settings. study by Viera et al. (1980), spontaneous abortion was observed in rats at 1000 ppm or more of N2O In summary with possible exception of N2O the conc of halogenated agents should be some 100 times more than what is present usually
  • 7. Infertility : Studies found higher than expected rates of involuntary infertility among exposed Birth Defects : Studies show that conc of inhalation agents are well above those found in even unscavenged OR Impaired Skilled performance : many studies failed to establish , but one study showed that neuropsychological symptoms and tiredness were reported more by individuals in OR that are less scavenged
  • 8. Carcinogenicity : A large study found higher risk of cancer in females than males who are exposed, but data has been questioned. Liver disease : recurrent hepatitis – halothane reported in few individuals Renal Disease Hematological : N2O Inactivates Vit B12 – impaired DNA sysnthesis Cardiac Disease : higher Freq of HTN and dysrthythmias
  • 9. Scavenging and Waste Anesthetic Gases (WAGs) SCAVENGING Definition  Scavenging is the collection and removal of vented anesthetic gases from the OR.  Since the amount of anesthetic gas supplied usually far exceeds the amount necessary for the patient, OR pollution is decreased by scavenging. Scavenger and operating room ventilation efficiency are the two most important factors in reduction of waste anesthetic gases (WAGs).
  • 10. The use of scavenging devices with anesthesia delivery systems is the most effective way to decrease waste anesthetic gases.  An efficient scavenging system is capable of reducing ambient concentrations of waste gases by up to 90%.  Anesthesia machines and breathing systems delivering halogenated hydrocarbon anesthetics and/or nitrous oxide should not be operated unless they are equipped with a functional scavenging system. To be effective, a scavenging system must not leak and must control the concentration of trace anesthetics in ambient air.  Waste gases should not be discharged into the outside air in an area where reentry into the building is likely.
  • 11. Types Scavenging may be active (suction applied) or passive (waste gases proceed passively down corrugated tubing through the room ventilation exhaust grill of the OR).  Active systems require a means to protect the patient's airway from the application of suction, or buildup of positive pressure.  Passive systems require that the patient be protected from positive pressure buildup only.
  • 12. The 2 major causes of waste gas contamination in the O.R : 1. The anesthetic technique used and 2. The equipment issues Regarding anesthesia technique the following will increase anesthetic gas pollution when inhalant anesthetics are used: Administering inhalant anesthetics by open drop (eg, periodically dripping liquid volatile anesthetic onto a gauze sponge) or insufflation (eg, delivery of a relatively high flow of anesthetic in oxygen into the trachea or pharynx through a catheter) techniques. Turning on flowmeters and vaporizers before attaching the breathing system to the patient. Allowing flowmeters and vaporizers to remain on after the patient is disconnected from the breathing system.
  • 13. Using uncuffed endotracheal tubes that do not create a completely sealed airway or using cuffed tubes without inflating the cuff. Disconnecting a patient from a breathing system without eliminating as much of the residual gases as reasonably possible through the scavenging system . Spilling liquid anesthetic during the filling of vaporizers, especially during an anesthetic procedure. Poorly fitting masks. Flushing the circuit. Use of breathing circuits that are difficult to scavenge, such as Jackson- Rees.
  • 14. NIOSH recommendations for trace gas levels anesthetic gas Max. TWA conc.[ppm] Halogenated agent alone 2 Nitrous oxide 25 Halogenated gas +nitrous oxide 0.5 + 25 Dental facilities[nitrous oxide alone] 50
  • 15. Components of the scavenger system: Gas collection assembly. Transfer tubing. Scavenging interface . Gas disposal tubing . Gas disposal assembly.
  • 16. 1.Gas collecting assembly Captures excess anesthetic gases and delivers it to the transfer tubing. WAG are vented from anesthesia system through either adjustable pressure limiting valve or ventilator relief valve. Gas passing through these valves accumulates in the gas collecting assembly and is directed to the transfer means.
  • 17. 2.Transfer tubing The transfer tubes carries excess gas from gas collecting assembly to scavenging interface. The tubes must be either 19 or 30 mm, sometimes yellow color-coded . The tubes should be sufficiently rigid to prevent kinking and as short as possible to minimize the chance of occlusion. Separate tubes from the APL valve and ventilator relief valve merge into a single hose before they enter scavenging interface. If the transfer means is occluded, baseline breathing circuit pressure will increase and barotrauma can occur.
  • 18. 3.Scavenging interface The scavenger interface is the most important component because it protects the breathing circuit or ventilator from excess positive or negative pressure. Positive-pressure relief is mandatory, irrespective of the type of disposable system used, to vent excess gas in case of occlusion distal to interface. scavenger interfaces may be open closed
  • 19. B&C closed interfaces A&B are active systems C is passive system
  • 20. OPEN INTERFACE: It contains no valves and is open to the atmosphere, allowing positive and negative pressure relief. Open interfaces should be used only with active disposable systems that have a central evacuation system.  open interfaces require a reservoir because waste gases are intermittently discharged in surges whereas flow from the evacuation system is continuous.
  • 21. The efficiency of it depends on several factors: A.The vacuum flow rate per min must equal or exceed the minute volume of excess gases to prevent spillage. B.Spillage will occur if the volume of a single exhaled breath exceeds the capacity of reservoir. Open interfaces are safer for the patients.
  • 22. CLOSED INTERFACE: It communicates with the atmosphere through valves. Two types of closed interfaces are commercially available: positive pressure relief only positive and negative pressure relief POSITIVE PRESSURE RELIEF ONLY:  It has a single positive pressure relief valve and is designated to be used only with passive disposable systems. Transfer of the waste gas from the interface to the disposable system relies on weight or pressure of the waste gas itself because a negative pressure evacuation system is not used. In this system reservoir bag is not required.
  • 23.
  • 24. POSITIVE AND NEGATIVE PRESSURE RELIEF:  It has positive and negative pressure relief valve in addition to a reservoir bag. It is used with active disposable systems. The effectiveness of a closed system in preventing spillage depends on: the rate of waste gas inflow the evacuation flow rate the size of the reservoir Leakage occurs only when the reservoir bag becomes fully inflated and pressure increases sufficiently to open the positive pressure relief valve . In contrast, the effectiveness of a open system in preventing spillage depends not only on the volume of the reservoir but also on the flow characteristics within the interface.
  • 25. 4.Gas disposal tubing The gas disposable tubing conducts waste gas from the Scavenging interface to the gas disposable assembly. It should be collapse proof and should run overhead, if possible to minimize the chance of accidental occlusion.
  • 26. 5.Gas disposal assembly It ultimately eliminates excess waste gas. It is of two types: Active Passive Active assembly: most commonly used. It uses central evacuation system. A vacuum pump serves as mechanical flow inducing device that removes the waste gases. An interface with a negative pressure relief valve is mandatory because the pressure within the system is negative. A reservoir is very desirable and the larger the reservoir, the lower the suction flow rate needed.
  • 27. - Piped Vacuum - Active duct system - employs flow inducing devices (fans , pumps , blowers etc..) to move even at low pressures Advantages  Convenient in large hospitals, where many machines are in use in different locations Disadvantages   Vacuum system and pipe work is a major expense   Needle valve may need continual adjustment
  • 28. Passive disposable system:  does not use a mechanical flow inducing device. The weight or pressure from the heavier-than-air anesthetic gases produces flow through the system. Positive pressure relief is mandatory, but a negative pressure relief and reservoir are unnecessary. Passive – simpler and less expensive but not effective - Room ventilation system - Nonrecirculating or – Recirculating - Piping direct to atmosphere - Adsorption devices - Catalyst decomposition
  • 29. Charcoal canisters : activated charcoal connected to the outlet of the breathing system removes halogenated anaesthetics by filtration Catalytic Decomposition : • Catalytic decomposition can be used to convert nitrous oxide to nitrogen and oxygen, reducing its contribution to the greenhouse effect
  • 30. Advantages   No set-up cost  Mobile - moves with the machine   Disadvantages   Continuing cost of replacement.  Continuos replacement is messy   Does not remove nitrous oxide   Heating the canister causes the release of the inhalational agents
  • 31. Evaluation of Anesthetic Equipment Each piece of equipment involved in the delivery of inhalant anesthetics should be evaluated regularly to assure its function and integrity. Procedures for checkout of anesthesia equipment, depending on the equipment to be used, should include the following: Status of the high-pressure system, including the oxygen supply and nitrous oxide supply - The nitrous oxide supply should not leak when the cylinder valve is on and the nitrous oxide flowmeter is off. Status of the low-pressure system (flowmeter function) - A negative-pressure leak test should be performed at the common gas outlet or the outlet of the vaporizer immediately upstream from the breathing system.
  • 32. Status of the breathing system - An appropriate leak test for a circle system and Noncircle systems can usually be done. The quantity of leakage can be measured by determining the flow rate of oxygen necessary to maintain a constant pressure in the system, and the leak rate should be less than 300 ml/min at 30 cm of h2O. Status of the scavenging system - The scavenging system should be properly attached at all connectors, and the appropriate vacuum should be assured for active systems. If charcoal canisters are employed for scavenging, they should be changed at appropriate intervals. Status of mechanical ventilators - Ventilators should be connected properly to the anesthesia machine, and an absence of leaks should be assured.
  • 33. Monitoring of the Effectiveness of Antipollution Techniques Monitoring trace-gas concentrations in the workplace provides a quantitative assessment of the effectiveness of a waste-gas control program. Measuring the concentration of anesthetic in the breathing zone of the most heavily exposed workers is the usual procedure. An air-monitoring program is most appropriately started after anesthesia delivery systems have been equipped with scavenging systems and after other techniques for minimizing waste gas pollution are in place.  An ideal approach would include frequent air monitoring, preferably at least semiannual evaluations.
  • 34. Effectiveness Unscavenged operating rooms show 10-70 ppm halothane, and 400-3000 ppm N2O.  scavenging brings these levels down to 1 and 60 ppm respectively.  Adding careful attention to leaks and technique can yield levels as low as 0.005 and 1 ppm.
  • 35. HAZARDS Scavenging system functionally extends the anesthesia circuit all the way from the anesthesia machine to the disposable site. Obstruction of scavenging pathway can cause excessive positive pressure in the breathing circuit and barotrauma can occur.  excessive vacuum applied can result in undesirable negative pressures within breathing system. Loss of Monitoring Input – it may mask the strong odor of a volatile anesthetic agent, delaying recognition of an overdose Alarm failure – neg. pressure from the scavenging system interface prevent the bellows from collapsing when breathing system is disconnected . Low airway pressure alarm not activated.
  • 36. Avoiding waste gas exposure Good mask fit. Avoid unscavengeable techniques if possible (insufflation). Prevent flow from breathing system into room air (only turn on agent and nitrous oxide after mask is on face, turn them off before suctioning). Washout anesthetics (into the breathing circuit) at the end of the anesthetia. Don’t spill liquid agent. Use low flows . Use cuffed tracheal tubes when possible. Check the machine regularly for leaks. Disconnect nitrous oxide pipeline connection at wall at the day’s end. Total intravenous anesthesia.