ELECTRICAL AND ELECTRONINCS
MEASUREMENTS
MEASUREMENT
The process of measuring is essentially that of
comparing some unknown value with a value which
is assumed to be known.
 1) latter one in standard.
 2) And measuring system is instrument.

 Characteristics

of INSTRUMENTS

 Calibration: All

the static characteristics are
obtained in one form or another by a process called
Calibration.
Calibration procedures involve a comparison of
the particular instrument with either
a) a primary standard.
b) a secondary standard with a higher accuracy
than the instruments be calibrated.
c) an instrument of known accuracy.

 Accuracy:

Accuracy may be defined as the ability of
a device or a system to respond to a true value of a
measured variable under reference conditions.
 Precision:

Precision is defined by the degree of exactness
for which an instrument is designed or intended to perform.

 Repeatability:

It is the closeness of agreement among a
number of consecutive measurements of the output for the
same value of the input, under the same operating
conditions.

 Reproducibility:

It is the closeness of agreement among
repeated measurements of the output for the same value of
the input, made under the same operating conditions over a
period of time.

 Drift:

It is an undesired change or a gradual variation in
output over a period of time that is unrelated to changes in
input and operating conditions.


Span: If in a measuring instrument the highest point of
calibration is y units and the lowest point x units.
Then the instrument range y units
The instrument span is given by
Span=(x-y)units



Sensitivity: Sensitivity can be defined as the ratio of a
change in output to the change in input which causes it.



Resolution: The smallest increment in input (the quantity
being measured) which can be detected with certainty by an
instrument is its resolution.



Dead zone: Dead zone is the largest range of values of a
measured variable to which the instrument does not
respond.
ERROR
 Types

of Error



Static Error : the numerical difference between the true
value of a quantity and its value as obtained by
measurement.



Mistakes: These are errors due to human mistakes such as
careless reading, mistakes in observations, incorrect
application of a correction, improper application of
instruments and computational errors.

Systematic Error:
 1)Instrumental Error: Instrumental errors are the errors
inherent in measuring instruments because of their
mechanical structure, such as friction in bearings of various
moving components, irregular spring tension.



2)Environmental Error: it is due to conditions external to
the measuring device including conditions in the area
surrounding the instrument such as the effect of change
in temp , humidity, barometric pressure , or magnetic or
electrostatic fields.



Random Error: the cause of such error is unknown or not
determinable in the ordinary process of making
measurements. Such errors are normally small and follows
the laws of chance.
SOURCES OF ERROR
Insufficient knowledge of process parameters and
design conditions.
 Poor design.
 Poor maintenance.
 Error caused by people who operate instrument
equipment.
 Certain design limitations.

TYPES OF INSTRUMENTS










Absolute Instruments:
Absolute instruments are those which give the value of the
electrical quantity to be measured, in terms of the constant of
the instruments and their deflection only, e.g. tangent
galvanometer.
Secondary Instruments:
Secondary instruments are those which have been
precalibrated by comparison with an absolute instrument. The
value of the electrical quantity to be measured in these
instruments can be determined from the deflection of the
instrument.
Without calibration of such an instrument, the deflection is
meaningless.
Different types of secondary instruments:
1)Indicating
2)Recording
3)Integrating


1)Indicating:
Indicating instruments are those which indicate the instantaneous
value of the electrical quantity being measured, at the time at
which it is being measured. Their indications are given by pointers
moving over calibrated dials(scale), e.g. ammeters,voltmeters and
wattmeters.



2)Recording:
Recording instruments are those which give a continuous record
of variations of the electrical quantity over a selected period of
time. The moving system of the instrument carries an inked pen
which rests tightly on a graph chart. E.g. recording voltmeters
used in supply station.



3)Integrating:
Integrating instruments are those which measure and register the
total quantity of electricity (in ampere-hour) or the total amount of
electrical energy(in watt-hours or kilowatt-hours) supplied to a
circuit over a period of time, e.g. ampere-hour meters, energy
meters.
ESSENTIALS OF INDICATING INSTRUMENTS
 Deflecting

Torque(Td):

It is the torque which deflects the pointer on a calibrated
scale according to the electrical quantity passing through the
instrument. The deflecting torque causes the moving system
and hence the pointer attached to it to move from zero
position to indicate on a graduated scale the value of
electrical quantity being measured.
 Controlling

Torque(Tc):

It is the torque which controls the movement of the pointer
on a particular scale according to the quantity of electricity
passing through it. If deflecting torque were acting alone, the
pointer would continue to move indefinitely and would swing
over to the maximum deflected position irrespective of the
magnitude of current (or voltage or power) to be measured.
1)

Spring Control:

In the spring control method, a hair-spring, usually of phosphorbronze, attached to the moving system is used. With the deflection of
the pointer, the spring is twisted in the opposite direction. This twist in
the spring produces a restoring torque which is directly proportional
to the angle of deflection of the moving system. The pointer comes to
a position of rest (or equilibrium) when the deflecting torque (Td) and
controlling torque (Tc) are equal.
Tc ∞ θ
To give a controlling torque which is directly proportional to the angle of
deflection of the moving system, the number of turns of the spring
should be fairly large so that the deformation per unit length is small.
The stress in the spring must be limited to such a value that there is no
permanent set. Springs are made of materials which are

Non magnetic

Not subject to much fatigue

Low in specific resistance

Have low temperature coefficient of resistance.
2)

Gravity Control:

Gravity control is obtained by attaching a small weight to the moving
system in such a way that it produces a restoring or controlling torque
when the system is deflected.
Tc ∞ Sinθ
Thus, controlling torque in a gravity control system is proportional to the sine
of the angle of deflection.
The degree of control is adjusted by screwing the weight up or down on the
carrying system.
Advantages:
The advantages of gravity control system, as compared to spring control,
are given below:

It is cheap

It is unaffected by temperature

It is not subjected to fatigue or distortion, with time.
Disadvantages:
The disadvantages of gravity control system, as compared to spring control,
are given below:

It gives a cramped scale.

The instrument has to be kept vertical.


Damping Torque:

If the moving system is acted upon by deflecting and controlling torques alone,
then pointer, due to inertia, will oscillate about its final deflected position for quite
sometime before coming to rest. This is often undesirable because it makes
difficult to obtain quick and accurate readings. In order to avoid these oscillations
of the pointer and to bring it quickly to its final deflected position, a damping
torque is provided in the indicating instruments.
There are three types of damping:


Air friction damping:

air friction damping uses either aluminium piston or vane, which is attached to or
mounted on the moving system and moves in an air chamber at one end.
 Fluid friction damping:
In fluid friction damping, a light vane (attached to the moving system) is dipped
into a pot of damping oil. The fluid produces the necessary opposing (or
damping) force to the vane. The vane should be completely submereged in the
oil.
The disadvantage of this type of damping is that it can only be used in the vertical
position.


Eddy Current Damping:

Eddy-current damping uses a conducting material which moves in a magnetic
field so as to cut through the lines of force, thus setting up eddy currents. Force
always exists between the eddy current and magnetic field which is always
opposite to the direction of motion. This is most efficient type of damping and is
largely used in permanent magnet moving coil instruments.
TYPES OF INDICATING INSTRUMENTS
PMMC(permanent magnet moving coil)
 MI(moving iron)

PMMC


PRINCIPLE:

This type of instrument based on the principle that
when a current carrying conductor is placed in a
magnetic field, a force acts on the conductor,
which tends to move it to one side and out of the
field.
CONSTRUCTION:

It consists of a powerful U-shaped permanent
magnet made of Alnico, and soft iron pole pieces
bored out cylindrically.
A soft iron core is fixed between the magnetic
poles whose functions are i) to make the field
uniform, and ii) to decrease the reluctance of the
air path between the poles and hence increase
the magnetic flux.
Surrounding the core is a rectangular coil of many
turns wound on a light aluminium or copper
frame, supported by delicate bearings.
A light pointer fixed to the frame moves on the
calibrated scale according to the amount of
electricity passed through the coil.
The aluminiun frame provides not only support for
the coil but also a damping torque by the eddy
currents induce in it.
WORKING:
When the instrument is connected in the
circuit to measure current or voltage,
the operating current flows through the
coil. Since the coil is carrying current
and is placed in the magnetic field of
the permanent magnet , a mechanical
force acts on it. As a result , the pointer
attached to the moving system moves
in a clockwise direction over the
graduated scale to indicate the value of
current or voltage being measured.
If the current in the coil is reversed, the
deflecting torque also be reversed
since the direction of the permanent
magnet is same. Consequently, the
pointer will try to deflect below zero.
Deflection in this direction is prevented
by a spring “stop”. Since the deflecting
torque reverses with the reversal of
current in the coil, such instrument can
be used to measure direct current and
voltage only.


ADVANTAGE:





Low power consumption.
Uniform scale extendable over an arc of 270º or so.
High torque weight ratio.
No hysteresis loss.
Very effective and efficient eddy current damping.
Not effected much by stray and magnetic fields due to strong operating field.



DISADVANTAGE:





Costlier compared to moving iron instruments, due to delicate construction and
accurate machining and assembly of various parts.
Some error arise due to the ageing of control springs and the permanent magnet.
Use limited to d.c. only.
Scale length of meter can be increased from 120º and 240º or even 270º or 300º.



APPLICATION:



PMMC instruments can be used as dc ammeter. And its range can be increased
by using a large number of turns in parallel with the instrument.
The range of this instrument, when used as a dc voltmeter, can be increased by
using a high resistance in series with it.










MOVING IRON INSTRUMENTS
Moving Iron instruments depend for their action upon
the magnetic effect of current, and are widely used as
indicating instruments. In this type of instrument , the
coil is stationery and the deflection is caused by a
soft-iron piece moving in the field produced by the
coil.
There are two types of moving iron instruments:
i)
Attraction type
ii)
Repulsion type
Attraction type:
 Construction:

Fig shows the constructional
details of an attraction type
MI instrument. It consists
of a cylindrical coil or
solenoid which is kept
fixed. An oval-shaped softiron piece. The controlling
torque is provided by an
aluminium vane, attached
to the spindle, which
moves in a closed air
chamber.
Working:

When the instrument is connected in the circuit to measure
current or voltage, the operating current flowing through
the coil sets up a magnetic field. In other words, the coil
behaves like a magnet and therefore it attracts the soft iron
piece towards it. The result is that the pointer attached to
the moving system moves from zero position. The pointer
will come to rest at a position where deflecting torque is
equal to the controlling torque. If current in the coil is
reversed, the direction of magnetic field also reverses and
so does the magnetism produce in the soft iron piece.
Hence, the direction of the deflecting torque remains
unchanged. For this reason, such instruments can be used
for both d.c. and a.c. measurements.
Repulsion type
 Construction:
Fig shows the constructional details
of
repulsion
type
moving-iron
instrument. It consists of two soft-iron
pieces or vanes surrounded by a
fixed cylindrical hollow coil which
carries the operating current. One of
these vanes is fixed and the other is
free to move. The movable vane is
cylindrical shape and is mounted
axially on a spindle to which a pointer
is attached. The fixed vane, which is
wedge-shaped and has a larger
radius, is attached to the stationery
coil. The controlling torque is provided
by one spiral spring at the top of the
instrument. It may be noted that in
this instrument, springs do not
provide the electrical connections.
Damping is provided by air friction
due to the motion of a piston in an air
chamber.
 Working:
When current to be measured or current proportional to the
voltage to be measured flows through the coil, a magnetic field is
set up by the coil. This magnetic field magnetises the two vanes
in the same direction i.e. similar polarities are developed at the
same ends of the vanes. Since the adjacent edges of the vanes
are of the same polarity, the two vanes repel each other. As the
fixed vane cannot move, the movable vane deflects and causes
the pointer to move from zero position. The pointer will come to
rest at a position where deflecting torque is equal to controlling
torque provided by the spring. If the current in the coil is
reversed, the direction of deflection remains unchanged. It is
because reversal of the field of the coil reverses the
magnetisation of both iron vanes so that they repel each other
regardless which way current flows through the coil. For this
reason, such instruments can be used for both d.c. and a.c.
applications.
o

Advantages:

ii)

Cheap, robust and give reliable service.
Usable in both a.c. and d.c. circuits.

o

Disadvantages:

i)

Have non-linear scale.
Cannot be calibrate with high degree of precision for d.c. on
account of the affect of hysteresis in the iron vanes.
Deflection up to 240º only may be obtained with this instrument.
This instrument will always have to be put in the vertical position
if it uses gravity control.

i)

ii)

iii)
iv)

o

Errors with MI instruments:

i)

Due to hysteresis when used in a.c. and d.c.
Due to stray magnetic fields when used both in a.c. and d.c.
Due to frequency variation when used in a.c.
Due to waveforms effect when used in a.c.

ii)

iii)
iv)
Applications of MI instruments:
As an ammeter:
It may be constructed for full-scale deflection of 0.1 to 30A
with out the use of shunts or current transformers. To
obtain full-scale deflection with currents less than 0.1A, it
requires a coil with a large number of fine wire turns, which
results in an ammeter with a high impedance.
As an voltmeter:
The MI voltmeter is a fairly low impedance instrument,
typically, 50Ω/V for a 100V instrument. The lowest full
scale is of the order of 50V.The range of the instrument,
when used as a voltmeter, can be extended by using a
high non-inductive resistance R connected in series with it.
This series resistance is known as „multiplier‟.
ELECTRODYNAMOMETER
TYPE INSTRUMENT
It works on dynamometer principle i.e. mechanical
force exists between two current carrying
conductors or coils.
 Similar to PMMC
 Portable, highest precision.
 Transfer instruments.

CONSTRUCTION


Fixed Coils: The operating field is
produced by the fixed coil which is
divided into two sections to give a
uniform field near the centre. The coil is
wound with fine wire when used as a
voltmeter and heavy wire when used as
an ammeter and wattmeter. The wire is
stranded, when necessary, to reduce
eddy current losses in conductors.



Moving Coil: The moving coil is
wound either as a self sustaining coil or
else on a non-metallic former. The use
of metallic former is avoided in order to
avoid the inducement of eddy currents
in it. Both moving coils and fixed coils
are air cored.








Moving System: The moving coil is supported by an aluminium spindle
and jewel bearings and carries a pointer moving over a graduated scale.
The entire movement is very solid and rigidly constructed in order to keep
mechanical dimensions stable and its calibration in tact.
Control system: The controlling torque is provided by two control
springs, which also act as leads to the moving coil.
Damping System: Air friction damping is used in these instruments and
may be either piston type or vane type. Eddy current damping cannot be
used in these instruments as introduction of a permanent magnet for the
purpose would distort the working magnetic field of the instrument.
Shielding: The operating magnetic field produced by the fixed coils in
these instruments is somewhat weaker (0.005-0.006T) in comparison to
that in instruments of other types. So it is essential to provide magnetic
shielding to this arrangement.
The complete assembly is surrounded by a laminated steel shield to
protect the instrument from external magnetic field which may affect the
operation of the instrument.
WORKING PRINCIPLE
The operating principle of dynamometer type instruments is
the interaction between the currents in the moving coil,
mounted on a shaft, and the fixed coils.
 W hen two coils are energized, their magnetic fields interact
and the resulting torque tends to rotate the moving coil. Since
there is no iron, the field strength is proportional to the current
in the fixed coil and, therefore, the deflecting torque is
proportional to the product of the currents in the fixed coil and
the moving coil.
 When used as a wattmeter and, the fixed coil is the current
coil and the moving coil is the pressure coil. Thus the current
in the latter is proportional to the voltage applied. Hence, the
deflecting torque is proportional to the product of the voltage
and current(that is power).

TORQUE EQUATION
If if and im are the currents flowing through the fixed and
moving coils respectively then
Deflecting torque, Td ∞ if im
In spring-controlled instruments
Controlling Torque, Tc ∞ θ, the deflection
But in steady deflected position
Tc= Td
θ ∞ if im
or θ ∞ i2 (if=im=i)

ERRORS
Frictional Error: Due to Heavy moving parts.
 Temperature Error: Temperature errors due to internal
heating.
 Error due to stray magnetic fields
 Frequency Error:

MERITS
As the instrument has square-law response, so can be
used on both dc as well as on ac.
 These instruments are free from hysteresis and eddy
current errors because of absence of iron in the
operating parts of the instrument.
 Ammeters up to 10A and voltmeters up to 600V can be
constructed with precision grade accuracy.
 Dynamometer type voltmeters are very useful for
accurate measurement of rms values of voltages
irrespective of waveforms.
 Because of precision grade accuracy and same
calibration for dc and ac measurements these
instruments are used as transfer and calibration
instruments.

DEMERITS
The scale is not uniform as the instrument has
square-law response.
 The magnetic field strength obtained in these
instruments, being small due to the absence of iron,
a large number of ampere turns are required on the
moving coil in order to obtain the necessary
deflecting torque. As a result, the moving system
becomes heavy and power loss becomes high.
 Small torque-weight ratio.
 High frictional losses.
 Low sensitivity.
 Expensive.

APPLICATION





As an Ammeter: When it is used as an ammeter of low range (about
0.2A), the moving and fixed coils are connected in series.
When this type of instrument is used to measure a heavy amount of
current(about 20A), the moving coil is usually connected in series with a
swamping resistance, across a shunt, together with the fixed coils. Fixed
coils are made of thick wires to carry larger currents.

As a Voltmeter: To use the instrument as a voltmeter, the coils are
wound with high resistance wire and fixed and moving coils are connected
in series with a high resistance.

Contenu connexe

Tendances

Resistive transducers and Strain Gauge
Resistive transducers and Strain GaugeResistive transducers and Strain Gauge
Resistive transducers and Strain Gauge
karoline Enoch
 

Tendances (20)

Tachogenerator
TachogeneratorTachogenerator
Tachogenerator
 
Resistive transducers and Strain Gauge
Resistive transducers and Strain GaugeResistive transducers and Strain Gauge
Resistive transducers and Strain Gauge
 
Electrical measurements
Electrical measurementsElectrical measurements
Electrical measurements
 
Electrodynamometer
ElectrodynamometerElectrodynamometer
Electrodynamometer
 
Digital voltmeter (DVM) and its Classification
Digital voltmeter (DVM) and its ClassificationDigital voltmeter (DVM) and its Classification
Digital voltmeter (DVM) and its Classification
 
Optical Encoder
Optical EncoderOptical Encoder
Optical Encoder
 
Permanent magnet brushless dc motors ppt
Permanent magnet brushless dc motors pptPermanent magnet brushless dc motors ppt
Permanent magnet brushless dc motors ppt
 
Linear Variable Differential Transducer(LVDT)
Linear Variable Differential Transducer(LVDT)Linear Variable Differential Transducer(LVDT)
Linear Variable Differential Transducer(LVDT)
 
Lvdt
LvdtLvdt
Lvdt
 
Actuators.ppt
Actuators.pptActuators.ppt
Actuators.ppt
 
Ppt of servomotor
Ppt of servomotorPpt of servomotor
Ppt of servomotor
 
Lecture 2 Servomotors - Basics & Working
Lecture 2   Servomotors - Basics & WorkingLecture 2   Servomotors - Basics & Working
Lecture 2 Servomotors - Basics & Working
 
Major electrical equipment in power plants
Major electrical equipment in power plantsMajor electrical equipment in power plants
Major electrical equipment in power plants
 
Pmmc instrument
Pmmc instrumentPmmc instrument
Pmmc instrument
 
Permanent magnet synchronous motor
Permanent magnet synchronous motorPermanent magnet synchronous motor
Permanent magnet synchronous motor
 
Functional elements of an instrument
Functional elements of an instrumentFunctional elements of an instrument
Functional elements of an instrument
 
Stepper motor ppt
Stepper motor pptStepper motor ppt
Stepper motor ppt
 
Stepper motor Presentation
Stepper motor Presentation Stepper motor Presentation
Stepper motor Presentation
 
Dc motor
Dc motorDc motor
Dc motor
 
Frequency Meter : Working principle
Frequency Meter : Working principleFrequency Meter : Working principle
Frequency Meter : Working principle
 

Similaire à Indicating instruments

chapter 2-measuring instrument.komkjsdbsjh
chapter 2-measuring instrument.komkjsdbsjhchapter 2-measuring instrument.komkjsdbsjh
chapter 2-measuring instrument.komkjsdbsjh
Saravanan Pazapughal
 
SSC JE Measurements - wifistudy (1).pdf
SSC JE Measurements - wifistudy (1).pdfSSC JE Measurements - wifistudy (1).pdf
SSC JE Measurements - wifistudy (1).pdf
Vibhugoyal6
 
Measuring instruments
Measuring instrumentsMeasuring instruments
Measuring instruments
Sayyed Raza
 

Similaire à Indicating instruments (20)

chp-10.pdf
chp-10.pdfchp-10.pdf
chp-10.pdf
 
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
ELECTRICAL MEASUREMENT & MEASURING INSTRUMENTS [Emmi- (NEE-302) -unit-1]
 
ELECTRICAL AND ELECTRONICS MEASUREMENT
ELECTRICAL AND ELECTRONICS MEASUREMENT ELECTRICAL AND ELECTRONICS MEASUREMENT
ELECTRICAL AND ELECTRONICS MEASUREMENT
 
IARE_EEI_PPT
IARE_EEI_PPTIARE_EEI_PPT
IARE_EEI_PPT
 
IARE_EEI_PPT.pdf
IARE_EEI_PPT.pdfIARE_EEI_PPT.pdf
IARE_EEI_PPT.pdf
 
EEI_PPT.pdf
EEI_PPT.pdfEEI_PPT.pdf
EEI_PPT.pdf
 
Unit-III.pptx
Unit-III.pptxUnit-III.pptx
Unit-III.pptx
 
UNIT 2.pptx
UNIT 2.pptxUNIT 2.pptx
UNIT 2.pptx
 
chapter 2-measuring instrument.komkjsdbsjh
chapter 2-measuring instrument.komkjsdbsjhchapter 2-measuring instrument.komkjsdbsjh
chapter 2-measuring instrument.komkjsdbsjh
 
Measuring instrumentsppt2
Measuring instrumentsppt2Measuring instrumentsppt2
Measuring instrumentsppt2
 
SSC JE Measurements - wifistudy (1).pdf
SSC JE Measurements - wifistudy (1).pdfSSC JE Measurements - wifistudy (1).pdf
SSC JE Measurements - wifistudy (1).pdf
 
instruments and measurements
instruments and measurementsinstruments and measurements
instruments and measurements
 
class 18.08.2020.pptx
class 18.08.2020.pptxclass 18.08.2020.pptx
class 18.08.2020.pptx
 
Emi unit 1 ppt
Emi unit 1 pptEmi unit 1 ppt
Emi unit 1 ppt
 
BEEME UNIT III.ppt
BEEME UNIT III.pptBEEME UNIT III.ppt
BEEME UNIT III.ppt
 
Unit 2
Unit 2Unit 2
Unit 2
 
Measuring instruments
Measuring instrumentsMeasuring instruments
Measuring instruments
 
Measurements and Instrumentation
Measurements and InstrumentationMeasurements and Instrumentation
Measurements and Instrumentation
 
PMMC Instruments
PMMC InstrumentsPMMC Instruments
PMMC Instruments
 
1-170114040648.pdf
1-170114040648.pdf1-170114040648.pdf
1-170114040648.pdf
 

Dernier

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
QucHHunhnh
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
heathfieldcps1
 
Gardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch LetterGardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch Letter
MateoGardella
 
An Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdfAn Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdf
SanaAli374401
 

Dernier (20)

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Gardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch LetterGardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch Letter
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
An Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdfAn Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdf
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 

Indicating instruments

  • 2. MEASUREMENT The process of measuring is essentially that of comparing some unknown value with a value which is assumed to be known.  1) latter one in standard.  2) And measuring system is instrument. 
  • 3.  Characteristics of INSTRUMENTS  Calibration: All the static characteristics are obtained in one form or another by a process called Calibration. Calibration procedures involve a comparison of the particular instrument with either a) a primary standard. b) a secondary standard with a higher accuracy than the instruments be calibrated. c) an instrument of known accuracy.  Accuracy: Accuracy may be defined as the ability of a device or a system to respond to a true value of a measured variable under reference conditions.
  • 4.  Precision: Precision is defined by the degree of exactness for which an instrument is designed or intended to perform.  Repeatability: It is the closeness of agreement among a number of consecutive measurements of the output for the same value of the input, under the same operating conditions.  Reproducibility: It is the closeness of agreement among repeated measurements of the output for the same value of the input, made under the same operating conditions over a period of time.  Drift: It is an undesired change or a gradual variation in output over a period of time that is unrelated to changes in input and operating conditions.
  • 5.  Span: If in a measuring instrument the highest point of calibration is y units and the lowest point x units. Then the instrument range y units The instrument span is given by Span=(x-y)units  Sensitivity: Sensitivity can be defined as the ratio of a change in output to the change in input which causes it.  Resolution: The smallest increment in input (the quantity being measured) which can be detected with certainty by an instrument is its resolution.  Dead zone: Dead zone is the largest range of values of a measured variable to which the instrument does not respond.
  • 6. ERROR  Types of Error  Static Error : the numerical difference between the true value of a quantity and its value as obtained by measurement.  Mistakes: These are errors due to human mistakes such as careless reading, mistakes in observations, incorrect application of a correction, improper application of instruments and computational errors. Systematic Error:  1)Instrumental Error: Instrumental errors are the errors inherent in measuring instruments because of their mechanical structure, such as friction in bearings of various moving components, irregular spring tension. 
  • 7.  2)Environmental Error: it is due to conditions external to the measuring device including conditions in the area surrounding the instrument such as the effect of change in temp , humidity, barometric pressure , or magnetic or electrostatic fields.  Random Error: the cause of such error is unknown or not determinable in the ordinary process of making measurements. Such errors are normally small and follows the laws of chance.
  • 8. SOURCES OF ERROR Insufficient knowledge of process parameters and design conditions.  Poor design.  Poor maintenance.  Error caused by people who operate instrument equipment.  Certain design limitations. 
  • 9. TYPES OF INSTRUMENTS       Absolute Instruments: Absolute instruments are those which give the value of the electrical quantity to be measured, in terms of the constant of the instruments and their deflection only, e.g. tangent galvanometer. Secondary Instruments: Secondary instruments are those which have been precalibrated by comparison with an absolute instrument. The value of the electrical quantity to be measured in these instruments can be determined from the deflection of the instrument. Without calibration of such an instrument, the deflection is meaningless. Different types of secondary instruments: 1)Indicating 2)Recording 3)Integrating
  • 10.  1)Indicating: Indicating instruments are those which indicate the instantaneous value of the electrical quantity being measured, at the time at which it is being measured. Their indications are given by pointers moving over calibrated dials(scale), e.g. ammeters,voltmeters and wattmeters.  2)Recording: Recording instruments are those which give a continuous record of variations of the electrical quantity over a selected period of time. The moving system of the instrument carries an inked pen which rests tightly on a graph chart. E.g. recording voltmeters used in supply station.  3)Integrating: Integrating instruments are those which measure and register the total quantity of electricity (in ampere-hour) or the total amount of electrical energy(in watt-hours or kilowatt-hours) supplied to a circuit over a period of time, e.g. ampere-hour meters, energy meters.
  • 11. ESSENTIALS OF INDICATING INSTRUMENTS  Deflecting Torque(Td): It is the torque which deflects the pointer on a calibrated scale according to the electrical quantity passing through the instrument. The deflecting torque causes the moving system and hence the pointer attached to it to move from zero position to indicate on a graduated scale the value of electrical quantity being measured.  Controlling Torque(Tc): It is the torque which controls the movement of the pointer on a particular scale according to the quantity of electricity passing through it. If deflecting torque were acting alone, the pointer would continue to move indefinitely and would swing over to the maximum deflected position irrespective of the magnitude of current (or voltage or power) to be measured.
  • 12. 1) Spring Control: In the spring control method, a hair-spring, usually of phosphorbronze, attached to the moving system is used. With the deflection of the pointer, the spring is twisted in the opposite direction. This twist in the spring produces a restoring torque which is directly proportional to the angle of deflection of the moving system. The pointer comes to a position of rest (or equilibrium) when the deflecting torque (Td) and controlling torque (Tc) are equal. Tc ∞ θ To give a controlling torque which is directly proportional to the angle of deflection of the moving system, the number of turns of the spring should be fairly large so that the deformation per unit length is small. The stress in the spring must be limited to such a value that there is no permanent set. Springs are made of materials which are  Non magnetic  Not subject to much fatigue  Low in specific resistance  Have low temperature coefficient of resistance.
  • 13. 2) Gravity Control: Gravity control is obtained by attaching a small weight to the moving system in such a way that it produces a restoring or controlling torque when the system is deflected. Tc ∞ Sinθ Thus, controlling torque in a gravity control system is proportional to the sine of the angle of deflection. The degree of control is adjusted by screwing the weight up or down on the carrying system. Advantages: The advantages of gravity control system, as compared to spring control, are given below:  It is cheap  It is unaffected by temperature  It is not subjected to fatigue or distortion, with time. Disadvantages: The disadvantages of gravity control system, as compared to spring control, are given below:  It gives a cramped scale.  The instrument has to be kept vertical.
  • 14.  Damping Torque: If the moving system is acted upon by deflecting and controlling torques alone, then pointer, due to inertia, will oscillate about its final deflected position for quite sometime before coming to rest. This is often undesirable because it makes difficult to obtain quick and accurate readings. In order to avoid these oscillations of the pointer and to bring it quickly to its final deflected position, a damping torque is provided in the indicating instruments. There are three types of damping:  Air friction damping: air friction damping uses either aluminium piston or vane, which is attached to or mounted on the moving system and moves in an air chamber at one end.  Fluid friction damping: In fluid friction damping, a light vane (attached to the moving system) is dipped into a pot of damping oil. The fluid produces the necessary opposing (or damping) force to the vane. The vane should be completely submereged in the oil. The disadvantage of this type of damping is that it can only be used in the vertical position.  Eddy Current Damping: Eddy-current damping uses a conducting material which moves in a magnetic field so as to cut through the lines of force, thus setting up eddy currents. Force always exists between the eddy current and magnetic field which is always opposite to the direction of motion. This is most efficient type of damping and is largely used in permanent magnet moving coil instruments.
  • 15. TYPES OF INDICATING INSTRUMENTS PMMC(permanent magnet moving coil)  MI(moving iron) 
  • 16. PMMC  PRINCIPLE: This type of instrument based on the principle that when a current carrying conductor is placed in a magnetic field, a force acts on the conductor, which tends to move it to one side and out of the field. CONSTRUCTION: It consists of a powerful U-shaped permanent magnet made of Alnico, and soft iron pole pieces bored out cylindrically. A soft iron core is fixed between the magnetic poles whose functions are i) to make the field uniform, and ii) to decrease the reluctance of the air path between the poles and hence increase the magnetic flux. Surrounding the core is a rectangular coil of many turns wound on a light aluminium or copper frame, supported by delicate bearings. A light pointer fixed to the frame moves on the calibrated scale according to the amount of electricity passed through the coil. The aluminiun frame provides not only support for the coil but also a damping torque by the eddy currents induce in it.
  • 17. WORKING: When the instrument is connected in the circuit to measure current or voltage, the operating current flows through the coil. Since the coil is carrying current and is placed in the magnetic field of the permanent magnet , a mechanical force acts on it. As a result , the pointer attached to the moving system moves in a clockwise direction over the graduated scale to indicate the value of current or voltage being measured. If the current in the coil is reversed, the deflecting torque also be reversed since the direction of the permanent magnet is same. Consequently, the pointer will try to deflect below zero. Deflection in this direction is prevented by a spring “stop”. Since the deflecting torque reverses with the reversal of current in the coil, such instrument can be used to measure direct current and voltage only.
  • 18.  ADVANTAGE:   Low power consumption. Uniform scale extendable over an arc of 270º or so. High torque weight ratio. No hysteresis loss. Very effective and efficient eddy current damping. Not effected much by stray and magnetic fields due to strong operating field.  DISADVANTAGE:   Costlier compared to moving iron instruments, due to delicate construction and accurate machining and assembly of various parts. Some error arise due to the ageing of control springs and the permanent magnet. Use limited to d.c. only. Scale length of meter can be increased from 120º and 240º or even 270º or 300º.  APPLICATION:  PMMC instruments can be used as dc ammeter. And its range can be increased by using a large number of turns in parallel with the instrument. The range of this instrument, when used as a dc voltmeter, can be increased by using a high resistance in series with it.       
  • 19. MOVING IRON INSTRUMENTS Moving Iron instruments depend for their action upon the magnetic effect of current, and are widely used as indicating instruments. In this type of instrument , the coil is stationery and the deflection is caused by a soft-iron piece moving in the field produced by the coil. There are two types of moving iron instruments: i) Attraction type ii) Repulsion type
  • 20. Attraction type:  Construction: Fig shows the constructional details of an attraction type MI instrument. It consists of a cylindrical coil or solenoid which is kept fixed. An oval-shaped softiron piece. The controlling torque is provided by an aluminium vane, attached to the spindle, which moves in a closed air chamber.
  • 21. Working: When the instrument is connected in the circuit to measure current or voltage, the operating current flowing through the coil sets up a magnetic field. In other words, the coil behaves like a magnet and therefore it attracts the soft iron piece towards it. The result is that the pointer attached to the moving system moves from zero position. The pointer will come to rest at a position where deflecting torque is equal to the controlling torque. If current in the coil is reversed, the direction of magnetic field also reverses and so does the magnetism produce in the soft iron piece. Hence, the direction of the deflecting torque remains unchanged. For this reason, such instruments can be used for both d.c. and a.c. measurements.
  • 22. Repulsion type  Construction: Fig shows the constructional details of repulsion type moving-iron instrument. It consists of two soft-iron pieces or vanes surrounded by a fixed cylindrical hollow coil which carries the operating current. One of these vanes is fixed and the other is free to move. The movable vane is cylindrical shape and is mounted axially on a spindle to which a pointer is attached. The fixed vane, which is wedge-shaped and has a larger radius, is attached to the stationery coil. The controlling torque is provided by one spiral spring at the top of the instrument. It may be noted that in this instrument, springs do not provide the electrical connections. Damping is provided by air friction due to the motion of a piston in an air chamber.
  • 23.  Working: When current to be measured or current proportional to the voltage to be measured flows through the coil, a magnetic field is set up by the coil. This magnetic field magnetises the two vanes in the same direction i.e. similar polarities are developed at the same ends of the vanes. Since the adjacent edges of the vanes are of the same polarity, the two vanes repel each other. As the fixed vane cannot move, the movable vane deflects and causes the pointer to move from zero position. The pointer will come to rest at a position where deflecting torque is equal to controlling torque provided by the spring. If the current in the coil is reversed, the direction of deflection remains unchanged. It is because reversal of the field of the coil reverses the magnetisation of both iron vanes so that they repel each other regardless which way current flows through the coil. For this reason, such instruments can be used for both d.c. and a.c. applications.
  • 24. o Advantages: ii) Cheap, robust and give reliable service. Usable in both a.c. and d.c. circuits. o Disadvantages: i) Have non-linear scale. Cannot be calibrate with high degree of precision for d.c. on account of the affect of hysteresis in the iron vanes. Deflection up to 240º only may be obtained with this instrument. This instrument will always have to be put in the vertical position if it uses gravity control. i) ii) iii) iv) o Errors with MI instruments: i) Due to hysteresis when used in a.c. and d.c. Due to stray magnetic fields when used both in a.c. and d.c. Due to frequency variation when used in a.c. Due to waveforms effect when used in a.c. ii) iii) iv)
  • 25. Applications of MI instruments: As an ammeter: It may be constructed for full-scale deflection of 0.1 to 30A with out the use of shunts or current transformers. To obtain full-scale deflection with currents less than 0.1A, it requires a coil with a large number of fine wire turns, which results in an ammeter with a high impedance. As an voltmeter: The MI voltmeter is a fairly low impedance instrument, typically, 50Ω/V for a 100V instrument. The lowest full scale is of the order of 50V.The range of the instrument, when used as a voltmeter, can be extended by using a high non-inductive resistance R connected in series with it. This series resistance is known as „multiplier‟.
  • 26. ELECTRODYNAMOMETER TYPE INSTRUMENT It works on dynamometer principle i.e. mechanical force exists between two current carrying conductors or coils.  Similar to PMMC  Portable, highest precision.  Transfer instruments. 
  • 27. CONSTRUCTION  Fixed Coils: The operating field is produced by the fixed coil which is divided into two sections to give a uniform field near the centre. The coil is wound with fine wire when used as a voltmeter and heavy wire when used as an ammeter and wattmeter. The wire is stranded, when necessary, to reduce eddy current losses in conductors.  Moving Coil: The moving coil is wound either as a self sustaining coil or else on a non-metallic former. The use of metallic former is avoided in order to avoid the inducement of eddy currents in it. Both moving coils and fixed coils are air cored.
  • 28.     Moving System: The moving coil is supported by an aluminium spindle and jewel bearings and carries a pointer moving over a graduated scale. The entire movement is very solid and rigidly constructed in order to keep mechanical dimensions stable and its calibration in tact. Control system: The controlling torque is provided by two control springs, which also act as leads to the moving coil. Damping System: Air friction damping is used in these instruments and may be either piston type or vane type. Eddy current damping cannot be used in these instruments as introduction of a permanent magnet for the purpose would distort the working magnetic field of the instrument. Shielding: The operating magnetic field produced by the fixed coils in these instruments is somewhat weaker (0.005-0.006T) in comparison to that in instruments of other types. So it is essential to provide magnetic shielding to this arrangement. The complete assembly is surrounded by a laminated steel shield to protect the instrument from external magnetic field which may affect the operation of the instrument.
  • 29. WORKING PRINCIPLE The operating principle of dynamometer type instruments is the interaction between the currents in the moving coil, mounted on a shaft, and the fixed coils.  W hen two coils are energized, their magnetic fields interact and the resulting torque tends to rotate the moving coil. Since there is no iron, the field strength is proportional to the current in the fixed coil and, therefore, the deflecting torque is proportional to the product of the currents in the fixed coil and the moving coil.  When used as a wattmeter and, the fixed coil is the current coil and the moving coil is the pressure coil. Thus the current in the latter is proportional to the voltage applied. Hence, the deflecting torque is proportional to the product of the voltage and current(that is power). 
  • 30.
  • 31. TORQUE EQUATION If if and im are the currents flowing through the fixed and moving coils respectively then Deflecting torque, Td ∞ if im In spring-controlled instruments Controlling Torque, Tc ∞ θ, the deflection But in steady deflected position Tc= Td θ ∞ if im or θ ∞ i2 (if=im=i) 
  • 32. ERRORS Frictional Error: Due to Heavy moving parts.  Temperature Error: Temperature errors due to internal heating.  Error due to stray magnetic fields  Frequency Error: 
  • 33. MERITS As the instrument has square-law response, so can be used on both dc as well as on ac.  These instruments are free from hysteresis and eddy current errors because of absence of iron in the operating parts of the instrument.  Ammeters up to 10A and voltmeters up to 600V can be constructed with precision grade accuracy.  Dynamometer type voltmeters are very useful for accurate measurement of rms values of voltages irrespective of waveforms.  Because of precision grade accuracy and same calibration for dc and ac measurements these instruments are used as transfer and calibration instruments. 
  • 34. DEMERITS The scale is not uniform as the instrument has square-law response.  The magnetic field strength obtained in these instruments, being small due to the absence of iron, a large number of ampere turns are required on the moving coil in order to obtain the necessary deflecting torque. As a result, the moving system becomes heavy and power loss becomes high.  Small torque-weight ratio.  High frictional losses.  Low sensitivity.  Expensive. 
  • 35.
  • 36. APPLICATION    As an Ammeter: When it is used as an ammeter of low range (about 0.2A), the moving and fixed coils are connected in series. When this type of instrument is used to measure a heavy amount of current(about 20A), the moving coil is usually connected in series with a swamping resistance, across a shunt, together with the fixed coils. Fixed coils are made of thick wires to carry larger currents. As a Voltmeter: To use the instrument as a voltmeter, the coils are wound with high resistance wire and fixed and moving coils are connected in series with a high resistance.