Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Present solutions to the problem of electromagnetic interference final
1. Present Solutions to the Problem of
Electromagnetic Interference: Are
They Good Enough?
Sergio L. Pinski, MD
Cleveland Clinic Florida
Weston, FL, USA
4. Pacemaker and ICD Responses to EMI
Pacing inhibition
Triggering of rapid or premature
pacing
Spurious tachyarrhythmia detection
Noise reversion mode
Electric (power-on) reset
Closure of the reed-switch
Damage to the generator or the
electrode-myocardial interface
12. Results
Episode Adjudication
5,248 Episodes
with Rhythm
Classification
3678 (70%)
1570 (30%) Estimated incidence
Appropriate
Inappropriate 1 year 5 year
(MVT / PVT /
(non-VT / VF) 6% 17%
VF)
820 (27.4%) 134 (2.6%)
Estimated incidence
Atrial Fib, SVT, Noise, Artifact,
1 year 5 year
sinus tach, or Oversensing
1.1% 1.8%
non-sustained
Page 12
13. Results
Episode Classification
Percent Percent of
of all NAO
Classification Episodes Patients Episodes Episodes
External noise / EMI 76 56 1.4% 56.7%
Lead / Connector 37 30 0.7% 27.6%
Muscle noise 11 11 0.2% 8.2%
Ventricular lead
7 3 0.1% 5.2%
oversensing of atrium
T wave oversensing 2 2 0.1% 1.5%
Other noise,
1 1 0.1% 0.7%
oversensing
Total 134 101 2.6% 100.0%
Page 13
14. Results
Incidence of NAO Resolution With Shock by Subtype
Category Episodes with Decrease in Noise
External noise / EMI 44 / 76 (58%)
Lead/Connector 13 / 37 (35%)
Muscle noise 3 / 11 (27%)
Other noise/oversensing 0 / 10 (0%)
Total 60 / 134 (45%)
P = .03 comparing External/EMI to Lead/Connector to Muscle – Fisher Exact test.
Example of Example with
noise reduction post shock no change post shock
Page 14
17. Design Constraints
• Need to sense very low level biological
signals
More ICDs than PMs, more in the atrial than
ventricular channel
• Small size is highly desirable by patients
and physicians for comfort and appearance
but limits size and number of components
• Low power
Power used to mitigate EMI reduces the life or
increases the battery size of the device
22. EMI Filter
Installation
In order to function properly at high frequencies, the EMI filter must be
installed (laser welded) so that it forms an integral part of the overall EMI
shield:
Filtered Hermetic Seal
Titanium Can
(EMI Shield)
23. Example of Passive
EMI Filter Performance
Cardiac Sense Lead
without EMI filter with EMI filter
100 MHz to 10 GHz 100 MHz to 10 GHz
24. Noise rejection algorithm
Clinical example of DNA in action: Identical noise and 5 mV R-wave in both devices
Legacy Device COGNIS & TELIGEN
Legacy devices could have sensed this noise as a COGNIS-TELIGEN recognizes the low level signal as noise
physiologic signal and appropriately adjusts sensitivity
25. – DNA uses the characteristics of a noise signal—frequency and
energy—to identify a signal as noise
– When noise is present, DNA keeps the AGC floor above the noise
– DNA is automatically active on all three sensing channels: atrium,
right ventricle and left ventricle
Note the presence of
electroconvulsive therapy
noise on the ventricular
rate sensing channel and
on the shocking egram
channel. In this Case
Study, DNA keeps AGC
sensing floor above noise.
Note: DNA will not make the Boston Scientific devices immune from sensing all noise. The device could still sense EMI or other sources of high amplitude noise.
Page 25
26.
27. Sorin Group Noise Management - Tachy
• Since 1996 Sorin Group/ELA has had a ventricular noise circuit in all
its ICDs
• Based on the premise that human beings cannot sustain intervals in
the 188-125 ms range. If intervals consistently in that range are
seen they are most likely the result of EMI (noise).
• After several retriggered windows Ventricular Sensitivity is
decreased by 0.2 mV on each retriggered window until not
retriggered. This process is done on a beat to beat basis
• The circuit is turned off for 15 cycles with an interval of
400 ms to 188 ms to assure arrhythmia detection
• On the 16th cycle, if there has not been an interval larger than 188
ms the noise circuit is turned on again
• Atrial noise circuit similar to Brady function
28. Sorin Group Noise Management - Tachy
R
P V
P
95 30
220 30
Figures show the addition of a 30 ms noise window to the 95 ms (sensed)
or the 220 ms (paced) absolute refractory
29. Sorin Group Noise Management - Tachy
AVD
95 ms
95 ms 95 ms
125 ms 125 ms
Noise level
0.2 mV steps
Sensitivity level Ventricular pacing can be inhibited
(parameter “V pacing on noise”)
Graphic representation of noise sensing to the point of decreasing V sensing
See example on the following slide
29
30. ICD Ventricular Noise Circuit Example
30
15 Cycles with noise circuit off after fast cycle
30
31. Noise Reversion Modes
Programmable
Manufacturer Noise detection window response Additional features
Boston Scientific XOO (nominal), Inhibit Non programmable
ICDs 40 ms retriggerable window pacing Dynamic Noise Algorithm
Programmed Non programmable
Medtronic PMs atrial/ventricular refractory asynchronous pacing
Medtronic ICDs Not available RV lead noise detection
Inhibit (nominal), pace Reduced sensitivity by 0.2
Sorin Paradym 125-188 ms asynchronous ms q 16 ms
St Jude PMs 30 Hz XOO (nomimal), off
XOO 50 BPM (nomimal),
St Jude ICDs 100 Hz off
32. Determinants of the Clinical
Consequences of EMI
Intensity of the field
Signal spectrum
Distance and position of the patient
Duration of exposure
Nonprogrammable device characteristics
Lead configuration
Programmed parameters
Sensitivity
Mode (baseline, noise reversion, committed)
Patient characteristics
Pacemaker-dependency
Susceptibility to asynchronous pacing
Susceptibility to rapid pacing rates
33. ANSI/AAMI PC69 standards -2207
Extensive guideline for in vitro
testing of pacemaker and ICDs
Typical settings, (eg cellular
phone operating at 6 inches)
Different frequencies
Tests also to rule out damage to
the generator from electrocautery
and external defibrillation
34. AAMI PC9- Testing for low-frequency EMI via
injected current
6
ISO 14708-2/EN 45502-2-1 Connection of tissue equivalent interface
circuit (left) and multichannel bipolar cardiac pacemaker (right).
36. “There are known knowns; there
are things we know we know. We
also know there are known
unknowns; that is to say we know
there are some things we do not
know. But there are also unknown
unknowns -- the ones we don't
know we don't know."
Donald Rumsfeld, US Secretary of Defense
37. Cell Phone EMI
• Cellular phone without amplification:
0.3 to 0.6 watts
• Cell phone with 3 watt after market
amplifier and 9 dB gain antenna:
23.8 watts
39. Kill the parasites that make you ill with
Para Zapper ™
improved Hulda Clark parasite zapper
In her book, The Cure For All Diseases, Dr. Clark describes how
certain parasites affect our health and how we can improve our
health by killing them using a parasite zapper . Since this book
was first published, many individuals have discovered that
zappers really do work. There are thousands of zappers sold
arround the world , most of them work, killing bacteria, protozoa,
and other parasites. We offer several different models and
options to suit your parasite zapper requirements. Our zappers
are offered to provide better quality at best price.
Better Zapper Made in the USA of high quality parts.
Better Dual Frequency zapper (30 kHz)and (2.5 kHz)
Better Top quality wires and plugs, more compact
Long battery life: Power LED
Best instruction manual available according to users.
Includes sample startup supplies. Sea Salt, etc.
Best zapper Price, Quality, Reliability, Performance .
44. Conclusions
Full electromagnetic compatibility
has not been achieved yet (and
may never be)
Multiple potential sources of EMI
exist in daily life, work and
medical environments
45. Conclusions
Improvement in sensing circuits and
algorithms together with better
awareness of sources of
electromagnetic interference have
already reduced EMI
Continous surveillance is needed as
new emitting sources are introduced
46. Conclusions
Sources of further minimization
Continuous improvements in
device engineering
Awareness of manufacturers of
emitters
Patient and public education
Notes de l'éditeur
Spurious mode-switch due to atrial oversensing of EMI in a patient with a dual-chamber Guidant ICD. Stored atrial (A), near-field (NF), and far-field (FF) electrograms in a patient who presented for routine ICD follow-up. Non-physiologic, high-frequency, pulsed activity is seen in the 3 channels, although with higher amplitude in the atrial and far-field. Very little ventricular oversensing occurs, so ventricular fibrillation is not detected. However, atrial oversensing results in transient mode-switch. The patient could not recall a potential source of EMI.
Signal amplitude versus frequency. This plot shows the approximate characteristics of the P and R waves that pacemakers and ICDs are intended to sense and the approximate characteristics of the electromagnetic interference (EMI, muscle potentials), T waves, and far-field R waves that they are intended not to sense. The sense amplifier's filters are designed to sense signals that are above the U-shaped amplifier threshold curve and to reject signals that are below the curve. P waves and R waves have similar frequency characteristics, but usually R waves have higher dominant frequency than P waves. Muscle potentials usually have higher-frequency components than intracardiac signals. T waves and far-field R waves have lower frequencies. As shown, there are some overlaps in these amplitude-frequency characteristics that cause oversensing or undersensing in particular situations. The ellipses representing the amplitude-frequency characteristics in this figure are conceptual and are not based on quantitative measurements.
Typically each catheter is protected by1) an EMI filtering capacitor (~ 500pF) in reference to the Titanium case ;2) 2 Zener diodes mounted in opposition to limit the voltage inputs to ~ +/- 9V.