The document provides an overview of pacemaker components, physiology, and programming. It discusses the basic hardware components of pacemakers including the pulse generator, leads, and electrodes. It then covers pacing and sensing principles such as capture, impedance, and sensing thresholds. The remainder summarizes various pacing modes and algorithms for managing arrhythmias, rate response, and minimizing ventricular pacing.
5. Pulse Generator
SINGLE CHAMBER PACEMAKER
Block Diagram of Basic Components
Input
Amplifier
Output
Circuits
Noise
Detector
Run-away
Protection
Control Unit
CPU
Analog.ue
Circuits
Storage
ROM / RAM
Transmitting and Receiving
logic
Rate Control
Circuit
Rate Response
Sensor Telemetry Coil
ERI
Detection
Power Source
DEVICE
ENCLOSURE
HEADER
BLOCK
Set screws
LiI <100microA
Li Ag Va in ICD in Amp
3V battery
256 KB to 1MB ROM
1-16 MB RAM
20-40 Hz
Timing Cycle by Crystal
Magnet
mode
Magnet close
reed switch
(VOO)
6. Pacemaker
leads
ELECTRODES
Steroid eluding
Low polarity (Titanium Nitride)
Elgiloy
CONDUCTORS
highly conductive Ag core
MP35N for mechanical stress
INSULATORS
Polyurethane> si rubber
CONNECTOR PIN
IS/ stainless steel
lead anchorage sleeve
of radio-dense MDX
PASSIVE/ TINES
ACTIVE/ SCREWS
(mannitol/ polyethelene)
FIXATION MECHANSIM
11. Strength duration curve
• Rheobase = lowest
stimulus voltage that
will electrically
stimulate the
myocardium at any
pulse duration.
• Chronaxie = threshold
pulse duration at a
stimulus amplitude that
is twice the rheobase
voltage
12. Threshold
• Minimum amplitude and duration required to
generate the self-propagating wave front that
results in cardiac activation
atrial pacing threshold of <1.5 V
and
ventricular threshold of <1 V
13. Wedensky Effect
• Stimulation thresholds that
are measured by
decrementing the stimulus
voltage until loss of
capture are usually 0.1–0.2
V lower than when the
stimulus intensity is
gradually increased from
sub-threshold until capture
is achieved
• The Wedensky effect may
be greater at narrow pulse
durations
14. Automated Capture
• AutoCapture in SJM (beat to beat with backup pacing)
• Capture Control in Biotronik (no backup pacing)
• Ventricular Capture Management in Medtronik (once/ day)
15. Current of injury
• Partially confirm acute tissue-electrode contact
• (intracardiac EGM)
16. Impedance
• V = I.R
• I is inversely proportional to R
• R = R1+R2+R3
• R1 across lead conductors
• R2 across electrode/ myocardium interface (max)
smaller diameter of electrode increases resistance
• R3 due to polarization
shorter duration of impulse minimizes polarization
larger surface area minimizes polarization thus resistance
17. Unipolar Sensing
30-50 cms
Sensing:
- Less affected by change of
ventricular activation
- Easily influenced by
electric interferences
Pacing
- larger spike _
18. Bipolar Sensing
3-5 cm
Sensing:
- Easily affected by change
of ventricular activation
- Less influenced by electric
interferences
Pacing
- smaller spike
19. Accurate Sensing...
• Ensures that undersensing will not occur –
the pacemaker will not miss P or R waves that
should have been sensed
• Ensures that oversensing will not occur – the
pacemaker will not mistake extra-cardiac
activity for intrinsic cardiac events
• Provides for proper timing of the pacing pulse
– an appropriately sensed event resets the
timing sequence of the pacemaker
21. Automated Sensing
• traditionally a fixed sensitivity
• Most common prob is with sensing
• Better if Regularly determined
Medtronic Sensing Assurance:
• atrial is maintained within a range that is 4.0–5.6 times
ventricular is maintained within a range that is 2.8-4 times
OF the programmed sensitivity.
22. Electromagnetic Interference (EMI)
• Interference is caused by electromagnetic
energy with a source that is outside the body
• Electromagnetic fields that may affect
pacemakers are radio-frequency waves
– 50-60 Hz are most frequently associated with
pacemaker interference
• Few sources of EMI are found in the home or
office but several exist in hospitals
23. EMI May Result in the Following Problems:
• Oversensing
– Rates will accelerate if sensed as P waves in dual-
chamber systems (P waves are “tracked”)
– Rates will be low or inhibited if sensed in single-
chamber systems, or on ventricular lead in dual-
chamber systems
• Transient mode change (noise reversion)
25. EMI
Fear
– Electrocautery
– Transthoracic
defibrillation
– Extracorporeal shock-
wave lithotripsy
– Therapeutic radiation
– RF ablation
– TENS units
– MRI
Fear Not
• Home, office, and shopping
environments
• Industrial environments with very
high electrical outputs
• Transportation systems with high
electrical energy exposure or with
high-powered radar and radio
transmission
– Engines or subway braking
systems
– Airport radar
– Airplane engines
• TV and radio transmission sites
27. Lead Maturation Process
• Fibrotic “capsule” develops around the
electrode following lead implantation
28. Time Changeth Everything
Impedence
• Falls within 1-2 wks
• Then rises to 15%
more
• Low impedence
reflects failure of
conductor insulation
• High impedence
suggest conductor
fracture or loose set
screws
Threshold
Active fixation
after complete
deployment threshold is
lesser
Steroid eluting electrodes
threshold almost
unchanged
Passive fixation:
P/R decreases within days
Normalizes in 6-8 weeks
Less in SEL
Active fixation:
Attaches to myocardium
P/R decreases within mins
Normalises in 20-30 mins
Sensing
29. Rate Responsive Pacing
• When the need for oxygenated blood increases,
the pacemaker ensures that the heart rate
increases to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive Pacing
Fixed-Rate Pacing
Daily Activities
31. • Stimulate cardiac depolarization
• Sense intrinsic cardiac function
• Respond to increased metabolic demand by
providing rate responsive pacing
• Provide diagnostic information stored by the
pacemaker
Most Pacemakers Perform Four Functions
53. Cross talk window
Pacing spike earlier
than programmed AVI
•
Safety pacing is designed to prevent
ventricular asystole
if cross-talk were to occur
in a pacemaker-dependent patient
54. Far field R wave oversensing
Ventricular
channel
55. Safety Pacing
Due To Atrial Undersensing Due To Atrial oversensing
Unsesed
P Uncaptured
A
R within
CSW
79. UPPER RATE BEHAVIOUR
AVI 125 ms
PVARP 225 ms
TARP 350 ms
MTR 350 ms (170 bpm)
URI 400 ms (150 bpm)
Wenchebach interval
URI-TARP = 50 ms
PP>TARP
RATE < MTR
(157)
PP>TARP
RATE > MTR
(330)
wenchebach
2:1
84. Rate smoothing
Smoothing
9% up to 6% down
RR: 800-72=728 ms
800+48=848 ms
AA: 728-150 = 578 ms
848 -150 = 698 ms
MTR 100 bpm
6% = 36 ms
b=a+36 msa b
85. Rate modulation acting as
rate smoothening
480 ms
(125bpm)
810 ms
(74bpm)
SIR
545 ms
(110 bpm)
URL
480 ms
(125bpm)
Difference
330 ms
Difference
65 ms
MAXIMUM LENGHTHEING = MTR - SIR
86. Ventricular rate stabilization
1 = VPC
2 = AV sequential pacing at the previous V–V interval plus interval increment
3 = gradual prolongation of AV sequential pacing
Prevents pause dependent VT
93. Advanced Rate Hysteresis
pacing is suspended for the pacemaker to “search” for
the intrinsic lower rate.
If the lower rate is greater than the hysteresis rate,
pacing is inhibited until the rate again falls below the
hysteresis rate.
99. Atrial Flutter algorithm, BS
An atrial pace will only occur if the AFR
window expires at least 50 ms before the
scheduled VP. This prevents competitive
pacing
UNIPOLAR – TIP CATHODE PG ANODE
ONLY ONE CONDUCTOR SURROUNDED BY INSULATION
BIPOLAR – TIP CATHODE RING ANODE
CO AXIAL – COMMONER, SIZE INCREASES
CORADIAL – SIZE DECRASES
PASSIVE TINES – NOT SUITABLE FOR PLACING IN NONTRADITIONAL LOCATION, MAY TRAPPED IN TC valve apparatus, FIBROSED EARLY BY 6M MAKING HARD TO EXTRACT IF NEEDED
ACTIVE SCREWS – SUITABLE FOR PLACING IN NONTRADITIONAL LOCATION, EASILY RETRACTABLE, current of injury is promonent as it fixes to myocardium
DRAWN BRAZED STRAND
DRAWN BRAZED TUBE
Si rubber – melts in heat/ high friction/ low tensile strength..however biostable and biocompatible
Polyurethane – now choice for 10 years
Sleeve to be reinforced by nonabsorbable suture like ethilon
Electrode: small radius but high surface area . . . Minimizes polarization . . And improves sensing
Smalled radius – higher current density – lower pacing threshold – high resistance of electrode/ myocardium interphase – high sensing impedence and polarization artifact
The strength-duration curve illustrates the tradeoff of amplitude (intensity of stimulation, measured in volts) and duration (the length of time the stimulation is applied, measured in milliseconds).
Capture occurs when the stimulus causes the tissue to react. On the graph, capture occurs on or above the curve. Anything below the curve will not capture.
The lowest voltage on a stimulation curve which still results in capture at an infinitely long pulse duration is called rheobase. A voltage programmed below is inefficient and will lead to non-capture.
The pulse duration at twice the rheobase value is defined as the chronaxie. Chronaxie time is typically considered most efficient in terms of battery consumption.
With pulse durations greater than the chronaxie, there is relatively little reduction in the threshold voltage. Rather, the wider pulse duration results in the wasting of stimulation energy without providing a substantial increase in safety margin
doubling the pulse duration results in two-fold increase in stimulation energy, whereas doubling the stimulus voltage results in a four-fold increase, which has practical implications for the effect of programming pacing output on battery longevity
Safety margins have traditionally been selected as follows, when a threshold is determined by decrementing the pulse width at a fixed voltage:
At a given voltage where the pulse width value is < .30 ms: Tripling the pulse width will provide a two-time voltage safety margin.
Pulse widths at a given voltage value which are >.30 ms are not typically selected, because they are less efficient (expend more energy), while not providing further safety. In this case, the voltage should be doubled to provide a two-time safety margin.
Adequate safety margins must be selected due to daily fluctuations in threshold that can occur due to eating, sleeping, exercise, or other factors that affect thresholds.
Also, a patient with an acute pacing system will typically be programmed to a setting allowing a higher safety margin, due to the lead maturation process, which can occur within the first 6-8 weeks following implant.
increasing the pulse width beyond 0.6 msec usually does not decrease the voltage threshold. At implant, an atrial pacing threshold of <1.5 V and ventricular threshold of <1 V should be obtained
set the pulse amplitude at two to three times that of the pacing threshold close to the chronaxie pulse duration.
The threshold commonly increases over the next 2 to 4 weeks, reaches a peak, and then decreases to a chronic threshold level after 6 to 8 weeks. With steroid-eluting leads, the acute rise in threshold is ameliorated and the chronic threshold is significantly lower than in nonsteroid-eluting leads.
During initial programing, the pacing output is programed at 3 to 5 times the threshold voltage with a pulse width of 0.4 to 0.5 msec.
At the 2-to 3-month follow up visit, the output is decreased to no less than twice the threshold to maintain an adequate safety margin and prevent battery drain
The amplitude of the impulse must be large enough to cause depolarization ( i.e., to “capture” the heart)
The amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin
Threshold kept at atleast two-three times
Programming stimulus intensity to greater than 2.8 V results in a marked increase in current drain from the battery
Generally, a three-fold safety Generally, a three-fold safety margin may be employed at the time of initial implant, followed by reprogramming to a lower value at the 6–8-week follow-up visit
The AutoCaptureTM algorithm used in St. Jude pacemakers assures capture on a beat-tobeat basis. This algorithm first verifies that the evoked response can be differentiated from the polarization signal by delivering paired pulses, with the second pulse occurring in the absolute refractory period and thus causing only polarization artifact. Via this mechanism, the algorithm is able to set a reliable evoked response sensitivity. If there is insufficient difference between the evoked response signal and the polarization artifact, the algorithm may not function adequately because it runs the risk of oversensing polarization signal and under-estimating the threshold. In these cases, the automatic capture feature has to be turned off. After determination of the threshold, the pacemaker adjusts the output to stimulate at 0.3 V (for single chamber) or 0.25 V (for dual chamber) above the prevailing threshold. The algorithm works by reducing the pacing output in 0.25-V steps over consecutive pairs of beats until loss of capture occurs on two consecutive beats, with a 5-V back-up pulse delivered with each loss of capture event. The output is then increased in 0.125-V steps until two consecutive capture beats are seen, thereby defining capture threshold. The 0.25-V margin is then added to this threshold for subsequent pacing until loss of capture is seen or the next threshold test is performed. It is important to remember that capture confirmation occurs on a beat-to-beat basis. In several multicenter studies, the algorithm was found to be safe and effective and did not result in loss of capture, exit block, or other adverse events.20,21 However, the algorithm requires use of a bipolar pacing lead with low polarization properties. Furthermore, patients with this feature turned “on” may exhibit unusual pacing behavior on telemetry or ECG when the device is performing automatic threshold testing. The Capture ControlTM algorithm used in Biotronik pacemakers similarly depends on the use of a bipolar lead with low polarization characteristics. The differences from the St. Jude Medical algorithm include the absence of a back-up safety pulse during loss of capture and an output increase in 2-V steps if there is persistent loss of capture. Then, after a programmable length of time, the output is decreased to the original value to see if capture can be obtained with a lower output.
D duration of injury current
ST amplitude of injury current
Impedance reading values range from 300 to 1,000 W
High impedance leads will show impedance reading values greater than 1,000 ohms
Most companies 600-800 ohms except biotronik which has high impedence 1500-1600 ohm
Higher the slew rate higher the frequencey content
Sense amplifiers most sensitive to signals at a range of 30-40Hz
Steroid eluding leads
T wave is sensed; RV>VV…solution: increase RP
SECOND R waVE IS NOT SENSED; RV<VV..solution: decrease RP
atrial pacing artifact inappropriately sensed by the ventricular-sensing amplifier could result in ventricular pacing inhibition, referred to as cross-talk.
To prevent cross-talk, atrial pacing also initiates a PAVB to avoid ventricular oversensing of atrial paced events
The blanking period is traditionally of short duration because it is important for the ventricular-sensing circuit to be returned to the “alert” state relatively early
the trailing edge of the atrial pacing artifact occurring after the PAVB can occasionally be sensed on the ventricular channel.
In a pacemaker-dependent patient, inhibition of ventricular output by cross-talk results in ventricular asystole.
To prevent such a catastrophic outcome, DDD pacing mode has a safety mechanism called the “ventricular triggering period” or the “cross-talk sensing window”.
If activity is sensed on the ventricular-sensing amplifier during the AVI immediately after the PAVB, it is assumed that cross-talk cannot be differentiated from intrinsic ventricular activity.
Sensing during this window will result in a triggered rather than an inhibited output.
Safety pacing is designed to prevent ventricular asystole if cross-talk were to occur in a pacemaker-dependent patient
20-60ms
(A) Paradoxical mechanism of pacemakermediated tachycardia (PMT) initiation by an inappropriately long post-ventricular atrial refractory period (PVARP). Early premature ventricular contraction (PVC) re-initiates PVARP and VA interval (VAI). A sinus P wave falls within the PVARP (without resetting VAI), followed by ineffective atrial pacing (AP) due to atrial refractoriness. Ventricular pacing (VP) occurs after the atrioventricular interval (AVI) is completed, causing a retrograde P wave (rP) due to unopposed ventriculoatrial
(VA) conduction (dashed red arrows) with subsequent initiation of PMT. (B) Repetitive non–re-entrant VA synchrony. To avoid PMT, a long PVARP is programmed. A rP after an early PVC falls within the PVARP, followed by an ineffective AP (after the VAI is completed), initiating an AVI. VP occurs after completion of the AVI, and finds unopposed VA conduction (retrograde P wave), falling again within the PVARP, causing repetitive ineffective AP and VP with VA conduction.
1000-150+200 = 1050
800 850
Atrial based = compensatory pause
Atrial based = VPC in AVI wont reset..as AA is fixed
Atrial indiacted rate
Sensor indicated rate
Dash dot line = intrinsic rate (P synchronous pacing)
Dot line = sensor rate
Heavy line =paced ventricuar rate
P tracking to AV pacing thru a period of wenchebaching
Period of wenchebaching is shortened due to ‘sensor driven rate smoothing’
6% == 36 ms
MDT
ONLY IN PAC OR PVC
NOT OPERATIONAL IN SUSTAINED TACHYARRYTHMIAS
Not operational if hysteresis on or mode switch on
For that ventricular rate regulatisation is there…but that is only operational during MODE switch
60000/172=52 bpm
As no fall back enabled
Fallback for 15 sec
Once the atrial rate exceeds the mode switch rate (also referred as the fallback onset rate), pacing mode will switch from DDD to a programmable non-tracking pacing mode (DDI, VDI, VVI). The fallback (FB) feature prevents sudden drop on ventricular pacing (VP) from the MTR to the lower rate limit (LRL). When fallback time is enabled, the device will slowly decrease the VP rate from the MTR to the LRL in the timeframe instructed. Once atrial rate decreases below the mode switch rate, the device will return back to DDD pacing mode
Sleep rate = 30 mins after bedtime
Night rate = after 10 pm
Rest rate = every 7days