Chapter 2
The Generation
of X-ray:
X-ray Tubes
The Generation of X-ray
•X-rays are produced
whenever electrons,
traveling at high
speeds, collide with
matter in any form.
The Generation of X-ray
• There are three essentials that must be
fulfilled before x-rays can be produced. They
are:
•1. A source of electrons
•2. A means of accelerating & controlling
their movement (via a difference in
potential)
•3. A place to stop them with great
suddenness (a point of impact)
The Generation of X-ray
• The Law of Conservation of Energy -
energy can neither be created or
destroyed.
• Binding Energies - The forces that hold
electrons in orbit around the nucleus.
The nucleus (+) attracts the electron (-),
but the spin of the electron keeps them
from collapsing toward the nucleus.
The Generation of X-ray
•Ground State - Inner shells have
more energy available due to
their proximity to the nucleus
(stronger interaction). Outer
shells have less energy available
because they have weaker
interaction with the nucleus.
The Coolidge Hot-Cathode Tube
In modern x-ray tube the glass bulb is
exhausted to as complete a vacuum as
is possible to attain. The cathode (-) is
composed of a small spiral filament of
tungsten wire, about 1 cm long & 0.2 to
0.3 cm in diameter, which is housed in a
focusing cup. The anode (+) is a solid
rod of copper & molybdenum in the
opposite end of the tube.
The Coolidge Hot-Cathode Tube
The surface of the anode (+)
facing the cathode (-) filament is
beveled at between7 and 17
degrees& has a block of
tungsten set into it; it is only a
few centimeters from the
filament.
How The Three Essentials are Fulfilled in
the Coolidge Hot-Cathode Tube
•The source of electrons in modern
x-ray tubes is the tungsten
filament. This is connected to a
step-down transformer & heated to
incandescence by current from it.
The heating of the wire alters its
atomic stability in that the electrons
are less firmly combined with the
nucleus of the atom.
How The Three Essentials are Fulfilled in
the Coolidge Hot-Cathode Tube
•These loosely bound electrons
hover about the cathode like a
cloud. The greater the heat to the
filament, the more electrons
available at the face of the cathode.
•This is called Thermionic Emission
or Boiling Off of Electrons (the
source of electrons)
How The Three Essentials are Fulfilled in
the Coolidge Hot-Cathode Tube
•If a high voltage current is applied
to the tube (negative to the
cathode) these electrons will be
repelled from the cathode (like
charges repelling) towards the
anode (unlike charges attracting)
with one-third to one-half the
speed of light (a means of
acceleration).
How The Three Essentials are Fulfilled in
the Coolidge Hot-Cathode Tube
• These electrons will strike the anode
with great force (a point of impact) & be
converted into x-rays & heat.
• The energy of the speeding electrons is
converted into two type of energy:
• Greater than 99% Heat
• Less than 1% X-Ray
(exothermic reaction - heat > energy)
The Focal Spot of the X-Ray Tube
•Most of the electrons bombard the
target over a small area near its
center. This is the actual focal spot
of the tube. The actual focal spot
has an area nearly equal in size to
the overall dimensions of the
filament.
The Focal Spot of the X-Ray Tube
• The smaller the focal spot, the greater is
the detail produced on the radiograph,
but the smaller is the capacity of the
tube to produce x-rays. This is
explained by the fact that all the energy
is expended at the focal spot .The
smaller the focal spot, the more intense
will be the heat developed. Longer
exposure time is necessary with such
tubes.
The Focal Spot of the X-Ray Tube
•To control the size of the focal
spot a sleeve of molybdenum
can be placed around the
filament & given a negative
charge which repels the
electrons from all directions,
this will form a more narrow
stream.
The Focal Spot of the X-Ray Tube
• The principle of line focus is a method
used to give a smaller focal spot with a
larger target area. The actual focal spot
is a rectangle approximately 3 times as
long as it is wide.
The Focal Spot of the X-Ray Tube
•When the beveled anode (+) face is
viewed from the patients point of view,
the focal spot appears to be nearly
square. This is the effective projected
focal spot. It serves to bring the x-ray
source closer to being one point while
still maintaining the larger area for
impact.
Double Focus Tubes
•Are those tubes having 2 focal
spots - one fine (0.3 mm) focus
for maximum detail & one large
(2.0 mm) for heavier exposures.
A mechanical switch includes
the focus of choice in the circuit.
Most x-ray machines have this
built in.
Double Focus Tubes
•Penumbra - blurring of the edge of
an organ or bone due to the size of
the focal spot.
•Today the large focal spots are 0.8
mm - 1.0 mm, and are actually
smaller than the old small focal
spots. This is accomplished via new
ways to cool the tubes.
Double Focus Tubes
Summary
•Large Filament = larger effective
focal spot - used for larger body
parts - looses some detail.
•Small Filament = smaller effective
focal spot - used for small body
parts such as extremity fractures
(hair line) and fine detail.(this
causes incredible heat build up)
Methods of Cooling the Anode
• Sufficient heat is generated in the
operation of an x-ray tube to melt the
tungsten target (3370 C) that methods
had to be developed to dissipate the
heat & protect the tube.
• Construction of the anode with two
metals - one with a very high melting
point (tungsten) & the other with a high
conductivity for heat (copper) - was an
important first step.
Methods of Cooling the Anode
•The main cooling methods that are
used or have been used are as
follows:
•Natural Radiation - In this method
heat is lost via the glass tube into
the air. Low capacity tubes may be
used for short periods without any
means of cooling.
Methods of Cooling the Anode
•Air Cooling by Radiation - A
radiator (series of metal discs) may
be attached to the extreme end of
the anode to increase the surface
area which can give off heat into
the air.
•Water Cooling - Obsolete today -
water was carried away through a
hollowed out anode stem.
Methods of Cooling the Anode
• Oil Cooling - Almost all x-ray tubes in
use today are surrounded by oil. The oil
insulates as well as cools. oil & air
cooling may be combined.
• The Rotating Anode Tube - As the
name implies the anode target rotates
during the exposure. This allows us to
increase the exposure because of the
tremendous ability to dissipate heat.
The Rotating Anode
• The anode in the tube is a beveled
tungsten disc attached to a rotor that
revolves when the tube is on. The
cathode filament is offset to one side so
that the electron stream hits near the
edge of the revolving disc.
• The rotating anode continually presents
a different area on the target to the
electron stream.
The Rotating Anode
• The focal spot remains fixed in space while
the circular anode rotates during the
exposure to provide a cooler surface for
the electron stream to strike.
• The heat is distributed over a broad band,
thus maintaining the temperature rise well
within safe limits. As the capacity of the
tube to withstand heat is increased, the
capacity of the tube to produce x-rays is
increased.
The Rotating Anode
•It also permits manufacturers to
produce tubes with smaller
effective focal spots.
•The disadvantages are:
»the tube is very delicate
»special lubricants are necessary for the
motor which will not produce volatile
gases
Tube Capacity
•The tube capacity or their
ability to produce x-rays is
affected by the rotating anode.
•Tubes are rated in terms of:
•Kilovoltage (kV)
•Miliamperage (MA)
•Time of exposure (S)
Tube Capacity
• These factors are dependent on the
rectification system, cooling method and
focal spot size.
• kV - capacity is determined by the distance
from the filament to the target.
• MA - capacity is determined by size of the
focal spot & the rectification system used.
• S - capacity is determined by the anode &
tube cooling rates.
The Heel Effect
• Heel effect is the term applied to the fact
that x-ray radiation does not exit the long
axis of the tube in uniform intensities.
• The intensity of the beam is equal to the
number of rays & diminishes fairly
rapidly from the central ray to the anode
side of the patient, while increasing
slightly toward the cathode side of the
patient.
The Heel Effect
• We can use this to our advantage in
radiography if we remember to position the
tube with the anode end of the tube towards
the more easily penetrated body part.
• Originally it was thought that the heel effect
was due to the angulation of the anode &
was based upon the theory of refraction.
The Heel Effect
• Today we believe that the electrons which are
traveling at high speeds, bombard the target and
xray is produced in 360 degrees of direction. 180
degrees of this x ray produced is absorbed by
the anode itself.
• Therefore, only 180 degrees of x-rays leave the
anode.
• Of those, some are absorbed by the lead shutters
of the collimator, allowing only those travelling
in the desired direction to exit the tube.
The Heel Effect
• X-ray is not uniform along the film surface.
Therefore the effects of the heel effect are
seen toward the edges of the film.
• To decrease the heel effect, increase the
collimation & decrease the film size.
• With increased collimation you get less heel
effect.
The Heel Effect
•For example, in taking a
radiograph of the cervical-
thoracic area, the neck area
should receive the rays from the
anode portion of the beam as
the neck is thinner than the
thoracic region.

Contenu connexe

Tendances

X ray beam rest
X ray beam restX ray beam rest
X ray beam restBuandari
 
Faults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarFaults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarSandeshMagar4
 
Production of x rays
Production of x raysProduction of x rays
Production of x rayspraveen kumar
 
X ray production and interaction
X ray production and interactionX ray production and interaction
X ray production and interactionharibudke
 
X ray generator
X ray generatorX ray generator
X ray generatoraslam bs
 
Characteristic and-bremsstrahlung-x-rays
Characteristic and-bremsstrahlung-x-raysCharacteristic and-bremsstrahlung-x-rays
Characteristic and-bremsstrahlung-x-raysJoaquinLuisRinehart1
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray TubeAlisha Karmali
 
Computed radiography AND ITS ADVANTAGES
Computed radiography AND ITS ADVANTAGESComputed radiography AND ITS ADVANTAGES
Computed radiography AND ITS ADVANTAGESFirdousDar4
 
X rays generators and transformers
X rays   generators  and transformersX rays   generators  and transformers
X rays generators and transformersshariq ahmad shah
 
X ray production and properties
X ray production and propertiesX ray production and properties
X ray production and propertiesHanuman Doke
 
CHAPTER-7 - QUALITY OF X-RAY BEAM
 CHAPTER-7 - QUALITY OF X-RAY BEAM CHAPTER-7 - QUALITY OF X-RAY BEAM
CHAPTER-7 - QUALITY OF X-RAY BEAMKanhu Charan
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...DrKajalLimbad
 
X ray tube
X ray tubeX ray tube
X ray tubeRad Tech
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakeshSylvia Diamond
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptxGMC Anantnag
 

Tendances (20)

X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
Faults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarFaults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magar
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
X ray production and interaction
X ray production and interactionX ray production and interaction
X ray production and interaction
 
X ray generator
X ray generatorX ray generator
X ray generator
 
Characteristic and-bremsstrahlung-x-rays
Characteristic and-bremsstrahlung-x-raysCharacteristic and-bremsstrahlung-x-rays
Characteristic and-bremsstrahlung-x-rays
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray Tube
 
Grids
GridsGrids
Grids
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Computed radiography AND ITS ADVANTAGES
Computed radiography AND ITS ADVANTAGESComputed radiography AND ITS ADVANTAGES
Computed radiography AND ITS ADVANTAGES
 
X ray filters
X ray filtersX ray filters
X ray filters
 
Radiation detectors
Radiation detectorsRadiation detectors
Radiation detectors
 
X rays generators and transformers
X rays   generators  and transformersX rays   generators  and transformers
X rays generators and transformers
 
X ray production and properties
X ray production and propertiesX ray production and properties
X ray production and properties
 
CHAPTER-7 - QUALITY OF X-RAY BEAM
 CHAPTER-7 - QUALITY OF X-RAY BEAM CHAPTER-7 - QUALITY OF X-RAY BEAM
CHAPTER-7 - QUALITY OF X-RAY BEAM
 
X ray generator new.ppt
X ray generator new.pptX ray generator new.ppt
X ray generator new.ppt
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
X ray tube
X ray tubeX ray tube
X ray tube
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakesh
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptx
 

Similaire à PHYSICS2.PPT

Similaire à PHYSICS2.PPT (20)

X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptx
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptx
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
ProductionOfXrays.pptx
ProductionOfXrays.pptxProductionOfXrays.pptx
ProductionOfXrays.pptx
 
ProductionOfXrays.pptx
ProductionOfXrays.pptxProductionOfXrays.pptx
ProductionOfXrays.pptx
 
X-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique LectureX-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique Lecture
 
X-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique LectureX-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique Lecture
 
X-RAY TUBE.pptx
X-RAY TUBE.pptxX-RAY TUBE.pptx
X-RAY TUBE.pptx
 
X-RAY TUBE.pptx
X-RAY TUBE.pptxX-RAY TUBE.pptx
X-RAY TUBE.pptx
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
 
Neelam
NeelamNeelam
Neelam
 
Neelam
NeelamNeelam
Neelam
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1
 
Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1
 

Dernier

Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpinRaunakKeshri1
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...fonyou31
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...Pooja Nehwal
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesFatimaKhan178732
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...anjaliyadav012327
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 

Dernier (20)

Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpin
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and Actinides
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 

PHYSICS2.PPT

  • 1. Chapter 2 The Generation of X-ray: X-ray Tubes
  • 2. The Generation of X-ray •X-rays are produced whenever electrons, traveling at high speeds, collide with matter in any form.
  • 3. The Generation of X-ray • There are three essentials that must be fulfilled before x-rays can be produced. They are: •1. A source of electrons •2. A means of accelerating & controlling their movement (via a difference in potential) •3. A place to stop them with great suddenness (a point of impact)
  • 4. The Generation of X-ray • The Law of Conservation of Energy - energy can neither be created or destroyed. • Binding Energies - The forces that hold electrons in orbit around the nucleus. The nucleus (+) attracts the electron (-), but the spin of the electron keeps them from collapsing toward the nucleus.
  • 5. The Generation of X-ray •Ground State - Inner shells have more energy available due to their proximity to the nucleus (stronger interaction). Outer shells have less energy available because they have weaker interaction with the nucleus.
  • 6. The Coolidge Hot-Cathode Tube In modern x-ray tube the glass bulb is exhausted to as complete a vacuum as is possible to attain. The cathode (-) is composed of a small spiral filament of tungsten wire, about 1 cm long & 0.2 to 0.3 cm in diameter, which is housed in a focusing cup. The anode (+) is a solid rod of copper & molybdenum in the opposite end of the tube.
  • 7.
  • 8. The Coolidge Hot-Cathode Tube The surface of the anode (+) facing the cathode (-) filament is beveled at between7 and 17 degrees& has a block of tungsten set into it; it is only a few centimeters from the filament.
  • 9.
  • 10. How The Three Essentials are Fulfilled in the Coolidge Hot-Cathode Tube •The source of electrons in modern x-ray tubes is the tungsten filament. This is connected to a step-down transformer & heated to incandescence by current from it. The heating of the wire alters its atomic stability in that the electrons are less firmly combined with the nucleus of the atom.
  • 11. How The Three Essentials are Fulfilled in the Coolidge Hot-Cathode Tube •These loosely bound electrons hover about the cathode like a cloud. The greater the heat to the filament, the more electrons available at the face of the cathode. •This is called Thermionic Emission or Boiling Off of Electrons (the source of electrons)
  • 12. How The Three Essentials are Fulfilled in the Coolidge Hot-Cathode Tube •If a high voltage current is applied to the tube (negative to the cathode) these electrons will be repelled from the cathode (like charges repelling) towards the anode (unlike charges attracting) with one-third to one-half the speed of light (a means of acceleration).
  • 13. How The Three Essentials are Fulfilled in the Coolidge Hot-Cathode Tube • These electrons will strike the anode with great force (a point of impact) & be converted into x-rays & heat. • The energy of the speeding electrons is converted into two type of energy: • Greater than 99% Heat • Less than 1% X-Ray (exothermic reaction - heat > energy)
  • 14.
  • 15. The Focal Spot of the X-Ray Tube •Most of the electrons bombard the target over a small area near its center. This is the actual focal spot of the tube. The actual focal spot has an area nearly equal in size to the overall dimensions of the filament.
  • 16.
  • 17. The Focal Spot of the X-Ray Tube • The smaller the focal spot, the greater is the detail produced on the radiograph, but the smaller is the capacity of the tube to produce x-rays. This is explained by the fact that all the energy is expended at the focal spot .The smaller the focal spot, the more intense will be the heat developed. Longer exposure time is necessary with such tubes.
  • 18. The Focal Spot of the X-Ray Tube •To control the size of the focal spot a sleeve of molybdenum can be placed around the filament & given a negative charge which repels the electrons from all directions, this will form a more narrow stream.
  • 19.
  • 20. The Focal Spot of the X-Ray Tube • The principle of line focus is a method used to give a smaller focal spot with a larger target area. The actual focal spot is a rectangle approximately 3 times as long as it is wide.
  • 21.
  • 22. The Focal Spot of the X-Ray Tube •When the beveled anode (+) face is viewed from the patients point of view, the focal spot appears to be nearly square. This is the effective projected focal spot. It serves to bring the x-ray source closer to being one point while still maintaining the larger area for impact.
  • 23. Double Focus Tubes •Are those tubes having 2 focal spots - one fine (0.3 mm) focus for maximum detail & one large (2.0 mm) for heavier exposures. A mechanical switch includes the focus of choice in the circuit. Most x-ray machines have this built in.
  • 24. Double Focus Tubes •Penumbra - blurring of the edge of an organ or bone due to the size of the focal spot. •Today the large focal spots are 0.8 mm - 1.0 mm, and are actually smaller than the old small focal spots. This is accomplished via new ways to cool the tubes.
  • 25. Double Focus Tubes Summary •Large Filament = larger effective focal spot - used for larger body parts - looses some detail. •Small Filament = smaller effective focal spot - used for small body parts such as extremity fractures (hair line) and fine detail.(this causes incredible heat build up)
  • 26. Methods of Cooling the Anode • Sufficient heat is generated in the operation of an x-ray tube to melt the tungsten target (3370 C) that methods had to be developed to dissipate the heat & protect the tube. • Construction of the anode with two metals - one with a very high melting point (tungsten) & the other with a high conductivity for heat (copper) - was an important first step.
  • 27. Methods of Cooling the Anode •The main cooling methods that are used or have been used are as follows: •Natural Radiation - In this method heat is lost via the glass tube into the air. Low capacity tubes may be used for short periods without any means of cooling.
  • 28. Methods of Cooling the Anode •Air Cooling by Radiation - A radiator (series of metal discs) may be attached to the extreme end of the anode to increase the surface area which can give off heat into the air. •Water Cooling - Obsolete today - water was carried away through a hollowed out anode stem.
  • 29. Methods of Cooling the Anode • Oil Cooling - Almost all x-ray tubes in use today are surrounded by oil. The oil insulates as well as cools. oil & air cooling may be combined. • The Rotating Anode Tube - As the name implies the anode target rotates during the exposure. This allows us to increase the exposure because of the tremendous ability to dissipate heat.
  • 30. The Rotating Anode • The anode in the tube is a beveled tungsten disc attached to a rotor that revolves when the tube is on. The cathode filament is offset to one side so that the electron stream hits near the edge of the revolving disc. • The rotating anode continually presents a different area on the target to the electron stream.
  • 31.
  • 32. The Rotating Anode • The focal spot remains fixed in space while the circular anode rotates during the exposure to provide a cooler surface for the electron stream to strike. • The heat is distributed over a broad band, thus maintaining the temperature rise well within safe limits. As the capacity of the tube to withstand heat is increased, the capacity of the tube to produce x-rays is increased.
  • 33.
  • 34. The Rotating Anode •It also permits manufacturers to produce tubes with smaller effective focal spots. •The disadvantages are: »the tube is very delicate »special lubricants are necessary for the motor which will not produce volatile gases
  • 35. Tube Capacity •The tube capacity or their ability to produce x-rays is affected by the rotating anode. •Tubes are rated in terms of: •Kilovoltage (kV) •Miliamperage (MA) •Time of exposure (S)
  • 36. Tube Capacity • These factors are dependent on the rectification system, cooling method and focal spot size. • kV - capacity is determined by the distance from the filament to the target. • MA - capacity is determined by size of the focal spot & the rectification system used. • S - capacity is determined by the anode & tube cooling rates.
  • 37. The Heel Effect • Heel effect is the term applied to the fact that x-ray radiation does not exit the long axis of the tube in uniform intensities. • The intensity of the beam is equal to the number of rays & diminishes fairly rapidly from the central ray to the anode side of the patient, while increasing slightly toward the cathode side of the patient.
  • 38.
  • 39. The Heel Effect • We can use this to our advantage in radiography if we remember to position the tube with the anode end of the tube towards the more easily penetrated body part. • Originally it was thought that the heel effect was due to the angulation of the anode & was based upon the theory of refraction.
  • 40.
  • 41. The Heel Effect • Today we believe that the electrons which are traveling at high speeds, bombard the target and xray is produced in 360 degrees of direction. 180 degrees of this x ray produced is absorbed by the anode itself. • Therefore, only 180 degrees of x-rays leave the anode. • Of those, some are absorbed by the lead shutters of the collimator, allowing only those travelling in the desired direction to exit the tube.
  • 42. The Heel Effect • X-ray is not uniform along the film surface. Therefore the effects of the heel effect are seen toward the edges of the film. • To decrease the heel effect, increase the collimation & decrease the film size. • With increased collimation you get less heel effect.
  • 43. The Heel Effect •For example, in taking a radiograph of the cervical- thoracic area, the neck area should receive the rays from the anode portion of the beam as the neck is thinner than the thoracic region.