DEFINITION
Defibrillation involves delivering a
high intensity electrical charge inorder
to depolarize the entire myocardium at
one time so that the fastest normal
pacemaker can regain control of the
pacing function of the heart.
INDICATION
•Ventricular fibrillation
•Asystole
DIFFERENCES BETWEEN MONOPHASIC AND BIPHASIC
SYSTEMS
•In monophasic systems, the current travels only in one direction -
from one paddle to the other.
•In biphasic systems, the current travels towards the positive
paddle and then reverse and go back; this occurs several times.
•Biphasic shocks deliver one cycle every 10 milliseconds.
•They are associated with fewer burns and less myocardial
damage.
•With monophasic shocks, the rate of first shock success in
cardiac arrests due to a shockable rhythm is only 60%, whereas
with biphasic shocks, this increases to 90%.
However, this efficacy of biphasic defibrillators over monophasic
defibrillators has not been consistently reported.
TYPES OF
DEFIBRILLATORS
Automated external defibrillators (AEDs):
•These are useful, as their use does not require
special medical training.
•They are found in public places - e.g., offices,
airports, train stations, shopping centers.
•They analyze the heart rhythm and then charge
and deliver a shock if appropriate.
•However, they cannot be overridden manually
and can take 10-20 seconds to determine
arrhythmias.
Semi-automated AEDs:
•These are similar to AEDs but can be overridden
and usually have an ECG display.
•They tend to be used by paramedics.
•They also have the ability to pace.
Standard defibrillators with monitor- may be
monophasic or biphasic.
Transvenousor implanted.
•An implantable cardioverter-defibrillator (ICD) is a device
combining a cardioverter and a defibrillator into one
implantable unit.
•It is thus a small battery-powered electrical impulse
generator that is implanted in patients who are at risk of
sudden cardiac death due to ventricular fibrillation and
ventricular tachycardia.
•The device is programmed to detect cardiac arrhythmia and
correct it by delivering a brief electrical impulse to the heart.
•In current variants, the ability to revert ventricular fibrillation
has been extended to include both atrial and ventricular
arrhythmias.
•There also exists the ability to perform biventricular pacing
in patients with congestive heart failure or bradycardia.
•The process of implantation of an ICD is similar to
implantation of a pacemaker.
PADDLES VERSUS ADHESIVE PATCHES
•Paddles were originally used but their use is being
superseded by adhesive patches.
•Adhesive patches are placed most commonly anterio-
apically - the anterior patch goes under the right clavicle
and the apical patch is placed at the apex.
•Adhesive electrodes are better, as they stick to the
chest wall, so there is no mess with gels.
•Paddles require at least 25 lbs of pressure, which is not
needed with adhesive electrodes.
•Adhesive electrodes also allow good ECG trace without
interference.
•They are also safer, as no operator required - although,
before discharging a shock, it is important to ensure
everyone is clear of the patient.
ENERGY LEVELS FOR DEFIBRILLATION
Monophasic - the cardiopulmonary resuscitation (CPR)
algorithm recommends single shocks started at and
repeated at 360 J.
Biphasic - the CPR algorithm recommends shocks
initially of 150-200 J and subsequent shocks of 150-360
J.
The Biphasic Trial in 2007 compared lower fixed (150,
150, 150 J) and gradually increasing energy (200, 300,
360 J) shocks for out-of-hospital cardiac arrests.
Escalating energy shocks were associated with more
frequent conversion and termination of VF as opposed
to low-level fixed shocks. This applied to patients who
remained in VF after the first shock.
PARTS OF DEFIBRILLATOR
•Machine
•Paddles- adult &pediatric
•ECG printout
•Alarm knobe for selecting energy
•Synchronizer
PROCEDURE
•It is an emergency procedure. The main principle is to
ACT QUICKLY. The steps of procedure involves
•Bring defibrillator to bedside
•Confirm diagnosis by palpating carotid pulse, or
analyzing the ECG
•Turn on the defibrillator
•Squeeze generous amount of jelly on to defibrillator
paddles, coat entire surface of the paddles with jelly by
rubbing paddles together, remove the excess jelly if it
spreads to the sides.
PROCEDURE
•Select the correct electrical charge on the defibrillator.
The average charge for an adult is 200 to 300 joules or
watt per second.
•Press the charge button to charge the capacitor. When
the intensity meter displays the required energy level.
Place the paddles on the patient’s chest. Put one paddle
to the right of the sternum between the second and third
intercostal space and the other at the fifth intercostal
space on the left side of the chest near the apex of the
heart.
•Make sure the paddles touch flat against the patient’s
body. If not it may cause burns in the patient
•Before delivering shock, tell everyone to stand clear off
the patient and his bed. If the patient is getting oxygen, it
should be turned off.
PROCEDURE
•Operator should not stand on wet floornor leaning
against the bed of the patient.
•Discharge the energy by pressing the discharge buttons
on the paddles simultaneously.
•Examine the ECG strip to see whether defibrillation has
altered arrhythmia and restored normal rhythm.
•Ask co-worker to start CPR. If desired rhythm not
restored, defibrillation is repeated with a higher energy
level.
•If the defibrillation is unnecessary, discharge the
machine by turning it off.Once the machine is
discharged, clean the paddles with soap and water,
making sure to remove all conductive jelly. Any jelly that
remains on the paddles will corrode the metal paddles.
CAUSES OF FAILURE OF DEFIBRILLATION
•Poor patient condition
Valvular diseases
Massive myocardial infarction
Cardiomegaly
Myocardial rupture and cardiac tamponade
Pulmonary embolism or infarction
Respiratory disease or trauma
•Prolonged cardiac arrest
•Inadequate cardio-respiratory resuscitative measures
Hypoxia in the patient
Acidosis
Alkalosis
Electrolyte imbalance
•Drug toxicity
•Inadequate sympathetic tone
•Inadequate electrical current delivered to the heart
Defective equipment
Poor paddle position
Excessive conductive jelly between paddles
Barrel shaped chest of the victim
•Inexperienced operator
COMPLICATIONS
•Damage to myocardium due to repeated
high energy electrical shocks
•Chest burns due to repeated high-energy
discharges and poor contract between the
paddles and the skin
•Electrocution of the by-standers
•Formation of short circuits between
paddles due to excessive amount of
conduction jelly applied on the paddles. This
causes loss of electrical energy.
NURSING IMPLICATIONS
•As defibrillation is an emergency procedure, the
equipment should be ready at all times.
•It should be kept functioning all the time, and
should be checked before each shift.
•Each staff should be aware of its functioning
•The defibrillator should be tested daily for its
proper functioning. To test the defibrillator, follow
the steps as described below
•Set the defibrillator at 300 joules
•Depress the charge button on the defibrillator
until the display number matches the joules
setting
•Leave the paddles in their resting place on the
defibrillator and simultaneously press the
discharge buttons on both paddles
•When someone is getting ready for defibrillation,
the co-worker should begin with basic life
support.
•Patient should be continuously monitored after
defibrillation. He should never be left alone for the
first 24hr. vital signs are to be recorded very
frequently until they are stabilized. Oxygen should
be continuously given to the patient. Drugs
should be administered as ordered. Intake and
output chart to be maintained.
THANK YOU

Contenu connexe

Tendances (20)

Infusion Pump
Infusion PumpInfusion Pump
Infusion Pump
 
Endotracheal intubation
Endotracheal intubationEndotracheal intubation
Endotracheal intubation
 
Cardiac monitor ppt
Cardiac monitor pptCardiac monitor ppt
Cardiac monitor ppt
 
Care of patient on ventilator
Care of patient on ventilatorCare of patient on ventilator
Care of patient on ventilator
 
Hemodynamic monitoring ppt
Hemodynamic monitoring pptHemodynamic monitoring ppt
Hemodynamic monitoring ppt
 
Bag and mask Ventilation
Bag and mask VentilationBag and mask Ventilation
Bag and mask Ventilation
 
Cvp
CvpCvp
Cvp
 
Et tube suctioning ppt
Et tube suctioning pptEt tube suctioning ppt
Et tube suctioning ppt
 
Intubation ppt
Intubation pptIntubation ppt
Intubation ppt
 
Syringe Pump
Syringe PumpSyringe Pump
Syringe Pump
 
Cardiac monitoring
Cardiac monitoringCardiac monitoring
Cardiac monitoring
 
Intercostal drainage tube insertion
Intercostal drainage tube insertionIntercostal drainage tube insertion
Intercostal drainage tube insertion
 
Defibrillator power point presentation for medical students
Defibrillator power point presentation for medical studentsDefibrillator power point presentation for medical students
Defibrillator power point presentation for medical students
 
Endotracheal tubes
Endotracheal tubesEndotracheal tubes
Endotracheal tubes
 
Intravenous Cannulation
Intravenous CannulationIntravenous Cannulation
Intravenous Cannulation
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 
Defibrillator
DefibrillatorDefibrillator
Defibrillator
 
INFUSION PUMPS
INFUSION PUMPSINFUSION PUMPS
INFUSION PUMPS
 
cvp monitoring
cvp monitoringcvp monitoring
cvp monitoring
 
Cvp line
Cvp lineCvp line
Cvp line
 

En vedette

En vedette (8)

Defibrillator (ppt)
Defibrillator (ppt)Defibrillator (ppt)
Defibrillator (ppt)
 
2. defibrillators
2. defibrillators2. defibrillators
2. defibrillators
 
Biomedical Engineering
Biomedical EngineeringBiomedical Engineering
Biomedical Engineering
 
Biomedical engineering
Biomedical engineeringBiomedical engineering
Biomedical engineering
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Defibrillator
Defibrillator Defibrillator
Defibrillator
 
Defibrillation presentation
Defibrillation presentationDefibrillation presentation
Defibrillation presentation
 
Biomedical engineering (BME)
Biomedical engineering (BME)Biomedical engineering (BME)
Biomedical engineering (BME)
 

Similaire à Defibrillation

Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentationlinsstalex
 
Defibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJDefibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJDharmendra Joshi
 
Electrical therapy in bls 2018
Electrical therapy  in bls 2018Electrical therapy  in bls 2018
Electrical therapy in bls 2018IntanBaiduriBadri
 
Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01tarek klai
 
Defibrillator & cardioversion
Defibrillator & cardioversion Defibrillator & cardioversion
Defibrillator & cardioversion RAJESH EAPEN
 
Defibrillation & Cardioversion
Defibrillation & CardioversionDefibrillation & Cardioversion
Defibrillation & CardioversionHIRANGER
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices Mohamed Elwakil
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.pptDHANAPALR7
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.pptDHANAPALR7
 
Advanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptxAdvanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptxRebilHeiru2
 

Similaire à Defibrillation (20)

Acls himanshu
Acls   himanshuAcls   himanshu
Acls himanshu
 
DEFIBRILLATION.pptx
DEFIBRILLATION.pptxDEFIBRILLATION.pptx
DEFIBRILLATION.pptx
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Defibrillation and cardioversion
Defibrillation and cardioversionDefibrillation and cardioversion
Defibrillation and cardioversion
 
Defibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJDefibrillation & cardioversion by DJ
Defibrillation & cardioversion by DJ
 
DEFIBRILLATOR
DEFIBRILLATORDEFIBRILLATOR
DEFIBRILLATOR
 
Electrical therapy in bls 2018
Electrical therapy  in bls 2018Electrical therapy  in bls 2018
Electrical therapy in bls 2018
 
Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01
 
defibrillator.pptx
defibrillator.pptxdefibrillator.pptx
defibrillator.pptx
 
Defibrillator & cardioversion
Defibrillator & cardioversion Defibrillator & cardioversion
Defibrillator & cardioversion
 
Defibrillation & Cardioversion
Defibrillation & CardioversionDefibrillation & Cardioversion
Defibrillation & Cardioversion
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Advanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptxAdvanced Cardiovascular Life Support (ACLS).pptx
Advanced Cardiovascular Life Support (ACLS).pptx
 
Dc shock
Dc shock  Dc shock
Dc shock
 
update in electrical burn management.pptx
update in electrical burn management.pptxupdate in electrical burn management.pptx
update in electrical burn management.pptx
 

Plus de resmigs

Pulmonary embolism ppt
Pulmonary embolism pptPulmonary embolism ppt
Pulmonary embolism pptresmigs
 
Nursing care of ventilated patient
Nursing care of ventilated patient Nursing care of ventilated patient
Nursing care of ventilated patient resmigs
 
Five year plan new (2)
Five year plan new (2)Five year plan new (2)
Five year plan new (2)resmigs
 
Conflict management
Conflict managementConflict management
Conflict managementresmigs
 
CARDIO-PULMONARY RESUSCITATION
CARDIO-PULMONARY RESUSCITATIONCARDIO-PULMONARY RESUSCITATION
CARDIO-PULMONARY RESUSCITATIONresmigs
 
Communication and interpersonal relationships ppt
Communication and interpersonal relationships ppt Communication and interpersonal relationships ppt
Communication and interpersonal relationships ppt resmigs
 

Plus de resmigs (6)

Pulmonary embolism ppt
Pulmonary embolism pptPulmonary embolism ppt
Pulmonary embolism ppt
 
Nursing care of ventilated patient
Nursing care of ventilated patient Nursing care of ventilated patient
Nursing care of ventilated patient
 
Five year plan new (2)
Five year plan new (2)Five year plan new (2)
Five year plan new (2)
 
Conflict management
Conflict managementConflict management
Conflict management
 
CARDIO-PULMONARY RESUSCITATION
CARDIO-PULMONARY RESUSCITATIONCARDIO-PULMONARY RESUSCITATION
CARDIO-PULMONARY RESUSCITATION
 
Communication and interpersonal relationships ppt
Communication and interpersonal relationships ppt Communication and interpersonal relationships ppt
Communication and interpersonal relationships ppt
 

Dernier

💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...India Call Girls
 
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*Mumbai Call girl
 
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...Rashmi Entertainment
 
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...Sheetaleventcompany
 
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhCall Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhSheetaleventcompany
 
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Sheetaleventcompany
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...dilpreetentertainmen
 
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...Sheetaleventcompany
 
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...Sheetaleventcompany
 
2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in Rheumatology2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in RheumatologySidney Erwin Manahan
 
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...India Call Girls
 
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...Sheetaleventcompany
 
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...Sheetaleventcompany
 
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...Sheetaleventcompany
 
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...Sheetaleventcompany
 
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...Sheetaleventcompany
 
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...Sheetaleventcompany
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...daljeetkaur2026
 
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Sheetaleventcompany
 

Dernier (20)

💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Bhopal 🧿 9332606886 🧿 High Class Call Gir...
 
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*
Ulhasnagar Call girl escort *88638//40496* Call me monika call girls 24*
 
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9815457724☎️ Call Girl service in Chandigarh☎️ C...
 
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...
❤️Call Girl In Chandigarh☎️9814379184☎️ Call Girl service in Chandigarh☎️ Cha...
 
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhCall Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
 
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
 
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...
Call Girls Service 11 Phase Mohali {7435815124} ❤️ MONA Call Girl in Mohali P...
 
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...
Erotic Call Girls Bangalore {7304373326} ❤️VVIP SIYA Call Girls in Bangalore ...
 
2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in Rheumatology2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in Rheumatology
 
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
 
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...
❤️Chandigarh Escort Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ C...
 
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...
❤️Amritsar Escort Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amrit...
 
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...
Low Rate Call Girls Nagpur {9xx000xx09} ❤️VVIP NISHA Call Girls in Nagpur Mah...
 
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9142599079} ❤️VVIP NISHA Call Girls in Pune Maharas...
 
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...
❤️Zirakpur Escorts☎️7837612180☎️ Call Girl service in Zirakpur☎️ Zirakpur Cal...
 
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...
Low Rate Call Girls Udaipur {9xx000xx09} ❤️VVIP NISHA CCall Girls in Udaipur ...
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
 
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
 

Defibrillation

  • 1.
  • 2. DEFINITION Defibrillation involves delivering a high intensity electrical charge inorder to depolarize the entire myocardium at one time so that the fastest normal pacemaker can regain control of the pacing function of the heart.
  • 4. DIFFERENCES BETWEEN MONOPHASIC AND BIPHASIC SYSTEMS •In monophasic systems, the current travels only in one direction - from one paddle to the other. •In biphasic systems, the current travels towards the positive paddle and then reverse and go back; this occurs several times. •Biphasic shocks deliver one cycle every 10 milliseconds. •They are associated with fewer burns and less myocardial damage. •With monophasic shocks, the rate of first shock success in cardiac arrests due to a shockable rhythm is only 60%, whereas with biphasic shocks, this increases to 90%. However, this efficacy of biphasic defibrillators over monophasic defibrillators has not been consistently reported.
  • 6. Automated external defibrillators (AEDs): •These are useful, as their use does not require special medical training. •They are found in public places - e.g., offices, airports, train stations, shopping centers. •They analyze the heart rhythm and then charge and deliver a shock if appropriate. •However, they cannot be overridden manually and can take 10-20 seconds to determine arrhythmias.
  • 7. Semi-automated AEDs: •These are similar to AEDs but can be overridden and usually have an ECG display. •They tend to be used by paramedics. •They also have the ability to pace.
  • 8. Standard defibrillators with monitor- may be monophasic or biphasic.
  • 9. Transvenousor implanted. •An implantable cardioverter-defibrillator (ICD) is a device combining a cardioverter and a defibrillator into one implantable unit. •It is thus a small battery-powered electrical impulse generator that is implanted in patients who are at risk of sudden cardiac death due to ventricular fibrillation and ventricular tachycardia. •The device is programmed to detect cardiac arrhythmia and correct it by delivering a brief electrical impulse to the heart. •In current variants, the ability to revert ventricular fibrillation has been extended to include both atrial and ventricular arrhythmias. •There also exists the ability to perform biventricular pacing in patients with congestive heart failure or bradycardia. •The process of implantation of an ICD is similar to implantation of a pacemaker.
  • 10. PADDLES VERSUS ADHESIVE PATCHES •Paddles were originally used but their use is being superseded by adhesive patches. •Adhesive patches are placed most commonly anterio- apically - the anterior patch goes under the right clavicle and the apical patch is placed at the apex. •Adhesive electrodes are better, as they stick to the chest wall, so there is no mess with gels. •Paddles require at least 25 lbs of pressure, which is not needed with adhesive electrodes. •Adhesive electrodes also allow good ECG trace without interference. •They are also safer, as no operator required - although, before discharging a shock, it is important to ensure everyone is clear of the patient.
  • 11. ENERGY LEVELS FOR DEFIBRILLATION Monophasic - the cardiopulmonary resuscitation (CPR) algorithm recommends single shocks started at and repeated at 360 J. Biphasic - the CPR algorithm recommends shocks initially of 150-200 J and subsequent shocks of 150-360 J. The Biphasic Trial in 2007 compared lower fixed (150, 150, 150 J) and gradually increasing energy (200, 300, 360 J) shocks for out-of-hospital cardiac arrests. Escalating energy shocks were associated with more frequent conversion and termination of VF as opposed to low-level fixed shocks. This applied to patients who remained in VF after the first shock.
  • 12. PARTS OF DEFIBRILLATOR •Machine •Paddles- adult &pediatric •ECG printout •Alarm knobe for selecting energy •Synchronizer
  • 13. PROCEDURE •It is an emergency procedure. The main principle is to ACT QUICKLY. The steps of procedure involves •Bring defibrillator to bedside •Confirm diagnosis by palpating carotid pulse, or analyzing the ECG •Turn on the defibrillator •Squeeze generous amount of jelly on to defibrillator paddles, coat entire surface of the paddles with jelly by rubbing paddles together, remove the excess jelly if it spreads to the sides.
  • 14. PROCEDURE •Select the correct electrical charge on the defibrillator. The average charge for an adult is 200 to 300 joules or watt per second. •Press the charge button to charge the capacitor. When the intensity meter displays the required energy level. Place the paddles on the patient’s chest. Put one paddle to the right of the sternum between the second and third intercostal space and the other at the fifth intercostal space on the left side of the chest near the apex of the heart. •Make sure the paddles touch flat against the patient’s body. If not it may cause burns in the patient •Before delivering shock, tell everyone to stand clear off the patient and his bed. If the patient is getting oxygen, it should be turned off.
  • 15. PROCEDURE •Operator should not stand on wet floornor leaning against the bed of the patient. •Discharge the energy by pressing the discharge buttons on the paddles simultaneously. •Examine the ECG strip to see whether defibrillation has altered arrhythmia and restored normal rhythm. •Ask co-worker to start CPR. If desired rhythm not restored, defibrillation is repeated with a higher energy level. •If the defibrillation is unnecessary, discharge the machine by turning it off.Once the machine is discharged, clean the paddles with soap and water, making sure to remove all conductive jelly. Any jelly that remains on the paddles will corrode the metal paddles.
  • 16. CAUSES OF FAILURE OF DEFIBRILLATION •Poor patient condition Valvular diseases Massive myocardial infarction Cardiomegaly Myocardial rupture and cardiac tamponade Pulmonary embolism or infarction Respiratory disease or trauma •Prolonged cardiac arrest
  • 17. •Inadequate cardio-respiratory resuscitative measures Hypoxia in the patient Acidosis Alkalosis Electrolyte imbalance •Drug toxicity •Inadequate sympathetic tone •Inadequate electrical current delivered to the heart Defective equipment Poor paddle position Excessive conductive jelly between paddles Barrel shaped chest of the victim •Inexperienced operator
  • 18. COMPLICATIONS •Damage to myocardium due to repeated high energy electrical shocks •Chest burns due to repeated high-energy discharges and poor contract between the paddles and the skin •Electrocution of the by-standers •Formation of short circuits between paddles due to excessive amount of conduction jelly applied on the paddles. This causes loss of electrical energy.
  • 19. NURSING IMPLICATIONS •As defibrillation is an emergency procedure, the equipment should be ready at all times. •It should be kept functioning all the time, and should be checked before each shift. •Each staff should be aware of its functioning •The defibrillator should be tested daily for its proper functioning. To test the defibrillator, follow the steps as described below •Set the defibrillator at 300 joules •Depress the charge button on the defibrillator until the display number matches the joules setting
  • 20. •Leave the paddles in their resting place on the defibrillator and simultaneously press the discharge buttons on both paddles •When someone is getting ready for defibrillation, the co-worker should begin with basic life support. •Patient should be continuously monitored after defibrillation. He should never be left alone for the first 24hr. vital signs are to be recorded very frequently until they are stabilized. Oxygen should be continuously given to the patient. Drugs should be administered as ordered. Intake and output chart to be maintained.