This document summarizes a presentation given by Dr. Janine Mauzeroll on scanning electrochemical microscopy (SECM). SECM is introduced, including operating modes and principles. Applications of SECM in studying multdrug resistance in cancer cells, electrochemical properties of battery materials, and corrosion of alloys are discussed. SECM allows visualization of heterogeneous electron transfer kinetics and mass transport at micro and nanoscale.
The Wonderful World of Scanning Electrochemical Microscopy (SECM)
1. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Janine Mauzeroll, PhD
Professor of Chemistry
Department of Chemistry
McGill University
The Wonderful World of
Scanning Electrochemical
Microscopy (SECM)
2. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
The Wonderful World of
Scanning Electrochemical
Microscopy (SECM)
Dr. Janine Mauzeroll discusses the
fundamentals, critical experimental
parameters and recent applications for
Scanning Electrochemical Microscopy (SECM).
3. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Janine Mauzeroll, PhD
Professor of Chemistry
Department of Chemistry
McGill University
The Wonderful World of
Scanning Electrochemical
Microscopy (SECM)
Copyright 2021 J. Mauzeroll and InsideScientific. All Rights Reserved.
4. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Piled Higher and Deeper by Jorge Cham www.phdcomics.com
title: "Thesis writing" - originally published 11/5/1999
4
Strange what also applies to talks…..
7. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Bulk Diffusion
Feedback mode
(Insulating surface)
Feedback mode
(Conductive surface)
Substrate generation –
tip collection
Tip generation –
substrate collection
Redox competition Direct mode Potentiometric
mode
7
Janine Mauzeroll
Modes of SECM
8. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
x
z
y
8
Janine Mauzeroll
Scanning Electrochemical Microscope
9. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
• The rate of electrochemical reactions (v) is monitored through
currents measured at the microelectrode (i).
• We can study the substrate’s electrochemical reactivity in this way.
9
Janine Mauzeroll
SECM Principle
10. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
• In electrochemistry, we control ΔGRXN using Eapplied
Product
Reactant
Oxidation A A+ + e
Transition
State
10
Janine Mauzeroll
Controlling Charge Transfer
11. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
v = Rate of electrochemical reactions
i = Microelectrode currents
n = # of electrons
F = Faraday’s constant
A = electrode area
1) Rate of electron transfer
2) Rate of mass transport
11
Janine Mauzeroll
What Do I Need to Know to Quantify the Current?
12. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
F = Faraday’s constant
A = Electrode area
k0 = Standard rate constant
CO(0,t) ; CR(0,t) = Surface concentrations of species O,R
α = Transfer coefficient
E – E0’ = Overpotential (driving force applied)
12
Janine Mauzeroll
• Reaction kinetics at an electrode
surface follow the Butler-Volmer
relationship
Heterogeneous Electron Transfer Rate
13. Laboratory for Electrochemical Reactive
Imaging of Biological Systems 13
Janine Mauzeroll
Mass Transport in the Electrolyte
14. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Mass transport in electrolyte
Electrode kinetics
Butler-Volmer:
First order kinetics:
Concentration boundaries
14
Janine Mauzeroll
Finite Element Modeling is Required in SECM
15. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Normalized Electrode current is a function of L, Rg, κ
κ
C. Lefrou and R. Cornut, ChemPhysChem, 11, 547 (2010)
Probe Approach Curve Analytical Approximations
16. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Positive FB
Negative FB
16
Janine Mauzeroll
Tracking Substrate Reactivity using SECM
17. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Red Ox
L (tip-substrate distance)
NiT (tip current)
e-
e-
17
Janine Mauzeroll
Substrates with Finite Kinetics
18. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Classical Probes
Multifunctional Probes
Danis, L.; Mauzeroll, J. et. al.
Anal. Chem. 2015. p 2565
Danis, L.; Mauzeroll, J. et. al.
Anal. Chem. 2015. p 2565
Katemann, B. B.; Schuhmann W.
ElectroAnalysis 2002. p 22
Walsh, D. A.; Bard, A. J.; et. al.
Anal. Chem. 2005. p 5182
19. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Polcari, D.; Dauphin-Ducharme, P.; Mauzeroll J., Chem. Rev., 116, 13234 (2016)
Application Fields of SECM
20. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Cancer Cells
Batteries
Corrosion
20
Janine Mauzeroll
Three Short Stories
21. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Expression and Functional Activity of Multidrug
Resistance-Associated Protein 1 using SECM
Polcari, D.; Mauzeroll, J. et al., Anal. Chem., 89, 8988 (2017).
Kuss, S.; Mauzeroll, J. et al., Anal. Chem., 87, 8096 (2015).
Kuss, S.; Mauzeroll, J. et al., Anal. Chem., 87, 8102 (2015).
Kuss1, Polcari1 et al., PNAS 110, 9249 (2013)
22. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Drug-Sensitive Cancer Cell
(Ex: HeLa Cells)
Multidrug Resistant Cancer Cell
(Ex: HeLa-R Cells)
22
Janine Mauzeroll
Multidrug Resistance (MDR)
23. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Drug-Sensitive Cancer Cell
(Ex: HeLa Cells)
MDR Cancer Cell
(Ex: HeLa-R Cells)
0.05 µM 0.10 µM
Untreated 0.05 µM 0.10 µM
Untreated
Drug Challenge
Janine Mauzeroll 25
25. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
No Interaction Negative Feedback
Diffusion into Cell Positive Feedback
Measuring MRP1 Activity
Janine Mauzeroll 25
26. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
• Extract tip to substrate distance, L,
at each image pixel
• Then extract heterogeneous kinetics , κ,
at each image pixel
Two Step Process Convolution
Janine Mauzeroll 26
27. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
(cm
s
-1
)
Effect of Pattern Size
Polcari, D.; Mauzeroll, J. et al., Anal. Chem., 89, 8988 (2017)
28. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
(cm
s
-1
)
Time-Lapse Imaging
Polcari, D.; Mauzeroll, J. et al., Anal. Chem., 89, 8988 (2017)
29. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Polcari, D.; Mauzeroll, J. et al., Anal. Chem., 89, 8988 (2017)
Activity of Six Different Cell Populations
30. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Cancer Cells
Batteries
Corrosion
30
Janine Mauzeroll
Three Short Stories
31. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Localized Investigations of the Electrochemical
Properties of Lithium Battery Materials Using
Micropipette
Dayeh, M.; Snowden, M. E.; Ghavidel, M.; Payne, N.;
Gervais, S.; Mauzeroll, J.; Schougaard, S. B.
Journal of Power Sources 2016, 325, 682-689.
ChemElectroChem, 2019, 6, 195–201
Analytical Chemistry 2019, 91(24), 15718-15725
Analytical Chemistry 2020, 92, 10908-10912.
32. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Malak Dayeh
1 μm 1 μm
1 μm
Are all battery particles created equal?
32
33. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
20 µm
Single particle battery
Measuring Isolated Active Particles
33
34. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2
-400
-200
0
200
400
600
i
(pA)
E (V) vs. Li/Li+
distance (μm)
i
(μA)
distance (μm)
i
(μA)
Approach
Land
Retract
Measure
Scanning Micropipette Contact Method
34
35. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2
-400
-200
0
200
400
600
i
(pA)
E (V) vs. Li/Li+
distance (μm)
i
(μA)
distance (μm)
i
(μA)
Approach
Land
Retract
Measure
Scanning Micropipette Contact Method
35
36. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Malak Dayeh
- Very low water and oxygen content (~ 1 ppm)
- The only gas present is Argon
Electrode
Pipette
Substrate
Development of SMCM in
Anaerobic Conditions
36
37. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Malak Dayeh
2μm
Journal of Power Sources 2016, 325, 682–689.
LiFePO4 Li+ + e- + FePO4
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2
-20
-10
0
10
20 1 mV/s
5 mV/s
10 mV/s
20 mV/s
50 mV/s
Potential / V (vs. Li/Li+
)
Current
/
p
A
Epf
ipf
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2
-600
-400
-200
0
200
400
600
Current
/
p
A
Potential / V (vs. Li/Li+
)
SMCM on Isolated LiFePO4 Active Particles
38. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Electrochemistry and Electron Microscopy
Malak Dayeh
0 1 2 3 4 5 6 7
0
10
20
30
40
Particle Area (µm2
)
i
pf
(pA)
0.0
0.4
0.8
1.2
1.6
2.0
Q
f
(nC)
- Particle cross section areas obtained from SEM are compared to the forward (oxidation) peak
current and the integrated charge.
- A larger cross section area correlates to both a higher peak current and increased charge.
LiFePO4 Li+ + e- + FePO4
Journal of Power Sources 2016, 325, 682–689.
38
39. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
2 µm
i
2 µm
iii
2 µm
ii
Malak Dayeh
6%
27%
67%
0 100
Shifted
peak
Double
peak
Single
peak
Differences in
electrochemical
performances
Journal of Power Sources 2016, 325, 682–689.
Particles’ Heterogeneities
40. Laboratory for Electrochemical Reactive
Imaging of Biological Systems 40
Malak Dayeh
Summary
SMCM technique is:
Valuable tool for probing the localized electroactivity of
battery active material
Capable of detecting heterogeneity in material properties
distribution
❌Considered as a quality control tool for active material
fabrication, and monitoring batch-to-batch variations in
particle properties
� Particle volume still evaluated by external means
41. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Malak Dayeh
Future Direction: Two Diffusion Regimes
Journal of Power Sources 2010, 195(24), 7904-7929.
Li+
LFP
1
2
LiFePO4 Li+ + e- + FePO4
1: Transport of Li+ ions in electrolyte
2: Diffusion of Li+ ions within LFP particle
42. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Use SICM to measure porous film conductivity
Analytical Chemistry 2019, 91(24), 15718-15725
43. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
A composite cathode is a porous film
Analytical Chemistry 2019, 91(24), 15718-15725
44. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
SICM approach curve= film conductivity
Analytical Chemistry 2019, 91(24), 15718-15725
45. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
From Macro to Micro: Using Electrochemical
Methods to Investigate the Effect of Alloy
Chemistry on Corrosion
Gateman, S.M.; Stephens, L.I.; Perry, S.C.; Lacasse, R.; Schulz, R.;
Mauzeroll, J.
Nature Materials Degradation 2018, 2, 1-8.
Nature Materials Degradation 2019, 3, 25.
46. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
1. Stainless Steels have inclusions promoting corrosion
Janine Mauzeroll
Elemental Composition of SS 444
Fe Cr Mo Si Ni Nb Mn V Cu Ti
Supplier’s
Claim
79.9 17.07 1.89 0.17 0.24 0.307 0.28 0.1
1
0.063 0.15
Experimental
ICP-OES
79.54 17.3 1.93 -- 0.25 -- 0.3 0.1
1
0.08 --
Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
47. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Extracting Kinetic Rate Constants Using FEM
Janine Mauzeroll
2.85 x 10-2 cm/s
0.2 x 10-2 cm/s
Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
48. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
SS 444 is extremely corrosion resistant according to ASTM
standard PDP measurements, but is vulnerable to
localized corrosion.
Gathered experimental and theoretical evidence of a
microgalvanic coupling effect between the Ti/Nb rich
inclusions/precipitates and the surrounding metal matrix.
1. Conclusions
Janine Mauzeroll
1 cm
Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
49. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
2. HVOF Thermal Spray Coatings Corrosion?
Janine Mauzeroll
T= ~2000 K
100 µm
100 µm
Struers, 2005, Metallographic
preparation of thermal spray coatings
50. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Tafel Analysis of Coatings: Active Corrosion
All tests were performed in 3.5 wt% NaCl at a scan rate of 0.167 mV/s. All coatings are
tested as received without the use of grinding/ polishing.
Janine Mauzeroll
Scan direction
Ecorr Anodic
Cathodic
Anodic
Cathodic
Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
51. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Scanning Electrochemical Microscopy: Feedback Mode
Insulating substrate = Topography
Conductive Substrate = Topography + Reactivity
Insulator
Conductor
Janine Mauzeroll 61
X
52. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Minimal Passivation Detected using SECM
Janine Mauzeroll Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
53. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Micro Polarization of a Single Powder Particle
Janine Mauzeroll Gateman, S. M., et al., npj Mater. Degrad, 2018, 2, 5.
54. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Linked the stainless steel thermal spray coating’s weak
corrosion resistance to the precursor powder
Identified regions of vulnerability across a single powder
particle using scanning electrochemical probe methods
Highlighted the power of using macro and micro
electrochemical methods to characterize load bearing
materials
Showcased the importance of powder metallurgy in
coating technologies
2. Conclusions
Janine Mauzeroll 54
55. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll 55
Mineral oil:
Insulating
Hydrophobic
Colorless
OI-SMCM
3. Oil-Immersed Scanning Micropipette Contact Method
56. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll
Mineral Oil Reduces Background Noise
2.2 pA in air
5.2 pA in humidified cell
1.35 pA in mineral oil
Background
noise
(pA)
Time
Current
Threshold
Time
A low background noise will reduce the risk of breaking
micropipette when landing.
57. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll 57
OI-SMCM Ecorr Map Reveals Microscale Heterogeneities
20 μm
Al
20 μm
Fe
20 μm
58. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll 58
Predict Galvanic Couples
610
608
5 μm
Fe-rich inclusion exhibits cathodic behavior relative to the Al
matrix area, which implies the surrounding Al will be consumed
as the anode in the galvanic couple with the inclusion.
59. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll 59
OI-SMCM Icorr Map Exhibits Microscale Corrosion Kinetics
20 μm
60. Laboratory for Electrochemical Reactive
Imaging of Biological Systems Janine Mauzeroll 60
Microgalvanic Corrosion
13
12
30
29
Galvanic couples
Oxygen reduction on inclusion surface: cathodic branch (13 and 30).
Surrounding Al dissolution: the anodic branch (12 and 29).
Ecorr at the meeting points are more anodic relative to that of Al
matrix.
Therefore, the Al surrounding the Fe-rich inclusions is more
susceptible to corrosion.
61. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
3. Conclusions
Janine Mauzeroll 61
Oil-Immersed SMCM:
Allows for the use of highly evaporative electrolyte
solutions
Long-time stability for a large map
Expands the application of SMCM
63. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
Corrosion
Cancer Cells
Batteries
Spectroelectrochemistry
Catalysis
Biofilms
Enzyme Films
Polymer Films
Janine Mauzeroll
Final Thoughts…
63
65. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
BSc
(30)
Meng, B; Potts, K.; Robert, A.;
Lin M., Sifakis, J.; Gordon, J.
Chen, Y. ; Gateman, S.
Langlois-T., T; .Sangji, H.
Boudreau, C.; Vassileva, I.
Wei, X.; Mack, T.
Salvatore, D.; Kwan, A.
Yong, K.; St-Pierre, C.
Poirier, S.; Wezel, N.
Meyrignac, P.; Fabre, D.
Joliton, A.; Gariepy, V.
Bavencove, A.; Benlounes, K.
Dansereau, D.; Tieu, J.
Noel, J.; Pierre, J.
MSc/PhD
(18)
Dawkins, J.; Li, Y; Pan,
Y.; Moussa, S.
Odette, W.; Dayeh, M.
Danis, A.; Payne, N.
Mazurkiewicz, S.
Polcari, D;
Dauphin D., P.
Danis, L.; Kuss, S.
Mezour, A.;
Mayoral, M.
Beaulieu, I; Correia, L
Lukova, N.
PDF
(8)
Ghavidel, R.
Perry, S.
Noyhouzer, T.
Kuss, C
Snowden, M.
Mengesha , U.
Thrin, D.
Cornut, R.
Collaborators
(15)
S. Canesi (UQAM)
B Kraatz (UofT)
I. Halalay (GM)
E. Ruthazer, (McGill)
H. Sleiman, (McGill)
M. Geissler (NRC)
D. Bélanger (UQÀM)
M. Lafleur (UdeM)
M. Morin (UQÀM),
S. Schougaard (UQÀM)
D. Shoesmith, (UWO)
G. Botton, (McMaster)
R. Lacasse, (HQ)
R. Schulz, (HQ)
C. Heineman (HEKA)
66. Laboratory for Electrochemical Reactive
Imaging of Biological Systems
To Companies, Funding Agencies & Universities
$$ $$
Discovery, CRD
Strategic, Create &
UFA
Nouveau Chercheur
& Équipes
$$
68. Thanks for participating!
• Watch the webinar here:
The Wonderful World of Scanning
Electrochemical Microscopy
• Want to learn more about the technology?
Visit: www.elproscan.com
Notes de l'éditeur
In the SECM experiment, the tip and sample (substrate) are immersed in a solution containing electrolyte, and an electroactive species (e.g. substance O at concentration CO* and with diffusion coefficient DO). The cell also contains auxiliary and reference electrodes.
Steady state
Conductor/insulator and implications for negative/positive feedback
Fixed tip-substrate distance
Local substrate reactivity
Not ideal but blurred
Multiple probes, one device. Reduced time and error.
Tandem techniques: SECM-AFM, SECM-SICM.
Broad variety of probe geometries.
Do expression and activity correlate?
Patterning facilitates microelectrode positioning and reduces experiment time
Allows for reproducible measurements
Fc induces an increase in GSH
Blebbing and apoptosis
Due to the inclusion’s square geometry, a 3D model was created and implemented to extract more precise values for local rate constants over the conductive inclusions and the passive film.
Comparing to pure negative feedback, the values reported over the passive film show difficulty the passive film has to regenerate the redox medaitor.
When approaching an inclusion however, the PAC is very similar to the current obtained over a pure conductive substrate.
This experiment tells us that our passive film is discontinuous over the inclusions and is therefore not protecting this vulnerable sites.
Important to extract local rate constants to compare values of that and the thermal spray coatings in later studies to come.
thermal spraying is a process in which molten, semi-molten or solid particles are deposited on a substrate. Consequently, the spraying technique is a way of generating a ‘stream’ of such particles. Coatings can be generated if the particles can plastically deform at impact with the substrate, which may only happen if they are molten or solid and sufficiently rapid. Their heating and/or acceleration are practical if they occur in a stream of gas.1 Thus, an academic classification of spray techniques is based on the way of generation of such streams.
Left SE-SEM image: shows gradient of porosity within the coatings
right BSE-SEM image: shows the hetergeneity of the coatings with dark oxide strings and lighter metallic particles.
Cooling is very fast in comparison to typical cooling rates
In High Velocity Oxy-Fuel Combustion spraying (HVOF) fuel gas and oxygen are fed into a chamber in which combustion produces a supersonic flame, which is forced down a nozzle increasing its veloc- ity. Powder of coating material is fed into this stream and the extreme velocity of the particles when hitting the substrate creates a very dense, strong coating (Fig. 5).
LEAD Question: The very high kinetic energy of the particles when striking the substrate ensures an adequate mechanical bond even without the particles being fully molten.
Hardness vs. bulk substrate
Hardness increases due to oxidization of the metal
Good wear applications
Low porosity
Minimal alteration of mechanical properties of substrate
**label cathodic and anodic branches**
When comparing the Tafel plots of the bulk vs. a coating, however, the coatings are much more active with a more negative Ecorr (~-470 mV vs. SCE) and show very different kinetic limitations in their cathodic/anodic branches.
The coatings seem to actively corrode, indicating the degradation mechanism of uniform corrosion.
We can recall that mechanical failure of components due to localized corrosion is catastrophic because the failure is usually rapid and unexpected. Uniform corrosion, in which the surface gradually degrades, is preferred because it is simpler to detect, predict, and often, to control.
We are interested as to why thermal spray coatings undergo a different degradation mechanism in comparison to the bulk material used to make the coatings. Undergo a significant change in thermal oxidization and forming Fe/Cr oxides along the surface, but also the material is in powder form.
Is this due to the HVOF thermal spray process or from powder fabrication processes?
3 electrode setup: with working electrode’s position controlled using 3D piezoelectric motors.
When approach the surface of the substrate, and in the presence of a redox mediator, information about local reactivity can be found.
Insulator: cannot regenerate the mediator, only information about topography
Conductor: regenerates the mediator, creates feedback loop. Topography and reactivity
The potential was scanned from -2 V to 0 V or until droplet instability (loss of surface tension) was observed via a loss of electrical connection at a scan rate of 100 mV/s. The micropipette was then retracted from the substrate’s surface by 10 μm, leaving the droplet of electrolyte residue from the previous measurement behind. A fresh micro droplet at the end of the tip is then exposed and moved to the next designated area where it is approached towards the surface at a rate of 1 μm/s until a spike in the WE’s current was measured.
The micro electrochemical behavior of the pure Mg and Al were first tested using SMCM. A minimum of five PDP measurements were collected at different points on each metal far away (>100 μm) from the intermetallic regions.
Solvent is ethylene glycol
The Mg-Al diffusion couple’s chemical composition was investigated using EPMA Back-Scattered Electron (BSE) image and WDS line scanning measurements (Figure 3). The interface between the two metals was imaged, where the darker part is lighter metal, Mg (right) and the brighter is heavier metal, Al (left). Rather than a smooth transition between Mg and Al, WDS line scans across the Mg-Al interface expose distinct intermetallic layers.
The diffusion depth of Al in the bulk Mg is only about 50 μm. The concentration of Al increases linearly from 0 to about 10 wt.%, which is the solubility limit of Al in Mg at 673 K.39 Two intermetallic phases, Mg17Al12 and Mg2Al3, are observed in 87 and 286 m thickness, respectively. As expected from the phase diagram, Al concentration in Mg17Al12 layer is progressively increasing with distance, while Al concentration in Mg2Al3 layer is nearly constant. Mg is diffused to bulk Al up to about 100 m depth.
Although all experiments were performed with caution to prevent noise, fluctuations in the pico range current were still present due to instrumental noise. All data was smoothed using a 100-point moving boxcar average (Figure 4A). All experiments were performed within the time limit of 40 min, i.e., the amount of time Ag/AgCl QRCEs have been reported to be stable for in 0.1 M LiCl electrolyte.42
Add the picture of david to the acknowldgemens as well as sam perry
In fact, this award is really a reflection of my team’s hard work.
I want to thank them all.
I am lucky to work with such creative, dynamic and politically incorrect people. We work hard, we have a lot of fun together and I could not ask for a better team.
I also acknowledge my precious collaborators and their students. We learn so much from their expertise. So we thanks them for this continued scientific education.
I want to acknowledge the companies, Funding Agencies and Universities that have provided the generous financial support to our team.
Without their sustained support, none of the infrastructure, operation funds, student and postdoctoral fellowships would be possible.
Without their sustained support none of our work would come to fuition.
But before I delve into the science, I would like to start off with a few acknowledgments.