Classical Brachytherapy System of
Interstitial Implants : Manchester
system
Dr. Supriya Sonaje
Henri Becquerel discovers
radioactivity on February 26, 1896
15 dec 1852 -25 august1908
RADIUM :Marie Curie in 1898.
RADIUM DECAY
α α α βγ βγ
Uranium Ra Rn RaA RaB RaC Pb
1620yrs 3.83D 3.05Min 26.8min 19.7min stable
Gamma rays of different energies are emmited – 0.6, 1.1 and 2.2
MeV , Average energy being – 1 MeV .
Ra B emits beta ray of 0.65 MeV .
Ra C emits beta ray of 3.17 MeV .
• AND SOON BECAME THERADIOACTIVE
QUACKERY !!!
• Few examples –
• Radiothor – salt containing radium endorsed
to have curative properties.
• Toothpastes contianing radium – to make the
teeth shiny .
• In bathing water , drinking water .
• Danlos and Bloc performed the first
radioactive implant in 1901.
• First schools of brachytherapy were at Holt
radium institute , Paris.
Brachytherapy consists of placing
sealed radioactive sources very close
to or in contact with the target tissue.
Types of brachytherapy
1. Intracavitary insertion consists of positioning
applicators (bearing RAS) into a body cavity in close
proximity to target tissue
2. Interstitial insertion consists of surgically implanting
small RAS directly into the target tissues
3. Surface dose ( mould ) applications consist of an
applicator containing an array of RAS usually designed
to deliver an uniform dose distribution to a
skin/mucosal surface
4. Transluminal application consists of inserting a single
line source into a body lumen to treat its surface &
adjacent tissues
• Dose rate
 LOW DOSE RATE (LDR)
 0.4-2 Gy/hr
 MEDIUM DOSE RATE (MDR)
 2-12 Gy/hr
 HIGH DOSE RATE (HDR)
 > 12 Gy/hr
 ULTRA LOW DOSE RATE
 0.01-0.3 Gy/hr
• Loading method-
• Preloading
• afterloading
Clinical advantages of brachytherapy -
• High biological efficacy
• Rapid dose fall off
• Tolerable acute reactions
• Decreased risk of tumor population
• High control rate
• Better cosmesis
Limitations of brachytherapy
• Difficult for inaccessible regions
• Limited for small tumor size
• Invasive
• Greater conformation
• Needs skilled personnel
• Radioactivity :The activity, A, of an amount of
radioactive nuclide, in a particular energy
state and at a given time,is defined by
equation: A= d N/ d T
• where dN is the expectation value of the
number of spontaneous nuclear transitions
from that energy state in the time interval dt.
• Problems with the radium
• Daughter element – radon (gas)- possibility of
leakage which is not detected by visual check .
• The practical maximum activity concentration
(the specific activity) of radium salt is low
(approximately 50 MBq mm~3 of active volume).
Sources of higher activity are therefore bulky and
not suitable for afterloading systems.
• Need for heavy fitration and shielding .
Properties of ideal brachytherapy
sources
• Pure γ emitter – less α/β emission.
• Medium γ energy – high enough to target tumor with
homogenous dose & low enough to avoid normal
tissues & reduce shielding needs
• High specific activity – small size & suitability for HDR
• Stable ( not liquid/gaseous) daughter product
• Long t ½ /medium t ½ for permanent/ temporary
implant
• Should be available in a form which doesn’t powder or
disperse if source is damaged or dispensed
RADIUM SUBSTITUTES
NAME ORIGIN T1/2 γ
ENERG
Y-MeV
β
ENERGY
β
FILTRATIO
N
HVL
(Pb -
mm)
SPECI.
ACTI.
DECAY
PRODUC
T
Rn 222 NATURAL 3.83
days
0.83 Stainless
steel
12 Pb 206
Cs 137 FISSION 30.1
7
yrs
0.662 0.512
1.17
- do - 6.5 87 Ba 137
Co 60 NEUTRON
ACTIVAT.
5.26
yrs
1.17,
1.33
0.38 - do- 11 1020 Ni 60
Ir 192 - do - 73.8
day
0.136-
1.06
0.079-
0.068
Platinum 4 7760 Pt 192
Tn 182 - do - 115
yrs
0.67 - Platinum 12 - -
Au 198 - do - 2.7
days
1.088-
0.412
0.96 St. steel 3.3 - Hg 198
I 125 - do - 59.4
days
0.274,
0.314
No Titanium 0.01
10th
- Te 125
Pd 103 - do - 16.9 0.21 No Platinum 0.03 - Ru 103
Systems in Interstitial brachytherapy
• “System” denotes a set of rules which takes into
account the source types & strengths, geometry &
method of application to obtain suitable dose
distributions over the volume(s) to be treated.
• Implant should follow both source distribution rules
and method of dose prescription & specification of a
system
• Consist of
1. Distribution rules
2. Dose specification & implant optimization criteria
3. Dose calculation aids
Manchester system
Need for system
• Two questions
• How much radium will be required?
• and how must it be arranged?
How much of radium ?
• A unit of dosage rather than of intensity was
preferred.
• Because -total dose is the most important single
factor in therapy (provided due consideration is
given to the duration of radiation).
• the clinician finds it very much easier to think in
terms of
• total dose delivered rather than in terms of
intensity and time, separately
How much of radium?
• For dosage unit of exposure – Roentgen (r) is
chosen .
• Roentgen is ‘the amount of Xray or gamma
radiation such that the associated corpuscular
emmision , per 0.001293 gramme of air,
produces in air ions carrying 1 electrostatic
unit charge of either sign ‘
• To determine exposure rate ‘R’ at a given point from
a point source ‘A’ mCi whose specific gamma ray
constant is ‘Γ’ at a distance ‘d’
• R = Γ x A (Roentgen/mCi hr.)
d2
• Thus exposure rate of 1 mg of Ra at 1 cm with
standard 0.5mm filtration thru Pt is found to be
8.4 R/ hr.
• Line source :
We divide the line into multiple point sources and the exposure rate simply added.
> 1 cm
1 cm
> 1 cm
1. Amount of Ra depends on
1. Value of desired exposure
2. Area being treated
3. Treating distance
2. Clinical working unit was 1000 Roentgens
3. Exposure rate was in R/hr
4. Exposure Rate constant Γ= 8.4 R/mg hr
5. Table compiled for Mg hr/1000R for a given
area/actual implant volume at a given distance
6. Filtered by 0.5mm of platinum (2% filtration
correction factor for every 1mm)
7. A radiation field could be described as ‘uniform’, if
variation is not more than +/-10 %.
8. R to cGy factor = 0.957
• The amount of radium needed for is
determined – for a perticular area , treating
distance , total exposure and treatment time
for the designed dosage chart.
For example-
• If an area of 20 cm2 is to receive 6000R over 50 hrs
from a mould for which h=1.0 cm , taking the values
from dosage chart-
• The milgramme hours(mgh)per 1000Rfor 20 cm2 and
h=1 cm is 641 .
• Therefore , 6000R: 641x6=3846mgh will be needed .
• And radium required for 50 hr treatment will be
3846/50=76.9 mg .
The distribution-governed by the
inverse square law .
• “for a point source , the radiation intensity at
any place varies inversely as the square of the
distance from the source to the place at which
intensity is being considered”.
• Dose varies more rapidly in 1st cm of distance
Geometrical terms :
• The plane formed by
the parallel sources -
radioactive plane.
• The centre of the
radioactive plane is
called the centre of
the implant.
• The central plane
through this point is
located at right angles
to the needles and
thus at right angles to
the radioactive plane.
A third plane parallel to the sources and at
right angles to the radioactive plane, will
be referred to as the coronal plane
• The length of the implant is the length of the radioactive
plane in the direction of the sources .
• The width of the implant is the dimension at right angles
to the length, in the radioactive plane or parallel to it.
• The thickness of the implant is the dimension at right
angles to the length and to the radioactive plane, in the
central plane or parallel to the central plane.
MOULDS:rules
PLANAR MOULDS
• The word ‘planar’implies that the radium
sources are mounted on a flat or curved
surface parallel to the area being treated ,the
curved surface being less than a semicircle .
• Mould arrangement-Circular/square/rectagle
/random
• Treatment surface-Flat/convex/concave
For moulds
The term mould is used to describe the situation in which the
radioactive sources are positioned external to the patient, usually at a
distance from the patient's skin known as the treatment distance and
represented by the letter d.
The treatment dose is prescribed to the plane which is at distance d
from the sources and the dose in this plane will be delivered with a 10%
accuracy if the rules are followed.
For circular mould
• The distribution of radium depends upon the
D/d. (D- diameter of the treatment area , d-
treatment distance )
• For ‘ideal circle’ central and periferal doses are
equal and the variation elsewhere across the
circle is minimal.
• If D/d is less than 3 , a single circle of radium will
produce uniform irradiation over the treatment
area , within the defination of ‘uniformity’
• Most uniform distribution is achived with D/d
=2.8 .
%Radium D/d
<3 3- <6 6-
<7.5
7.5<1
0
10
In outer
circle
100 95 80 75 70
In inner
circle
- - 17 22 27
In central
spot
- 5 3 3 3
For square mould :
• Source is devided in to periphery and in lines,
called as bars.
• The number of bars is such that area is
devided in to strips of width not greater than
2 d .
the amount of radiumdistribution depends upon
the number of bars .
For square moulds
• If one bar suffices , its linear density
(miligrammes per cm) should be half of that in
periphery.
• If two or more bars are needed , linear density
should be 2/3rd of that of the periphery.
• Attempt is made to maintain the space
between two active ends of the sources less
than the treating distance (d)
For rectangular moulds
• The bars are taken in a line parallel to the
longer side .
• To get the vallue of mgh per 1000R ,
elongation correction factor is applied
Ratio of
sides of
rectangle
1.5:1 2:1 3:1 4:1
Elongatio
n
correctio
n factor
1.025 1.05 1.09 1.12
For curved surface
The smaller area is chosen.
Dose below the treated surface
• As tumors treated by moulds are superficial
and thin , the dose at the distance d’ below
the surface can be calculated by adding d’ and
d, and looking for the value of mghper 1000R
for treatment distance (d’+d) in the dosage
chart .
Interstitial implant
Planar interstitial implant
• Considered as direct development of planar
mould with d=0.5 cm , inside the tissue .
• They are implanted inside the tissue to get the
uniform dose distribution on either side
within distance of 0.5 cm .
• For the values of mgh per 1000R , values from
the charts for mould can be used .
Distribution rule for single planar
implant
• The proportion of the radium used on
periphery depends upon the area of the
treatment itself .
Area Under 25
cm2
25-100
cm2
Over 100
cm 2
Peripheral
fraction
2/3 1/2 1/3
45
Importance of Differential Loading
1. All sources equal
2. Weaker sources in interior
prescribed dose does
not conform to target
volume
Dose conforms to
target volume
2. The needles should be arranged in parallel rows 1
cm apart with the ends crossed(active ends < 1 cm
from crossed needles)
Areas of under dosage
End source fills in under dosed areas
Shape of isodose line
around a linear source
3.. If the ends of the implant are uncrossed, the area should
reduced by 10 % for each uncrossed end for table reading
purposes
A. Both ends crossed. The area treated is (a x b)
B. One end uncrossed. The area treated should be considered as 0.90 (a x b)
C. Both ends uncrossed. The area treated should be considered as 0.80 (a x b)
4. If two planes are to be used, the separate planes should
be arranged as for single planes, parallel to each other, and
if they differ in area, then the average area is used to
determine the mg-hrs and the activity is proportioned to
each plane.
Volume interstitial implants
• To calculate the amount of radium required , it
follows the various set of charts .
• M=34.1 V2/3 x f
• M- miligramme per 1000R
• V- implanted volume in cubic cm
• f – correction factor
Rules for distribution in a
cylindrical implant
CORE
RIND/BELT
UPPER END
LOWER END
1. Any volume can be visualized as having two
components : outer surface(RIND) and CORE.
Total amt. of Radium is divided into 8 parts and
distributed for the various shapes:
If these rules followed exposure throughout treatment volume
will not drop by > 10% or rise by > 15% above stated exposure
2. The sources on each surface should be spaced as
evenly as possible, with 1 -1.5 cm separation
3. For the core, the sources should be spread as
evenly as possible throughout the volume and not all
of it at the center.
4. In cylinders, belt should consist of not less than 8
needles, the core not less than 4.
5. If crossing not possible, volume of implant must be
reduced by 7.5% for each uncrossed end. Crossing
needles inserted at active ends, not at tips.
INTRACAVITARY BRACHYTHERAPY
• Introduced the points at which dosage can be
calculated .
• Such a point must be anatomically
comparable in the patients and should be at
which dosage is not highly sensitive to small ,
clinically unimportant alteration in the source
geometry and should be in the region of
limiting radiosensitivity.
• Introduced – ovoids and intrauterine tubes.
Point A- a point 2 cm latersl to the centre of uterine canal and 2 cm
superior to the mucous membrane of lateral fornix , in the plane of uteru
Point B - 5 cm laterally
to midline at the same level as the A points
• Point A - The dose at point A is representative
of the dose throughout much of the malignant tissue.
Corresponds to the the point where uterine artery
crosses the uterus .
Point B – anatomically proximal to the obturator glands .
Represents the lateral fall off of the dose .
• The dose at point B is approximately 20-25% of the
dose at point A and is of importance when calculating
the total dose when brachytherapy is combined with
external-beam irradiation
• Not more than about a third of total exposure
rate at Pint A should be delivered from the
vaginal sources
BIOLOGICAL ASPECTS GIVEN BY
MANCHESTER SYSTEM
Showed that biological changes in tumor as well as normal tissue
not only depend upon the dose of radiation but also on the
duration over which this dose is spread.
• The dose required to produce lethal change rises as the total
period of radiation increases.
• “It has been found that, subject to the proviso regarding
duration of exposure already discussed, a dose of 6,000 ‘r,’ if
delivered to the whole of tumour and tumour-bearing zone,
causes permanent resolution of the great majority of
epitheliomata. This figure may, therefore, be taken as a
serviceable in vivo ‘tumour-lethal’ dose”.
• Data regarding the other tumors is not extensive.
Suggested readings :
• 1. Patterson, R. and Parker, H.M. (1934) A dosage
system for gamma ray therapy. Br.J. Radiol., 7, 592.
• Fundamental Physics of Radiology by W.J. Meredith
& J.B. Massey (dosage calculation for plesiotherapy)
• Principles and practice of brachytherapy –CAF Joslin
THANK YOU

Contenu connexe

Tendances

Time, dose and fractionation
Time, dose and fractionationTime, dose and fractionation
Time, dose and fractionationDr. Ankita Pandey
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in RadiotherapySantam Chakraborty
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranKiran Ramakrishna
 
Icru reports in external beam radiotherapy
Icru reports in external beam radiotherapyIcru reports in external beam radiotherapy
Icru reports in external beam radiotherapyDeepika Malik
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devicesVibhay Pareek
 
Brachytherapy.ar
Brachytherapy.arBrachytherapy.ar
Brachytherapy.arraazvarma
 
TIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONTIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONIsha Jaiswal
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYSathish Kumar
 
Brachytherapy msc lecture sam copy
Brachytherapy msc lecture sam copyBrachytherapy msc lecture sam copy
Brachytherapy msc lecture sam copyAyodele Dinyo
 
Total skin electron irradiation
Total skin electron irradiation Total skin electron irradiation
Total skin electron irradiation Rupon Bhowmik
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesSantam Chakraborty
 
LDR and HDR Brachytherapy: A Primer for non radiation oncologists
LDR and HDR Brachytherapy: A Primer for non radiation oncologistsLDR and HDR Brachytherapy: A Primer for non radiation oncologists
LDR and HDR Brachytherapy: A Primer for non radiation oncologistsSantam Chakraborty
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsDrAyush Garg
 
Brachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .RBrachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .Rraazvarma
 

Tendances (20)

Time, dose and fractionation
Time, dose and fractionationTime, dose and fractionation
Time, dose and fractionation
 
Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in Radiotherapy
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr Kiran
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Icru reports in external beam radiotherapy
Icru reports in external beam radiotherapyIcru reports in external beam radiotherapy
Icru reports in external beam radiotherapy
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devices
 
ICRU 38 nayana
ICRU 38 nayanaICRU 38 nayana
ICRU 38 nayana
 
Brachytherapy.ar
Brachytherapy.arBrachytherapy.ar
Brachytherapy.ar
 
TIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONTIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATION
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPY
 
Isodose lines
Isodose linesIsodose lines
Isodose lines
 
Brachytherapy msc lecture sam copy
Brachytherapy msc lecture sam copyBrachytherapy msc lecture sam copy
Brachytherapy msc lecture sam copy
 
Total skin electron irradiation
Total skin electron irradiation Total skin electron irradiation
Total skin electron irradiation
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 
LDR and HDR Brachytherapy: A Primer for non radiation oncologists
LDR and HDR Brachytherapy: A Primer for non radiation oncologistsLDR and HDR Brachytherapy: A Primer for non radiation oncologists
LDR and HDR Brachytherapy: A Primer for non radiation oncologists
 
Gap correction
Gap correctionGap correction
Gap correction
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 Radiophysics
 
Brachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .RBrachytherapy dosimetry systems .R
Brachytherapy dosimetry systems .R
 
Icru 58.
Icru 58.Icru 58.
Icru 58.
 

Similaire à Manchester system_Dr.Supriya

Interstitial BT Principles
Interstitial BT PrinciplesInterstitial BT Principles
Interstitial BT PrinciplesYamini Baviskar
 
Evolution of gynaecological brachytherapy
Evolution of gynaecological brachytherapyEvolution of gynaecological brachytherapy
Evolution of gynaecological brachytherapyRitam Joarder
 
Evolution of Intra-cavitary Brachytherapy
Evolution of Intra-cavitary BrachytherapyEvolution of Intra-cavitary Brachytherapy
Evolution of Intra-cavitary BrachytherapyDr Boaz Vincent
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapyVinay Desai
 
Brachytherapy in Carcinoma Cervix
Brachytherapy in Carcinoma Cervix Brachytherapy in Carcinoma Cervix
Brachytherapy in Carcinoma Cervix Nidhil Krishna
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Arya Prasad
 
Brachytherapy in carcinoma cervix - history to present
Brachytherapy in carcinoma cervix - history to presentBrachytherapy in carcinoma cervix - history to present
Brachytherapy in carcinoma cervix - history to presentSACHINS700327
 
Thermoluminescent dosimeter
Thermoluminescent dosimeterThermoluminescent dosimeter
Thermoluminescent dosimeterDr Vijay Raturi
 
Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalmedicalHunter2
 
Physical and clinical aspects of brachytherapy
Physical and clinical aspects of  brachytherapyPhysical and clinical aspects of  brachytherapy
Physical and clinical aspects of brachytherapyRam Pukar Bharat
 
RAMAN-Chapter-18-part-2-useful for us.ppt
RAMAN-Chapter-18-part-2-useful for us.pptRAMAN-Chapter-18-part-2-useful for us.ppt
RAMAN-Chapter-18-part-2-useful for us.pptChristopherJeyakumar3
 

Similaire à Manchester system_Dr.Supriya (20)

Interstitial BT Principles
Interstitial BT PrinciplesInterstitial BT Principles
Interstitial BT Principles
 
Evolution of gynaecological brachytherapy
Evolution of gynaecological brachytherapyEvolution of gynaecological brachytherapy
Evolution of gynaecological brachytherapy
 
Evolution of Intra-cavitary Brachytherapy
Evolution of Intra-cavitary BrachytherapyEvolution of Intra-cavitary Brachytherapy
Evolution of Intra-cavitary Brachytherapy
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapy
 
Icru 38
Icru   38Icru   38
Icru 38
 
Brachytherapy in Carcinoma Cervix
Brachytherapy in Carcinoma Cervix Brachytherapy in Carcinoma Cervix
Brachytherapy in Carcinoma Cervix
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
 
Brachytherapy in carcinoma cervix - history to present
Brachytherapy in carcinoma cervix - history to presentBrachytherapy in carcinoma cervix - history to present
Brachytherapy in carcinoma cervix - history to present
 
Co60 vs Ir192
Co60 vs Ir192Co60 vs Ir192
Co60 vs Ir192
 
ICRU 71
ICRU 71ICRU 71
ICRU 71
 
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTSUnit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
 
Icru seminar 38
Icru seminar 38Icru seminar 38
Icru seminar 38
 
co 60.ppt
co 60.pptco 60.ppt
co 60.ppt
 
Thermoluminescent dosimeter
Thermoluminescent dosimeterThermoluminescent dosimeter
Thermoluminescent dosimeter
 
Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopital
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Physical and clinical aspects of brachytherapy
Physical and clinical aspects of  brachytherapyPhysical and clinical aspects of  brachytherapy
Physical and clinical aspects of brachytherapy
 
Instruments of NMR
Instruments of NMR Instruments of NMR
Instruments of NMR
 
RAMAN-Chapter-18-part-2-useful for us.ppt
RAMAN-Chapter-18-part-2-useful for us.pptRAMAN-Chapter-18-part-2-useful for us.ppt
RAMAN-Chapter-18-part-2-useful for us.ppt
 
BT.pptx
BT.pptxBT.pptx
BT.pptx
 

Dernier

All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...jageshsingh5554
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Dernier (20)

All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 

Manchester system_Dr.Supriya

  • 1. Classical Brachytherapy System of Interstitial Implants : Manchester system Dr. Supriya Sonaje
  • 2. Henri Becquerel discovers radioactivity on February 26, 1896 15 dec 1852 -25 august1908
  • 4.
  • 5. RADIUM DECAY α α α βγ βγ Uranium Ra Rn RaA RaB RaC Pb 1620yrs 3.83D 3.05Min 26.8min 19.7min stable Gamma rays of different energies are emmited – 0.6, 1.1 and 2.2 MeV , Average energy being – 1 MeV . Ra B emits beta ray of 0.65 MeV . Ra C emits beta ray of 3.17 MeV .
  • 6. • AND SOON BECAME THERADIOACTIVE QUACKERY !!! • Few examples – • Radiothor – salt containing radium endorsed to have curative properties. • Toothpastes contianing radium – to make the teeth shiny . • In bathing water , drinking water .
  • 7. • Danlos and Bloc performed the first radioactive implant in 1901. • First schools of brachytherapy were at Holt radium institute , Paris.
  • 8. Brachytherapy consists of placing sealed radioactive sources very close to or in contact with the target tissue.
  • 9. Types of brachytherapy 1. Intracavitary insertion consists of positioning applicators (bearing RAS) into a body cavity in close proximity to target tissue 2. Interstitial insertion consists of surgically implanting small RAS directly into the target tissues 3. Surface dose ( mould ) applications consist of an applicator containing an array of RAS usually designed to deliver an uniform dose distribution to a skin/mucosal surface 4. Transluminal application consists of inserting a single line source into a body lumen to treat its surface & adjacent tissues
  • 10. • Dose rate  LOW DOSE RATE (LDR)  0.4-2 Gy/hr  MEDIUM DOSE RATE (MDR)  2-12 Gy/hr  HIGH DOSE RATE (HDR)  > 12 Gy/hr  ULTRA LOW DOSE RATE  0.01-0.3 Gy/hr
  • 11. • Loading method- • Preloading • afterloading
  • 12. Clinical advantages of brachytherapy - • High biological efficacy • Rapid dose fall off • Tolerable acute reactions • Decreased risk of tumor population • High control rate • Better cosmesis
  • 13. Limitations of brachytherapy • Difficult for inaccessible regions • Limited for small tumor size • Invasive • Greater conformation • Needs skilled personnel
  • 14. • Radioactivity :The activity, A, of an amount of radioactive nuclide, in a particular energy state and at a given time,is defined by equation: A= d N/ d T • where dN is the expectation value of the number of spontaneous nuclear transitions from that energy state in the time interval dt.
  • 15. • Problems with the radium • Daughter element – radon (gas)- possibility of leakage which is not detected by visual check . • The practical maximum activity concentration (the specific activity) of radium salt is low (approximately 50 MBq mm~3 of active volume). Sources of higher activity are therefore bulky and not suitable for afterloading systems. • Need for heavy fitration and shielding .
  • 16. Properties of ideal brachytherapy sources • Pure γ emitter – less α/β emission. • Medium γ energy – high enough to target tumor with homogenous dose & low enough to avoid normal tissues & reduce shielding needs • High specific activity – small size & suitability for HDR • Stable ( not liquid/gaseous) daughter product • Long t ½ /medium t ½ for permanent/ temporary implant • Should be available in a form which doesn’t powder or disperse if source is damaged or dispensed
  • 17. RADIUM SUBSTITUTES NAME ORIGIN T1/2 γ ENERG Y-MeV β ENERGY β FILTRATIO N HVL (Pb - mm) SPECI. ACTI. DECAY PRODUC T Rn 222 NATURAL 3.83 days 0.83 Stainless steel 12 Pb 206 Cs 137 FISSION 30.1 7 yrs 0.662 0.512 1.17 - do - 6.5 87 Ba 137 Co 60 NEUTRON ACTIVAT. 5.26 yrs 1.17, 1.33 0.38 - do- 11 1020 Ni 60 Ir 192 - do - 73.8 day 0.136- 1.06 0.079- 0.068 Platinum 4 7760 Pt 192 Tn 182 - do - 115 yrs 0.67 - Platinum 12 - - Au 198 - do - 2.7 days 1.088- 0.412 0.96 St. steel 3.3 - Hg 198 I 125 - do - 59.4 days 0.274, 0.314 No Titanium 0.01 10th - Te 125 Pd 103 - do - 16.9 0.21 No Platinum 0.03 - Ru 103
  • 18. Systems in Interstitial brachytherapy • “System” denotes a set of rules which takes into account the source types & strengths, geometry & method of application to obtain suitable dose distributions over the volume(s) to be treated. • Implant should follow both source distribution rules and method of dose prescription & specification of a system • Consist of 1. Distribution rules 2. Dose specification & implant optimization criteria 3. Dose calculation aids
  • 20.
  • 21. Need for system • Two questions • How much radium will be required? • and how must it be arranged?
  • 22. How much of radium ? • A unit of dosage rather than of intensity was preferred. • Because -total dose is the most important single factor in therapy (provided due consideration is given to the duration of radiation). • the clinician finds it very much easier to think in terms of • total dose delivered rather than in terms of intensity and time, separately
  • 23. How much of radium? • For dosage unit of exposure – Roentgen (r) is chosen . • Roentgen is ‘the amount of Xray or gamma radiation such that the associated corpuscular emmision , per 0.001293 gramme of air, produces in air ions carrying 1 electrostatic unit charge of either sign ‘
  • 24. • To determine exposure rate ‘R’ at a given point from a point source ‘A’ mCi whose specific gamma ray constant is ‘Γ’ at a distance ‘d’ • R = Γ x A (Roentgen/mCi hr.) d2 • Thus exposure rate of 1 mg of Ra at 1 cm with standard 0.5mm filtration thru Pt is found to be 8.4 R/ hr. • Line source : We divide the line into multiple point sources and the exposure rate simply added. > 1 cm 1 cm > 1 cm
  • 25. 1. Amount of Ra depends on 1. Value of desired exposure 2. Area being treated 3. Treating distance 2. Clinical working unit was 1000 Roentgens 3. Exposure rate was in R/hr 4. Exposure Rate constant Γ= 8.4 R/mg hr 5. Table compiled for Mg hr/1000R for a given area/actual implant volume at a given distance 6. Filtered by 0.5mm of platinum (2% filtration correction factor for every 1mm) 7. A radiation field could be described as ‘uniform’, if variation is not more than +/-10 %. 8. R to cGy factor = 0.957
  • 26. • The amount of radium needed for is determined – for a perticular area , treating distance , total exposure and treatment time for the designed dosage chart.
  • 27. For example- • If an area of 20 cm2 is to receive 6000R over 50 hrs from a mould for which h=1.0 cm , taking the values from dosage chart- • The milgramme hours(mgh)per 1000Rfor 20 cm2 and h=1 cm is 641 . • Therefore , 6000R: 641x6=3846mgh will be needed . • And radium required for 50 hr treatment will be 3846/50=76.9 mg .
  • 28. The distribution-governed by the inverse square law . • “for a point source , the radiation intensity at any place varies inversely as the square of the distance from the source to the place at which intensity is being considered”. • Dose varies more rapidly in 1st cm of distance
  • 29. Geometrical terms : • The plane formed by the parallel sources - radioactive plane. • The centre of the radioactive plane is called the centre of the implant. • The central plane through this point is located at right angles to the needles and thus at right angles to the radioactive plane. A third plane parallel to the sources and at right angles to the radioactive plane, will be referred to as the coronal plane
  • 30. • The length of the implant is the length of the radioactive plane in the direction of the sources . • The width of the implant is the dimension at right angles to the length, in the radioactive plane or parallel to it. • The thickness of the implant is the dimension at right angles to the length and to the radioactive plane, in the central plane or parallel to the central plane.
  • 32. PLANAR MOULDS • The word ‘planar’implies that the radium sources are mounted on a flat or curved surface parallel to the area being treated ,the curved surface being less than a semicircle . • Mould arrangement-Circular/square/rectagle /random • Treatment surface-Flat/convex/concave
  • 33. For moulds The term mould is used to describe the situation in which the radioactive sources are positioned external to the patient, usually at a distance from the patient's skin known as the treatment distance and represented by the letter d. The treatment dose is prescribed to the plane which is at distance d from the sources and the dose in this plane will be delivered with a 10% accuracy if the rules are followed.
  • 34. For circular mould • The distribution of radium depends upon the D/d. (D- diameter of the treatment area , d- treatment distance ) • For ‘ideal circle’ central and periferal doses are equal and the variation elsewhere across the circle is minimal. • If D/d is less than 3 , a single circle of radium will produce uniform irradiation over the treatment area , within the defination of ‘uniformity’ • Most uniform distribution is achived with D/d =2.8 .
  • 35. %Radium D/d <3 3- <6 6- <7.5 7.5<1 0 10 In outer circle 100 95 80 75 70 In inner circle - - 17 22 27 In central spot - 5 3 3 3
  • 36.
  • 37. For square mould : • Source is devided in to periphery and in lines, called as bars. • The number of bars is such that area is devided in to strips of width not greater than 2 d . the amount of radiumdistribution depends upon the number of bars .
  • 38. For square moulds • If one bar suffices , its linear density (miligrammes per cm) should be half of that in periphery. • If two or more bars are needed , linear density should be 2/3rd of that of the periphery. • Attempt is made to maintain the space between two active ends of the sources less than the treating distance (d)
  • 39. For rectangular moulds • The bars are taken in a line parallel to the longer side . • To get the vallue of mgh per 1000R , elongation correction factor is applied Ratio of sides of rectangle 1.5:1 2:1 3:1 4:1 Elongatio n correctio n factor 1.025 1.05 1.09 1.12
  • 40. For curved surface The smaller area is chosen.
  • 41. Dose below the treated surface • As tumors treated by moulds are superficial and thin , the dose at the distance d’ below the surface can be calculated by adding d’ and d, and looking for the value of mghper 1000R for treatment distance (d’+d) in the dosage chart .
  • 43. Planar interstitial implant • Considered as direct development of planar mould with d=0.5 cm , inside the tissue . • They are implanted inside the tissue to get the uniform dose distribution on either side within distance of 0.5 cm . • For the values of mgh per 1000R , values from the charts for mould can be used .
  • 44. Distribution rule for single planar implant • The proportion of the radium used on periphery depends upon the area of the treatment itself . Area Under 25 cm2 25-100 cm2 Over 100 cm 2 Peripheral fraction 2/3 1/2 1/3
  • 45. 45 Importance of Differential Loading 1. All sources equal 2. Weaker sources in interior prescribed dose does not conform to target volume Dose conforms to target volume
  • 46. 2. The needles should be arranged in parallel rows 1 cm apart with the ends crossed(active ends < 1 cm from crossed needles) Areas of under dosage End source fills in under dosed areas Shape of isodose line around a linear source
  • 47. 3.. If the ends of the implant are uncrossed, the area should reduced by 10 % for each uncrossed end for table reading purposes A. Both ends crossed. The area treated is (a x b) B. One end uncrossed. The area treated should be considered as 0.90 (a x b) C. Both ends uncrossed. The area treated should be considered as 0.80 (a x b)
  • 48. 4. If two planes are to be used, the separate planes should be arranged as for single planes, parallel to each other, and if they differ in area, then the average area is used to determine the mg-hrs and the activity is proportioned to each plane.
  • 49. Volume interstitial implants • To calculate the amount of radium required , it follows the various set of charts . • M=34.1 V2/3 x f • M- miligramme per 1000R • V- implanted volume in cubic cm • f – correction factor
  • 50.
  • 51. Rules for distribution in a cylindrical implant
  • 52. CORE RIND/BELT UPPER END LOWER END 1. Any volume can be visualized as having two components : outer surface(RIND) and CORE. Total amt. of Radium is divided into 8 parts and distributed for the various shapes:
  • 53. If these rules followed exposure throughout treatment volume will not drop by > 10% or rise by > 15% above stated exposure 2. The sources on each surface should be spaced as evenly as possible, with 1 -1.5 cm separation 3. For the core, the sources should be spread as evenly as possible throughout the volume and not all of it at the center. 4. In cylinders, belt should consist of not less than 8 needles, the core not less than 4. 5. If crossing not possible, volume of implant must be reduced by 7.5% for each uncrossed end. Crossing needles inserted at active ends, not at tips.
  • 55. • Introduced the points at which dosage can be calculated . • Such a point must be anatomically comparable in the patients and should be at which dosage is not highly sensitive to small , clinically unimportant alteration in the source geometry and should be in the region of limiting radiosensitivity. • Introduced – ovoids and intrauterine tubes.
  • 56. Point A- a point 2 cm latersl to the centre of uterine canal and 2 cm superior to the mucous membrane of lateral fornix , in the plane of uteru Point B - 5 cm laterally to midline at the same level as the A points
  • 57. • Point A - The dose at point A is representative of the dose throughout much of the malignant tissue. Corresponds to the the point where uterine artery crosses the uterus . Point B – anatomically proximal to the obturator glands . Represents the lateral fall off of the dose . • The dose at point B is approximately 20-25% of the dose at point A and is of importance when calculating the total dose when brachytherapy is combined with external-beam irradiation
  • 58.
  • 59.
  • 60. • Not more than about a third of total exposure rate at Pint A should be delivered from the vaginal sources
  • 61.
  • 62. BIOLOGICAL ASPECTS GIVEN BY MANCHESTER SYSTEM
  • 63. Showed that biological changes in tumor as well as normal tissue not only depend upon the dose of radiation but also on the duration over which this dose is spread. • The dose required to produce lethal change rises as the total period of radiation increases. • “It has been found that, subject to the proviso regarding duration of exposure already discussed, a dose of 6,000 ‘r,’ if delivered to the whole of tumour and tumour-bearing zone, causes permanent resolution of the great majority of epitheliomata. This figure may, therefore, be taken as a serviceable in vivo ‘tumour-lethal’ dose”. • Data regarding the other tumors is not extensive.
  • 64. Suggested readings : • 1. Patterson, R. and Parker, H.M. (1934) A dosage system for gamma ray therapy. Br.J. Radiol., 7, 592. • Fundamental Physics of Radiology by W.J. Meredith & J.B. Massey (dosage calculation for plesiotherapy) • Principles and practice of brachytherapy –CAF Joslin THANK YOU