TRS 398
CODE OF PRACTICE FOR HEAVY
ION BEAMS
This Code of Practice is based on a
calibration factor in terms of absorbed dose
to water of an ionization chamber in a
reference beam which is taken to be 60Co
gamma rays.
 The Code of Practice applies to heavy
ion beams with atomic numbers between
2(He) and 18(Ar) which have ranges of 2–
30 g/cm2 in water.
 The depth dose distribution of a mono
energetic heavy ion beam in water is shown
in figure
Introduction
For clinical applications of heavy ion beams
we have to generate ‘spread-out Bragg peaks
‘(SOBP) in order to include the complete
target volume inside the SOBP.
 In clinical applications it is common to use
a biological effective dose instead of a physical
dose (absorbed dose to water) because of the
uniformity in its dose distributions.
 The difference between the two kinds of
distributions can be compared from the figures
. In the physical dose distribution we can see a
lack of uniformity in the SOBP.
 The use of a biological effective dose makes
it possible to compare results obtained with
conventional radiotherapy to those using heavy
ion radiotherapy.
 Unlike the clinical applications ,the dosimetry of heavy ion is carried out by
determination of the physical dose using an ionization chamber calibrated in terms
of absorbed dose to water, ND,w,Qo.
 The goal behind this approach is to achieve international consistency in
dosimetry by adopting the same formalism and procedures for all the radiotherapy
beams used throughout the world.
Unfortunately, Task Group 20 in 1984 by the Association of American Physicists
in Medicine is the only protocol available regarding dosimetry recommendations .
Thus there is a need for a new protocol which will be able to establish a global
consistency in the determination of absorbed dose to water with heavy ions and is
common to dosimetry protocols.
 For an accurate determination of absorbed dose using an ionization chamber, we
need to know the energy spectra of the incident heavy ion beam, the projectile
fragments as well as that of the target fragmented nuclei.
Dosimetry equipments
Ionization chambers
Cylindrical and plane-parallel ionization chambers are recommended for use as
reference instruments in clinical heavy ion beams.
 cylindrical ionization chambers are preferred for reference dosimetry for SOBP
width ≥ 2.0 g/cm2 and plane-parallel chambers are preferred for SOBP width < 2.0
g/cm2.
The reason for this preferences is the higher combined standard uncertainty on
Dw,Q for plane-parallel ionization chambers due to their higher uncertainty for pwall
in the 60Co reference beam quality.
In the case of cylindrical chambers ,graphite walled are preferred to plastic
walled chambers because of their better long term stability and smaller chamber to
chamber variations.
 Since the depth dose distribution in the SOBP for heavy ion beams is not flat and
the slope depends on the width of the SOBP , measurement of an effective point of
the chamber, Peff, is important.
 The reference point of the cylindrical chamber should be positioned at a distance
0.75 rcyl deeper than the point of interest in the phantom, where rcyl is the inner
radius of the chamber.
For plane-parallel ionization chambers, the reference point is taken to be on the
inner surface of the entrance window, at the centre of the window. This point is
positioned at the point of interest in the phantom.
The cavity diameter of the plane-parallel ionization chamber or the cavity length
of the cylindrical ionization chamber should not be larger than approximately half
the reference field size.
Phantoms and chamber sleeves
 Water is recommended as the reference medium for measurements of absorbed
dose.
The extension of phantom should be at least 5 cm beyond all four sides of the
field size at the depth of measurement and should extend to at least 5 g/cm2
beyond the maximum depth of measurement.
In horizontal beams, phantom should have plastic window with a thickness twin
between 0.2 and 0.5 cm. In the case of non-waterproof chambers, a waterproofing
sleeve made of PMMA, with a thickness than 1.0 mm have to be used.
To provide equilibrium air pressure in the chamber ,the air gap between the
chamber wall and the waterproofing sleeve should be within (0.1–0.3 mm).
If it is possible, the same waterproofing sleeve that was used for calibration of user
chamber should also be used for reference dosimetry. Otherwise another sleeve of
the same material and of similar thickness should be used.
Use of plastic phantoms for reference dosimetry in heavy ion beams is not
recommended.
 One of the reason for this is difficulty in availability of water to plastic fluence
correction factors, hpl, and other reason is difference in the fluence of heavy ions
including fragmented particles in a plastic phantom and that in a water phantom.
 However, for routine quality assurance measurements plastic phantoms can be
used, provided a transfer factor between plastic and water has been established.
BEAM QUALITY SPECIFICATION
Unavailability of both experimental and theoretical data regarding spectral
distributions of heavy ion beams makes beam quality specification more
impractical.
 The current practice for characterizing a heavy ion beam is to use the atomic
number, mass number, energy of the incident heavy ion beam, width of SOBP and
range.
DETERMINATION OF ABSORBED DOSE TO WATER
From the spread out Bragg peak of a heavy ion figure we have seen that
The depth dose distribution is not flat, and the dose at the distal end of the SOBP
is smaller than that at the proximal part.
The slope near the centre of a broad SOBP is rather small whereas that of a
narrow SOBP is steep.
 The reference depth for calibration should be taken at the centre of the SOBP, at
the centre of the target volume.
Reference conditions for the determination of absorbed dose to water are given as
Determination of absorbed dose under reference conditions
The absorbed dose to water at the reference depth zref in water in a heavy ion beam
of quality Q and in the absence of the chamber is given by
where MQ is the meter reading of the dosimeter in accordance with the
reference conditions corrected for the influence quantities like
temperature and pressure, electrometer calibration, polarity effect
and ion recombination etc
ND,w,Qo is the calibration factor in terms of absorbed dose to water for the
dosimeter at the reference quality Qo.
kQ,Qo is a chamber specific factor which corrects for differences between
the reference beam quality Qo and the actual beam quality being
used, Q
Recombination correction in heavy ion beams
Dose rate will be very high when beams are generated by pulsed scanning
techniques and so recombination effects must be taken into account.
The correction factor for general (Volume) recombination is obtained
experimentally by the two voltage method.
 When general recombination is negligible, initial recombination should be taken
into account for heavy ion beams, especially when the dose is measured using
plane-parallel ionization chambers.
 The collected ionization current should be fitted by the linear relation
l/icol = 1/i∞ + b/V
where V is the polarizing voltage applied to the chamber. The correction factor is
given by ks
ini = i∞ /icol.
Part of worksheet regarding calculation for Initial ion recombination
Values of kQ,Qo
 Beam quality specifications are not currently used for the dosimetry of heavy-ion
beams and so kQ values depend only on the chamber type used.
 Experimental values of the factor kQ,Qo are not readily available and, therefore, in
this report only theoretical values will be used.
The correction factor is defined as.
 Since there is no primary standard of absorbed dose to water for heavy ion
beams is available, all values for kQ,Qo given in this Code of Practice are derived by
calculation and are based on 60Co gamma radiation as the reference beam quality Qo
at 60Co.
 There is no information available on perturbation factors for ion chambers in
heavy ion beams, and so is assumed as unity.
The stopping-power ratios and W values for heavy ion beams are taken to be
independent of the beam quality, because of lack of experimental data.
The contribution of fragmented nuclei to stopping-power ratios and W values
are also assumed to be negligible.
 Constant values of the stopping-power ratio and W value are therefore
adopted here for all heavy ion beams — these are 1.130 and 34.50 eV,
respectively.
 As the stopping power ratio sw,air of heavy ions is so close to that of 60Co, the
kQ values for heavy ions are dominated by the ratio of Wair values and the
chamber specific perturbation factors
Values of kQ for various cylindrical and plane-parallel ionization chambers
in common use are given. Some of the chambers listed in this table fail to
meet some of the minimum requirements .However, they have been
included because of their current clinical use.
Thank you

Contenu connexe

Tendances

Shielding calculation
Shielding calculationShielding calculation
Shielding calculation
BBAdhikari
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
Santam Chakraborty
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
fondas vakalis
 

Tendances (20)

Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
Session 2 radiotherapy
Session 2 radiotherapySession 2 radiotherapy
Session 2 radiotherapy
 
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapy
 
Shielding calculation
Shielding calculationShielding calculation
Shielding calculation
 
Dose Distribution Measurement (part 1)
Dose Distribution Measurement (part 1)Dose Distribution Measurement (part 1)
Dose Distribution Measurement (part 1)
 
Multileaf Collimator
Multileaf CollimatorMultileaf Collimator
Multileaf Collimator
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPY
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 
Brachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .RBrachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .R
 
Dose volume histogram
Dose volume histogramDose volume histogram
Dose volume histogram
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
 
Beam modification
Beam modification Beam modification
Beam modification
 
Physical and clinical aspects of brachytherapy
Physical and clinical aspects of  brachytherapyPhysical and clinical aspects of  brachytherapy
Physical and clinical aspects of brachytherapy
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in Radiotherapy
 
Icru 58.
Icru 58.Icru 58.
Icru 58.
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy Planning
 
Icru – 83 dr. upasna
Icru – 83  dr. upasnaIcru – 83  dr. upasna
Icru – 83 dr. upasna
 

En vedette (9)

TRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xraysTRS 398 code of practice for low KV xrays
TRS 398 code of practice for low KV xrays
 
Dosis absorbida
Dosis absorbidaDosis absorbida
Dosis absorbida
 
218 2 - radioterapia
218 2 - radioterapia218 2 - radioterapia
218 2 - radioterapia
 
Mejora de la seguridad en radioterapia PhD Sandra Guzman 2015
Mejora de la seguridad en radioterapia PhD Sandra Guzman 2015Mejora de la seguridad en radioterapia PhD Sandra Guzman 2015
Mejora de la seguridad en radioterapia PhD Sandra Guzman 2015
 
Clase Villareal
Clase VillarealClase Villareal
Clase Villareal
 
Radioterapia
RadioterapiaRadioterapia
Radioterapia
 
Proteccion radiologica en radioterapia
Proteccion radiologica en radioterapiaProteccion radiologica en radioterapia
Proteccion radiologica en radioterapia
 
CONTROL DE CALIDAD DE EQUIPOS DE DOSIMETRIA
CONTROL DE CALIDAD DE EQUIPOS DE DOSIMETRIACONTROL DE CALIDAD DE EQUIPOS DE DOSIMETRIA
CONTROL DE CALIDAD DE EQUIPOS DE DOSIMETRIA
 
Fundamentos De Radioterapia
Fundamentos De  RadioterapiaFundamentos De  Radioterapia
Fundamentos De Radioterapia
 

Similaire à Trs 398 code of practice for heavy ion

PVSC2015 final paper
PVSC2015 final paperPVSC2015 final paper
PVSC2015 final paper
Abhishek Iyer
 
Brent Godau Chem 499 Thesis
Brent Godau Chem 499 ThesisBrent Godau Chem 499 Thesis
Brent Godau Chem 499 Thesis
Brent Godau
 
INTERNSHIP-REPORT-CHIRAS
INTERNSHIP-REPORT-CHIRASINTERNSHIP-REPORT-CHIRAS
INTERNSHIP-REPORT-CHIRAS
Dimitris Chiras
 
TG-51 Ramesh presentation
TG-51 Ramesh presentationTG-51 Ramesh presentation
TG-51 Ramesh presentation
K.K.D.RAMESH -
 
The International Journal of Engineering and Science
The International Journal of Engineering and ScienceThe International Journal of Engineering and Science
The International Journal of Engineering and Science
theijes
 
CHEME 5650 FINAL REPORT rc695
CHEME 5650 FINAL REPORT rc695CHEME 5650 FINAL REPORT rc695
CHEME 5650 FINAL REPORT rc695
Ricki Chairil
 
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
David Parsons
 

Similaire à Trs 398 code of practice for heavy ion (20)

Basics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptxBasics of Electron Dosimetry presentation.pptx
Basics of Electron Dosimetry presentation.pptx
 
Out put calibration of co60
Out put calibration of co60Out put calibration of co60
Out put calibration of co60
 
PVSC2015 final paper
PVSC2015 final paperPVSC2015 final paper
PVSC2015 final paper
 
Brent Godau Chem 499 Thesis
Brent Godau Chem 499 ThesisBrent Godau Chem 499 Thesis
Brent Godau Chem 499 Thesis
 
Analysis and Design of Lead Salt PbSe/PbSrSe Single Quantum Well In the Infra...
Analysis and Design of Lead Salt PbSe/PbSrSe Single Quantum Well In the Infra...Analysis and Design of Lead Salt PbSe/PbSrSe Single Quantum Well In the Infra...
Analysis and Design of Lead Salt PbSe/PbSrSe Single Quantum Well In the Infra...
 
INTERNSHIP-REPORT-CHIRAS
INTERNSHIP-REPORT-CHIRASINTERNSHIP-REPORT-CHIRAS
INTERNSHIP-REPORT-CHIRAS
 
Dosimetry 2
Dosimetry  2Dosimetry  2
Dosimetry 2
 
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVALCE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
 
TG-51 Ramesh presentation
TG-51 Ramesh presentationTG-51 Ramesh presentation
TG-51 Ramesh presentation
 
Well logs
Well logsWell logs
Well logs
 
Sonochemical Reactor with Phase Control
Sonochemical Reactor with Phase ControlSonochemical Reactor with Phase Control
Sonochemical Reactor with Phase Control
 
The International Journal of Engineering and Science
The International Journal of Engineering and ScienceThe International Journal of Engineering and Science
The International Journal of Engineering and Science
 
CHEME 5650 FINAL REPORT rc695
CHEME 5650 FINAL REPORT rc695CHEME 5650 FINAL REPORT rc695
CHEME 5650 FINAL REPORT rc695
 
Mafuz ppt
Mafuz pptMafuz ppt
Mafuz ppt
 
EAGE 2013
EAGE 2013EAGE 2013
EAGE 2013
 
Dosimetry concepts and dosimeters
Dosimetry concepts and dosimetersDosimetry concepts and dosimeters
Dosimetry concepts and dosimeters
 
Study of semiconductor with positron
Study of semiconductor with positronStudy of semiconductor with positron
Study of semiconductor with positron
 
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
Beam generation and planar imaging at energies below 2.40 MeV with carbon and...
 
Cavity theory-Radiation physics
Cavity theory-Radiation physicsCavity theory-Radiation physics
Cavity theory-Radiation physics
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopy
 

Dernier

The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
seri bangash
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
levieagacer
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 
Conjugation, transduction and transformation
Conjugation, transduction and transformationConjugation, transduction and transformation
Conjugation, transduction and transformation
Areesha Ahmad
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Sérgio Sacani
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
PirithiRaju
 

Dernier (20)

Dubai Call Girls Beauty Face Teen O525547819 Call Girls Dubai Young
Dubai Call Girls Beauty Face Teen O525547819 Call Girls Dubai YoungDubai Call Girls Beauty Face Teen O525547819 Call Girls Dubai Young
Dubai Call Girls Beauty Face Teen O525547819 Call Girls Dubai Young
 
module for grade 9 for distance learning
module for grade 9 for distance learningmodule for grade 9 for distance learning
module for grade 9 for distance learning
 
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit flypumpkin fruit fly, water melon fruit fly, cucumber fruit fly
pumpkin fruit fly, water melon fruit fly, cucumber fruit fly
 
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort ServiceCall Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
 
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verifiedSector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
Sector 62, Noida Call girls :8448380779 Model Escorts | 100% verified
 
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptxCOST ESTIMATION FOR A RESEARCH PROJECT.pptx
COST ESTIMATION FOR A RESEARCH PROJECT.pptx
 
300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx
 
The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
 
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
9999266834 Call Girls In Noida Sector 22 (Delhi) Call Girl Service
 
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
Human & Veterinary Respiratory Physilogy_DR.E.Muralinath_Associate Professor....
 
Clean In Place(CIP).pptx .
Clean In Place(CIP).pptx                 .Clean In Place(CIP).pptx                 .
Clean In Place(CIP).pptx .
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 
Site Acceptance Test .
Site Acceptance Test                    .Site Acceptance Test                    .
Site Acceptance Test .
 
Conjugation, transduction and transformation
Conjugation, transduction and transformationConjugation, transduction and transformation
Conjugation, transduction and transformation
 
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
 
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate ProfessorThyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
 

Trs 398 code of practice for heavy ion

  • 1. TRS 398 CODE OF PRACTICE FOR HEAVY ION BEAMS
  • 2. This Code of Practice is based on a calibration factor in terms of absorbed dose to water of an ionization chamber in a reference beam which is taken to be 60Co gamma rays.  The Code of Practice applies to heavy ion beams with atomic numbers between 2(He) and 18(Ar) which have ranges of 2– 30 g/cm2 in water.  The depth dose distribution of a mono energetic heavy ion beam in water is shown in figure Introduction
  • 3. For clinical applications of heavy ion beams we have to generate ‘spread-out Bragg peaks ‘(SOBP) in order to include the complete target volume inside the SOBP.  In clinical applications it is common to use a biological effective dose instead of a physical dose (absorbed dose to water) because of the uniformity in its dose distributions.  The difference between the two kinds of distributions can be compared from the figures . In the physical dose distribution we can see a lack of uniformity in the SOBP.  The use of a biological effective dose makes it possible to compare results obtained with conventional radiotherapy to those using heavy ion radiotherapy.
  • 4.  Unlike the clinical applications ,the dosimetry of heavy ion is carried out by determination of the physical dose using an ionization chamber calibrated in terms of absorbed dose to water, ND,w,Qo.  The goal behind this approach is to achieve international consistency in dosimetry by adopting the same formalism and procedures for all the radiotherapy beams used throughout the world. Unfortunately, Task Group 20 in 1984 by the Association of American Physicists in Medicine is the only protocol available regarding dosimetry recommendations . Thus there is a need for a new protocol which will be able to establish a global consistency in the determination of absorbed dose to water with heavy ions and is common to dosimetry protocols.  For an accurate determination of absorbed dose using an ionization chamber, we need to know the energy spectra of the incident heavy ion beam, the projectile fragments as well as that of the target fragmented nuclei.
  • 5. Dosimetry equipments Ionization chambers Cylindrical and plane-parallel ionization chambers are recommended for use as reference instruments in clinical heavy ion beams.  cylindrical ionization chambers are preferred for reference dosimetry for SOBP width ≥ 2.0 g/cm2 and plane-parallel chambers are preferred for SOBP width < 2.0 g/cm2. The reason for this preferences is the higher combined standard uncertainty on Dw,Q for plane-parallel ionization chambers due to their higher uncertainty for pwall in the 60Co reference beam quality. In the case of cylindrical chambers ,graphite walled are preferred to plastic walled chambers because of their better long term stability and smaller chamber to chamber variations.
  • 6.  Since the depth dose distribution in the SOBP for heavy ion beams is not flat and the slope depends on the width of the SOBP , measurement of an effective point of the chamber, Peff, is important.  The reference point of the cylindrical chamber should be positioned at a distance 0.75 rcyl deeper than the point of interest in the phantom, where rcyl is the inner radius of the chamber. For plane-parallel ionization chambers, the reference point is taken to be on the inner surface of the entrance window, at the centre of the window. This point is positioned at the point of interest in the phantom. The cavity diameter of the plane-parallel ionization chamber or the cavity length of the cylindrical ionization chamber should not be larger than approximately half the reference field size.
  • 7. Phantoms and chamber sleeves  Water is recommended as the reference medium for measurements of absorbed dose. The extension of phantom should be at least 5 cm beyond all four sides of the field size at the depth of measurement and should extend to at least 5 g/cm2 beyond the maximum depth of measurement. In horizontal beams, phantom should have plastic window with a thickness twin between 0.2 and 0.5 cm. In the case of non-waterproof chambers, a waterproofing sleeve made of PMMA, with a thickness than 1.0 mm have to be used. To provide equilibrium air pressure in the chamber ,the air gap between the chamber wall and the waterproofing sleeve should be within (0.1–0.3 mm).
  • 8. If it is possible, the same waterproofing sleeve that was used for calibration of user chamber should also be used for reference dosimetry. Otherwise another sleeve of the same material and of similar thickness should be used. Use of plastic phantoms for reference dosimetry in heavy ion beams is not recommended.  One of the reason for this is difficulty in availability of water to plastic fluence correction factors, hpl, and other reason is difference in the fluence of heavy ions including fragmented particles in a plastic phantom and that in a water phantom.  However, for routine quality assurance measurements plastic phantoms can be used, provided a transfer factor between plastic and water has been established.
  • 9. BEAM QUALITY SPECIFICATION Unavailability of both experimental and theoretical data regarding spectral distributions of heavy ion beams makes beam quality specification more impractical.  The current practice for characterizing a heavy ion beam is to use the atomic number, mass number, energy of the incident heavy ion beam, width of SOBP and range. DETERMINATION OF ABSORBED DOSE TO WATER From the spread out Bragg peak of a heavy ion figure we have seen that The depth dose distribution is not flat, and the dose at the distal end of the SOBP is smaller than that at the proximal part. The slope near the centre of a broad SOBP is rather small whereas that of a narrow SOBP is steep.  The reference depth for calibration should be taken at the centre of the SOBP, at the centre of the target volume. Reference conditions for the determination of absorbed dose to water are given as
  • 10.
  • 11. Determination of absorbed dose under reference conditions The absorbed dose to water at the reference depth zref in water in a heavy ion beam of quality Q and in the absence of the chamber is given by where MQ is the meter reading of the dosimeter in accordance with the reference conditions corrected for the influence quantities like temperature and pressure, electrometer calibration, polarity effect and ion recombination etc ND,w,Qo is the calibration factor in terms of absorbed dose to water for the dosimeter at the reference quality Qo. kQ,Qo is a chamber specific factor which corrects for differences between the reference beam quality Qo and the actual beam quality being used, Q
  • 12. Recombination correction in heavy ion beams Dose rate will be very high when beams are generated by pulsed scanning techniques and so recombination effects must be taken into account. The correction factor for general (Volume) recombination is obtained experimentally by the two voltage method.  When general recombination is negligible, initial recombination should be taken into account for heavy ion beams, especially when the dose is measured using plane-parallel ionization chambers.  The collected ionization current should be fitted by the linear relation l/icol = 1/i∞ + b/V where V is the polarizing voltage applied to the chamber. The correction factor is given by ks ini = i∞ /icol.
  • 13. Part of worksheet regarding calculation for Initial ion recombination
  • 14. Values of kQ,Qo  Beam quality specifications are not currently used for the dosimetry of heavy-ion beams and so kQ values depend only on the chamber type used.  Experimental values of the factor kQ,Qo are not readily available and, therefore, in this report only theoretical values will be used. The correction factor is defined as.  Since there is no primary standard of absorbed dose to water for heavy ion beams is available, all values for kQ,Qo given in this Code of Practice are derived by calculation and are based on 60Co gamma radiation as the reference beam quality Qo at 60Co.
  • 15.  There is no information available on perturbation factors for ion chambers in heavy ion beams, and so is assumed as unity. The stopping-power ratios and W values for heavy ion beams are taken to be independent of the beam quality, because of lack of experimental data. The contribution of fragmented nuclei to stopping-power ratios and W values are also assumed to be negligible.  Constant values of the stopping-power ratio and W value are therefore adopted here for all heavy ion beams — these are 1.130 and 34.50 eV, respectively.  As the stopping power ratio sw,air of heavy ions is so close to that of 60Co, the kQ values for heavy ions are dominated by the ratio of Wair values and the chamber specific perturbation factors
  • 16. Values of kQ for various cylindrical and plane-parallel ionization chambers in common use are given. Some of the chambers listed in this table fail to meet some of the minimum requirements .However, they have been included because of their current clinical use.
  • 17.
  • 18.
  • 19.
  • 20.

Notes de l'éditeur

  1. bcs there is a lack of primary standards for heavy ions
  2. Forthe commonly used plastics PMMA and clear polystyrene, the nominal values ñ PMMA = 1.19 g/cm3 and ñ polystyrene = 1.06 g/cm3 may be used for the calcula-tion of the water equivalent thickness of the window.