FAULTS IN X-RAY TUBE
AND ITS CARE
PRESENTED BY
Saru Gosain
BSc.MIT 2nd year
(2017)
COMMON CAUSES OF
TUBE FAILURE
• Extremely high voltage
• Excessive heat generation
• Poor cooling system
• Rough-handling/ Careless
• Aging
COMMON FAULTS, THEIR
CAUSES & REMEDY
A. FAULTS IN X-RAY TUBE
HOUSING
1. Cracking of x-ray tube
housing
 Escape of oil from the housing
 Insulation level &Cooling rate
are reduced
Causes
• Expansion of air & oil present
inside it and increased in its
internal pressure.
Remedy
• Excluding air
• Use of metal bellows to allow oil
expansion as well as to operate
as “microswitch”
Leakage oil testing devices
Cont.…
2. Radiation leakage
 Increases patient’s as well as
personnel radiation dose.
 Should not exceed 100 mR/hr
at 1m
Causes
• Cracking of the tube housing
• Improper shielding
Remedy
• Proper shielding should be
done .
• Replace the tube with new
one as soon as possible.
Cont.…
B. FAULTS IN GLASS/METAL
ENVELOPE
1. Loss of vacuum
 Production of secondary electrons
 Result in the variation in the number,
reduced speed of electrons impinging
on the target, energy of x-ray
produced
 Production couldn’t be controlled
independently.
Causes
• Crazing of the glass enclosure due to
difference in coefficients of expansion
of glass and metal
• Inactivity
Cont.…
Remedy
 Use of special alloys having the
same coefficients of linear
expansion as Pyrex glass are used.
C. FAULTS IN FILAMENT
1. Vaporization of filament
a) Breakage of filament
 No exposure takes place
Causes
• Excessive heating of filament for
the longer period of time
Cont.…
Remedy
• Use of automatic filament-boosting
circuit
• Use of 1-2 % Thorium with tungsten
filament
b) Tube arcing
 Produces a bronzed
-colored “sunburn” color
 Filters the x-ray beam
 “Crazing” of the tube
Causes
• Evaporation of the filament
Remedy
• Use of metal enclosure instead of
glass enclosure
Cont.…
D. FAULTS IN ANODE
1. Kinking of the target material
 Results in distortion of the anode angle, narrowing of the
beam, greater geometric unsharpness.
Causes
• Difference in thermal expansion of metal on the surface
& metal immediately beneath the surface.
Remedy
• 10% Rhenium is added to 90% of tungsten target
• Manufacture anode disc with slits in the target surface to
allow the material in the focal track to expand without
producing the mechanical tension.
Cont….
2. Roughening and pitting of anode surface
 Scattering of the x-rays
 Absorption of x-rays by the
target itself.
 Results in diminish x-ray output
Cause
• Excessive thermal stress
Remedy
• Use of lowest necessary power
• Largest possible focal spot
• Longer exposures at reduced power
For
FIGURE 6-14 Comparison of smooth, shiny appearances of rotating anodes when new (A) versus
their appearance after failure
(B–D). Examples of anode separation and surface melting shown were caused by slow rotation
caused by bearing damage (B),
repeated overload (C), and exceeding of maximum heat storage capacity (D). (Courtesy Philips
Medical Systems.)
A B
C D
Cont….
3. Imbalance ball bearings
 Jamming
 Increased bearing noise
 Anode stops rotating
Causes
• Expansion due to excessive
heating of the rotor bearings
• Increased rotational friction
Remedy
• Poor conductor of heat(Mo) is
used as anode stem
• Metallic Lubricants such as Silver
& liquid Gallium are used.
Cont…
4. Broken stator windings
 No power supply to make the
anode rotate.
Causes
• Heavy radiography exposures
• Excessive heat in the target
material, rotor and its bearings,
• Difference in thermal expansion of
the various components used in the
anode.
Remedy
• Provide safety circuit to terminate
exposure.
OTHER FAULTS
A. DEFICIENCIES IN MANUFACTURING
1. Unlike tubes
2. Weed out by test:- Check high voltage stability, cathode
emission, focal spot size, thermal loading and for
rotating anode tubes additional tests such as noise,
vibration, coast time etc. are tested. Not meeting these
specifications are rejected.
3. Hold period:- If tubes are held for 2-4 weeks vacuum
may leak
4. Improper materials:- O2 free Cu bar can cause vacuum
leaks
Cont….
B. APPLICATION MISMATCH:- e.g., poor diagnosis and
radiation burns while using standard diagnostic tubes to
produce mammograms
C. IMPROPER DRIVE BY THE POWER SUPPLY:-
1. Filament life is reduced 2-10 times by DC filament than
AC filament.
2. Magnetic materials ( e.g., iron, nickel, cobalt used for
glass-to-metal seal) can be subjected to magnetic
hysteresis, eddy currents etc.)
3. Filament limit:- maximum output current of the filament
is limited
CARE OF THE X-RAY
TUBE
1. WARM UP THE ANODE FOLLOWING MANUFACTURER’S
RECOMMENDATIONS
 Performed if the system not been energized for ≥ 2 hrs
 To prevent tube damage from thermal shock
WARM UP PROCEDURES
Make 1st 3-exposures waiting approximately 30s between
each exposures @
75kVp 200mA 1s
Wait 60s, then take 1 exposure @
85kVp 200mA ½ s
Wait 60s, then take the last exposure @
100kVp 200mA ¼ s
Cont….
2. MINIMIZE FILAMENT BOOST(“PREP”) TIME
 Current applied for too long will shorten filament life and
will lead to unstable operation.
Cont…
3. USE LOWER TUBE CURRENT (mA)
 High tube current will cause evaporation of the tungsten
 Whenever possible, use a lower mA station and a longer
exposure time to arrive at the desired mAs.
Cont….
4. FOLLOW RATING CHARTS AND ANODE AND
HOUSING HEATING/COOLING CHARTS
 Operation beyond published ratings will result in
premature focal track wear and tear.
TUBE RATING CHART
o The chart which provides the information about the
details of the electrical loads which may be applied to the
tube without damaging it & the safe duration of such
loads is called tube rating chart.
Cont….
o In considering the tube rating three characteristics are
encountered:
 Ability of x-ray tube to withstand a single exposure
 Ability of x-ray tube to function despite rapid exposure
 Ability of x-ray tube to withstand multiple exposure during
several hours of heavy use.
o There are 3 types
A. RADIOGRAPHIC RATING CHARTS
B. ANODE HEATING/COOLING CHART
C. HOUSING HEATING/COOLING CHART
RADIOGRAPHIC
RATING CHART
• The safe limit within
which x-ray tubes can be
operated for a single
exposure can be easily
determined by it.
• It is the most imp
because it conveys
which radiographic
techniques are safe or
unsafe for tube
operation.
• Microprocessors are
used to allow only safe
exposures.
Rating chart of x-ray room
no.1
Rating chart of x-ray room
no.1
ANODE
HEATING/COOLING
CHART
• Determines the
maximum heat
capacity of the
anode.
• Determines the
length of time the
tube must be
allowed to cool
before additional
exposures can be
made.
HOUSING
HEATING/COOLING
CHART
• Determines the
maximum heat
capacity(1-1.5
million HU) of the
tube housing.
• Complete cooling
after maximum
heat capacity
may take 1-2 hrs.
Cont.…
• Thermal rating can be determined by calculating
Energy deposited in the target(E)
=kVp × mA × exposure time in sec
• Unit is heat units(HU)
• SI unit is Joule(J)
1 HU= 0.71 J
1 J= 1.4 HU
• Now the Energy deposited in the target(E)
=kVp × mA × s J
CARE OF THE X-RAY
TUBE
5. LIMIT OPERATION TO 80% OF MAXIMUM SINGLE
EXPOSURE RATINGS
 It will minimize the reduction in radiation output
associated with a roughened focal track.
6. DO NOT EXCEED ANODE THERMAL CAPACITY OR
DISSIPATION RATE OF THE TARGET.
7. DO NOT MAKE HIGH mA EXPOSURES ON A COLD
TARGET.
8. AVOID LONG INTERVALS BETWEEN SPOT-FILMS.
o Most common “holdover period”(during which the rotor is
kept at high speed before the rotor brake cycle is
allowed to occur) is approximately 25 s between spot
films
Cont.….
9. LIMIT ROTOR START/STOP OPERATIONS
o Minimum of 30-40 s should be between starts
10.DONOT ROTATE THE TUBE HOUSING RAPIDLY
o Tension or pressure will produce to the x-ray tube.
o Gyroscopic effect( resistance to turns in any direction)
may crack or otherwise damage the rotor.
Cont.…
11.SOUND COOLING SYSTEM SHOULD BE PROVIDED
o Enormous heat(99%) is generated in the anode during
the exposure.
o Heat can be dissipated one of the 3 ways:-
RADIATION
CONDUCTION
CONVECTION.
o During exposures , heat is primarily dissipated by
radiation whereas between exposures it is primarily
dissipated by conduction.
 General life of x-ray tubes is ten of thousands of
exposures.
 And CT tubes are now guaranteed for 50,000 exposures
ANY
QUESTIONS???
Reference
• Christensen’s PHYSICS OF DIAGNOSTIC
RADIOLOGY(4TH EDITION)
• Stewart Carlyle Bushong’s RADIOLOGIC SCIENCE
FOR TECHNOLOGISTS (10TH EDITION)
• En.m.Wikipedia.org
• www.studentradiographer.com
• www.Radiopaedia.org
• www.spellmahv.com
• www.scribd.com
• www.slideshare.com
• www.sprawls.com
Faults in x ray tube and its care

Contenu connexe

Tendances

Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beamSUJAN KARKI
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...DrKajalLimbad
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,InosRagan
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictorsvishwanath0908
 
X ray generator, circuits
X  ray generator, circuitsX  ray generator, circuits
X ray generator, circuitsAnjan Dangal
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancementYashawant Yadav
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generatorsRakesh Ca
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Sarin Vidya
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettesSudil Paudyal
 
Radiographic Intensifying Screen
Radiographic Intensifying ScreenRadiographic Intensifying Screen
Radiographic Intensifying ScreenTarique Ajij
 
Image quality
Image qualityImage quality
Image qualityairwave12
 

Tendances (20)

Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beam
 
X ray generators
X ray generatorsX ray generators
X ray generators
 
Fluroscopy
FluroscopyFluroscopy
Fluroscopy
 
Filters
FiltersFilters
Filters
 
Grids
GridsGrids
Grids
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
X ray generator, circuits
X  ray generator, circuitsX  ray generator, circuits
X ray generator, circuits
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancement
 
radiographic Image quality by Astuti Mishra
 radiographic Image quality by Astuti Mishra radiographic Image quality by Astuti Mishra
radiographic Image quality by Astuti Mishra
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Production of x rays & generators
Production of x rays & generatorsProduction of x rays & generators
Production of x rays & generators
 
X-RAY GENERATORS(1).pdf
X-RAY GENERATORS(1).pdfX-RAY GENERATORS(1).pdf
X-RAY GENERATORS(1).pdf
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettes
 
Radiographic Intensifying Screen
Radiographic Intensifying ScreenRadiographic Intensifying Screen
Radiographic Intensifying Screen
 
Image quality
Image qualityImage quality
Image quality
 
Grids
GridsGrids
Grids
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 

Similaire à Faults in x ray tube and its care

Faults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarFaults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarSandeshMagar4
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............justinfan550
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screensTarun Goyal
 
7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).ppt7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).pptDavy40
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdfgametheory686sharech
 
Production of x rays
Production of x raysProduction of x rays
Production of x rayspraveen kumar
 
Conventional X-RAY Tube
 Conventional X-RAY Tube Conventional X-RAY Tube
Conventional X-RAY TubeVibhaJoshi12
 
Introduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxIntroduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxbabaaijaz01
 
Digital testing of high voltage circuit breaker
Digital testing of high voltage circuit breakerDigital testing of high voltage circuit breaker
Digital testing of high voltage circuit breakerneeraj prasad
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray TubeAlisha Karmali
 
X ray tube
X ray tubeX ray tube
X ray tubeRad Tech
 
3 cp corrosion control
3 cp corrosion control3 cp corrosion control
3 cp corrosion controlhimanshu jain
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptxGMC Anantnag
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptxGMC Anantnag
 

Similaire à Faults in x ray tube and its care (20)

Faults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magarFaults in xray tube by Sandesh magar
Faults in xray tube by Sandesh magar
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Cttubeanddetectors
Cttubeanddetectors Cttubeanddetectors
Cttubeanddetectors
 
CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............CT TUBES AND ITS TYPES.pptx...............
CT TUBES AND ITS TYPES.pptx...............
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screens
 
7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).ppt7. Faults in Xray tube (1).ppt
7. Faults in Xray tube (1).ppt
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
X-Ray Tube.pptx
X-Ray Tube.pptxX-Ray Tube.pptx
X-Ray Tube.pptx
 
X ray tube
X ray tube X ray tube
X ray tube
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Conventional X-RAY Tube
 Conventional X-RAY Tube Conventional X-RAY Tube
Conventional X-RAY Tube
 
Introduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxIntroduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptx
 
Digital testing of high voltage circuit breaker
Digital testing of high voltage circuit breakerDigital testing of high voltage circuit breaker
Digital testing of high voltage circuit breaker
 
Components of X-Ray Tube
Components of X-Ray TubeComponents of X-Ray Tube
Components of X-Ray Tube
 
X ray tube
X ray tubeX ray tube
X ray tube
 
3 cp corrosion control
3 cp corrosion control3 cp corrosion control
3 cp corrosion control
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptx
 
X-ray Tube & Production.pptx
X-ray Tube & Production.pptxX-ray Tube & Production.pptx
X-ray Tube & Production.pptx
 

Plus de SaruGosain

Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skullSaruGosain
 
Jugular venography
Jugular venographyJugular venography
Jugular venographySaruGosain
 
Frontal venography
 Frontal venography Frontal venography
Frontal venographySaruGosain
 
Abdomen & Pelvis II
Abdomen & Pelvis IIAbdomen & Pelvis II
Abdomen & Pelvis IISaruGosain
 
Abdomen & pelvis part I
Abdomen & pelvis part IAbdomen & pelvis part I
Abdomen & pelvis part ISaruGosain
 
Oral cholecystography
Oral cholecystographyOral cholecystography
Oral cholecystographySaruGosain
 

Plus de SaruGosain (7)

Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skull
 
Jugular venography
Jugular venographyJugular venography
Jugular venography
 
Frontal venography
 Frontal venography Frontal venography
Frontal venography
 
Abdomen & Pelvis II
Abdomen & Pelvis IIAbdomen & Pelvis II
Abdomen & Pelvis II
 
Abdomen & pelvis part I
Abdomen & pelvis part IAbdomen & pelvis part I
Abdomen & pelvis part I
 
Oral cholecystography
Oral cholecystographyOral cholecystography
Oral cholecystography
 
X ray filters
X ray filtersX ray filters
X ray filters
 

Dernier

Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Bangalore Call Girls Mg Road ⟟  9332606886 ⟟ Call Me For Genuine Se...Top Rated Bangalore Call Girls Mg Road ⟟  9332606886 ⟟ Call Me For Genuine Se...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine Se...narwatsonia7
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Genuine Call Girls
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 

Dernier (20)

Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Bangalore Call Girls Mg Road ⟟  9332606886 ⟟ Call Me For Genuine Se...Top Rated Bangalore Call Girls Mg Road ⟟  9332606886 ⟟ Call Me For Genuine Se...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 

Faults in x ray tube and its care

  • 1. { FAULTS IN X-RAY TUBE AND ITS CARE PRESENTED BY Saru Gosain BSc.MIT 2nd year (2017)
  • 2. COMMON CAUSES OF TUBE FAILURE • Extremely high voltage • Excessive heat generation • Poor cooling system • Rough-handling/ Careless • Aging
  • 3. COMMON FAULTS, THEIR CAUSES & REMEDY A. FAULTS IN X-RAY TUBE HOUSING 1. Cracking of x-ray tube housing  Escape of oil from the housing  Insulation level &Cooling rate are reduced Causes • Expansion of air & oil present inside it and increased in its internal pressure. Remedy • Excluding air • Use of metal bellows to allow oil expansion as well as to operate as “microswitch”
  • 5. Cont.… 2. Radiation leakage  Increases patient’s as well as personnel radiation dose.  Should not exceed 100 mR/hr at 1m Causes • Cracking of the tube housing • Improper shielding Remedy • Proper shielding should be done . • Replace the tube with new one as soon as possible.
  • 6. Cont.… B. FAULTS IN GLASS/METAL ENVELOPE 1. Loss of vacuum  Production of secondary electrons  Result in the variation in the number, reduced speed of electrons impinging on the target, energy of x-ray produced  Production couldn’t be controlled independently. Causes • Crazing of the glass enclosure due to difference in coefficients of expansion of glass and metal • Inactivity
  • 7. Cont.… Remedy  Use of special alloys having the same coefficients of linear expansion as Pyrex glass are used. C. FAULTS IN FILAMENT 1. Vaporization of filament a) Breakage of filament  No exposure takes place Causes • Excessive heating of filament for the longer period of time
  • 8. Cont.… Remedy • Use of automatic filament-boosting circuit • Use of 1-2 % Thorium with tungsten filament b) Tube arcing  Produces a bronzed -colored “sunburn” color  Filters the x-ray beam  “Crazing” of the tube Causes • Evaporation of the filament Remedy • Use of metal enclosure instead of glass enclosure
  • 9. Cont.… D. FAULTS IN ANODE 1. Kinking of the target material  Results in distortion of the anode angle, narrowing of the beam, greater geometric unsharpness. Causes • Difference in thermal expansion of metal on the surface & metal immediately beneath the surface. Remedy • 10% Rhenium is added to 90% of tungsten target • Manufacture anode disc with slits in the target surface to allow the material in the focal track to expand without producing the mechanical tension.
  • 10. Cont…. 2. Roughening and pitting of anode surface  Scattering of the x-rays  Absorption of x-rays by the target itself.  Results in diminish x-ray output Cause • Excessive thermal stress Remedy • Use of lowest necessary power • Largest possible focal spot • Longer exposures at reduced power For
  • 11. FIGURE 6-14 Comparison of smooth, shiny appearances of rotating anodes when new (A) versus their appearance after failure (B–D). Examples of anode separation and surface melting shown were caused by slow rotation caused by bearing damage (B), repeated overload (C), and exceeding of maximum heat storage capacity (D). (Courtesy Philips Medical Systems.) A B C D
  • 12. Cont…. 3. Imbalance ball bearings  Jamming  Increased bearing noise  Anode stops rotating Causes • Expansion due to excessive heating of the rotor bearings • Increased rotational friction Remedy • Poor conductor of heat(Mo) is used as anode stem • Metallic Lubricants such as Silver & liquid Gallium are used.
  • 13. Cont… 4. Broken stator windings  No power supply to make the anode rotate. Causes • Heavy radiography exposures • Excessive heat in the target material, rotor and its bearings, • Difference in thermal expansion of the various components used in the anode. Remedy • Provide safety circuit to terminate exposure.
  • 14. OTHER FAULTS A. DEFICIENCIES IN MANUFACTURING 1. Unlike tubes 2. Weed out by test:- Check high voltage stability, cathode emission, focal spot size, thermal loading and for rotating anode tubes additional tests such as noise, vibration, coast time etc. are tested. Not meeting these specifications are rejected. 3. Hold period:- If tubes are held for 2-4 weeks vacuum may leak 4. Improper materials:- O2 free Cu bar can cause vacuum leaks
  • 15. Cont…. B. APPLICATION MISMATCH:- e.g., poor diagnosis and radiation burns while using standard diagnostic tubes to produce mammograms C. IMPROPER DRIVE BY THE POWER SUPPLY:- 1. Filament life is reduced 2-10 times by DC filament than AC filament. 2. Magnetic materials ( e.g., iron, nickel, cobalt used for glass-to-metal seal) can be subjected to magnetic hysteresis, eddy currents etc.) 3. Filament limit:- maximum output current of the filament is limited
  • 16. CARE OF THE X-RAY TUBE 1. WARM UP THE ANODE FOLLOWING MANUFACTURER’S RECOMMENDATIONS  Performed if the system not been energized for ≥ 2 hrs  To prevent tube damage from thermal shock WARM UP PROCEDURES Make 1st 3-exposures waiting approximately 30s between each exposures @ 75kVp 200mA 1s Wait 60s, then take 1 exposure @ 85kVp 200mA ½ s Wait 60s, then take the last exposure @ 100kVp 200mA ¼ s
  • 17. Cont…. 2. MINIMIZE FILAMENT BOOST(“PREP”) TIME  Current applied for too long will shorten filament life and will lead to unstable operation.
  • 18. Cont… 3. USE LOWER TUBE CURRENT (mA)  High tube current will cause evaporation of the tungsten  Whenever possible, use a lower mA station and a longer exposure time to arrive at the desired mAs.
  • 19. Cont…. 4. FOLLOW RATING CHARTS AND ANODE AND HOUSING HEATING/COOLING CHARTS  Operation beyond published ratings will result in premature focal track wear and tear. TUBE RATING CHART o The chart which provides the information about the details of the electrical loads which may be applied to the tube without damaging it & the safe duration of such loads is called tube rating chart.
  • 20. Cont…. o In considering the tube rating three characteristics are encountered:  Ability of x-ray tube to withstand a single exposure  Ability of x-ray tube to function despite rapid exposure  Ability of x-ray tube to withstand multiple exposure during several hours of heavy use. o There are 3 types A. RADIOGRAPHIC RATING CHARTS B. ANODE HEATING/COOLING CHART C. HOUSING HEATING/COOLING CHART
  • 21. RADIOGRAPHIC RATING CHART • The safe limit within which x-ray tubes can be operated for a single exposure can be easily determined by it. • It is the most imp because it conveys which radiographic techniques are safe or unsafe for tube operation. • Microprocessors are used to allow only safe exposures.
  • 22. Rating chart of x-ray room no.1
  • 23. Rating chart of x-ray room no.1
  • 24. ANODE HEATING/COOLING CHART • Determines the maximum heat capacity of the anode. • Determines the length of time the tube must be allowed to cool before additional exposures can be made.
  • 25. HOUSING HEATING/COOLING CHART • Determines the maximum heat capacity(1-1.5 million HU) of the tube housing. • Complete cooling after maximum heat capacity may take 1-2 hrs.
  • 26. Cont.… • Thermal rating can be determined by calculating Energy deposited in the target(E) =kVp × mA × exposure time in sec • Unit is heat units(HU) • SI unit is Joule(J) 1 HU= 0.71 J 1 J= 1.4 HU • Now the Energy deposited in the target(E) =kVp × mA × s J
  • 27. CARE OF THE X-RAY TUBE 5. LIMIT OPERATION TO 80% OF MAXIMUM SINGLE EXPOSURE RATINGS  It will minimize the reduction in radiation output associated with a roughened focal track. 6. DO NOT EXCEED ANODE THERMAL CAPACITY OR DISSIPATION RATE OF THE TARGET. 7. DO NOT MAKE HIGH mA EXPOSURES ON A COLD TARGET. 8. AVOID LONG INTERVALS BETWEEN SPOT-FILMS. o Most common “holdover period”(during which the rotor is kept at high speed before the rotor brake cycle is allowed to occur) is approximately 25 s between spot films
  • 28. Cont.…. 9. LIMIT ROTOR START/STOP OPERATIONS o Minimum of 30-40 s should be between starts 10.DONOT ROTATE THE TUBE HOUSING RAPIDLY o Tension or pressure will produce to the x-ray tube. o Gyroscopic effect( resistance to turns in any direction) may crack or otherwise damage the rotor.
  • 29. Cont.… 11.SOUND COOLING SYSTEM SHOULD BE PROVIDED o Enormous heat(99%) is generated in the anode during the exposure. o Heat can be dissipated one of the 3 ways:- RADIATION CONDUCTION CONVECTION. o During exposures , heat is primarily dissipated by radiation whereas between exposures it is primarily dissipated by conduction.
  • 30.
  • 31.  General life of x-ray tubes is ten of thousands of exposures.  And CT tubes are now guaranteed for 50,000 exposures
  • 33. Reference • Christensen’s PHYSICS OF DIAGNOSTIC RADIOLOGY(4TH EDITION) • Stewart Carlyle Bushong’s RADIOLOGIC SCIENCE FOR TECHNOLOGISTS (10TH EDITION) • En.m.Wikipedia.org • www.studentradiographer.com • www.Radiopaedia.org • www.spellmahv.com • www.scribd.com • www.slideshare.com • www.sprawls.com