2. An Introduction to
Chromatography
What IS chromatography?
The separation of a mixture by distribution of its components between
a mobile and stationary phase over time
mobile phase = solvent
stationary phase = column packing material
The basis of gas chromatography is the distribution of a sample
between two phases namely stationary phase & gas phase ( mobile
phase )
The sample being vaporized & injected into the head of
chromatographic column
3. An Introduction to
Chromatography
A gas chromatograph uses a flow through a narrow tube known as
column through which different chemicals constituents of the sample
pass in a gas stream known as carrier gas or mobile phase at different
rates .
Depending on their various chemical and physical prosperities
Their interaction with specific column filling called stationary
phase
The sample is transported into the column by the flow of the inert
gaseous mobile phase
4. Components of a Chromatographic
System
● Source of Carrier Flow (mobile phase)
Cylinder of carrier gas or solvent bottles
● sample inlet
● Column with stationary phase
● Detector(s)
● Signal Transducers & Data Analyzers
Recorders, integrators
Computers for library matching
● Controllers
Temperature controls for injectors, columns and detector
Flow controllers and pressure regulators
6. Basic Chromatography
A basic Gas Chromatograph (GC) consists of the
following parts:
Carrier Gas
Carrier
Gas
Carrier
Regulator
Sample In
Sample out
SV Column Sense
Ref.
Atmos
Vents
Carries the sample through the column
and detector to an atmospheric vent.
7. Basic Chromatography
A basic Gas Chromatograph (GC) consists of the
following parts:
Carrier Regulator
Carrier
Gas
Carrier
Regulator
Sample In
Sample out
SV Column Sense
Ref.
Atmos
Vents
Maintains a constant pressure of carrier
gas which results in a constant carrier
flow rate.
8. Basic Chromatography
A basic Gas Chromatograph (GC) consists of the
following parts:
Sample Valve
Carrier
Gas
Carrier
Regulator
Sample In
Sample out
SV Column Sense
Ref.
Atmos
Vents
Injects a measured amount of sample.
9. Basic Chromatography
A basic Gas Chromatograph (GC) consists of the
following parts:
Column
Carrier
Gas
Carrier
Regulator
Sample In
Sample out
SV Column Sense
Ref.
Atmos
Vents
Separates the sample into individual
components.
10. Basic Chromatography
A basic Gas Chromatograph (GC) consists of the
following parts:
Detector
Carrier
Gas
Carrier
Regulator
Sample In
Sample out
SV Column Sense
Ref.
Atmos
Vents
Senses the individual components as
they elute off the column.
11. Carrier Gas-Supply
Carrier gases, which must be chemically inert, include
helium, nitrogen, and hydrogen. Associated with the
gas supply are pressure regulators, gauges, and flow
meters. In addition, the carrier gas system often
contains a molecular sieve to remove water or other
impurities.
12. A carrier gas should have the following properties:
1. Highly pure (> 99.9%)
2. Inert so that no reaction with stationary phase or instrumental
components can take place, especially at high temperatures.
3. When analyzing gas sample , the carrier is sometimes selected
based on the sample matrix. For example:
When analyzing a mix. In argon an argon carrier is preferred
because the argon in the sample doesn’t show up on the
chromatogram.
4. Compatible with the detector since some detectors require the
use of a specific carrier gas.
5. A cheap and available carrier gas is an advantage.
13. Sample Injection System
For quantitative work, more reproducible sample sizes
for both liquids and gases are obtained by means of a
rotary sample valve.
Errors due to sample size can be reduced to 0.5% to 2%
relative.
The sampling loop is filled by injection of an excess of
sample.
14.
15. Chromatographic separation involves the use of a
stationary phase and a mobile phase.
Components of a mixture carried in the mobile phase
are differentially attracted to the stationary phase and
thus move through the stationary phase at different
rates.
Columns
16. As the carrier gas sweeps the analyte molecules through the
column this motion is inhibited
by the adsorption of the analyte molecules either into the
column wall or into packing materials in the column
the rate at which molecules progress along the column depends
on the strength of adsorption which is depend on :
The type of molecules
The stationary phase material
Since each type of molecules has a different rates of progression ,
the various components of the analyte mixture are separated as
they progress along the column and reach the end of the column
at different times ( retention time )
Retention time
Retention time of an analyte is defined as the time it takes
after sample injection for the analyte to elute and reach the
detector.
Columns
18. Velocity of a compound through the column depends
upon affinity for the stationary phase
Area under curve is
______ of compound
adsorbed to stationary
phase
Gas phase concentration
Carrier gas
mass
19. Column Configurations
Two general types of columns are encountered in gas
chromatography, packed and open tubular, or
capillary.
Chromatographic columns vary in length from less
than 2 m to 50 m or more. They are constructed of
stainless steel, glass, fused silica, or Teflon. In order to
fit into an oven for thermostating, they are usually
formed as coils having diameters of 10 to 30 cm.
20. Packed Columns
These columns are fabricated from glass, stainless steel, copper, or other
suitable tubes.
Stainless steel is the most common tubing used with internal diameters
from 1-4 mm because it is most inert and easy to work with.
The column is packed with finely divided particles (<100-300 mm
diameter)
21. Capillary Columns
The most frequently used capillary column, nowadays, is the fused silica
open tubular column (FSOT), which is a WCOT column.
• Wall-coated open tubular (WCOT) <1 mm thick liquid coating on inside
of silica tube
22. Capillary vs. Packed Columns
Capillary Columns:
Higher resolution (R)
Shorter analysis time
Greater sensitivity
Most common in analytical
laboratory GC instruments
Smaller sample capacity
Higher cost/column
Columns more susceptible to
damage
Packed Columns
Greater sample capacity
Lower cost (can make your own)
More rugged
Most common in process labs or
separating/determining major
components in a sample (prep GC)
Limited lengths reduces R and N
Not compatible with some GC
detectors
23. Retention Time
Sample molecules spend part of the time in the mobile phase & the
other part in the stationary phase during the passage through the
column.
Column dead time tm
It’s the time for unretained solute to reach the detector from
the point of injection.
Solute retention time tr
Is the time difference between sample injection and the
detector sensing the maximum of the peak
Adjusted retention time tr’
The time solute molecules spend in the stationary
phase
24. Temperature and Pressure Programming
- Temperature is raised during the separation
(gradient)
- increases solute vapor pressure and decrease
retention time
Temperature gradient improves
resolution while also decreasing
retention time
25. Temperature and Pressure Programming
Separating components with a wide range of boiling points is
accomplished by starting at a low oven temperature and
increasing the temperature over time to elute the high-boiling
point components
26. Temperature and Pressure Programming
Programmed 30 to 180°
Isothermal at 145°
Isothermal at 45°
First, a temperature suitable for the separation of the first eluting
component is selected, and then the temperature is increased so that
the second component is separated and so on.
28. Gas Chromatography
Increase Stationary Phase Thickness
Increase resolution of early eluting compounds
Also, increase in
capacity factor and
reduce peak tailing
But also decreases
stability of stationary
phase
Increasing Resolution
29. Gas Chromatograph Output
Peak ____ proportional to mass of compound injected
Peak time dependent on ______ through column
time (s)
detector
output
area
velocity
30. For capillary GC columns….
Increased length = greater N, therefore a greater R
expense is possible band broadening if analytes are on the column
too long!
Increased length leads to longer separations. Do you have the time?
Increased stationary phase thickness and column diameter
provides increased sample capacity and can provide
increased resolution
tradeoffs are a longer analysis time and more column bleed with
thicker stationary phases
For most analytical work, a best “compromise” column is
chosen and other variables (temp, etc.) are altered to
optimize the separation.
31. GC Detectors
Separated components of the mixture must be
detected as they exit the GC column
Thermal-conduc. (TCD) and flame ionization
(FID) detectors - two most common detectors on
commercial GCs.
32. a. Thermal Conductivity Detector (TCD)
This is a nondestructive detector which is used for the
separation and collection of solutes to further
perform some other experiments on each purely
separated component.
The heart of the detector is a heated filament which is
cooled by helium carrier gas. Any solute passes across
the filament will not cool it as much as helium does
because helium has the highest thermal conductivity.
This results in an increase in the temperature of the
filament which is related to concentration.
The detector is simple, nondestructive, and universal
but is not very sensitive and is flow rate sensitive.
35. 35
b. Flame Ionization Detector (FID)
This is one of the most sensitive and reliable
destructive detectors. Separate two gas
cylinders, one for fuel and the other for O2 or
air are used in the ignition of the flame of the
FID. The fuel is usually hydrogen gas. The
flow rate of air and hydrogen should be
carefully adjusted in order to successfully
ignite the flame.
38. 38
The FID detector is a mass sensitive detector where
solutes are ionized in the flame and electrons
emitted are attracted by a positive electrode,
where a current is obtained.
Remember that FID characteristics include:
• Rugged
• Signal depends on number of carbon atoms in organic
analytes which is referred to as mass sensitive rather
than concentration sensitive
• Not sensitive to non-combustibles – H2O, CO2, SO2,
NOx
• Destructive
39.
40. Siemens: Systems Integration
Systems Integration Package
Features may include:
sun & rain protection
light & receptacle
120 or 480 VAC power circuits
cylinder rack
ventilation fan
electric heat or A/C
hazardous area classification