Very high myopic LASIK using new
     hybrid aspheric profiles

Dan Z Reinstein MD MA(Cantab) FRCSC FRCOphth1,2,3,4

1. London Vision Clinic, London, UK
2. St. Thomas’ Hospital - Kings College, London, UK
3. Weill Medical College of Cornell University, New York, USA
4. Centre Hospitalier National d’Ophtalmologie, (Pr. Laroche) , Paris, France
First Results: Munnerlyn Ablation Profile

• Early ablation profiles often induced:
   – Night Vision disturbances
   – Decreased contrast sensitivity
• Limited the range of treatable refractions
• PROBLEM: Induction of spherical aberration




        Eur J Ophthalmol. 1994 Jan-Mar;4(1):43-51. Night vision after
        excimer laser photorefractive keratectomy: haze and halos. O'Brart
        DP, Lohmann CP, Fitzke FW, Smith SE, Kerr-Muir MG, Marshall J.
                                                                    ©DZ Reinstein 2009
                                                             dzr@londonvisionclinic.com
Ablation Profile Design: Larger Optical Zone
    Example: 5-mm Munnerlyn ablation for -6.00 D (1993 Summit Laser)
       Topography            Wavefront


                                                  Z(4,0) (OSA)

                                                    1.18 µm




                 J Refract Corneal Surg. 1994 Mar-Apr;10(2):87-94.
                 Excimer laser photorefractive keratectomy for myopia:
                 comparison of 4.00- and 5.00-millimeter ablation zones.
                 O'Brart DP, Gartry DS, Lohmann CP, Muir MG, Marshall J.

                 Arch Ophthalmol. 1995 Apr;113(4):438-43. The effects of
                 ablation diameter on the outcome of excimer laser
                 photorefractive keratectomy. A prospective, randomized,
                 double-blind study. O'Brart DP, Corbett MC, Lohmann CP,
                 Kerr Muir MG, Marshall J.
                                                                        ©DZ Reinstein 2009
                                                                 dzr@londonvisionclinic.com
Ablation Profile Design: Aspheric Profiles
• Barraquer 1980
   – Suggested parabolic keratomileusis




• Seiler 1993 – PRK aspheric profiles
   – Less starburst & halos
   – Larger effective clear optical zone size




                                                       ©DZ Reinstein 2009
                                                dzr@londonvisionclinic.com
Why was spherical aberration
increasing?
Optimization: Fluence correction
• Fluence correction: Topography
  – Beam reflection compensation
  – Beam projection compensation

                               J Refract Surg 2001;17(5):S584-7.
                               Influence of corneal curvature on
                               calculation of ablation patterns
                               used in photorefractive laser
                               surgery. Mrochen M, Seiler T.




                                                          ©DZ Reinstein 2009
                                                   dzr@londonvisionclinic.com
Optimization

• Biomechanics




                 VHF digital ultrasound




                                                 ©DZ Reinstein 2009
                                          dzr@londonvisionclinic.com
Reinstein et al. Journal of Refractive Surgery
Artemis C12 Display                                              2000 Jul-Aug;16:414-30




                                        VHF digital ultrasound




  Roberts C. The cornea is not a piece of plastic.
  JRS 2000; 16:407-413

                                                                                                      ©DZ Reinstein 2009
                                                                                               dzr@londonvisionclinic.com
Examples of Peripheral Stromal Thickening




                                                      Peripheral Stromal
                                                      Thickening



                                                                           Central Flattening
   Roberts C. The cornea is not a piece of plastic.

                                                                                                       ©DZ Reinstein 2009
                                                                                                dzr@londonvisionclinic.com
Corneal Biomechanical Trade-off

• Hyperopic shift induced by
   – Central flattening due to peripheral tissue removal



• Myopic shift induced by
                                                                           Pre-Op
   – Epithelial thickening
   – Bowing of the back surface
                                     Epithelial thickening
                                                                         Post-Op



                                                 Back surface bowing




                                                                      ©DZ Reinstein 2009
                                                               dzr@londonvisionclinic.com
Free lunch?
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider)




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
Free Lunch?

• Increasing ablation zone diameter
• Adding asphericity


                                • Increases central
                                  ablation depth
                                • No “Free Lunch”




                                                 ©DZ Reinstein 2009
                                          dzr@londonvisionclinic.com
550 µm Pachymetry: Forces Compromise
• Modern aspheric ablation profiles still induce spherical
  aberration
• Problem: high myopic corrections may result in NVDs
                         Attempted Spherical Equivalent vs. Change in Z(4,0) Coefficient
                                               ASA Treatments
               0.7

                          y = -0.059x - 0.0136
               0.6
           )                   R² = 0.6444
           A
           S
           O   0.5
           ,
           m
           µ
           (
           t   0.4
           n
           e
           i
           c
           i
           f
           f   0.3
           e
           o
           C   0.2
           )
           0
           ,
           4
           (
           Z   0.1
           n
           i
           e
           g   0.0
           n
           a
           h
           C   -0.1


               -0.2
                  0.00       -1.00    -2.00      -3.00   -4.00   -5.00   -6.00   -7.00    -8.00   -9.00   -10.00

                                              Attempted Spherical Equivalent (Diopters)



                                                                                                                          ©DZ Reinstein 2009
                                                                                                                   dzr@londonvisionclinic.com
Wavefront-Guided Treatment
of Spherical Aberration
Correlation of Contrast with Wavefront
           Spherical Aberration                                              Contrast Sensitivity
                                                         p
     900                                                                       1.2
     800                                                              t        1.1
                                                                      s
     700                                                              a o
                                                                      r i
                                                                      t t      1.0
     600                                                              n a
                                                                      o R
                                                                      C y      0.9
     500                                                                t
                                                                      d i
                                                                      e v
                                                                      z i
                                                                        t
2    400                                                              i i
                                                                      l s
                                                                      a n
                                                                               0.8
 m   300                                                              m
                                                                      r e
 µ                                                                    o S
                                                                               0.7
     200                                                              N
                                                                               0.6
     100
                                                                               0.5
       0                                                                               3 cpd   6 cpd    12 cpd        18 cpd
                                           Pre CRS-M   Post CRSM-
              Control Pre   Control Post                            Control Pre        1.02    1.02      1.03          1.04
                                             Repair      Repair
                                                                    Control Post       1.04    1.01      1.03          1.01
Sph Ab Area      122            276          563          410       Pre CRSM-Repair    0.85    0.84      0.77          0.75
                                                                    Post CRSM-Repair   1.04    1.02      1.02          1.00




                  • 27% Gross Reduction
                  • 53% Net Reduction (cf tolerable level)
                  • Tolerable level ~0.56 µm @ 6mm



                                                                                                              ©DZ Reinstein 2009
                                                                                                       dzr@londonvisionclinic.com
Pre-Compensate for Spherical Aberration

• “Q-slider”
   – (WaveLight)
• Wavefront-guided ablation
   – Includes pre-op spherical aberration
   – Effect dependent on pre-op spherical aberration


• Our Approach: Include an “artificial” wavefront
   – Isolate spherical aberration: Z(4,0) as the only coefficient
   – Z(4,0) coefficient proportional to expected induction
   – Increase Z(4,0) coefficient: wavefront only 20% effective



                                                             ©DZ Reinstein 2009
                                                      dzr@londonvisionclinic.com
Patient 1, OD

                   -7.13 D Corrected




6mm       OSA               6mm         OSA
Coma     0.09 µm            Coma       0.04 µm
Sph Ab   0.48 µm            Sph Ab     0.42 µm
HO RMS   0.59 µm            HO RMS     0.52 µm


                                                        ©DZ Reinstein 2009
                                                 dzr@londonvisionclinic.com
Patient 1, OS

                   -9.00 D Corrected




6mm       OSA               6mm         OSA
Coma     0.03 µm            Coma       0.05 µm
Sph Ab   0.49 µm            Sph Ab     0.55 µm
HO RMS   0.57 µm            HO RMS     0.60 µm


                                                        ©DZ Reinstein 2009
                                                 dzr@londonvisionclinic.com
Patient 1, Night Vision

                  Pre Op                   Post Op




                      Rx Treated
             OD     -6.50 -1.25 x 178
             OS      -8.25 -1.50 x 17

                                               ©DZ Reinstein 2009
                                        dzr@londonvisionclinic.com
Induction of Spherical Aberration




                                    Complaint of NVD post RS1




 1


                                                       ©DZ Reinstein 2009
                                                dzr@londonvisionclinic.com
Limits to SA Pre-Compensation
• Excess spherical aberration pre-compensation can
  lead to “central islands”


         TMS          WASCA (zonal)       Epithelium



OD




                                                    ©DZ Reinstein 2009
                                             dzr@londonvisionclinic.com
CENTRAL ISLANDS:
-5.50 D ablation
Diplopia first week
Slow resolution over 2 weeks




Slides courtesy Gordon Balazsi, MD
                                            ©DZ Reinstein 2009
                                     dzr@londonvisionclinic.com
CENTRAL ISLANDS:
-5.00 D ablation




Slides courtesy Gordon Balazsi, MD
                                            ©DZ Reinstein 2009
                                     dzr@londonvisionclinic.com
Ablation Depth with SA Pre-Compensation




                                        ©DZ Reinstein 2009
                                 dzr@londonvisionclinic.com
New Profile for High Myopia

• Non-linear aspheric ablation profile:
   – Increased peripheral ablation (not ↑ z(4,0))
   – Reduced induction of spherical aberration
   – Free lunch: some myopia corrected due to central
     flattening

   Extend this concept further to promote central flattening
   Ability to correct high myopia without risk of NVDs

                                                       Peripheral Stromal
                                                       Thickening



                                                                            Central Flattening
    Roberts C. The cornea is not a piece of plastic.




                                                                                                        ©DZ Reinstein 2009
                                                                                                 dzr@londonvisionclinic.com
New Profile: “Free Lunch”
• Over-corrected by +0.50 D compared with theory
• Ablation depth lower than expected
                            Attempted vs. Achieved Spherical Equivalent
              -14

                    y = 0.9958x - 0.5106
            ) -13        R² = 0.9291
            s
            r
            e
            t
            p-12
            o
            i
            D
            (
            t -11
            n
            e
            l
            a
            v -10
            i
            u
            q
            E -9
            l
            a
            c
            i
            r
            e -8
            h
            p
            S
            d -7
            e
            v
            e
            i
            h -6
            c
            A
               -5
                       -6       -7     -8    -9    -10   -11    -12   -13   -14

                        Attempted Spherical Equivalent (Diopters)

                                                                                         ©DZ Reinstein 2009
                                                                                  dzr@londonvisionclinic.com
Ablation Depth for New Profile




                                        ©DZ Reinstein 2009
                                 dzr@londonvisionclinic.com
Femtosecond Lasers
• Femtosecond lasers have improved flap thickness
  reproducibility (VisuMax SD: 8 µm)
• We can create thinner flaps (VisuMax: 80 µm)
• Thinner flaps extends the range of myopia in LASIK




                      Pre-release online




                                                     ©DZ Reinstein 2009
                                              dzr@londonvisionclinic.com
Example RST Planning



     Refraction                 -10.75 D sph

     Pachymetry                 509 µm

     Flap Thickness (VisuMax)   80 µm

     Ablation Depth             135 µm

     Predicted RST              296 µm




                                                ©DZ Reinstein 2009
                                         dzr@londonvisionclinic.com
Outcomes
New Profile for High Myopia

• Patients
  –   Myopia SEQ -9.51 ± 1.32 D -8.00 up to -14.50 D
  –   Myopia max merid -10.18 ± 1.48 D -8.00 up to -16.00 D
  –   Cylinder -1.32 ± 1.10 D up to -6.25 D
  –   220 eyes
  –   1 year follow up
• Retreatments
  – 45% eyes treated as “two-stage”
  – Enhancement rate (non two-stage): 35%




                                                         ©DZ Reinstein 2009
                                                  dzr@londonvisionclinic.com
Advantages of Two Stage Procedure
• Increased safety
   – Greater RST for primary treatment
   – Artemis measured RST to calculate retreatment
   – Option to retreat using topography-guided profile
• More accurate result
• Patient has lower expectations




                                                           ©DZ Reinstein 2009
                                                    dzr@londonvisionclinic.com
Topography Guided Retreatment
                               Pre       Post      Reduced

                     Sph Ab   0.48 µm   0.28 µm        41%

                     HO RMS   0.72 µm   0.57 µm        21%




                                                   ©DZ Reinstein 2009
                                            dzr@londonvisionclinic.com
MEL80 High Myopia: Accuracy
                     Attempted vs. Achieved Spherical Equivalent
         -14

     )              y = 1.0726x + 0.8394
     s
     r
     e
     t
         -13             R² = 0.8759
     p
     o
     i
     D
     (
     t   -12
     n
     e
     l
     a
     v
     i   -11
     u
     q
     E
     l
     a
     c
     i   -10
     r
     e
     h
     p
     S    -9
     d
     e
     v
     e
     i
     h    -8
     c
     A

          -7
               -7       -8          -9       -10       -11       -12     -13       -14
                             Attempted Spherical Equivalent (Diopters)


                                                                                      ©DZ Reinstein 2009
                                                                               dzr@londonvisionclinic.com
MEL80 High Myopia: Accuracy
Within ±0.50
                   71%
D
Within ±1.00
                   94%      Accuracy: Within Range of Intended
D
                   35%                            33%

                   30%
                                                          25%
               s   25%
               e
               y
               E   20%
               e                          15%
               g   15%                                            13%
               a
               t
               n
               e   10%
               c
               r                                                           7%
               e
               P   5%             3%
                           0%                                                       1%        1%
                   0%
                          -2.00   -1.50   -1.00   -0.50   -0.13    0.14   +0.51   +1.01     +1.51
                           To -    To -    To -    To -     To      To     To      To        To
                           1.51    1.01    0.51    0.14    0.13   +0.50   +1.00   +1.50     +2.00
               Accuracy   0%      3%      15%     33%     25%     13%      7%       1%        1%
                                          Accuracy of Spherical Equivalent


                                                                                         ©DZ Reinstein 2009
                                                                                  dzr@londonvisionclinic.com
Monocular Efficacy
(excluding eyes not intended plano)

n=176
                             Efficacy: Monocular UCVA
                                                    100%          99%     99%           99%
          100%                                          97%
                                              90%
                                        83%
          80%
      s
      e
      y
      E   60%
      e                           47%
      g
      a
      t
      n   40%
      e                     28%
      c
      r
      e
      P   20%
                      11%
                 1%
           0%
                 20/12.5     20/16       20/20       20/25    20/32     20/40      20/63
     Pre BSCVA    1%         28%         83%         100%
     Efficacy     11%        47%         90%          97%     99%       99%         99%

                                                 Monocular UCVA


                                                                                       ©DZ Reinstein 2009
                                                                                dzr@londonvisionclinic.com
MEL80 High Myopia: Safety – BSCVA
n=220                        Safety: Lines Change BSCVA

        60%
                                                            52%


   s                                               40%
   e
   y 40%
   E
   e
   g
   a
   t
   n
   e 20%
   c
   r
   e
   P                                      N=4
                                                                       6%
                                           2%
                  0.0%        0.0%
         0%
                 Loss 3 or                           No              Gain 2 or
                              Loss 2      Loss 1            Gain 1
                   More                            Change             More
        Safety    0.0%        0.0%         2%       40%     52%        6%

                                       Lines Change BSCVA




                                                                                ©DZ Reinstein 2009
                                                                         dzr@londonvisionclinic.com
MEL80 High Myopia: Contrast Sensitivity




                                         *




  * Statistically significant (p<0.05)
                                                    ©DZ Reinstein 2009
                                             dzr@londonvisionclinic.com
Stability
           2.00



         ) 0.00
         D
         (
         t
         n
         e
         l -2.00
         a
         v
         i
         u
         q
         E -4.00
         l
         a
         c
         i
         r
         e -6.00
         h
         p
         S

          -8.00


         -10.00


         -12.00
                              3 Mo      6 Mo             12 Mo                           24 Mo

                     Pre-op          1 Day     1 Month      3 Months     6 Months      1 Year           2 Years

Mean±SD            -9.60±1.39   +0.41±0.82   +0.01±0.82    -0.18±0.86   -0.22±0.91   -0.04±0.91       -0.06±1.07

# eyes                220            199        201           188          158          124                45




                                                                                                     ©DZ Reinstein 2009
                                                                                              dzr@londonvisionclinic.com
Take Home Message
• Know your spherical aberration induction per dioptre
• Measure pre-op spherical aberration
• Check whether spherical aberration is going to go
  beyond the threshold
   – Use SA pre-compensation
   – Use a 2-stage procedure (wavefront / topography guided
     repair if necessary as second treatment)
• Caution with predicted RST
   –   Reduce potential errors
   –   Measure pachymetry with high repeatability instrument
   –   Use high reproducibility flap creation technique
   –   Always include flap thickness bias
                                                           ©DZ Reinstein 2009
                                                    dzr@londonvisionclinic.com
Thank You




 Very high myopic LASIK using new
 hybrid aspheric profiles
 Dan Z Reinstein MD MA(Cantab) FRCSC FRCOphth1,2,3,4

Contenu connexe

Tendances

Individualized Patient Care_Comparison of Different Variants of Blended Vision
Individualized Patient Care_Comparison of Different Variants of Blended VisionIndividualized Patient Care_Comparison of Different Variants of Blended Vision
Individualized Patient Care_Comparison of Different Variants of Blended VisionBreyer, Kaymak & Klabe Augenchirurgie
 
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...Breyer, Kaymak & Klabe Augenchirurgie
 
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract Surgery
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract SurgeryVector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract Surgery
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract SurgeryRyan Alfonso
 
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser Ophthalmoscope
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser OphthalmoscopeStevenson Eye Tracking With The Adaptive Optics Scanning Laser Ophthalmoscope
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser OphthalmoscopeKalle
 
New refractive optical MIOL concepts _ 
Comparison of different optical systems
New refractive optical MIOL concepts _ 
Comparison of different optical systemsNew refractive optical MIOL concepts _ 
Comparison of different optical systems
New refractive optical MIOL concepts _ 
Comparison of different optical systemsBreyer, Kaymak & Klabe Augenchirurgie
 
Aspheric IOLs for CRGH
Aspheric IOLs for CRGHAspheric IOLs for CRGH
Aspheric IOLs for CRGHperfectvision
 
Aspheric vs Spheric IOL
Aspheric vs Spheric IOLAspheric vs Spheric IOL
Aspheric vs Spheric IOLOlly Congga
 
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...Breyer, Kaymak & Klabe Augenchirurgie
 
A Comparative Study Of Surgically Induced Astigmatism In Superior, Supero-Te...
A Comparative  Study Of Surgically Induced Astigmatism In Superior, Supero-Te...A Comparative  Study Of Surgically Induced Astigmatism In Superior, Supero-Te...
A Comparative Study Of Surgically Induced Astigmatism In Superior, Supero-Te...Dr. Jagannath Boramani
 
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university)
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university) femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university)
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university) Hind Safwat
 
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...Breyer, Kaymak & Klabe Augenchirurgie
 
Lenstar 900 'Improving Outcomes' Brochure
Lenstar 900 'Improving Outcomes' BrochureLenstar 900 'Improving Outcomes' Brochure
Lenstar 900 'Improving Outcomes' BrochureHaag-Streit UK (HS-UK)
 
IOL power calculation formulae
IOL power calculation formulaeIOL power calculation formulae
IOL power calculation formulaepujarai
 
Biometry: Iol calculation
Biometry: Iol calculation Biometry: Iol calculation
Biometry: Iol calculation Noor Munirah Aab
 
Laser in periodontology
Laser in periodontologyLaser in periodontology
Laser in periodontologyMuthenna Rajab
 
Anterior Segment Company Showcase - ReVision Optics
Anterior Segment Company Showcase - ReVision OpticsAnterior Segment Company Showcase - ReVision Optics
Anterior Segment Company Showcase - ReVision OpticsHealthegy
 
Surgical induced astigmatism
Surgical induced astigmatismSurgical induced astigmatism
Surgical induced astigmatismNamrata Gupta
 

Tendances (20)

Individualized Patient Care_Comparison of Different Variants of Blended Vision
Individualized Patient Care_Comparison of Different Variants of Blended VisionIndividualized Patient Care_Comparison of Different Variants of Blended Vision
Individualized Patient Care_Comparison of Different Variants of Blended Vision
 
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...
Long-term Results of Refractive Lenticule Extraction ReLEx SMILE compared to ...
 
MIOL Treatment – Quo Vadis?
MIOL Treatment – Quo Vadis?MIOL Treatment – Quo Vadis?
MIOL Treatment – Quo Vadis?
 
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract Surgery
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract SurgeryVector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract Surgery
Vector-Adjusted NAPA Limbal Relaxing Incisions at the Time of Cataract Surgery
 
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser Ophthalmoscope
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser OphthalmoscopeStevenson Eye Tracking With The Adaptive Optics Scanning Laser Ophthalmoscope
Stevenson Eye Tracking With The Adaptive Optics Scanning Laser Ophthalmoscope
 
Acunex Vario – Neue Hydrophobe EDOF IOL mit Blaulichtfilter
Acunex Vario – Neue Hydrophobe EDOF IOL mit Blaulichtfilter Acunex Vario – Neue Hydrophobe EDOF IOL mit Blaulichtfilter
Acunex Vario – Neue Hydrophobe EDOF IOL mit Blaulichtfilter
 
New refractive optical MIOL concepts _ 
Comparison of different optical systems
New refractive optical MIOL concepts _ 
Comparison of different optical systemsNew refractive optical MIOL concepts _ 
Comparison of different optical systems
New refractive optical MIOL concepts _ 
Comparison of different optical systems
 
Aspheric IOLs for CRGH
Aspheric IOLs for CRGHAspheric IOLs for CRGH
Aspheric IOLs for CRGH
 
Aspheric vs Spheric IOL
Aspheric vs Spheric IOLAspheric vs Spheric IOL
Aspheric vs Spheric IOL
 
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...
Refractive versus Diffractive Optics for Enhanced Depth of Focus Intraocular ...
 
A Comparative Study Of Surgically Induced Astigmatism In Superior, Supero-Te...
A Comparative  Study Of Surgically Induced Astigmatism In Superior, Supero-Te...A Comparative  Study Of Surgically Induced Astigmatism In Superior, Supero-Te...
A Comparative Study Of Surgically Induced Astigmatism In Superior, Supero-Te...
 
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university)
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university) femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university)
femtosecond laser in ophthalmology by Dr. Hind Safwat (Al Azhr university)
 
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...
Evaluation of Stereopsis for Blended Vision Variants with Refractive MIOL com...
 
Lenstar 900 'Improving Outcomes' Brochure
Lenstar 900 'Improving Outcomes' BrochureLenstar 900 'Improving Outcomes' Brochure
Lenstar 900 'Improving Outcomes' Brochure
 
NW2011 Optic of human eye
NW2011 Optic of human eyeNW2011 Optic of human eye
NW2011 Optic of human eye
 
IOL power calculation formulae
IOL power calculation formulaeIOL power calculation formulae
IOL power calculation formulae
 
Biometry: Iol calculation
Biometry: Iol calculation Biometry: Iol calculation
Biometry: Iol calculation
 
Laser in periodontology
Laser in periodontologyLaser in periodontology
Laser in periodontology
 
Anterior Segment Company Showcase - ReVision Optics
Anterior Segment Company Showcase - ReVision OpticsAnterior Segment Company Showcase - ReVision Optics
Anterior Segment Company Showcase - ReVision Optics
 
Surgical induced astigmatism
Surgical induced astigmatismSurgical induced astigmatism
Surgical induced astigmatism
 

Similaire à High Myopia Final CZM Dubai_2011

MS Textile Chemistry Lecture 3- 4 Advanced Analytical Techniques.pptx
MS Textile Chemistry Lecture 3- 4  Advanced Analytical Techniques.pptxMS Textile Chemistry Lecture 3- 4  Advanced Analytical Techniques.pptx
MS Textile Chemistry Lecture 3- 4 Advanced Analytical Techniques.pptxChaudharyWaseemWasee
 
4. Lasers in Periodontics.ppt
4. Lasers in Periodontics.ppt4. Lasers in Periodontics.ppt
4. Lasers in Periodontics.pptViola Esther
 
Nuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingNuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingDr ATHUL CHANDRA.M
 
LASERS IN PROSTHODONTICS
LASERS IN PROSTHODONTICSLASERS IN PROSTHODONTICS
LASERS IN PROSTHODONTICSshari kurup
 
Contact lens by navid
Contact lens by navidContact lens by navid
Contact lens by navidNavidRahman6
 
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.senaimais
 
X ray diffraction or braggs diffraction or
X ray diffraction or braggs diffraction orX ray diffraction or braggs diffraction or
X ray diffraction or braggs diffraction orbpati5271
 
XRD-calculations and characterization.pdf
XRD-calculations and characterization.pdfXRD-calculations and characterization.pdf
XRD-calculations and characterization.pdfEmadElsehly
 
10. x raydiffraction jntu pharmacy
10. x raydiffraction jntu pharmacy10. x raydiffraction jntu pharmacy
10. x raydiffraction jntu pharmacyDr. Suman Pattanayak
 
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...Dr Sylvain Chamberland
 
8.1. microtech ion implant,1,2
8.1. microtech ion implant,1,28.1. microtech ion implant,1,2
8.1. microtech ion implant,1,2Bhargav Veepuri
 
Optical forces for assembling complex plasmonic nanostructures
Optical forces for assembling complex plasmonic nanostructuresOptical forces for assembling complex plasmonic nanostructures
Optical forces for assembling complex plasmonic nanostructuresNitish Chandra
 
advance material science
advance material scienceadvance material science
advance material scienceMalika Nischal
 

Similaire à High Myopia Final CZM Dubai_2011 (20)

MS Textile Chemistry Lecture 3- 4 Advanced Analytical Techniques.pptx
MS Textile Chemistry Lecture 3- 4  Advanced Analytical Techniques.pptxMS Textile Chemistry Lecture 3- 4  Advanced Analytical Techniques.pptx
MS Textile Chemistry Lecture 3- 4 Advanced Analytical Techniques.pptx
 
4. Lasers in Periodontics.ppt
4. Lasers in Periodontics.ppt4. Lasers in Periodontics.ppt
4. Lasers in Periodontics.ppt
 
Nuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingNuclear Magnification Resonance Imaging
Nuclear Magnification Resonance Imaging
 
LASERS IN PROSTHODONTICS
LASERS IN PROSTHODONTICSLASERS IN PROSTHODONTICS
LASERS IN PROSTHODONTICS
 
Degi2019.pdf
Degi2019.pdfDegi2019.pdf
Degi2019.pdf
 
Contact lens by navid
Contact lens by navidContact lens by navid
Contact lens by navid
 
WDS
WDSWDS
WDS
 
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.
Palestra 5 - Aplicação do laser como ferramenta de fabricação de moldes.
 
X ray diffraction or braggs diffraction or
X ray diffraction or braggs diffraction orX ray diffraction or braggs diffraction or
X ray diffraction or braggs diffraction or
 
XRD-calculations and characterization.pdf
XRD-calculations and characterization.pdfXRD-calculations and characterization.pdf
XRD-calculations and characterization.pdf
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
10. x raydiffraction jntu pharmacy
10. x raydiffraction jntu pharmacy10. x raydiffraction jntu pharmacy
10. x raydiffraction jntu pharmacy
 
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...
Aao 109th annual_session_boston_2009_closer look at the stability of sarpe 20...
 
8.1. microtech ion implant,1,2
8.1. microtech ion implant,1,28.1. microtech ion implant,1,2
8.1. microtech ion implant,1,2
 
X ray diffraction
X ray diffractionX ray diffraction
X ray diffraction
 
x-raydiffraction
x-raydiffractionx-raydiffraction
x-raydiffraction
 
Optical forces for assembling complex plasmonic nanostructures
Optical forces for assembling complex plasmonic nanostructuresOptical forces for assembling complex plasmonic nanostructures
Optical forces for assembling complex plasmonic nanostructures
 
advance material science
advance material scienceadvance material science
advance material science
 
Bm presentation
Bm presentationBm presentation
Bm presentation
 
Particle Shape
Particle ShapeParticle Shape
Particle Shape
 

Plus de London Vision Clinic

Insight Magazine Interview with Dan Reinstein
Insight Magazine Interview with Dan ReinsteinInsight Magazine Interview with Dan Reinstein
Insight Magazine Interview with Dan ReinsteinLondon Vision Clinic
 
Financial Times Article - Turning Politics into a Spectacular Event
Financial Times Article - Turning Politics into a Spectacular EventFinancial Times Article - Turning Politics into a Spectacular Event
Financial Times Article - Turning Politics into a Spectacular EventLondon Vision Clinic
 
Men's Health Laser Eye Surgery Article
Men's Health Laser Eye Surgery ArticleMen's Health Laser Eye Surgery Article
Men's Health Laser Eye Surgery ArticleLondon Vision Clinic
 
Target Sport Laser Eye Surgery Article
Target Sport Laser Eye Surgery ArticleTarget Sport Laser Eye Surgery Article
Target Sport Laser Eye Surgery ArticleLondon Vision Clinic
 
Excimer Laser for Hyperopia: Wider Limits
Excimer Laser for Hyperopia: Wider LimitsExcimer Laser for Hyperopia: Wider Limits
Excimer Laser for Hyperopia: Wider LimitsLondon Vision Clinic
 
Introduction to Refractive Eye Surgery
Introduction to Refractive Eye SurgeryIntroduction to Refractive Eye Surgery
Introduction to Refractive Eye SurgeryLondon Vision Clinic
 
6 Month Results of Topography Guided Repair
6 Month Results of Topography Guided Repair6 Month Results of Topography Guided Repair
6 Month Results of Topography Guided RepairLondon Vision Clinic
 
LASIK Surgery is Safe in the Long-Term
LASIK Surgery is Safe in the Long-TermLASIK Surgery is Safe in the Long-Term
LASIK Surgery is Safe in the Long-TermLondon Vision Clinic
 
The Pathophysiology of Regression Following LASIK
The Pathophysiology of Regression Following LASIKThe Pathophysiology of Regression Following LASIK
The Pathophysiology of Regression Following LASIKLondon Vision Clinic
 

Plus de London Vision Clinic (12)

Insight Magazine Interview with Dan Reinstein
Insight Magazine Interview with Dan ReinsteinInsight Magazine Interview with Dan Reinstein
Insight Magazine Interview with Dan Reinstein
 
Body Language Presbyopia Article
Body Language Presbyopia ArticleBody Language Presbyopia Article
Body Language Presbyopia Article
 
Financial Times Article - Turning Politics into a Spectacular Event
Financial Times Article - Turning Politics into a Spectacular EventFinancial Times Article - Turning Politics into a Spectacular Event
Financial Times Article - Turning Politics into a Spectacular Event
 
Men's Health Laser Eye Surgery Article
Men's Health Laser Eye Surgery ArticleMen's Health Laser Eye Surgery Article
Men's Health Laser Eye Surgery Article
 
Target Sport Laser Eye Surgery Article
Target Sport Laser Eye Surgery ArticleTarget Sport Laser Eye Surgery Article
Target Sport Laser Eye Surgery Article
 
Excimer Laser for Hyperopia: Wider Limits
Excimer Laser for Hyperopia: Wider LimitsExcimer Laser for Hyperopia: Wider Limits
Excimer Laser for Hyperopia: Wider Limits
 
Introduction to Refractive Eye Surgery
Introduction to Refractive Eye SurgeryIntroduction to Refractive Eye Surgery
Introduction to Refractive Eye Surgery
 
6 Month Results of Topography Guided Repair
6 Month Results of Topography Guided Repair6 Month Results of Topography Guided Repair
6 Month Results of Topography Guided Repair
 
The Routine LASIK Procedure
The Routine LASIK Procedure The Routine LASIK Procedure
The Routine LASIK Procedure
 
LASIK Surgery is Safe in the Long-Term
LASIK Surgery is Safe in the Long-TermLASIK Surgery is Safe in the Long-Term
LASIK Surgery is Safe in the Long-Term
 
Reinstein visumax
Reinstein visumaxReinstein visumax
Reinstein visumax
 
The Pathophysiology of Regression Following LASIK
The Pathophysiology of Regression Following LASIKThe Pathophysiology of Regression Following LASIK
The Pathophysiology of Regression Following LASIK
 

Dernier

Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Genuine Call Girls
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...jageshsingh5554
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Dernier (20)

Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 

High Myopia Final CZM Dubai_2011

  • 1. Very high myopic LASIK using new hybrid aspheric profiles Dan Z Reinstein MD MA(Cantab) FRCSC FRCOphth1,2,3,4 1. London Vision Clinic, London, UK 2. St. Thomas’ Hospital - Kings College, London, UK 3. Weill Medical College of Cornell University, New York, USA 4. Centre Hospitalier National d’Ophtalmologie, (Pr. Laroche) , Paris, France
  • 2. First Results: Munnerlyn Ablation Profile • Early ablation profiles often induced: – Night Vision disturbances – Decreased contrast sensitivity • Limited the range of treatable refractions • PROBLEM: Induction of spherical aberration Eur J Ophthalmol. 1994 Jan-Mar;4(1):43-51. Night vision after excimer laser photorefractive keratectomy: haze and halos. O'Brart DP, Lohmann CP, Fitzke FW, Smith SE, Kerr-Muir MG, Marshall J. ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 3. Ablation Profile Design: Larger Optical Zone Example: 5-mm Munnerlyn ablation for -6.00 D (1993 Summit Laser) Topography Wavefront Z(4,0) (OSA) 1.18 µm J Refract Corneal Surg. 1994 Mar-Apr;10(2):87-94. Excimer laser photorefractive keratectomy for myopia: comparison of 4.00- and 5.00-millimeter ablation zones. O'Brart DP, Gartry DS, Lohmann CP, Muir MG, Marshall J. Arch Ophthalmol. 1995 Apr;113(4):438-43. The effects of ablation diameter on the outcome of excimer laser photorefractive keratectomy. A prospective, randomized, double-blind study. O'Brart DP, Corbett MC, Lohmann CP, Kerr Muir MG, Marshall J. ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 4. Ablation Profile Design: Aspheric Profiles • Barraquer 1980 – Suggested parabolic keratomileusis • Seiler 1993 – PRK aspheric profiles – Less starburst & halos – Larger effective clear optical zone size ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 5. Why was spherical aberration increasing?
  • 6. Optimization: Fluence correction • Fluence correction: Topography – Beam reflection compensation – Beam projection compensation J Refract Surg 2001;17(5):S584-7. Influence of corneal curvature on calculation of ablation patterns used in photorefractive laser surgery. Mrochen M, Seiler T. ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 7. Optimization • Biomechanics VHF digital ultrasound ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 8. Reinstein et al. Journal of Refractive Surgery Artemis C12 Display 2000 Jul-Aug;16:414-30 VHF digital ultrasound Roberts C. The cornea is not a piece of plastic. JRS 2000; 16:407-413 ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 9. Examples of Peripheral Stromal Thickening Peripheral Stromal Thickening Central Flattening Roberts C. The cornea is not a piece of plastic. ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 10. Corneal Biomechanical Trade-off • Hyperopic shift induced by – Central flattening due to peripheral tissue removal • Myopic shift induced by Pre-Op – Epithelial thickening – Bowing of the back surface Epithelial thickening Post-Op Back surface bowing ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 12. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 13. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 14. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 15. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 16. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 17. ESCRS 2002, DZ Reinstein: Z4,0-Slider (aka Q-slider) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 18. Free Lunch? • Increasing ablation zone diameter • Adding asphericity • Increases central ablation depth • No “Free Lunch” ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 19. 550 µm Pachymetry: Forces Compromise • Modern aspheric ablation profiles still induce spherical aberration • Problem: high myopic corrections may result in NVDs Attempted Spherical Equivalent vs. Change in Z(4,0) Coefficient ASA Treatments 0.7 y = -0.059x - 0.0136 0.6 ) R² = 0.6444 A S O 0.5 , m µ ( t 0.4 n e i c i f f 0.3 e o C 0.2 ) 0 , 4 ( Z 0.1 n i e g 0.0 n a h C -0.1 -0.2 0.00 -1.00 -2.00 -3.00 -4.00 -5.00 -6.00 -7.00 -8.00 -9.00 -10.00 Attempted Spherical Equivalent (Diopters) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 21. Correlation of Contrast with Wavefront Spherical Aberration Contrast Sensitivity p 900 1.2 800 t 1.1 s 700 a o r i t t 1.0 600 n a o R C y 0.9 500 t d i e v z i t 2 400 i i l s a n 0.8 m 300 m r e µ o S 0.7 200 N 0.6 100 0.5 0 3 cpd 6 cpd 12 cpd 18 cpd Pre CRS-M Post CRSM- Control Pre Control Post Control Pre 1.02 1.02 1.03 1.04 Repair Repair Control Post 1.04 1.01 1.03 1.01 Sph Ab Area 122 276 563 410 Pre CRSM-Repair 0.85 0.84 0.77 0.75 Post CRSM-Repair 1.04 1.02 1.02 1.00 • 27% Gross Reduction • 53% Net Reduction (cf tolerable level) • Tolerable level ~0.56 µm @ 6mm ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 22. Pre-Compensate for Spherical Aberration • “Q-slider” – (WaveLight) • Wavefront-guided ablation – Includes pre-op spherical aberration – Effect dependent on pre-op spherical aberration • Our Approach: Include an “artificial” wavefront – Isolate spherical aberration: Z(4,0) as the only coefficient – Z(4,0) coefficient proportional to expected induction – Increase Z(4,0) coefficient: wavefront only 20% effective ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 23. Patient 1, OD -7.13 D Corrected 6mm OSA 6mm OSA Coma 0.09 µm Coma 0.04 µm Sph Ab 0.48 µm Sph Ab 0.42 µm HO RMS 0.59 µm HO RMS 0.52 µm ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 24. Patient 1, OS -9.00 D Corrected 6mm OSA 6mm OSA Coma 0.03 µm Coma 0.05 µm Sph Ab 0.49 µm Sph Ab 0.55 µm HO RMS 0.57 µm HO RMS 0.60 µm ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 25. Patient 1, Night Vision Pre Op Post Op Rx Treated OD -6.50 -1.25 x 178 OS -8.25 -1.50 x 17 ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 26. Induction of Spherical Aberration Complaint of NVD post RS1 1 ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 27. Limits to SA Pre-Compensation • Excess spherical aberration pre-compensation can lead to “central islands” TMS WASCA (zonal) Epithelium OD ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 28. CENTRAL ISLANDS: -5.50 D ablation Diplopia first week Slow resolution over 2 weeks Slides courtesy Gordon Balazsi, MD ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 29. CENTRAL ISLANDS: -5.00 D ablation Slides courtesy Gordon Balazsi, MD ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 30. Ablation Depth with SA Pre-Compensation ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 31. New Profile for High Myopia • Non-linear aspheric ablation profile: – Increased peripheral ablation (not ↑ z(4,0)) – Reduced induction of spherical aberration – Free lunch: some myopia corrected due to central flattening Extend this concept further to promote central flattening Ability to correct high myopia without risk of NVDs Peripheral Stromal Thickening Central Flattening Roberts C. The cornea is not a piece of plastic. ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 32. New Profile: “Free Lunch” • Over-corrected by +0.50 D compared with theory • Ablation depth lower than expected Attempted vs. Achieved Spherical Equivalent -14 y = 0.9958x - 0.5106 ) -13 R² = 0.9291 s r e t p-12 o i D ( t -11 n e l a v -10 i u q E -9 l a c i r e -8 h p S d -7 e v e i h -6 c A -5 -6 -7 -8 -9 -10 -11 -12 -13 -14 Attempted Spherical Equivalent (Diopters) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 33. Ablation Depth for New Profile ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 34. Femtosecond Lasers • Femtosecond lasers have improved flap thickness reproducibility (VisuMax SD: 8 µm) • We can create thinner flaps (VisuMax: 80 µm) • Thinner flaps extends the range of myopia in LASIK Pre-release online ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 35. Example RST Planning Refraction -10.75 D sph Pachymetry 509 µm Flap Thickness (VisuMax) 80 µm Ablation Depth 135 µm Predicted RST 296 µm ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 37. New Profile for High Myopia • Patients – Myopia SEQ -9.51 ± 1.32 D -8.00 up to -14.50 D – Myopia max merid -10.18 ± 1.48 D -8.00 up to -16.00 D – Cylinder -1.32 ± 1.10 D up to -6.25 D – 220 eyes – 1 year follow up • Retreatments – 45% eyes treated as “two-stage” – Enhancement rate (non two-stage): 35% ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 38. Advantages of Two Stage Procedure • Increased safety – Greater RST for primary treatment – Artemis measured RST to calculate retreatment – Option to retreat using topography-guided profile • More accurate result • Patient has lower expectations ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 39. Topography Guided Retreatment Pre Post Reduced Sph Ab 0.48 µm 0.28 µm 41% HO RMS 0.72 µm 0.57 µm 21% ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 40. MEL80 High Myopia: Accuracy Attempted vs. Achieved Spherical Equivalent -14 ) y = 1.0726x + 0.8394 s r e t -13 R² = 0.8759 p o i D ( t -12 n e l a v i -11 u q E l a c i -10 r e h p S -9 d e v e i h -8 c A -7 -7 -8 -9 -10 -11 -12 -13 -14 Attempted Spherical Equivalent (Diopters) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 41. MEL80 High Myopia: Accuracy Within ±0.50 71% D Within ±1.00 94% Accuracy: Within Range of Intended D 35% 33% 30% 25% s 25% e y E 20% e 15% g 15% 13% a t n e 10% c r 7% e P 5% 3% 0% 1% 1% 0% -2.00 -1.50 -1.00 -0.50 -0.13 0.14 +0.51 +1.01 +1.51 To - To - To - To - To To To To To 1.51 1.01 0.51 0.14 0.13 +0.50 +1.00 +1.50 +2.00 Accuracy 0% 3% 15% 33% 25% 13% 7% 1% 1% Accuracy of Spherical Equivalent ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 42. Monocular Efficacy (excluding eyes not intended plano) n=176 Efficacy: Monocular UCVA 100% 99% 99% 99% 100% 97% 90% 83% 80% s e y E 60% e 47% g a t n 40% e 28% c r e P 20% 11% 1% 0% 20/12.5 20/16 20/20 20/25 20/32 20/40 20/63 Pre BSCVA 1% 28% 83% 100% Efficacy 11% 47% 90% 97% 99% 99% 99% Monocular UCVA ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 43. MEL80 High Myopia: Safety – BSCVA n=220 Safety: Lines Change BSCVA 60% 52% s 40% e y 40% E e g a t n e 20% c r e P N=4 6% 2% 0.0% 0.0% 0% Loss 3 or No Gain 2 or Loss 2 Loss 1 Gain 1 More Change More Safety 0.0% 0.0% 2% 40% 52% 6% Lines Change BSCVA ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 44. MEL80 High Myopia: Contrast Sensitivity * * Statistically significant (p<0.05) ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 45. Stability 2.00 ) 0.00 D ( t n e l -2.00 a v i u q E -4.00 l a c i r e -6.00 h p S -8.00 -10.00 -12.00 3 Mo 6 Mo 12 Mo 24 Mo Pre-op 1 Day 1 Month 3 Months 6 Months 1 Year 2 Years Mean±SD -9.60±1.39 +0.41±0.82 +0.01±0.82 -0.18±0.86 -0.22±0.91 -0.04±0.91 -0.06±1.07 # eyes 220 199 201 188 158 124 45 ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 46. Take Home Message • Know your spherical aberration induction per dioptre • Measure pre-op spherical aberration • Check whether spherical aberration is going to go beyond the threshold – Use SA pre-compensation – Use a 2-stage procedure (wavefront / topography guided repair if necessary as second treatment) • Caution with predicted RST – Reduce potential errors – Measure pachymetry with high repeatability instrument – Use high reproducibility flap creation technique – Always include flap thickness bias ©DZ Reinstein 2009 dzr@londonvisionclinic.com
  • 47. Thank You Very high myopic LASIK using new hybrid aspheric profiles Dan Z Reinstein MD MA(Cantab) FRCSC FRCOphth1,2,3,4