Electrical impedance tomography simulator
Krammer P, Waldmann AD, Zogg M, Róka PL, Brunner JX ,
Bohm SH; 16th International Conference on Biomedical
Applications of Electrical Impedance Tomography, Neuchâtel
Switzerland, June 2-5, 2015
40 SESSION 5. HARDWARE I
Electrical impedance tomography simulator
Péter Krammer1
, Andreas D. Waldmann1
, Michel Zogg1
, Péter L. Róka1,2
, Josef X. Brunner1
and
Stephan H. Bohm1
1
Swisstom AG, Landquart, Switzerland, pkr@swisstom.com
2
Budapest University of Technology and Economics, Budapest, Hungary
Abstract: Assessing the performance of electrical
impedance tomography (EIT) devices usually requires a
phantom for validation, calibration or comparative
measurements. To bridge the gap between the typical data
obtained in patients wearing a chest belt ventilation- and
cardiac- related impedance changes should also be
simulated either manually or automatically. In this abstract
we describe a novel EIT simulator based on a dynamic
controllable resistor mesh.
1 Introduction
Electrical impedance tomography is a novel method for
monitoring lung and heart providing moving real-time
images of function rather than structures. For this purpose,
16 or 32 electrodes are typically placed around the human
thorax. Weak alternating currents are applied via two
electrodes and the resulting potentials are measured via
the remaining electrodes. From the measured voltages,
images are calculated which show the distribution of
electrical impedance within the body [1]. Intra-thoracic
impedance changes with ventilation [2] and within each
cardiac cycle [3].
Although some groups have proposed resistor mesh
phantoms to test and present EIT devices an EIT simulator
allowing the user to dynamically adjust impedance
changes is still missing.
2 Methods
Therefore, we developed an EIT simulator consisting of a
resistor phantom with 160 resistors, details can be found
in [4], [5], a microcontroller (ATmega64), to program
various simulation scenarios, a 10 bit digital to analogue
converter (MAX5723) and 7 field-effect transistors
(IRLL2705) to induce impedance changes. Each field-
effect transistor (FET) is located parallel to one of the 160
resistor and induces an impedance change proportional to
the applied voltage at its gate. 6 FETs are placed in such a
way that they produce impedance changes within the
demarcated expected lung contours of the EIT image and
one in the heart region. Breath and heart rate as well as
their respective amplitudes can be controlled by four
Figure 1: Components of EIT Simulator
knobs. Furthermore, the EIT simulator is equipped with a
pressure sensor (MPX5010) to input airway pressures
from the ventilator circuit. In this case the FETs can be
driven according to the measured pressure signal. Failing
electrodes can also be simulated simply by pressing one of
the switches, see Figure 1.
First measurements were performed with the
Swisstom BB2
EIT device. Impedance changes were
induced in both, the lung region only and in the lung and
heart regions. Power images were generated to describe
the impedance distribution while the global impedance
curves were plotted to show the signals along the time.
3 Results
Representative sample results are presented in Figure 2. In
A the amplitude of only the lung signals can be seen, in B
the cardiac related impedance change and the lung signal
is shown and C depicts a measurement on a healthy
volunteer. D depicts the simulated and measured
ventilation and cardiac related impedance changes.
Figure 2: Simulation results: A amplitudes of only the lung
signal, B amplitudes of lung and heart signal, C amplitudes of a
healthy volunteer, D time signals for case A, B and C.
4 Conclusions
Both, EIT images as well as global impedance curves
proved representative of what can be obtained from living
beings. Compared to other resistor mesh phantoms our
system can also simulate dynamic impedance changes at
different anatomy-related locations within the resistor
mesh phantom. The novel EIT simulator is used for
performance testing, comparative measurements and for
demonstration purposes.
References
[1] Costa EL, Curr.Opin.Crit.Care, vol 15, pp. 18-24, 2009
[2] Wolf GK, Critical Care Medicine, vol. 41, pp. 509-515, 2013
[3] Frerichs I, IEEE Trans. Med. Imaging, vol. 21 pp. 646-52, 2002
[4] Gagnon H, IEEE Trans. Bio Med. Eng, vol. 57, pp. 2257-2266, 2010
[5] Hahn G, Physiol. Meas, vol. 29, pp. 163-72, 2008
A
D
B
B
D
A
C
A: Control of heart and
ventilation
B: resistor mesh phantom
C: Connection to EIT device
D: External signal input
E: Failing electrode control
F: Modus selection control
E
F
C
SWI SSTO M SCIE NT IFIC LIBRARY
Content:Dr.StephanBöhm;Concept&Design:ZweizeitBrandDevelopment
Made in Switzerland
electrical
impedance
tomography
Real-time tomographic images for organ
function monitoring and diagnosis
Contact us!
call: + 41 (0) 81 330 09 72
mail: info@swisstom.com
visit: www.swisstom.com
Swisstom AG
Schulstrasse 1, CH-7302
Landquart, Switzerland
Swisstom AG
Swisstom AG, located in Landquart, Switzerland,
develops and manufactures innovative medical devices.
Our new lung function monitor enables life-saving
treatments for patients in intensive care and during
general anesthesia.
Unlike traditional tomography, Swisstom´s bedside
imaging is based on non-radiating principles: Electrical
Impedance Tomography (EIT). To date, no comparable
devices can show such regional organ function
continuously and in real-time at the patient’s bedside.
Swisstom creates its competitive edge by passionate
leadership in non-invasive tomography with the goal
to improve individual lives and therapies.
© Swisstom AG

Contenu connexe

Tendances

Reconstruction of electrical impedance tomography images based on the expecta...
Reconstruction of electrical impedance tomography images based on the expecta...Reconstruction of electrical impedance tomography images based on the expecta...
Reconstruction of electrical impedance tomography images based on the expecta...ISA Interchange
 
Implementation and design of new low-cost foot pressure sensor module using p...
Implementation and design of new low-cost foot pressure sensor module using p...Implementation and design of new low-cost foot pressure sensor module using p...
Implementation and design of new low-cost foot pressure sensor module using p...IJECEIAES
 
Embedded system for upper-limb exoskeleton based on electromyography control
Embedded system for upper-limb exoskeleton based on electromyography controlEmbedded system for upper-limb exoskeleton based on electromyography control
Embedded system for upper-limb exoskeleton based on electromyography controlTELKOMNIKA JOURNAL
 
Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)Anurag Tewari MD
 
Mecta ppt product overview
Mecta ppt product overviewMecta ppt product overview
Mecta ppt product overviewodojam
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)nishanthi p
 
Hybrid neural networks in cyber physical system interface control systems
Hybrid neural networks in cyber physical system interface control systemsHybrid neural networks in cyber physical system interface control systems
Hybrid neural networks in cyber physical system interface control systemsjournalBEEI
 
Electromyography Analysis for Person Identification
Electromyography Analysis for Person IdentificationElectromyography Analysis for Person Identification
Electromyography Analysis for Person IdentificationCSCJournals
 
Real time ecg signal analysis by using new data reduction algorithm for
Real time ecg signal analysis by using new data reduction algorithm forReal time ecg signal analysis by using new data reduction algorithm for
Real time ecg signal analysis by using new data reduction algorithm forIAEME Publication
 
Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Amir Ziai
 
medical terminology presentation
medical terminology presentationmedical terminology presentation
medical terminology presentationayodeji adeyemi
 
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...SomeBitchIKnow PhD
 
NET 2014-Myoelectric Prosthetic Hand with Air muscles
NET 2014-Myoelectric Prosthetic Hand with Air musclesNET 2014-Myoelectric Prosthetic Hand with Air muscles
NET 2014-Myoelectric Prosthetic Hand with Air musclesRosemary James T
 
Utilizing ECG Waveform Features as New Biometric Authentication Method
Utilizing ECG Waveform Features as New Biometric Authentication Method Utilizing ECG Waveform Features as New Biometric Authentication Method
Utilizing ECG Waveform Features as New Biometric Authentication Method IJECEIAES
 
Parameters affecting the selectivity of an electrical insecticide
Parameters affecting the selectivity of an electrical insecticideParameters affecting the selectivity of an electrical insecticide
Parameters affecting the selectivity of an electrical insecticideIJECEIAES
 

Tendances (18)

Reconstruction of electrical impedance tomography images based on the expecta...
Reconstruction of electrical impedance tomography images based on the expecta...Reconstruction of electrical impedance tomography images based on the expecta...
Reconstruction of electrical impedance tomography images based on the expecta...
 
ELECTROMYOGRAM
ELECTROMYOGRAM ELECTROMYOGRAM
ELECTROMYOGRAM
 
I047020950101
I047020950101I047020950101
I047020950101
 
Implementation and design of new low-cost foot pressure sensor module using p...
Implementation and design of new low-cost foot pressure sensor module using p...Implementation and design of new low-cost foot pressure sensor module using p...
Implementation and design of new low-cost foot pressure sensor module using p...
 
Embedded system for upper-limb exoskeleton based on electromyography control
Embedded system for upper-limb exoskeleton based on electromyography controlEmbedded system for upper-limb exoskeleton based on electromyography control
Embedded system for upper-limb exoskeleton based on electromyography control
 
Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)
 
Mecta ppt product overview
Mecta ppt product overviewMecta ppt product overview
Mecta ppt product overview
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
 
Hybrid neural networks in cyber physical system interface control systems
Hybrid neural networks in cyber physical system interface control systemsHybrid neural networks in cyber physical system interface control systems
Hybrid neural networks in cyber physical system interface control systems
 
Whitepaper
WhitepaperWhitepaper
Whitepaper
 
Electromyography Analysis for Person Identification
Electromyography Analysis for Person IdentificationElectromyography Analysis for Person Identification
Electromyography Analysis for Person Identification
 
Real time ecg signal analysis by using new data reduction algorithm for
Real time ecg signal analysis by using new data reduction algorithm forReal time ecg signal analysis by using new data reduction algorithm for
Real time ecg signal analysis by using new data reduction algorithm for
 
Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...
 
medical terminology presentation
medical terminology presentationmedical terminology presentation
medical terminology presentation
 
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
 
NET 2014-Myoelectric Prosthetic Hand with Air muscles
NET 2014-Myoelectric Prosthetic Hand with Air musclesNET 2014-Myoelectric Prosthetic Hand with Air muscles
NET 2014-Myoelectric Prosthetic Hand with Air muscles
 
Utilizing ECG Waveform Features as New Biometric Authentication Method
Utilizing ECG Waveform Features as New Biometric Authentication Method Utilizing ECG Waveform Features as New Biometric Authentication Method
Utilizing ECG Waveform Features as New Biometric Authentication Method
 
Parameters affecting the selectivity of an electrical insecticide
Parameters affecting the selectivity of an electrical insecticideParameters affecting the selectivity of an electrical insecticide
Parameters affecting the selectivity of an electrical insecticide
 

En vedette

Fpga based motor controller
Fpga based motor controllerFpga based motor controller
Fpga based motor controllerUday Wankar
 
Single phase grid connected fuel system based on boost inverter
Single phase grid connected fuel system based on boost inverterSingle phase grid connected fuel system based on boost inverter
Single phase grid connected fuel system based on boost inverterRohithasangaraju
 
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...Field Programmable Gate Array(FPGA) Application In Instrument Landing System...
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...Mal Mai
 
Speed estimation error of sensorless induction motor drives using soft comput...
Speed estimation error of sensorless induction motor drives using soft comput...Speed estimation error of sensorless induction motor drives using soft comput...
Speed estimation error of sensorless induction motor drives using soft comput...IAEME Publication
 
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...Efficient Motor Control Solutions: High Performance Servo Control (Design Con...
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...Analog Devices, Inc.
 
Optimization of distributed generation of renewable energy sources by intelli...
Optimization of distributed generation of renewable energy sources by intelli...Optimization of distributed generation of renewable energy sources by intelli...
Optimization of distributed generation of renewable energy sources by intelli...Beniamino Murgante
 
Project presentation - battery charger 24V/33A
Project presentation - battery charger 24V/33AProject presentation - battery charger 24V/33A
Project presentation - battery charger 24V/33APrathamesh Deshpande
 
electronics and robotics
electronics and roboticselectronics and robotics
electronics and roboticsRajat Dhiman
 
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...Analysis of Power Factor Improvement Techniques in case of Non Linear System ...
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...Kaustubh Nande
 
Using a Field Programmable Gate Array to Accelerate Application Performance
Using a Field Programmable Gate Array to Accelerate Application PerformanceUsing a Field Programmable Gate Array to Accelerate Application Performance
Using a Field Programmable Gate Array to Accelerate Application PerformanceOdinot Stanislas
 
An industrial automation satyaaaa
An industrial automation  satyaaaaAn industrial automation  satyaaaa
An industrial automation satyaaaasatyarout9090
 
Application of electric and electronic drives in robotics
Application of electric and electronic drives in roboticsApplication of electric and electronic drives in robotics
Application of electric and electronic drives in roboticsAshutosh Shahi
 
Resonant Converter
Resonant ConverterResonant Converter
Resonant Converterpilurout
 
ANTI COLLISION SYSTEM IN CARS
ANTI COLLISION SYSTEM IN CARSANTI COLLISION SYSTEM IN CARS
ANTI COLLISION SYSTEM IN CARSRakesh Reddy
 
Tachometer using AT89S52 microcontroller with motor control
Tachometer using AT89S52 microcontroller with motor controlTachometer using AT89S52 microcontroller with motor control
Tachometer using AT89S52 microcontroller with motor controlSushil Mishra
 

En vedette (20)

PPTSIRISHPROPOSAL
PPTSIRISHPROPOSALPPTSIRISHPROPOSAL
PPTSIRISHPROPOSAL
 
2009 11 35838829
2009 11 358388292009 11 35838829
2009 11 35838829
 
Fpga based motor controller
Fpga based motor controllerFpga based motor controller
Fpga based motor controller
 
Ac2 09-anti windup
Ac2 09-anti windupAc2 09-anti windup
Ac2 09-anti windup
 
Single phase grid connected fuel system based on boost inverter
Single phase grid connected fuel system based on boost inverterSingle phase grid connected fuel system based on boost inverter
Single phase grid connected fuel system based on boost inverter
 
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...Field Programmable Gate Array(FPGA) Application In Instrument Landing System...
Field Programmable Gate Array(FPGA) Application In Instrument Landing System...
 
Speed estimation error of sensorless induction motor drives using soft comput...
Speed estimation error of sensorless induction motor drives using soft comput...Speed estimation error of sensorless induction motor drives using soft comput...
Speed estimation error of sensorless induction motor drives using soft comput...
 
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...Efficient Motor Control Solutions: High Performance Servo Control (Design Con...
Efficient Motor Control Solutions: High Performance Servo Control (Design Con...
 
Ballas1
Ballas1Ballas1
Ballas1
 
Optimization of distributed generation of renewable energy sources by intelli...
Optimization of distributed generation of renewable energy sources by intelli...Optimization of distributed generation of renewable energy sources by intelli...
Optimization of distributed generation of renewable energy sources by intelli...
 
Project presentation - battery charger 24V/33A
Project presentation - battery charger 24V/33AProject presentation - battery charger 24V/33A
Project presentation - battery charger 24V/33A
 
electronics and robotics
electronics and roboticselectronics and robotics
electronics and robotics
 
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...Analysis of Power Factor Improvement Techniques in case of Non Linear System ...
Analysis of Power Factor Improvement Techniques in case of Non Linear System ...
 
Using a Field Programmable Gate Array to Accelerate Application Performance
Using a Field Programmable Gate Array to Accelerate Application PerformanceUsing a Field Programmable Gate Array to Accelerate Application Performance
Using a Field Programmable Gate Array to Accelerate Application Performance
 
An industrial automation satyaaaa
An industrial automation  satyaaaaAn industrial automation  satyaaaa
An industrial automation satyaaaa
 
Application of electric and electronic drives in robotics
Application of electric and electronic drives in roboticsApplication of electric and electronic drives in robotics
Application of electric and electronic drives in robotics
 
Resonant Converter
Resonant ConverterResonant Converter
Resonant Converter
 
ANTI COLLISION SYSTEM IN CARS
ANTI COLLISION SYSTEM IN CARSANTI COLLISION SYSTEM IN CARS
ANTI COLLISION SYSTEM IN CARS
 
Automobile collision avoidance system
Automobile collision avoidance systemAutomobile collision avoidance system
Automobile collision avoidance system
 
Tachometer using AT89S52 microcontroller with motor control
Tachometer using AT89S52 microcontroller with motor controlTachometer using AT89S52 microcontroller with motor control
Tachometer using AT89S52 microcontroller with motor control
 

Similaire à Krammer P. et al.: Electrical impedance tomography Simulator.

White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) Hauke Sann
 
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...Jose Francisco Saenz Cogollo
 
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Hauke Sann
 
A comparative study of wavelet families for electromyography signal classific...
A comparative study of wavelet families for electromyography signal classific...A comparative study of wavelet families for electromyography signal classific...
A comparative study of wavelet families for electromyography signal classific...journalBEEI
 
A portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacA portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacArhamSheikh1
 
Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Alexander Decker
 
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Hauke Sann
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptopijics
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsIJERA Editor
 
Identification of Myocardial Infarction from Multi-Lead ECG signal
Identification of Myocardial Infarction from Multi-Lead ECG signalIdentification of Myocardial Infarction from Multi-Lead ECG signal
Identification of Myocardial Infarction from Multi-Lead ECG signalIJERA Editor
 
Eeg importance and challenges
Eeg importance and challengesEeg importance and challenges
Eeg importance and challengesHoopeer Hoopeer
 
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...IJECEIAES
 
ecsa-1_2423_manuscript
ecsa-1_2423_manuscriptecsa-1_2423_manuscript
ecsa-1_2423_manuscriptLucas Mendes
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - FullEmeel Gayed
 
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemRight-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemJewel Haque
 
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMDENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMIJEEE
 
Ijaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xIjaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xsalborunda
 
ECG Signal Analysis for Myocardial Infarction Detection
ECG Signal Analysis for Myocardial Infarction DetectionECG Signal Analysis for Myocardial Infarction Detection
ECG Signal Analysis for Myocardial Infarction DetectionUzair Akbar
 

Similaire à Krammer P. et al.: Electrical impedance tomography Simulator. (20)

White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB) White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
White Paper on novel EIT interfaces, English version (PDF, 1.2 MB)
 
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...
Saenz Cogollo et al. - 2011 - A new integrated system combining atomic force ...
 
Seminar
SeminarSeminar
Seminar
 
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
Braun F. et al.: Comparing belt positions for monitoring the descending aorta...
 
A comparative study of wavelet families for electromyography signal classific...
A comparative study of wavelet families for electromyography signal classific...A comparative study of wavelet families for electromyography signal classific...
A comparative study of wavelet families for electromyography signal classific...
 
A portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiacA portable electrocardiogram for real‑time monitoring of cardiac
A portable electrocardiogram for real‑time monitoring of cardiac
 
EMG final report
EMG final reportEMG final report
EMG final report
 
Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...Switch matrix logic implementation of electrical impedance tomography measure...
Switch matrix logic implementation of electrical impedance tomography measure...
 
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
Waldmann A. et al.: Position-dependent distribution of lung Ventilation – A f...
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart Ailments
 
Identification of Myocardial Infarction from Multi-Lead ECG signal
Identification of Myocardial Infarction from Multi-Lead ECG signalIdentification of Myocardial Infarction from Multi-Lead ECG signal
Identification of Myocardial Infarction from Multi-Lead ECG signal
 
Eeg importance and challenges
Eeg importance and challengesEeg importance and challenges
Eeg importance and challenges
 
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...
CMOS Temperature Sensor with Programmable Temperature Range for Biomedical Ap...
 
ecsa-1_2423_manuscript
ecsa-1_2423_manuscriptecsa-1_2423_manuscript
ecsa-1_2423_manuscript
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - Full
 
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemRight-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
 
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORMDENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
 
Ijaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195xIjaest volume 2-number-2pp-190-195x
Ijaest volume 2-number-2pp-190-195x
 
ECG Signal Analysis for Myocardial Infarction Detection
ECG Signal Analysis for Myocardial Infarction DetectionECG Signal Analysis for Myocardial Infarction Detection
ECG Signal Analysis for Myocardial Infarction Detection
 

Plus de Hauke Sann

Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Hauke Sann
 
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Hauke Sann
 
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...Hauke Sann
 
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Hauke Sann
 
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Hauke Sann
 
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...Hauke Sann
 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Hauke Sann
 
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hauke Sann
 
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)Hauke Sann
 
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Hauke Sann
 
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Hauke Sann
 
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Hauke Sann
 
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Hauke Sann
 

Plus de Hauke Sann (19)

Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
Wolf G.K. et al.: Mechanical Ventilation Guided by Electrical Impedance Tomog...
 
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
Proença M. et al.: Influence of heart motion on EIT-based stroke volume estim...
 
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
März A. et al.:Detektion ventilatorischer Umverteilung durch Lageänderung mit...
 
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.Roka P.L. et al.: Software tool for analysing ventilation EIT data.
Roka P.L. et al.: Software tool for analysing ventilation EIT data.
 
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
 
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...
McCall K.E. et al.: Using EIT to determine the role of gestational age on lun...
 
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
Grychtol B. et al.: 3D EIT image reconstruction with GREIT.
 
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
Hager C. et al.: Einsatz der elektrischen Impedanztomograpie im klinischen Al...
 
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
White Paper on Swisstom BB2 Silent Spaces, English version (PDF, 1.7 MB)
 
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
Short instruction Swisstom BB2, EIT, Portuguese version (PDF, 651 KB)
 
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
Short instruction Swisstom BB2, EIT, French version (PDF, 649 KB)
 
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Japanese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
Product brochure Swisstom BB2, French version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Chinese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Italian version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Finnish version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Danish version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Portuguese version (PDF, 1.8 MB)
 
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
Product brochure Swisstom BB2, Spanish version (PDF, 1.8 MB)
 

Dernier

Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRupali Sharma
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...mahaiklolahd
 
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Sheetaleventcompany
 
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...Sheetaleventcompany
 
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsEscorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsAhmedabad Call Girls
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreDeny Daniel
 
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real Meet
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real MeetVip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real Meet
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real MeetAhmedabad Call Girls
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Sheetaleventcompany
 
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetvadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...Escorts In Kolkata
 
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort ServiceSexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Servicejaanseema653
 
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Sheetaleventcompany
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...Ahmedabad Call Girls
 
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetsurat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAhmedabad Call Girls
 
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...Joya Singh
 
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...dilpreetentertainmen
 
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhCall Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhSheetaleventcompany
 
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...Sheetaleventcompany
 

Dernier (20)

Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
 
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
Independent Call Girls Service Chandigarh Sector 17 | 8868886958 | Call Girl ...
 
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...
Low Rate Call Girls Pune {9xx000xx09} ❤️VVIP NISHA Call Girls in Pune Maharas...
 
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsEscorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
 
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real Meet
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real MeetVip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real Meet
Vip Call Girls Makarba 👙 6367187148 👙 Genuine WhatsApp Number for Real Meet
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
 
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetvadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
vadodara Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...
Ludhiana Call Girls Service Just Call 6367187148 Top Class Call Girl Service ...
 
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort ServiceSexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
 
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetsurat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
 
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
 
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...
❤️Ludhiana Call Girls ☎️98157-77685☎️ Call Girl service in Ludhiana☎️Ludhiana...
 
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service ChandigarhCall Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
Call Now ☎ 8868886958 || Call Girls in Chandigarh Escort Service Chandigarh
 
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...
Call Girls In Indore 📞9235973566📞Just Call Inaaya📲 Call Girls Service In Indo...
 

Krammer P. et al.: Electrical impedance tomography Simulator.

  • 1. Electrical impedance tomography simulator Krammer P, Waldmann AD, Zogg M, Róka PL, Brunner JX , Bohm SH; 16th International Conference on Biomedical Applications of Electrical Impedance Tomography, Neuchâtel Switzerland, June 2-5, 2015
  • 2. 40 SESSION 5. HARDWARE I Electrical impedance tomography simulator Péter Krammer1 , Andreas D. Waldmann1 , Michel Zogg1 , Péter L. Róka1,2 , Josef X. Brunner1 and Stephan H. Bohm1 1 Swisstom AG, Landquart, Switzerland, pkr@swisstom.com 2 Budapest University of Technology and Economics, Budapest, Hungary Abstract: Assessing the performance of electrical impedance tomography (EIT) devices usually requires a phantom for validation, calibration or comparative measurements. To bridge the gap between the typical data obtained in patients wearing a chest belt ventilation- and cardiac- related impedance changes should also be simulated either manually or automatically. In this abstract we describe a novel EIT simulator based on a dynamic controllable resistor mesh. 1 Introduction Electrical impedance tomography is a novel method for monitoring lung and heart providing moving real-time images of function rather than structures. For this purpose, 16 or 32 electrodes are typically placed around the human thorax. Weak alternating currents are applied via two electrodes and the resulting potentials are measured via the remaining electrodes. From the measured voltages, images are calculated which show the distribution of electrical impedance within the body [1]. Intra-thoracic impedance changes with ventilation [2] and within each cardiac cycle [3]. Although some groups have proposed resistor mesh phantoms to test and present EIT devices an EIT simulator allowing the user to dynamically adjust impedance changes is still missing. 2 Methods Therefore, we developed an EIT simulator consisting of a resistor phantom with 160 resistors, details can be found in [4], [5], a microcontroller (ATmega64), to program various simulation scenarios, a 10 bit digital to analogue converter (MAX5723) and 7 field-effect transistors (IRLL2705) to induce impedance changes. Each field- effect transistor (FET) is located parallel to one of the 160 resistor and induces an impedance change proportional to the applied voltage at its gate. 6 FETs are placed in such a way that they produce impedance changes within the demarcated expected lung contours of the EIT image and one in the heart region. Breath and heart rate as well as their respective amplitudes can be controlled by four Figure 1: Components of EIT Simulator knobs. Furthermore, the EIT simulator is equipped with a pressure sensor (MPX5010) to input airway pressures from the ventilator circuit. In this case the FETs can be driven according to the measured pressure signal. Failing electrodes can also be simulated simply by pressing one of the switches, see Figure 1. First measurements were performed with the Swisstom BB2 EIT device. Impedance changes were induced in both, the lung region only and in the lung and heart regions. Power images were generated to describe the impedance distribution while the global impedance curves were plotted to show the signals along the time. 3 Results Representative sample results are presented in Figure 2. In A the amplitude of only the lung signals can be seen, in B the cardiac related impedance change and the lung signal is shown and C depicts a measurement on a healthy volunteer. D depicts the simulated and measured ventilation and cardiac related impedance changes. Figure 2: Simulation results: A amplitudes of only the lung signal, B amplitudes of lung and heart signal, C amplitudes of a healthy volunteer, D time signals for case A, B and C. 4 Conclusions Both, EIT images as well as global impedance curves proved representative of what can be obtained from living beings. Compared to other resistor mesh phantoms our system can also simulate dynamic impedance changes at different anatomy-related locations within the resistor mesh phantom. The novel EIT simulator is used for performance testing, comparative measurements and for demonstration purposes. References [1] Costa EL, Curr.Opin.Crit.Care, vol 15, pp. 18-24, 2009 [2] Wolf GK, Critical Care Medicine, vol. 41, pp. 509-515, 2013 [3] Frerichs I, IEEE Trans. Med. Imaging, vol. 21 pp. 646-52, 2002 [4] Gagnon H, IEEE Trans. Bio Med. Eng, vol. 57, pp. 2257-2266, 2010 [5] Hahn G, Physiol. Meas, vol. 29, pp. 163-72, 2008 A D B B D A C A: Control of heart and ventilation B: resistor mesh phantom C: Connection to EIT device D: External signal input E: Failing electrode control F: Modus selection control E F C
  • 3. SWI SSTO M SCIE NT IFIC LIBRARY Content:Dr.StephanBöhm;Concept&Design:ZweizeitBrandDevelopment Made in Switzerland electrical impedance tomography Real-time tomographic images for organ function monitoring and diagnosis Contact us! call: + 41 (0) 81 330 09 72 mail: info@swisstom.com visit: www.swisstom.com Swisstom AG Schulstrasse 1, CH-7302 Landquart, Switzerland Swisstom AG Swisstom AG, located in Landquart, Switzerland, develops and manufactures innovative medical devices. Our new lung function monitor enables life-saving treatments for patients in intensive care and during general anesthesia. Unlike traditional tomography, Swisstom´s bedside imaging is based on non-radiating principles: Electrical Impedance Tomography (EIT). To date, no comparable devices can show such regional organ function continuously and in real-time at the patient’s bedside. Swisstom creates its competitive edge by passionate leadership in non-invasive tomography with the goal to improve individual lives and therapies. © Swisstom AG