CLAUS SULFUR RECOVERY TAIL GAS
APPLYING 100 MILLION HOURS OF OPERATIONAL
TIME TO THE NEXT GENERATION ANALYZER
Randy Hauer, Stuart Simmonds, Ed Pavina / AMETEK Process Instruments
455 Corporate Blvd, Newark DE, 19702
Bob McCartney / PBF Energy Paulsboro Refinery
600 Billingsport Road, Paulsboro, NJ 08066
KEYWORDS
Sulfur Recovery, Modified Claus Process, H2S/SO2, Tail Gas Analyzer
ABSTRACT
The measurement of H2S/SO2 in Claus sulfur recovery unit (SRU) tail gas has been adequately
addressed by UV spectroscopy for over 40 years. Reliability of the analytical principle was
established in the first generation of analyzers and in the second generation sample handling was
improved to the point where automatic control if not universal is at least considered the norm.
With a deep understanding of the process and a population of 1,100 second generation analyzers
it was possible in the third generation to address failure modes external to the analyzer.
Reliability is not limited to the analyzer / sample system; it extends to the process and contains
elements of health and safety. Analyzer professionals are compelled to look beyond what we
consider our "deliverables", to address abnormal operations and bad piping design. The paper
combines extensive feedback from analyzer professionals and a survey of sulfur recovery
operators to address the external failure modes. Looking back on 100 million hours of
operational time and one year of field testing the third generation analyzer the paper discusses
reliability as viewed by the real end use customer; Operations.
THE PROCESS, DEVELOPMENT OF THE TAIL GAS ANALYZER
Sulfur recovery is the unit operation of converting hydrogen sulfide (H2S) to elemental sulfur
and is an essential block in an integrated refinery complex or sour gas processing plant. Referred
to as the Modified Claus process; in a sub stoichiometric first step 1/3rd
of the H2S undergoes
combustion to SO2 in a reaction furnace then passes through two or three catalytic converters to
produce elemental sulfur. The Modified Claus process is capable of ~97-98% recovery
efficiency(1)
. The primary consideration for high efficiency is control of the (feed forward)
air:gas ratio by trim (feedback) control using the H2S/SO2 tail gas analyzer.
FIGURE 1. MODIFIED CLAUS SRU PROCESS & INSTRUMENT DRAWING
Sulfur is primarily used for the production of sulphuric acid which is the largest traded chemical
commodity in the world. While sulfur does have a commodity value in most cases the driving
force for process measurement and control is largely an environmental concern. The US EPA
Clean Air Act of 1970 and the ground-breaking study by Alberta Environment on The Capability
of the Modified-Claus Process(1)
coincided with the first attempts to control the Modified-Claus
process using an on-line analyzer and closed loop control.
The first report of an on-line tail gas analyzer was a technical paper presented at the 1970
Analysis Division Symposium based on a gas chromatograph installed at the Dow Chemical
Freeport TX (USA) facility(2)
. Amoco Oil and DuPont Process Instruments published a paper at
the 1972 Analysis Division Symposium on an Ultraviolet (UV) based tail gas analyzer installed
at the Amoco Whiting IN (USA) refinery(3)
. Shortly after this in 1974 Western Research
participated in a pilot study using a prototype UV-based tail gas analyzer at the Shell Waterton
AB (Canada) gas plant.
It was already noted in the 1972 paper that “the non-continuous response of a process GC is
often inadequate to provide the close control”. Given the process dead time of 30 seconds (to 2
minutes or more at turndown) and the requirement of a continuous sampling subsequent
developments by various manufacturers were all based on UV spectroscopy with extractive
sampling techniques which remains the case today.
There is a collective experience from ~2,700 tail gas analyzers (DuPont, Western Research and
the amalgamation of these two; AMETEK Process Instruments); five models; two full
generations; and now moving into a 3rd
generation tail gas analyzer. The 3rd
generation analyzer
is the evolution of the close coupled 880 NSL (no sample line) analyzer with 1,100 units in
service and ~100 million hours of operational time.
It has been noted by experienced analyzer engineers that certain process analyzer applications
are much more difficult, among these being crude tower feed/bottoms and Claus tail gas (4)
.
Further to this point it has been estimated that something less than half of all process analyzers
are placed in control. Analyzers placed in control get a great deal of scrutiny from Operations.
Failure due to adverse conditions and unexpected (abnormal) results are not differentiated from
analyzer failure. Having a large population of analyzers dedicated to one process gave insight
beyond just the analyzer. A critical mass of information was acquired to look beyond the
“deliverables” to address failure modes that are external to the analyzer and sample handling.
TWO INDUSTRY SURVEYS
Looking back on 40 years of experience the 1st
generation (~1972-1997) proved UV worked, the
weak link was sample handling, steam jacketed pipe and a large volume of sample gas. The 2nd
generation and the application of a packed reflux type demister improved sample handling
reliability but closed loop control was still not universal with process problems blamed on the
analyzer. For the 3rd
generation the focus was on what could be addressed in the way of failure
modes external to the analyzer. With the aid of two industry surveys, a close study of the 2nd
generation analyzer was undertaken.
For feedback from hands on professionals a voice of customer (VOC) survey polled analyzer
technicians for their views. The 2nd
generation top of the pipe analyzer has a strong following
and many said “change nothing” in relation to the sample handling and “don’t change the size”
(meaning do not shrink it down and make the oven hard to access with gloves on). In terms of
desired new features the list included double block from the process (safety), a single steam
pressure, automatic flow control, over range measurement and improved digital communications.
The paper will expand on these aspects in detail.
A second survey was conducted to obtain valuable end user experience in addition to analyzer
professionals. Three distinct stakeholder groups in sulphur recovery were polled;
 Front end engineering design (FEED) and start up engineers
 SRU Operations
 End user analyzer specialists (engineers and technicians)
It was deemed important while measuring reliability not to restrict feedback to just the analyzer
profession, recognizing that on many occasions Operations has a different opinion as to “the
analyzer is working”. The survey was extended to Operations, start up engineers and process
design specialists for their perspective. The spread in the response was suitable with all three
individual groups representing between 20-40% of the responses. To the question of reliability
~90% rated tail gas analyzers “reliable most of the time / all of the time”. This response was
gratifying based on a comparison question asked when conducting process training for Operators
over the past 15 years; “Is the analyzer in control mode?” for which ~25% say “No”. The take
away being that even when the analyzer is considered reliable Operations decides to use it in
manual control.
FIGURE 2. SRU INDUSTRY SURVEY
ADDRESSING EXTERNAL MODES OF FAILURE, PREDICTING
ADVERSE PROCESS CONDITIONS
Risk can be managed and reduced by understanding the potential failure modes of an analyzer,
its sample conditioning system and the process. The first step in the process is to define all of the
critical conditions needed to make a valid measurement. The second step is to determine which
diagnostic methods can be used to evaluate each of the conditions. Diagnostic methods require
the collection or generation of data which is then evaluated against the conditional requirements
and to provide either a direct or indirect indication of the potential failure modes. There are three
types of diagnostics that are implemented in the analyzer. They are observational, model-based,
and functional. These three diagnostic types are differentiated by the means through which the
evaluated data is collected. This data is used to predict failure and manage reliability aspects
internal to the analyzer / sample handling system and was covered in a previous paper (5).
The following expands on the methods applied to predicting and reacting to failure modes which
are external to the analyzer and sample handling system. Having a long history in sulfur recovery
there was strong confidence to enter into the process world and address problems that were
previously beyond our control. Carefully considering feedback from the process stakeholders’
attention was placed on the three main external failure modes; one is process based, one is utility
based and one is environment based.
1. Failure Mode (Process): Entrained liquid sulfur (sulfur fog, sulfur mist)
SRUs are designed to operate at a certain capacity and refinery SRUs in particular go
through a wide range of turn down according to crude slate. When a SRU goes below
half load a phenomena occurs where the over cooling of the gas stream causes sulfur to
form a fog, small sub-micron drops of sulfur that defy agglomeration in the process
demister pads (6)
. The fog can then collect on the analyzer optics. Further complicating
this is the attendant increase in process pressure (from ~1 to ~2 psig) causing unintended
increase in the flow rate of the analyzer, drawing up more entrained sulfur.
FIGURE 3. SULFUR ENTRAINMENT – FOG AND MIST
The remedy is to adjust the sample flow rate relative to the process pressure. Simply
measuring the flow is not sufficient and placing a flow sensor in direct contact with SRU
process gas is not good practice. The 3rd
generation analyzer monitors the pressure
differential between the process and the sample aspiration and automatically adjusts the
flow rate relative to the process pressure during the zero cycle (every 90 minutes). During
start up, shut down, turn down, plugged rundowns and any time there is entrained sulfur
leaving the final condenser the flow rate can be minimized to reduce sulfur mist take up.
2. Failure Mode (Utilities): Poor quality steam, bad steam traps
Steam traps are known to commonly fail, at any given time 20 to 30% of steam traps are
defunct. The process connection point for a tail gas analyzer depends must be maintained
at 300F (sulfur freezes at 246F). To maintain the heat integrity a custom steam jacketed
ball valve is used. The problem of wet steam or a plugged steam trap is not usually a
sudden failure. Most often it is intermittent it can come and go with wet weather or from
night to day or result in “choking” which gives the appearance of slow response. The
remedy is an embedded RTD in the process connection flange which provides an alarm.
This advanced warning is unequivocal it indicates compromised heat integrity, calling
attention to and eliminates a problem that can be the cause of intermittent failure.
3. Failure Mode (Environment): Solar gain, latent heat, internal heat release
The 2nd
generation tail gas analyzer was the first to be close-coupled to the pipe. Placing
the entire analyzer on the pipe eliminated the need for a shelter but also exposed the
analyzer to solar gain and latent process heat. On a SRU the analyzer location is, as often
or not is temperature compromised and engineering procurement Contractors (EPCs)
rarely take care to add a sunshield to the analyzer.
While the existing analyzer is rated to 122o
F (50o
C) there is empirical evidence in the
form of premature component failures that the internal temperature was far beyond the
50o
C rating. Solar gain is the principal cause and in some cases latent process heat but
also the internal heat release from the oven was a contributing factor.
The effect of temperature on analyzer electronics was very well covered in a paper from
ISA-AD 2012 precisely quantifying how the life of electronic semiconductor circuits is
affected adversely by temperature as indicated by the Arrhenius Equation. The equation
predicts that for approximately each 10°C increase in operating temperature mean
electronic component operating life is reduced by 50% (7)
.
One of the objectives for the 3rd
generation analyzer was to increase the temperature
rating to 60o
C without the aid of any external cooling device (Vortec cooler, Peltier
device, air conditioning). Given that the oven must be capable of 315o
F (~160o
C)
operation it was somewhat in question if the heat release from the sample side onto the
electronics side could be mitigated while also overcoming solar gain. A highly effective
thermal insulating barrier was developed early in the design phase as well as a custom
extruded aluminum convection heater replacing the off shelf oven heater. Advanced
thermal management algorithms resulted in a net reduction of the internal temperature of
the electronics enclosure by 15°C (27°F). The outcome is an industry first 60°C
specification for ambient temperature and a 75% improvement in electronics life(7)
. The
analyzer is suitable for installation in hot climates with a sunshade being recommended.
EVOLUTIONARY IMPROVEMENTS
Digital communications and the HMI (human machine interface) were obviously due for
updating, benefiting from improvements in electronics over the intervening eighteen years. For
the sample handling and photometry great care was taken not to compromise the elements that
worked well, avoid change for the sake of change. The following describes both categories.
IMPROVING ON WHAT ALREADY WORKED WELL
Photometer: For this application the measurement of H2S/SO2 does not require a high degree of
precision or accuracy. Control is a simple binary action and too much can be made of comparing
different methods of UV spectroscopy. The final control element is typically a butter fly valve
sized to control 10% of the process air in the form as a trim air control loop which is manipulated
its full range by the tail gas analyzer (8)
. In steady state full load conditions the trim air (“air
demand”) is moving inside a peak to peak band of +/- 1% which results in a +/- 10% valve
movement. Because of process dead time the peak to peak period is ~30 seconds (at full load)
and the controller proportional gain is set low. The result is the time domain resolution of a
photo-diode UV analyzer is 10x the resolution of the valve (9)
. Add to this that the control
function is binary (“add air, cut air”, it never dwells at set point). The result is accuracy is not
paramount; simplicity, support, reliability are the prime requirements.
The photometer for the 3rd
generation retains the 2nd
generation Xenon flash lamp and improves
upon the photometer validation algorithms. The light from the sample cell is divided into four
separate channels using partially reflective and partially transmitting beam splitters. The UV
photometer uses four separate detectors each measuring a different narrow range of UV
wavelengths. The intensity information from each of the four detectors is used to calculate the
H2S and SO2 concentrations using a factory established calibration. The temperature of the
photometer is carefully controlled to optimize the accuracy and precision of the reported
concentrations. During the zero and span cycle an automatic multi-point photometric span
calibration is performed on each of the four channels. The source pulse sequencing is
automatically varied to create an intensity calibration for each detector channel. It is uniquely
capable of automatic intensity calibration but calibration with a (manual) removable filter was
retained as customer feedback indicated this was a widely accepted method for Operations to be
assured the analyzer is working.
The dynamic range was improved to the extent the analyzer can measure 100% over range while
maintaining linearity specifications. This is especially useful during an upset giving Operations a
widow beyond the normal air control parameter. Tail Gas is one of those applications where the
instrument data sheet stays slavishly attached to the standard range, not acknowledging the
ability of a next generation analyzer to provide this valuable information. It is up to the analyzer
profession to communicate improvements such as over-range to front end engineering (FEED)
designers and instrument engineers who stay rooted to past history and dated data sheets.
FIGURE 4. PHOTOMETER
Demister: What was well proven in the 2nd
generation design was the veracity of using
demisting pads to coalesce sulfur drops. Straight gravity drainage using an unpacked column
(“cold finger” probe) works well enough when there is minimal entrained sulfur or when the
drops are large and easily agglomerate. Adverse process conditions (plugged run down, failure of
the process demister pad, sulfur fog at turndown) require a packed column and reflux action. The
sulfur demister borrows heavily from the experience of dynamic reflux (pyrolysis) gas sampling.
The subject has been extensively covered at previous symposia with reference made to the ability
to prevent clouding of optical windows (10)
. The hallmarks of a reflux sampler are surface area,
vertical mounting and being close coupled but external to the process (11)
. One of the 3rd
generation improvements borrowed from the reflux sampler was mounting the demister pads on
a shaft with retention rings (11)
which in the case of the tail gas demister keeps the mesh pads in
firmly in place during the hot condensate steam blow back for ammonia salts.
FIGURE 5. DEMISTER
Incomplete combustion of ammonia in the SRU process reacts with sulfur compounds to form
ammonia salts, which can plug the demister. These salts are not removed by air blow-back but
are water soluble. Hot condensate steam blow-back provides a steam propelled blast of hot water
to dissolve any salts that have accumulated within the demister. An example of what these salts
look like in the process and propensity to cause pressure drop is shown in the following figure.
FIGURE 6. AMMONIA SALTS ON CONDENSER TUBE SHEET
The analyzer is also equipped with anti-clogging hot air blowback feature that is automatically
initiated if plugging is detected by the smart diagnostics. The 3rd
generation demister retains the
double demister pads (316 SS top and PTFE bottom). A change was made to connect up with
vacuum compression o-ring (VCO) fittings in place of tube fittings making it easy to access and
requiring minutes to disassemble. While not yet common in the process world VCO fittings have
a place as they allow technicians easy access components inside a confined space like an oven.
NEW IMPROVEMENTS
Digital Interface: In our industry fifteen years is a typical life span of an analyzer while
graphical interface and digital communications goes through several incarnations in the same
period. The challenge is to stay up to date while at the same time finding commonality across a
range of products for human engineering as well as cost of manufacturing and ongoing support.
The next generation tail gas analyzer followed closely behind the 5th
generation WDG oxygen
analyzers and the same local HMI was implemented. AMEVision is an icon driven graphical
interface with a color display for local communication with the analyzer. It provides screens
showing trending functions, predictive maintenance indicators, analog output verification, and
time stamped alarm/event log.
Remote communication is via a standard PC web browser enabled interface (no software need be
installed) over TCP/IP Ethernet. This point raises an anomaly that is perhaps unique to sulfur
recovery but in any case is worth revisiting on every new project. Remote digital communication
has been available on 2nd
generation tail gas analyzers (two models, ~1,800 units) for the past 17
years. Of that total fewer than 10% are tied into remote digital communications. At least half
have been new SRUs, grass roots project. The question to ask of EPCs (and end users) why they
deem it acceptable to go on to the sulfur unit when it could all be done from the safety of the
analyzer shop using a remote digital connection. This has hazardous operation implications and
should be on top of the list for any analyzer engineer on a project to say “and why not?”
UTILITIES
There was opportunity to reduce the utilities. The Division 2/Zone 2 analyzer does not require an
EExP purge, this reduces the instrument air volume and eliminates damage from wet instrument
air. Steam as a utility is reduced to a single pressure. The demister pads held firm in place on a
shaft can now withstand medium pressure (MP) steam and can utilize the same MP steam
required for the process connection valve to maintain 300o
F (150o
C).
SAFETY
Hazardous Operation (“HazOp”) review is becoming more prevalent in the world of process
analyzers. It has been said that sulfur recovery is not as well understood because it is a chemical
plant in a refinery world and that is largely true. Sulfur recovery while not any more inherently
dangerous than other refinery unit operations has the stigma of H2S. With that in mind safety
features were added for with future requirements in mind.
Three major refining companies in the USA often require double block isolation at their sites.
This was not entirely possible in the 2nd
generation analyzer but has been implemented in the 3rd
generation analyzer with two isolation valves at the process access valve on both the sample and
the vent. The sample conditioning components are external to the process and it is possible to
remove the probe from the process under live conditions with zero egress through the ConaxTM
fitting. The sample handling components are a safe distance from the process (500 mm/18”,
which is considered the boundary for “safe distance”)
To close this subject it is worthwhile to look at the safety implications of having to work with a
compromised sample point. A modern SRU and tail gas treating unit when built as a single
project is typically conjoined with a short vertical section of pipe. When analyzer professionals
get zero input into at the piping design stage they are tied into a situation that can compromise
both function and safety. For SRU-TGTUs the problems are confined quarters (egress), access,
and process heat amongst others (9)
. It is incumbent on our profession to interject ourselves at an
early stage of a project to take a leading role in sample point selection and to take advantage of
the available knowledge base
FIGURE 7. COMPROMISED SAMPLE POINT LOCATION
FIELD TRIAL OF 3rd
GENERATION ANALYZER
A search for a suitable site to conduct field trials commenced in late 2013; the prime requirement
being close proximity to the factory located in Newark DE. There are a number of refineries in
the area but SRUs are normally only in turn around every 4 years and a pool of suitable sites is
therefore quite limited. The PBF Paulsboro NJ refinery had an available sample point location
and consented to installing the new analyzer by hot-tapping a process connection valve. Like the
2nd
generation analyzer the 3rd generation analyzer can be installed while the process is hot (i.e.
ingress and egress of the probe via a Conax ™ gland) as long as a one inch full port ball valve is
in place as the process connection.
The hot tap was made and the model 888 analyzer installed at PBF Paulsboro in April 2014.
Operations considered keeping the other analyzer in service in case the new analyzer required
intervention. This would have added considerable cable and utility work. Given the additional
cost, the confidence in the new analyzer was high and the opinion of analyzer professionals that
“two watches” is never good the decision was made to go solo with the new analyzer. The cut
over was made in one day and the new analyzer was placed in control by Operations same day.
During the first 4 months of the trial there were no major outages for the analyzer. There was
concern that the process connection valve used for the hot-tap was not steam jacketed but as it
was close coupled there were no immediate problems. Later in July during a cold rain there was
intermittent plugging and this was remedied with an off the shelf piece of Contra-Trace ™ from
Controls Southeast Inc (CSI) to match this specific valve body. CSI specializes in sulfur
recovery applications but the product is well applied to any analyzer heat tracing requirement
and is much preferable to wrapping in SS tubing which in this service serves no useful purpose.
There was only minimum intervention in the first seven months, modifications were restricted to
software changes, availability was >99%. Operating conditions went through the normal range of
events and the sample handling system performed to expectations with zero complaints from
operations. Given the anomalies and the degree of difficulty associated with this type of process
analyzer application the testing program allowed for up to 12 months of evaluation to observe
long term effects on the components.
In early November a heater fuse failed, it was deemed to be under rated and a larger fuse
installed. In late November/early December condensed water was discovered in the sealed
photometer box as well as on the detector board and the mystery as to how water was getting in.
Wet instrument air was quickly eliminated. After some investigation a leaking check valve from
a well known supplier was determined to be the root cause, it was intermittently failing and
allowing small amounts of wet process gas to find a pathway into the photometer. The check
valve was replaced with one from another supplier.
In late early January there was observation of something off-gassing on the optics causing a slow
decay of the measured light (compensated for but eventually causing low light level alarms). The
photometer mirrors were replaced so the material could be extracted and analyzed. The root
cause was a small silicon based seal which was replaced with another material. At the time of
writing there were no other problems. The evaluation will run until March at which time the
analyzer design will be turned over to production and manufacturing will commence.
FIGURE 8. FIELD TRIAL PBF PAULSBORO REFINERY
DISCUSSION AND CONCLUSIONS
Process analytics is a young industry (approximately 50-60 years) with a product life cycle that
can be relatively long. Most vendors think in terms of 15-20 year operation before considering
replacement. Reviewing an application with a known degree of difficulty such as sulfur
recovery is not complete without addressing the external failure modes that potentially occur
within the process. As the analyzer results are the point of contact between the process and
operations, these external failure modes are more often ascribed as “analyzer failures” and in all
cases Operations will always have the last word; “we trust it/we don’t trust it”.
First and 2nd generation tail gas analyzers have successfully concluded the measurement
technique and closed loop control respectively. Now approaching the third generation of tail gas
analyzers affords the opportunity to move beyond the borders of the analyzer itself and step into
the process world, evaluating the process and responding to external failure modes. It takes a
step further by educating the operators on the analytical technique, the sample system as well as
the anomalies of the process to arrive at a holistic concept of reliability.
Finally, it is imperative when approaching the process industry with a 3rd generation of tail gas
analyzer that a long term field trial (minimum of 12 months) has been concluded. Even with
many years of experience it takes time for hidden flaws to be revealed. Only after a successful
field trial can an analyzer be introduced with confidence to an industry that expects a high level
of reliability.
ACKNOWLEDGMENTS
The authors wish to thank the PBF Energy Paulsboro refinery for installing the analyzer for the
field trial. In addition, to thank Bob McCartney and Ed Pavina for their support in the field
REFERENCES
1) Paskall, H.G. "Capability of the Modified Claus Process." 1979.
2) Jackson, J.L, P.V Maynard. “The Analysis & Control of the H2S/SO2 Ratio in a Sulfur
Recovery Plant Tail Gas Stream by Process Gas Chromatography”, ISA Analysis Division
Symposium, Pittsburgh PA, May, 1970
3) Saltzman, R., Dr. E.B. Hunt. “A Photometric Analyzer System for Monitoring & Control of
the [H2S]/[SO2] Ratio in Sulfur Recovery Plants”, ISA Analysis Division Symposium, San
Francisco CA, May 1972
4) Novak, D., “Process Analytics: Are There Dinosaurs Among Us? - The Stigma of Process
Analytics” and Dubois, R., “Process Analytics: Are There Dinosaurs Among Us? - Myths and
Mistakes That May Contribute to Our Extinction”, Panel Discussion, ISA Automation Week
Houston TX, October, 2010
5) Myles, T “Design Features of a Process Analyzer to Reduce Unscheduled Downtime”. The
59th
ISA Analysis Division Symposium, Baton Rouge, LA May 5-8, 2014
6) Bohme, G., J. Sames. "The Seven Deadly Sins of Sulphur Recovery". The British Sulphur
Conference, Calgary, Alberta, Canada, October 1999.
7) Farmer, B., U. Gellert, T. Campbell. “Ambient Temperature Testing Issues for Process
Analyzers”, The 57th
ISA Analysis Division Symposium, Anaheim, CA April 22-26, 2012
8) Neumeister, L., Controlling Sulphur Plants: Practical Issues Surrounding Control Strategies,
Controller Hardware, Control Equipment Performance and Tuning, September 2000
9) Al-Misfer, A., S.Vedula, R. Hauer, Z. Juddy, “Process Analyzer Best Practices for Sulfur
Recovery, Enhanced Claus and Tail Gas Treating”. Sour Oil & Gas Advanced Technology
(SOGAT) Conference, Abu Dhabi, UAE March, 2009
10) Clemons, J., T Dye. “Complex Sampling Problems and Real World Solutions: Dynamic
Reflux Sampling”. The 47th
ISA Analysis Division Symposium, Denver CO, April 2002
11) Smith, S. “Pyrolysis Gas Sampling”. The 56th
ISA Analysis Division Symposium, League
City TX, April 2011
12) Hauer, R., Z. Juddy, Y. Yoshikane, J. Sames “The Seven Deadly Sins of Process Analyzer
Applications”, SOGAT Conference Abu Dhabi, UAE March, 2011

Contenu connexe

Tendances

Calculation of an Ammonia Plant Energy Consumption:
Calculation of an Ammonia Plant Energy Consumption:  Calculation of an Ammonia Plant Energy Consumption:
Calculation of an Ammonia Plant Energy Consumption: Gerard B. Hawkins
 
Determination of Residue on Evaporation in Anhydrous Ammonia
Determination of Residue on Evaporation in Anhydrous AmmoniaDetermination of Residue on Evaporation in Anhydrous Ammonia
Determination of Residue on Evaporation in Anhydrous AmmoniaGerard B. Hawkins
 
Piping and Vessels Flushing and Cleaning Procedure
Piping and Vessels Flushing and Cleaning ProcedurePiping and Vessels Flushing and Cleaning Procedure
Piping and Vessels Flushing and Cleaning ProcedureGerard B. Hawkins
 
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTS
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTSSTEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTS
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTSGerard B. Hawkins
 
Shift Conversion Catalysts - Operating Manual
Shift Conversion Catalysts - Operating ManualShift Conversion Catalysts - Operating Manual
Shift Conversion Catalysts - Operating ManualGerard B. Hawkins
 
METHANOL PLANT ARC RETROFIT Case Study
METHANOL PLANT ARC RETROFIT Case StudyMETHANOL PLANT ARC RETROFIT Case Study
METHANOL PLANT ARC RETROFIT Case StudyGerard B. Hawkins
 
Other Separations Techniques for Suspensions
Other Separations Techniques for SuspensionsOther Separations Techniques for Suspensions
Other Separations Techniques for SuspensionsGerard B. Hawkins
 
Practical Analytical Instrumentation in On-line Applications
Practical Analytical Instrumentation in On-line ApplicationsPractical Analytical Instrumentation in On-line Applications
Practical Analytical Instrumentation in On-line ApplicationsLiving Online
 
Steam & water analysis system (swas)
Steam & water analysis system (swas)Steam & water analysis system (swas)
Steam & water analysis system (swas)khashayar Ataee
 
Steam and water analysis system 1
Steam and water analysis system 1Steam and water analysis system 1
Steam and water analysis system 1Sanjana Khound
 
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...IRJET Journal
 
Determination of Hydrogen Sulfide by Cadmium Sulfide Precipitation
Determination of Hydrogen Sulfide by Cadmium Sulfide PrecipitationDetermination of Hydrogen Sulfide by Cadmium Sulfide Precipitation
Determination of Hydrogen Sulfide by Cadmium Sulfide PrecipitationGerard B. Hawkins
 
High Temperature Shift Catalyst Reduction Procedure
High Temperature Shift Catalyst Reduction ProcedureHigh Temperature Shift Catalyst Reduction Procedure
High Temperature Shift Catalyst Reduction ProcedureGerard B. Hawkins
 
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDS
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDSSYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDS
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDSGerard B. Hawkins
 
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...Gerard B. Hawkins
 
Biological Systems - A Special Case
Biological Systems - A Special CaseBiological Systems - A Special Case
Biological Systems - A Special CaseGerard B. Hawkins
 

Tendances (20)

Calculation of an Ammonia Plant Energy Consumption:
Calculation of an Ammonia Plant Energy Consumption:  Calculation of an Ammonia Plant Energy Consumption:
Calculation of an Ammonia Plant Energy Consumption:
 
Determination of Residue on Evaporation in Anhydrous Ammonia
Determination of Residue on Evaporation in Anhydrous AmmoniaDetermination of Residue on Evaporation in Anhydrous Ammonia
Determination of Residue on Evaporation in Anhydrous Ammonia
 
Piping and Vessels Flushing and Cleaning Procedure
Piping and Vessels Flushing and Cleaning ProcedurePiping and Vessels Flushing and Cleaning Procedure
Piping and Vessels Flushing and Cleaning Procedure
 
Sedimentation
SedimentationSedimentation
Sedimentation
 
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTS
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTSSTEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTS
STEAMING PROCEDURE FOR VULCAN STEAM REFORMING CATALYSTS
 
Shift Conversion Catalysts - Operating Manual
Shift Conversion Catalysts - Operating ManualShift Conversion Catalysts - Operating Manual
Shift Conversion Catalysts - Operating Manual
 
METHANOL PLANT ARC RETROFIT Case Study
METHANOL PLANT ARC RETROFIT Case StudyMETHANOL PLANT ARC RETROFIT Case Study
METHANOL PLANT ARC RETROFIT Case Study
 
Other Separations Techniques for Suspensions
Other Separations Techniques for SuspensionsOther Separations Techniques for Suspensions
Other Separations Techniques for Suspensions
 
Practical Analytical Instrumentation in On-line Applications
Practical Analytical Instrumentation in On-line ApplicationsPractical Analytical Instrumentation in On-line Applications
Practical Analytical Instrumentation in On-line Applications
 
Steam & water analysis system (swas)
Steam & water analysis system (swas)Steam & water analysis system (swas)
Steam & water analysis system (swas)
 
Steam and water analysis system 1
Steam and water analysis system 1Steam and water analysis system 1
Steam and water analysis system 1
 
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...
IRJET - Minimizing Steam Loss through Steam Distribution Network using an Int...
 
Determination of Hydrogen Sulfide by Cadmium Sulfide Precipitation
Determination of Hydrogen Sulfide by Cadmium Sulfide PrecipitationDetermination of Hydrogen Sulfide by Cadmium Sulfide Precipitation
Determination of Hydrogen Sulfide by Cadmium Sulfide Precipitation
 
High Temperature Shift Catalyst Reduction Procedure
High Temperature Shift Catalyst Reduction ProcedureHigh Temperature Shift Catalyst Reduction Procedure
High Temperature Shift Catalyst Reduction Procedure
 
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDS
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDSSYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDS
SYNGAS CONDITIONING UNIT FEASIBILITY CASE STUDY: COAL-TO-LIQUIDS
 
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...
METHANOL PRODUCTION USING VULCAN SYSTEMS COMBINED REFORMING TECHNOLOGY (ATR) ...
 
Biological Systems - A Special Case
Biological Systems - A Special CaseBiological Systems - A Special Case
Biological Systems - A Special Case
 
CEMSI Cement Solutions
CEMSI Cement SolutionsCEMSI Cement Solutions
CEMSI Cement Solutions
 
Gas - Liquid Reactors
Gas - Liquid ReactorsGas - Liquid Reactors
Gas - Liquid Reactors
 
swas equipment
 swas equipment swas equipment
swas equipment
 

Similaire à Claus Sulfur Recovery Tail Gas Applying 100 Million Hours of Operational Time to the Next Generation Analyzer

Leak Detection And Repair
Leak Detection And RepairLeak Detection And Repair
Leak Detection And RepairAshutosh Soni
 
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)Ultrasonic vs. Mechanical Anemometers (Wind Sensors)
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)Alexandre Ferrete
 
Advanced gas leakage detection using acoustic sensors new
Advanced gas leakage detection using acoustic sensors newAdvanced gas leakage detection using acoustic sensors new
Advanced gas leakage detection using acoustic sensors newYatheesh Kaggere
 
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docxeugeniadean34240
 
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_Naimish Raval
 
EricKatzen_NEEG_Abstract_072616
EricKatzen_NEEG_Abstract_072616EricKatzen_NEEG_Abstract_072616
EricKatzen_NEEG_Abstract_072616Eric Katzen
 
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE AND
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE ANDUSING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE AND
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE ANDScot D. Abbott
 
An optimised approach for the inspection of aboveground
An optimised approach for the inspection of abovegroundAn optimised approach for the inspection of aboveground
An optimised approach for the inspection of abovegroundPatrick Schüller
 
Pressure Loss Optimization of a Sprinkler Valve
Pressure Loss Optimization of a Sprinkler ValvePressure Loss Optimization of a Sprinkler Valve
Pressure Loss Optimization of a Sprinkler ValveOmar Wannous
 
Part technical cleanliness/Millipore
Part technical cleanliness/MilliporePart technical cleanliness/Millipore
Part technical cleanliness/MilliporeSURAJ JATHAR
 
Dynamic Stand-Alone Gas Detection System
Dynamic Stand-Alone Gas Detection SystemDynamic Stand-Alone Gas Detection System
Dynamic Stand-Alone Gas Detection SystemIRJET Journal
 
Chakresh Tiwari Philips report
Chakresh Tiwari Philips reportChakresh Tiwari Philips report
Chakresh Tiwari Philips reportpranav kapoor
 
Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...eSAT Journals
 
Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...eSAT Publishing House
 

Similaire à Claus Sulfur Recovery Tail Gas Applying 100 Million Hours of Operational Time to the Next Generation Analyzer (20)

Leak Detection And Repair
Leak Detection And RepairLeak Detection And Repair
Leak Detection And Repair
 
Tube bundle gas monitoring system
Tube bundle gas monitoring systemTube bundle gas monitoring system
Tube bundle gas monitoring system
 
COSGC Symposium Paper
COSGC Symposium PaperCOSGC Symposium Paper
COSGC Symposium Paper
 
Rams 2007 paper pdf
Rams 2007 paper pdfRams 2007 paper pdf
Rams 2007 paper pdf
 
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)Ultrasonic vs. Mechanical Anemometers (Wind Sensors)
Ultrasonic vs. Mechanical Anemometers (Wind Sensors)
 
Advanced gas leakage detection using acoustic sensors new
Advanced gas leakage detection using acoustic sensors newAdvanced gas leakage detection using acoustic sensors new
Advanced gas leakage detection using acoustic sensors new
 
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx
2015-2016 MechanicalCivil Undergraduate Senior DesignWa.docx
 
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_
MASS MASTER METERING SOLUTION FOR USERS PAIN_ FCRI 2012 KS NR _2_
 
EricKatzen_NEEG_Abstract_072616
EricKatzen_NEEG_Abstract_072616EricKatzen_NEEG_Abstract_072616
EricKatzen_NEEG_Abstract_072616
 
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE AND
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE ANDUSING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE AND
USING MINE SITE GAS CHROMATOGRAPHY FOR ACCURATE AND
 
ultrametr
ultrametrultrametr
ultrametr
 
An optimised approach for the inspection of aboveground
An optimised approach for the inspection of abovegroundAn optimised approach for the inspection of aboveground
An optimised approach for the inspection of aboveground
 
Pressure Loss Optimization of a Sprinkler Valve
Pressure Loss Optimization of a Sprinkler ValvePressure Loss Optimization of a Sprinkler Valve
Pressure Loss Optimization of a Sprinkler Valve
 
Part technical cleanliness/Millipore
Part technical cleanliness/MilliporePart technical cleanliness/Millipore
Part technical cleanliness/Millipore
 
Dynamic Stand-Alone Gas Detection System
Dynamic Stand-Alone Gas Detection SystemDynamic Stand-Alone Gas Detection System
Dynamic Stand-Alone Gas Detection System
 
Chakresh Tiwari Philips report
Chakresh Tiwari Philips reportChakresh Tiwari Philips report
Chakresh Tiwari Philips report
 
NDT
NDTNDT
NDT
 
PPT-on-false-air.ppt
PPT-on-false-air.pptPPT-on-false-air.ppt
PPT-on-false-air.ppt
 
Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...
 
Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...Study the effect of using ultrasonic membrane anaerobic system in treating su...
Study the effect of using ultrasonic membrane anaerobic system in treating su...
 

Plus de Ives Equipment

United Electric 2018 Short Form Product Catalog
United Electric 2018 Short Form Product CatalogUnited Electric 2018 Short Form Product Catalog
United Electric 2018 Short Form Product CatalogIves Equipment
 
Worcester Controls Industrial Valve and Actuator Catalog 2018
Worcester Controls Industrial Valve and Actuator Catalog 2018Worcester Controls Industrial Valve and Actuator Catalog 2018
Worcester Controls Industrial Valve and Actuator Catalog 2018Ives Equipment
 
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...Ives Equipment
 
Valves and Automation for Hygienic Use
Valves and Automation for Hygienic UseValves and Automation for Hygienic Use
Valves and Automation for Hygienic UseIves Equipment
 
United Electric TX200 Pressure Transmitter
United Electric TX200 Pressure TransmitterUnited Electric TX200 Pressure Transmitter
United Electric TX200 Pressure TransmitterIves Equipment
 
SITRANS FS230 Ultrasonic Flow Measurement
SITRANS FS230 Ultrasonic Flow MeasurementSITRANS FS230 Ultrasonic Flow Measurement
SITRANS FS230 Ultrasonic Flow MeasurementIves Equipment
 
Lead Free Regulations Impact on Selection of Potable Water Valves
Lead Free Regulations Impact on Selection of Potable Water ValvesLead Free Regulations Impact on Selection of Potable Water Valves
Lead Free Regulations Impact on Selection of Potable Water ValvesIves Equipment
 
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...Ives Equipment
 
United Electric Controls Product Catalog
United Electric Controls Product CatalogUnited Electric Controls Product Catalog
United Electric Controls Product CatalogIves Equipment
 
UE One Series Safety Transmitter
UE One Series Safety TransmitterUE One Series Safety Transmitter
UE One Series Safety TransmitterIves Equipment
 
Industrial Level Instrument Selection Guide
Industrial Level Instrument Selection GuideIndustrial Level Instrument Selection Guide
Industrial Level Instrument Selection GuideIves Equipment
 
Use of Process Analyzers in Fossil Fuel Plants
Use of Process Analyzers in Fossil Fuel PlantsUse of Process Analyzers in Fossil Fuel Plants
Use of Process Analyzers in Fossil Fuel PlantsIves Equipment
 
Technolog Water Utility Data Logger Cello 4S
Technolog Water Utility Data Logger Cello 4STechnolog Water Utility Data Logger Cello 4S
Technolog Water Utility Data Logger Cello 4SIves Equipment
 
SensAlert ASI Advanced Safety Integrity Gas Transmitter
SensAlert ASI Advanced Safety Integrity Gas TransmitterSensAlert ASI Advanced Safety Integrity Gas Transmitter
SensAlert ASI Advanced Safety Integrity Gas TransmitterIves Equipment
 
Instrumentation, Automation and Control Solutions for the Grain Industry
Instrumentation, Automation and Control Solutions for the Grain IndustryInstrumentation, Automation and Control Solutions for the Grain Industry
Instrumentation, Automation and Control Solutions for the Grain IndustryIves Equipment
 
Siemens Sitrans F VA 250
Siemens Sitrans F VA 250 Siemens Sitrans F VA 250
Siemens Sitrans F VA 250 Ives Equipment
 
Double Offset High Performance Butterfly Valves
Double Offset High Performance Butterfly ValvesDouble Offset High Performance Butterfly Valves
Double Offset High Performance Butterfly ValvesIves Equipment
 
Material Compatibility / Corrosion Guide for Process Equipment
Material Compatibility / Corrosion Guide for Process EquipmentMaterial Compatibility / Corrosion Guide for Process Equipment
Material Compatibility / Corrosion Guide for Process EquipmentIves Equipment
 
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...Ives Equipment
 
Accurate and adaptable temperature measurement in a streamlined package
Accurate and adaptable temperature measurement in a streamlined packageAccurate and adaptable temperature measurement in a streamlined package
Accurate and adaptable temperature measurement in a streamlined packageIves Equipment
 

Plus de Ives Equipment (20)

United Electric 2018 Short Form Product Catalog
United Electric 2018 Short Form Product CatalogUnited Electric 2018 Short Form Product Catalog
United Electric 2018 Short Form Product Catalog
 
Worcester Controls Industrial Valve and Actuator Catalog 2018
Worcester Controls Industrial Valve and Actuator Catalog 2018Worcester Controls Industrial Valve and Actuator Catalog 2018
Worcester Controls Industrial Valve and Actuator Catalog 2018
 
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...
Explosion Proof/Flame Proof (Ex D) Sensors for Hazardous Area Now Available i...
 
Valves and Automation for Hygienic Use
Valves and Automation for Hygienic UseValves and Automation for Hygienic Use
Valves and Automation for Hygienic Use
 
United Electric TX200 Pressure Transmitter
United Electric TX200 Pressure TransmitterUnited Electric TX200 Pressure Transmitter
United Electric TX200 Pressure Transmitter
 
SITRANS FS230 Ultrasonic Flow Measurement
SITRANS FS230 Ultrasonic Flow MeasurementSITRANS FS230 Ultrasonic Flow Measurement
SITRANS FS230 Ultrasonic Flow Measurement
 
Lead Free Regulations Impact on Selection of Potable Water Valves
Lead Free Regulations Impact on Selection of Potable Water ValvesLead Free Regulations Impact on Selection of Potable Water Valves
Lead Free Regulations Impact on Selection of Potable Water Valves
 
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...
Vector System Image Processor for Water in Hydrocarbons and Particles in Hydr...
 
United Electric Controls Product Catalog
United Electric Controls Product CatalogUnited Electric Controls Product Catalog
United Electric Controls Product Catalog
 
UE One Series Safety Transmitter
UE One Series Safety TransmitterUE One Series Safety Transmitter
UE One Series Safety Transmitter
 
Industrial Level Instrument Selection Guide
Industrial Level Instrument Selection GuideIndustrial Level Instrument Selection Guide
Industrial Level Instrument Selection Guide
 
Use of Process Analyzers in Fossil Fuel Plants
Use of Process Analyzers in Fossil Fuel PlantsUse of Process Analyzers in Fossil Fuel Plants
Use of Process Analyzers in Fossil Fuel Plants
 
Technolog Water Utility Data Logger Cello 4S
Technolog Water Utility Data Logger Cello 4STechnolog Water Utility Data Logger Cello 4S
Technolog Water Utility Data Logger Cello 4S
 
SensAlert ASI Advanced Safety Integrity Gas Transmitter
SensAlert ASI Advanced Safety Integrity Gas TransmitterSensAlert ASI Advanced Safety Integrity Gas Transmitter
SensAlert ASI Advanced Safety Integrity Gas Transmitter
 
Instrumentation, Automation and Control Solutions for the Grain Industry
Instrumentation, Automation and Control Solutions for the Grain IndustryInstrumentation, Automation and Control Solutions for the Grain Industry
Instrumentation, Automation and Control Solutions for the Grain Industry
 
Siemens Sitrans F VA 250
Siemens Sitrans F VA 250 Siemens Sitrans F VA 250
Siemens Sitrans F VA 250
 
Double Offset High Performance Butterfly Valves
Double Offset High Performance Butterfly ValvesDouble Offset High Performance Butterfly Valves
Double Offset High Performance Butterfly Valves
 
Material Compatibility / Corrosion Guide for Process Equipment
Material Compatibility / Corrosion Guide for Process EquipmentMaterial Compatibility / Corrosion Guide for Process Equipment
Material Compatibility / Corrosion Guide for Process Equipment
 
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...
Simple Ways to Maximize the Efficiency of Your Process Control Application Wh...
 
Accurate and adaptable temperature measurement in a streamlined package
Accurate and adaptable temperature measurement in a streamlined packageAccurate and adaptable temperature measurement in a streamlined package
Accurate and adaptable temperature measurement in a streamlined package
 

Dernier

VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performancesivaprakash250
 
result management system report for college project
result management system report for college projectresult management system report for college project
result management system report for college projectTonystark477637
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130Suhani Kapoor
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxpranjaldaimarysona
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Call Girls in Nagpur High Profile
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordAsst.prof M.Gokilavani
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSISrknatarajan
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Christo Ananth
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 

Dernier (20)

VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
result management system report for college project
result management system report for college projectresult management system report for college project
result management system report for college project
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptx
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 

Claus Sulfur Recovery Tail Gas Applying 100 Million Hours of Operational Time to the Next Generation Analyzer

  • 1. CLAUS SULFUR RECOVERY TAIL GAS APPLYING 100 MILLION HOURS OF OPERATIONAL TIME TO THE NEXT GENERATION ANALYZER Randy Hauer, Stuart Simmonds, Ed Pavina / AMETEK Process Instruments 455 Corporate Blvd, Newark DE, 19702 Bob McCartney / PBF Energy Paulsboro Refinery 600 Billingsport Road, Paulsboro, NJ 08066 KEYWORDS Sulfur Recovery, Modified Claus Process, H2S/SO2, Tail Gas Analyzer ABSTRACT The measurement of H2S/SO2 in Claus sulfur recovery unit (SRU) tail gas has been adequately addressed by UV spectroscopy for over 40 years. Reliability of the analytical principle was established in the first generation of analyzers and in the second generation sample handling was improved to the point where automatic control if not universal is at least considered the norm. With a deep understanding of the process and a population of 1,100 second generation analyzers it was possible in the third generation to address failure modes external to the analyzer. Reliability is not limited to the analyzer / sample system; it extends to the process and contains elements of health and safety. Analyzer professionals are compelled to look beyond what we consider our "deliverables", to address abnormal operations and bad piping design. The paper combines extensive feedback from analyzer professionals and a survey of sulfur recovery operators to address the external failure modes. Looking back on 100 million hours of operational time and one year of field testing the third generation analyzer the paper discusses reliability as viewed by the real end use customer; Operations. THE PROCESS, DEVELOPMENT OF THE TAIL GAS ANALYZER Sulfur recovery is the unit operation of converting hydrogen sulfide (H2S) to elemental sulfur and is an essential block in an integrated refinery complex or sour gas processing plant. Referred to as the Modified Claus process; in a sub stoichiometric first step 1/3rd of the H2S undergoes combustion to SO2 in a reaction furnace then passes through two or three catalytic converters to
  • 2. produce elemental sulfur. The Modified Claus process is capable of ~97-98% recovery efficiency(1) . The primary consideration for high efficiency is control of the (feed forward) air:gas ratio by trim (feedback) control using the H2S/SO2 tail gas analyzer. FIGURE 1. MODIFIED CLAUS SRU PROCESS & INSTRUMENT DRAWING Sulfur is primarily used for the production of sulphuric acid which is the largest traded chemical commodity in the world. While sulfur does have a commodity value in most cases the driving force for process measurement and control is largely an environmental concern. The US EPA Clean Air Act of 1970 and the ground-breaking study by Alberta Environment on The Capability of the Modified-Claus Process(1) coincided with the first attempts to control the Modified-Claus process using an on-line analyzer and closed loop control. The first report of an on-line tail gas analyzer was a technical paper presented at the 1970 Analysis Division Symposium based on a gas chromatograph installed at the Dow Chemical Freeport TX (USA) facility(2) . Amoco Oil and DuPont Process Instruments published a paper at the 1972 Analysis Division Symposium on an Ultraviolet (UV) based tail gas analyzer installed at the Amoco Whiting IN (USA) refinery(3) . Shortly after this in 1974 Western Research participated in a pilot study using a prototype UV-based tail gas analyzer at the Shell Waterton AB (Canada) gas plant. It was already noted in the 1972 paper that “the non-continuous response of a process GC is often inadequate to provide the close control”. Given the process dead time of 30 seconds (to 2 minutes or more at turndown) and the requirement of a continuous sampling subsequent developments by various manufacturers were all based on UV spectroscopy with extractive sampling techniques which remains the case today. There is a collective experience from ~2,700 tail gas analyzers (DuPont, Western Research and the amalgamation of these two; AMETEK Process Instruments); five models; two full generations; and now moving into a 3rd generation tail gas analyzer. The 3rd generation analyzer is the evolution of the close coupled 880 NSL (no sample line) analyzer with 1,100 units in service and ~100 million hours of operational time.
  • 3. It has been noted by experienced analyzer engineers that certain process analyzer applications are much more difficult, among these being crude tower feed/bottoms and Claus tail gas (4) . Further to this point it has been estimated that something less than half of all process analyzers are placed in control. Analyzers placed in control get a great deal of scrutiny from Operations. Failure due to adverse conditions and unexpected (abnormal) results are not differentiated from analyzer failure. Having a large population of analyzers dedicated to one process gave insight beyond just the analyzer. A critical mass of information was acquired to look beyond the “deliverables” to address failure modes that are external to the analyzer and sample handling. TWO INDUSTRY SURVEYS Looking back on 40 years of experience the 1st generation (~1972-1997) proved UV worked, the weak link was sample handling, steam jacketed pipe and a large volume of sample gas. The 2nd generation and the application of a packed reflux type demister improved sample handling reliability but closed loop control was still not universal with process problems blamed on the analyzer. For the 3rd generation the focus was on what could be addressed in the way of failure modes external to the analyzer. With the aid of two industry surveys, a close study of the 2nd generation analyzer was undertaken. For feedback from hands on professionals a voice of customer (VOC) survey polled analyzer technicians for their views. The 2nd generation top of the pipe analyzer has a strong following and many said “change nothing” in relation to the sample handling and “don’t change the size” (meaning do not shrink it down and make the oven hard to access with gloves on). In terms of desired new features the list included double block from the process (safety), a single steam pressure, automatic flow control, over range measurement and improved digital communications. The paper will expand on these aspects in detail. A second survey was conducted to obtain valuable end user experience in addition to analyzer professionals. Three distinct stakeholder groups in sulphur recovery were polled;  Front end engineering design (FEED) and start up engineers  SRU Operations  End user analyzer specialists (engineers and technicians) It was deemed important while measuring reliability not to restrict feedback to just the analyzer profession, recognizing that on many occasions Operations has a different opinion as to “the analyzer is working”. The survey was extended to Operations, start up engineers and process design specialists for their perspective. The spread in the response was suitable with all three individual groups representing between 20-40% of the responses. To the question of reliability ~90% rated tail gas analyzers “reliable most of the time / all of the time”. This response was gratifying based on a comparison question asked when conducting process training for Operators over the past 15 years; “Is the analyzer in control mode?” for which ~25% say “No”. The take away being that even when the analyzer is considered reliable Operations decides to use it in manual control.
  • 4. FIGURE 2. SRU INDUSTRY SURVEY ADDRESSING EXTERNAL MODES OF FAILURE, PREDICTING ADVERSE PROCESS CONDITIONS Risk can be managed and reduced by understanding the potential failure modes of an analyzer, its sample conditioning system and the process. The first step in the process is to define all of the critical conditions needed to make a valid measurement. The second step is to determine which diagnostic methods can be used to evaluate each of the conditions. Diagnostic methods require the collection or generation of data which is then evaluated against the conditional requirements and to provide either a direct or indirect indication of the potential failure modes. There are three types of diagnostics that are implemented in the analyzer. They are observational, model-based, and functional. These three diagnostic types are differentiated by the means through which the evaluated data is collected. This data is used to predict failure and manage reliability aspects internal to the analyzer / sample handling system and was covered in a previous paper (5). The following expands on the methods applied to predicting and reacting to failure modes which are external to the analyzer and sample handling system. Having a long history in sulfur recovery there was strong confidence to enter into the process world and address problems that were previously beyond our control. Carefully considering feedback from the process stakeholders’ attention was placed on the three main external failure modes; one is process based, one is utility based and one is environment based. 1. Failure Mode (Process): Entrained liquid sulfur (sulfur fog, sulfur mist) SRUs are designed to operate at a certain capacity and refinery SRUs in particular go through a wide range of turn down according to crude slate. When a SRU goes below half load a phenomena occurs where the over cooling of the gas stream causes sulfur to form a fog, small sub-micron drops of sulfur that defy agglomeration in the process demister pads (6) . The fog can then collect on the analyzer optics. Further complicating this is the attendant increase in process pressure (from ~1 to ~2 psig) causing unintended increase in the flow rate of the analyzer, drawing up more entrained sulfur.
  • 5. FIGURE 3. SULFUR ENTRAINMENT – FOG AND MIST The remedy is to adjust the sample flow rate relative to the process pressure. Simply measuring the flow is not sufficient and placing a flow sensor in direct contact with SRU process gas is not good practice. The 3rd generation analyzer monitors the pressure differential between the process and the sample aspiration and automatically adjusts the flow rate relative to the process pressure during the zero cycle (every 90 minutes). During start up, shut down, turn down, plugged rundowns and any time there is entrained sulfur leaving the final condenser the flow rate can be minimized to reduce sulfur mist take up. 2. Failure Mode (Utilities): Poor quality steam, bad steam traps Steam traps are known to commonly fail, at any given time 20 to 30% of steam traps are defunct. The process connection point for a tail gas analyzer depends must be maintained at 300F (sulfur freezes at 246F). To maintain the heat integrity a custom steam jacketed ball valve is used. The problem of wet steam or a plugged steam trap is not usually a sudden failure. Most often it is intermittent it can come and go with wet weather or from night to day or result in “choking” which gives the appearance of slow response. The remedy is an embedded RTD in the process connection flange which provides an alarm. This advanced warning is unequivocal it indicates compromised heat integrity, calling attention to and eliminates a problem that can be the cause of intermittent failure. 3. Failure Mode (Environment): Solar gain, latent heat, internal heat release The 2nd generation tail gas analyzer was the first to be close-coupled to the pipe. Placing the entire analyzer on the pipe eliminated the need for a shelter but also exposed the analyzer to solar gain and latent process heat. On a SRU the analyzer location is, as often or not is temperature compromised and engineering procurement Contractors (EPCs) rarely take care to add a sunshield to the analyzer.
  • 6. While the existing analyzer is rated to 122o F (50o C) there is empirical evidence in the form of premature component failures that the internal temperature was far beyond the 50o C rating. Solar gain is the principal cause and in some cases latent process heat but also the internal heat release from the oven was a contributing factor. The effect of temperature on analyzer electronics was very well covered in a paper from ISA-AD 2012 precisely quantifying how the life of electronic semiconductor circuits is affected adversely by temperature as indicated by the Arrhenius Equation. The equation predicts that for approximately each 10°C increase in operating temperature mean electronic component operating life is reduced by 50% (7) . One of the objectives for the 3rd generation analyzer was to increase the temperature rating to 60o C without the aid of any external cooling device (Vortec cooler, Peltier device, air conditioning). Given that the oven must be capable of 315o F (~160o C) operation it was somewhat in question if the heat release from the sample side onto the electronics side could be mitigated while also overcoming solar gain. A highly effective thermal insulating barrier was developed early in the design phase as well as a custom extruded aluminum convection heater replacing the off shelf oven heater. Advanced thermal management algorithms resulted in a net reduction of the internal temperature of the electronics enclosure by 15°C (27°F). The outcome is an industry first 60°C specification for ambient temperature and a 75% improvement in electronics life(7) . The analyzer is suitable for installation in hot climates with a sunshade being recommended. EVOLUTIONARY IMPROVEMENTS Digital communications and the HMI (human machine interface) were obviously due for updating, benefiting from improvements in electronics over the intervening eighteen years. For the sample handling and photometry great care was taken not to compromise the elements that worked well, avoid change for the sake of change. The following describes both categories. IMPROVING ON WHAT ALREADY WORKED WELL Photometer: For this application the measurement of H2S/SO2 does not require a high degree of precision or accuracy. Control is a simple binary action and too much can be made of comparing different methods of UV spectroscopy. The final control element is typically a butter fly valve sized to control 10% of the process air in the form as a trim air control loop which is manipulated its full range by the tail gas analyzer (8) . In steady state full load conditions the trim air (“air demand”) is moving inside a peak to peak band of +/- 1% which results in a +/- 10% valve movement. Because of process dead time the peak to peak period is ~30 seconds (at full load) and the controller proportional gain is set low. The result is the time domain resolution of a photo-diode UV analyzer is 10x the resolution of the valve (9) . Add to this that the control function is binary (“add air, cut air”, it never dwells at set point). The result is accuracy is not paramount; simplicity, support, reliability are the prime requirements.
  • 7. The photometer for the 3rd generation retains the 2nd generation Xenon flash lamp and improves upon the photometer validation algorithms. The light from the sample cell is divided into four separate channels using partially reflective and partially transmitting beam splitters. The UV photometer uses four separate detectors each measuring a different narrow range of UV wavelengths. The intensity information from each of the four detectors is used to calculate the H2S and SO2 concentrations using a factory established calibration. The temperature of the photometer is carefully controlled to optimize the accuracy and precision of the reported concentrations. During the zero and span cycle an automatic multi-point photometric span calibration is performed on each of the four channels. The source pulse sequencing is automatically varied to create an intensity calibration for each detector channel. It is uniquely capable of automatic intensity calibration but calibration with a (manual) removable filter was retained as customer feedback indicated this was a widely accepted method for Operations to be assured the analyzer is working. The dynamic range was improved to the extent the analyzer can measure 100% over range while maintaining linearity specifications. This is especially useful during an upset giving Operations a widow beyond the normal air control parameter. Tail Gas is one of those applications where the instrument data sheet stays slavishly attached to the standard range, not acknowledging the ability of a next generation analyzer to provide this valuable information. It is up to the analyzer profession to communicate improvements such as over-range to front end engineering (FEED) designers and instrument engineers who stay rooted to past history and dated data sheets. FIGURE 4. PHOTOMETER Demister: What was well proven in the 2nd generation design was the veracity of using demisting pads to coalesce sulfur drops. Straight gravity drainage using an unpacked column (“cold finger” probe) works well enough when there is minimal entrained sulfur or when the drops are large and easily agglomerate. Adverse process conditions (plugged run down, failure of the process demister pad, sulfur fog at turndown) require a packed column and reflux action. The sulfur demister borrows heavily from the experience of dynamic reflux (pyrolysis) gas sampling. The subject has been extensively covered at previous symposia with reference made to the ability to prevent clouding of optical windows (10) . The hallmarks of a reflux sampler are surface area, vertical mounting and being close coupled but external to the process (11) . One of the 3rd generation improvements borrowed from the reflux sampler was mounting the demister pads on
  • 8. a shaft with retention rings (11) which in the case of the tail gas demister keeps the mesh pads in firmly in place during the hot condensate steam blow back for ammonia salts. FIGURE 5. DEMISTER Incomplete combustion of ammonia in the SRU process reacts with sulfur compounds to form ammonia salts, which can plug the demister. These salts are not removed by air blow-back but are water soluble. Hot condensate steam blow-back provides a steam propelled blast of hot water to dissolve any salts that have accumulated within the demister. An example of what these salts look like in the process and propensity to cause pressure drop is shown in the following figure. FIGURE 6. AMMONIA SALTS ON CONDENSER TUBE SHEET
  • 9. The analyzer is also equipped with anti-clogging hot air blowback feature that is automatically initiated if plugging is detected by the smart diagnostics. The 3rd generation demister retains the double demister pads (316 SS top and PTFE bottom). A change was made to connect up with vacuum compression o-ring (VCO) fittings in place of tube fittings making it easy to access and requiring minutes to disassemble. While not yet common in the process world VCO fittings have a place as they allow technicians easy access components inside a confined space like an oven. NEW IMPROVEMENTS Digital Interface: In our industry fifteen years is a typical life span of an analyzer while graphical interface and digital communications goes through several incarnations in the same period. The challenge is to stay up to date while at the same time finding commonality across a range of products for human engineering as well as cost of manufacturing and ongoing support. The next generation tail gas analyzer followed closely behind the 5th generation WDG oxygen analyzers and the same local HMI was implemented. AMEVision is an icon driven graphical interface with a color display for local communication with the analyzer. It provides screens showing trending functions, predictive maintenance indicators, analog output verification, and time stamped alarm/event log. Remote communication is via a standard PC web browser enabled interface (no software need be installed) over TCP/IP Ethernet. This point raises an anomaly that is perhaps unique to sulfur recovery but in any case is worth revisiting on every new project. Remote digital communication has been available on 2nd generation tail gas analyzers (two models, ~1,800 units) for the past 17 years. Of that total fewer than 10% are tied into remote digital communications. At least half have been new SRUs, grass roots project. The question to ask of EPCs (and end users) why they deem it acceptable to go on to the sulfur unit when it could all be done from the safety of the analyzer shop using a remote digital connection. This has hazardous operation implications and should be on top of the list for any analyzer engineer on a project to say “and why not?” UTILITIES There was opportunity to reduce the utilities. The Division 2/Zone 2 analyzer does not require an EExP purge, this reduces the instrument air volume and eliminates damage from wet instrument air. Steam as a utility is reduced to a single pressure. The demister pads held firm in place on a shaft can now withstand medium pressure (MP) steam and can utilize the same MP steam required for the process connection valve to maintain 300o F (150o C). SAFETY Hazardous Operation (“HazOp”) review is becoming more prevalent in the world of process analyzers. It has been said that sulfur recovery is not as well understood because it is a chemical plant in a refinery world and that is largely true. Sulfur recovery while not any more inherently
  • 10. dangerous than other refinery unit operations has the stigma of H2S. With that in mind safety features were added for with future requirements in mind. Three major refining companies in the USA often require double block isolation at their sites. This was not entirely possible in the 2nd generation analyzer but has been implemented in the 3rd generation analyzer with two isolation valves at the process access valve on both the sample and the vent. The sample conditioning components are external to the process and it is possible to remove the probe from the process under live conditions with zero egress through the ConaxTM fitting. The sample handling components are a safe distance from the process (500 mm/18”, which is considered the boundary for “safe distance”) To close this subject it is worthwhile to look at the safety implications of having to work with a compromised sample point. A modern SRU and tail gas treating unit when built as a single project is typically conjoined with a short vertical section of pipe. When analyzer professionals get zero input into at the piping design stage they are tied into a situation that can compromise both function and safety. For SRU-TGTUs the problems are confined quarters (egress), access, and process heat amongst others (9) . It is incumbent on our profession to interject ourselves at an early stage of a project to take a leading role in sample point selection and to take advantage of the available knowledge base FIGURE 7. COMPROMISED SAMPLE POINT LOCATION FIELD TRIAL OF 3rd GENERATION ANALYZER A search for a suitable site to conduct field trials commenced in late 2013; the prime requirement being close proximity to the factory located in Newark DE. There are a number of refineries in the area but SRUs are normally only in turn around every 4 years and a pool of suitable sites is therefore quite limited. The PBF Paulsboro NJ refinery had an available sample point location
  • 11. and consented to installing the new analyzer by hot-tapping a process connection valve. Like the 2nd generation analyzer the 3rd generation analyzer can be installed while the process is hot (i.e. ingress and egress of the probe via a Conax ™ gland) as long as a one inch full port ball valve is in place as the process connection. The hot tap was made and the model 888 analyzer installed at PBF Paulsboro in April 2014. Operations considered keeping the other analyzer in service in case the new analyzer required intervention. This would have added considerable cable and utility work. Given the additional cost, the confidence in the new analyzer was high and the opinion of analyzer professionals that “two watches” is never good the decision was made to go solo with the new analyzer. The cut over was made in one day and the new analyzer was placed in control by Operations same day. During the first 4 months of the trial there were no major outages for the analyzer. There was concern that the process connection valve used for the hot-tap was not steam jacketed but as it was close coupled there were no immediate problems. Later in July during a cold rain there was intermittent plugging and this was remedied with an off the shelf piece of Contra-Trace ™ from Controls Southeast Inc (CSI) to match this specific valve body. CSI specializes in sulfur recovery applications but the product is well applied to any analyzer heat tracing requirement and is much preferable to wrapping in SS tubing which in this service serves no useful purpose. There was only minimum intervention in the first seven months, modifications were restricted to software changes, availability was >99%. Operating conditions went through the normal range of events and the sample handling system performed to expectations with zero complaints from operations. Given the anomalies and the degree of difficulty associated with this type of process analyzer application the testing program allowed for up to 12 months of evaluation to observe long term effects on the components. In early November a heater fuse failed, it was deemed to be under rated and a larger fuse installed. In late November/early December condensed water was discovered in the sealed photometer box as well as on the detector board and the mystery as to how water was getting in. Wet instrument air was quickly eliminated. After some investigation a leaking check valve from a well known supplier was determined to be the root cause, it was intermittently failing and allowing small amounts of wet process gas to find a pathway into the photometer. The check valve was replaced with one from another supplier. In late early January there was observation of something off-gassing on the optics causing a slow decay of the measured light (compensated for but eventually causing low light level alarms). The photometer mirrors were replaced so the material could be extracted and analyzed. The root cause was a small silicon based seal which was replaced with another material. At the time of writing there were no other problems. The evaluation will run until March at which time the analyzer design will be turned over to production and manufacturing will commence.
  • 12. FIGURE 8. FIELD TRIAL PBF PAULSBORO REFINERY DISCUSSION AND CONCLUSIONS Process analytics is a young industry (approximately 50-60 years) with a product life cycle that can be relatively long. Most vendors think in terms of 15-20 year operation before considering replacement. Reviewing an application with a known degree of difficulty such as sulfur recovery is not complete without addressing the external failure modes that potentially occur within the process. As the analyzer results are the point of contact between the process and operations, these external failure modes are more often ascribed as “analyzer failures” and in all cases Operations will always have the last word; “we trust it/we don’t trust it”. First and 2nd generation tail gas analyzers have successfully concluded the measurement technique and closed loop control respectively. Now approaching the third generation of tail gas analyzers affords the opportunity to move beyond the borders of the analyzer itself and step into the process world, evaluating the process and responding to external failure modes. It takes a step further by educating the operators on the analytical technique, the sample system as well as the anomalies of the process to arrive at a holistic concept of reliability. Finally, it is imperative when approaching the process industry with a 3rd generation of tail gas analyzer that a long term field trial (minimum of 12 months) has been concluded. Even with many years of experience it takes time for hidden flaws to be revealed. Only after a successful field trial can an analyzer be introduced with confidence to an industry that expects a high level of reliability.
  • 13. ACKNOWLEDGMENTS The authors wish to thank the PBF Energy Paulsboro refinery for installing the analyzer for the field trial. In addition, to thank Bob McCartney and Ed Pavina for their support in the field REFERENCES 1) Paskall, H.G. "Capability of the Modified Claus Process." 1979. 2) Jackson, J.L, P.V Maynard. “The Analysis & Control of the H2S/SO2 Ratio in a Sulfur Recovery Plant Tail Gas Stream by Process Gas Chromatography”, ISA Analysis Division Symposium, Pittsburgh PA, May, 1970 3) Saltzman, R., Dr. E.B. Hunt. “A Photometric Analyzer System for Monitoring & Control of the [H2S]/[SO2] Ratio in Sulfur Recovery Plants”, ISA Analysis Division Symposium, San Francisco CA, May 1972 4) Novak, D., “Process Analytics: Are There Dinosaurs Among Us? - The Stigma of Process Analytics” and Dubois, R., “Process Analytics: Are There Dinosaurs Among Us? - Myths and Mistakes That May Contribute to Our Extinction”, Panel Discussion, ISA Automation Week Houston TX, October, 2010 5) Myles, T “Design Features of a Process Analyzer to Reduce Unscheduled Downtime”. The 59th ISA Analysis Division Symposium, Baton Rouge, LA May 5-8, 2014 6) Bohme, G., J. Sames. "The Seven Deadly Sins of Sulphur Recovery". The British Sulphur Conference, Calgary, Alberta, Canada, October 1999. 7) Farmer, B., U. Gellert, T. Campbell. “Ambient Temperature Testing Issues for Process Analyzers”, The 57th ISA Analysis Division Symposium, Anaheim, CA April 22-26, 2012 8) Neumeister, L., Controlling Sulphur Plants: Practical Issues Surrounding Control Strategies, Controller Hardware, Control Equipment Performance and Tuning, September 2000 9) Al-Misfer, A., S.Vedula, R. Hauer, Z. Juddy, “Process Analyzer Best Practices for Sulfur Recovery, Enhanced Claus and Tail Gas Treating”. Sour Oil & Gas Advanced Technology (SOGAT) Conference, Abu Dhabi, UAE March, 2009 10) Clemons, J., T Dye. “Complex Sampling Problems and Real World Solutions: Dynamic Reflux Sampling”. The 47th ISA Analysis Division Symposium, Denver CO, April 2002
  • 14. 11) Smith, S. “Pyrolysis Gas Sampling”. The 56th ISA Analysis Division Symposium, League City TX, April 2011 12) Hauer, R., Z. Juddy, Y. Yoshikane, J. Sames “The Seven Deadly Sins of Process Analyzer Applications”, SOGAT Conference Abu Dhabi, UAE March, 2011