MRI Hot Topics
Motion Correction for MR Imaging




                                   s
                                   medical
Motion Correction for MR Imaging
Kyle A. Salem, PhD



Motion Artifacts in a Clinical Setting                       or be scanned with another, less ideal, modality.
Patient motion is probably the most common cause             Integrated Parallel Acquisition Techniques (iPAT)
of artifacts in today’s MR imaging. This can range           significantly reduce scan times and are useful, but in
from physiological motion, such as respiratory or            a few cases strategies for motion control are needed.
cardiac movement, to physical movement by the                In the setting of the MR department, fast, effective
patient. Physical movement may be due to voluntary           motion control provides better care, serves more
movements or involuntary movements as in                     patients (which increases revenue), and makes
Parkinsonism. Physiologic motion can be controlled           scanning more efficient by eliminating repeat
by gating or in the sequence design, so it is really         scanning (which could potentially increase profit).
the physical patient movement that creates the
                                                             Current Strategies for Motion Control
largest concern and will be addressed here.
                                                             Today, the most common strategy for handling
The reality is that patients are often in pain and may       motion artifacts is to use retrospective motion
or may not be cooperative. Many times, the patient           correction. These post-processing approaches use
for whom the results are most needed is also the             a variety of algorithmic, iterative approaches applied
patient who can be the least cooperative. Beyond             in the image domain as well as in k-space.1 While
that, the longer a patient spends in the scanner the         they can be effective and are certainly applicable
more likely they are to become restless, agitated, or        to any acquired image, once the data is corrupted
nervous, which again leads to motion in the images.          by motion, you cannot recover ideal data. A better
                                                             strategy, as seen in some of the newer techniques
Patient Motion and the Clinical Exam
                                                             and research, is to perform the motion control
Since most exams in the chest and abdomen are
                                                             during the acquisition, or “inline”, so that the data
dominated by physiologic motion, the goal of most
                                                             is never corrupted.
motion correction algorithms is to produce ideal
images in the head or extremities. In the clinic a           Navigators
typical head examination consists of T1, T2 Turbo            The fastest “inline” method of controlling motion
Spin Echo (TSE), Dark-Fluid (FLAIR) inversion                is the “navigator” technique, 1D-PACE (Prospective
recovery (IR), and possibly a diffusion scan (EPI or         Acquisition Correction) to Siemens.2 This technique
TSE). Practical imaging times for these exams at             only adds about 30 ms to your scan and is typically
1.5T are about 3 minutes, 2 minutes, 3 minutes,              used for controlling physiologic motion, such as
and 1 minute, respectively. Notice that the sequences        respiratory motion in cardiac or abdominal exams.
with the longest scan times are also those where             It works by acquiring a single line of data from
the most motion artifact can occur, simply because           a pencil-shaped volume that crosses the boundary
the patient has more time to move. So the diffusion          of the diaphragm. This single line of data allows
scans, and to a lesser extent the T2 TSE sequences,          the scanner, in real time, to know the exact position
are not as susceptible to patient motion since they          of the diaphragm and trigger a scan only when
are also the fastest.                                        the diaphragm is at the appropriate position. So
But motion is still a very real problem that requires        misalignment is eliminated by only acquiring data
an efficient solution. Images that are corrupted by          when the anatomy is in a specific location.
motion artifacts can be rendered unreadable. Many            Siemens provides an enhancement to the traditional
sick patients are unable to cooperate long enough            navigator technique that makes it more robust and
to be imaged without artifacts. Some are simply              adds quality and reliability. The 2D-PACE technique
unable to complete an exam and must have the scan            acquires a low resolution gradient echo image in
repeated, be called back for another appointment,            approximately 100 ms. A low flip angle is used to




                                                         1
Figure 2: Finger tapping fMRI:
trained stimulus-correlated
1.5° rotation head motion.                             t-test:




                                             3D spatial filter
                                                    & t-test:


                                          Motion Correction:         None                   ART                    3D PACE




ensure that the magnetization is only minimally                       An alternative strategy is to use each acquired volume
effected maintaining final image homogeneity.                         to estimate any motion that may have occurred
Instead of a pencil beam, a box is positioned by                      using a similar rigid body technique. Each volume
the user and a 2D evaluation of the diaphragm                         is compared with the previous volume to calculate
position is used. The 2D-image provides much                          motion in six dimensions (3 translations and 3
more information making the technique broadly                         rotational axes). Then, in real time, the system uses
applicable. This lead to the emergence of excellent                   this information to adjust the acquisition parameters
quality “free breathing” or “multi-breath-hold” studies               (the imaged volume) so that the brain stays in the
like 3D TSE MRCP. (See Figure 1.)                                     same position in the image, no matter where the
                                                                      head has moved. Siemens is the only vendor to
                               It’s not hard to see,
                                                                      have implemented this strategy. It is called 3D-PACE
                               however, that navigator
                                                                      and is used with functional MRI exams.3 3D-PACE
                               techniques provide little
                                                                      prospectively adjusts gradient commands according
                               help in the brain where bulk
                                                                      to 6-dimensional head motion providing accurate
                               motion of the entire imaged
                                                                      fMRI results. (See Figure 2.)
                               volume is the problem.
                               Additionally, since this               Emerging Strategies
                               motion is not periodic and             In recent years there have been a number of new
Figure 1: T2 TSE 3D Restore;
                               there is no guarantee that             methods proposed for motion control, though few
PAT 2; 2D PACE; 1 mm x 1 mm    the head will ever move                have caught on. Two methods of note are PROPELLER
x 1.5 mm; Body Matrix,
Spine Matrix.
                               back to the original position,         and octant or cloverleaf navigators.
                               the user is left with only
                               post-processing techniques.            A technique termed “periodically rotated overlapping
                                                                      parallel lines with enhanced reconstruction”
3D Strategies                                                         (PROPELLER) was introduced by Pipe in 1999.4
Other methods have been developed specifically to                     Originally developed but not implemented on a
deal with this unpredictable bulk motion that is seen                 Siemens MR scanner, over the last 5 years this
in head examinations. Often rigid body registration                   method has received quite a bit of attention and
techniques are used as a post-processing strategy.                    has proven itself to be useful for motion correction
While from an ideal perspective, these should be                      in some cases. The PROPELLER technique collects
more than sufficient, practical issues such as partial                data in concentric rectangular strips rotated about
volume effects and through-plane signal changes                       the k-space origin. (See Figure 3.) The central region
can create tremendous problems if all of the motion                   of k-space is sampled for every strip, which
control is left until after the acquisition is complete.              (a) allows one to correct spatial inconsistencies




                                                                 2
The results, however, have been positive. In a number
                                                                                of studies, radiologists have preferred to view
                                                                                PROPELLER images over conventional T2-weighted TSE
                                                                                or EPI-DWI scans.6-8 One drawback, recently noted, was
                                                            Blade
                                                                                that EPI-DWI is superior to PROPELLER for quantitative
                                                            width               analysis (i.e. FA mapping) due to artifactual
                                                                                high-signal bands in the PROPELLER images.9
  X
                                                                                A concern for PROPELLER is total time required. In
                                                                                today’s clinical workflow, the goal is to decrease the
                Y                                                               amount of time needed to perform an examination,
                                                                                from patient set-up to scanning, etc. PROPELLER,
Figure 3: PROPELLER.                                                            however, increases the time it takes to perform a
                                                                                scan. Specifically, it adds a factor of pi/2 to the
in position, rotation, and phase between strips,                                acquisition due to the oversampling of the center
(b) allows one to reject data based on a correlation                            of k-space.4 To its credit, this does increase SNR due
measure indicating through-plane motion, and                                    to increased k-space sampling. Additionally, the series
(c) further decreases motion artifacts through                                  of corrections that are applied to the 2D slices in
an averaging effect for low spatial frequencies.                                PROPELLER to account for translational, rotation,
However, while PROPELLER does acquire data                                      and phase changes that may occur during acquisition
differently, all of the motion correction occurs                                each add to the reconstruction time. The intense
as a post-processing technique that attempts to                                 post-processing of the PROPELLER technique adds
correct corrupted data instead of inline correction                             about 2 to 3 minutes of additional reconstruction
to start with the best data.                                                    time per scan.6
In PROPELLER, you can acquire any 2D T2-weighted                                In all, while PROPELLER can reduce apparent motion
turbo spin echo (TSE) scan4 or a diffusion weighted                             artifacts that appear in T2 and diffusion weighted
multi-shot TSE scan.5 (See Figures 4A–4B.) These are                            TSE exams, it fails to seamlessly fit into today’s
                                                                                clinical workflow and is outperformed by traditional
                                                                                EPI techniques for quantitative diffusion analysis.
                                                                                Another new method for controlling motion is the
                                                                                octant or cloverleaf navigator approach. Developed
                                                                                by Dale and Van Der Kouwe in 2004, cloverleaf
                                                                                navigators are an improved k-space trajectory
                                                                                and associated mapping procedure that allows rapid,
                                                                                inline correction or rotations and translations with
 A                                    B
                                                                                minimal additional acquisition time.10, 11
Figures 4A–4B: Head with axial rotation in plane. Figure 4A head without        The cloverleaf trajectory covers each of the principal
propeller; Standard TSE. Figure 4B head with propeller; motion corrected.
                                                                                axes and three arcs, which connect each of the three
important images to be able to acquire without                                  axis pairs. A complete navigator set takes only 2.2 ms
image artifacts, however, this falls far short of being                         inside of each TR to be played out. The inherent 3D
able to provide a full clinical exam, with both T1                              nature of the navigator makes it appropriate for both
and IR, on an uncooperative patient. Additionally,                              2D and 3D acquisitions and it could be added to
the PROPELLER technique does not provide through-                               almost any imaging sequence. This provides distinct
plane motion correction.4,6 So while a patient can                              advantages since it can be used for any contrast,
rotate their head “in-plane”, through-plane motion                              any acquisition method, and could be applied to
is accounted for by eliminating data. It is rather                              an entire study for an uncooperative patient.
unrealistic to think that a real patient would only
                                                                                The navigator correction system works in concert
move their head in the imaging plane.
                                                                                with a map of k-space in the vicinity of the navigator




                                                                            3
acquired in a short preliminary mapping sequence of           1. Medley M and Yan H. Motion artifact suppression:
15 s. By playing out the navigator in each TR, a linear          A review of post-processing techniques. Magnetic
                                                                 Resonance Imaging. 10(4): 627-635, 1992.
mapping between the acquired navigator and the
preliminary map allows the angle of rotation to be            2. Ehman RL and Felmlee JP. Adaptive technique for high-
                                                                 definition MR imaging of moving structures. Radiology.
determined with a simple matrix multiplication.
                                                                 173(1): 255-263, 1989.
The changes due to translational and rotational
                                                              3. Thesen S, Heid O, Mueller E and Schad LR. Prospective
motion are then fed back to the MR system that                   acquisition corrections for head motion with image-
applies gradient corrections during the acquisition.             based tracking for real-time fMRI. Magnetic Resonance
This technique carries with it the advantage of                  in Medicine. 44(3): 457-465, 2000.
finding and correcting motion during the acquisition,         4. Pipe JG. Motion correction with PROPELLER MRI:
as opposed to using post-processing.                             application to head motion and free-breathing
                                                                 cardiac imaging. Magnetic Resonance in Medicine.
A very new and promising technique, cloverleaf                   42(5): 963-969, 1999.
navigators have shown great promise for inline                5. Pipe JG, Farthing VG and Forbes KP. Multi-shot diffusion
control of motion without the drawback of additional             weighted FSE using PROPELLER MRI. Magnetic
scan and reconstruction time. It is hoped that this              Resonance in Medicine. 47(1): 42-52, 2002.
will provide a robust method for motion control,              6. Forbes KP, Pipe JG, Bird CR and Heiserman JE. PROPELLER
applicable to today’s demanding clinical environment,            MRI: clinical testing of a novel technique for
but to date this technique is unproven clinically.               quantification and compensation of head motion.
                                                                 Journal of Magnetic Resonance Imaging. 14(3): 215-
Conclusion                                                       222, 2001.
A series of motion correction techniques have                 7. Forbes KP, Pipe JG, Karis JP, Farthing V and Heiserman JE.
                                                                 Brain imaging in the usedated pediatric patient:
been proposed over the years. There is common
                                                                 comparison of periodically rotated overlapping parallel
understanding that the underlying goal is to provide             lines with enhanced reconstruction and single-shot fast
MRI to everyone who needs a scan. While many                     spin-echo sequences. AJNR American Journal of
believe this is best achieved through motion control,            Neuroradiology. 24(5): 794-798, 2003.
there are still a number of retrospective correction          8. Forbes KP, Pipe JG, Karis JP and Heiserman JE. Improved
algorithms available. The new techniques, while                  image quality and detection of acute cerebral infarction
promising, have their drawbacks, as well. The hope               with PROPELLER diffusion-weighted imaging. Radiology.
                                                                 225(2): 551-555, 2002.
is that through further refinement and innovation,
                                                              9. Abe O, Mori H, Aoki S, Kunimatsu A, Hayashi N,
a robust method (or combination of methods) will
                                                                 Masumoto T, Yamada H, Masutani Y, Kabasawa H and
provide perfect motion control for all patients.                 Ohtomo K. Periodically rotated overlapping parallel lines
                                                                 with enhanced reconstruction-based diffusion tensor
                                                                 imaging: comparison with echo planar imaging-based
                                                                 diffusion tensor imaging. Journal of Computer Assisted
                                                                 Tomography. 28(5): 654-660, 2004.
                                                              10. van der Kouwe AJ and Dale AM. Rapid real-time
                                                                  prospective rigid body motion correction during
                                                                  imaging using clover-leaf navigators. Proceedings of
                                                                  the International Society of Magnetic Resonance in
                                                                  Medicine. 11: 95, 2004.
                                                              11. van der Kouwe AJ and Dale AM. Real-time motion
                                                                  correction using octant navigators. NeuroImage.
                                                                  13(6): S48, 2001.




                                                          4
On account of certain regional limitations of sales rights
and service availability, we cannot guarantee that all
products included in this brochure are available through
the Siemens sales organization worldwide. Availability
and packaging may vary by country and is subject to
change without prior notice. Some/All of the features
and products described herein may not be available in
the United States.

The information in this document contains general
technical descriptions of specifications and options as
well as standard and optional features which do not
always have to be present in individual cases.

Siemens reserves the right to modify the design,
packaging, specifications and options described herein
without prior notice. Please contact your local Siemens
sales representative for the most current information.

Note: Any technical data contained in this document
may vary within defined tolerances. Original images
always lose a certain amount of detail when reproduced.


Siemens Medical Solutions USA, Inc.                          Siemens Medical
Magnetic Resonance Division                                          Solutions that help
51 Valley Stream Parkway
Malvern, PA 19355-1406 USA
Telephone: 1-888-826-9702
www.usa.siemens.com/medical                                          © 2005 Siemens Medical Solutions
                                                                     Order No. A91004-M2220-M100-12-4A00
                                                                     Printed in USA

Contenu connexe

Tendances

DIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareDIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareCarestream
 
Neurorobotics and Advances in rehabilitation engineering
Neurorobotics and Advances in rehabilitation engineeringNeurorobotics and Advances in rehabilitation engineering
Neurorobotics and Advances in rehabilitation engineeringBhaskarBorgohain4
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...Carestream
 
NIH TGM Guo 38 Real-Time 3D
NIH TGM Guo 38 Real-Time 3DNIH TGM Guo 38 Real-Time 3D
NIH TGM Guo 38 Real-Time 3DLiheng Guo
 
Altaire Performance Paper
Altaire Performance PaperAltaire Performance Paper
Altaire Performance PaperAvneesh Nigam
 
Magnetom flash 51_scmr_issue_final-00308068
Magnetom flash 51_scmr_issue_final-00308068Magnetom flash 51_scmr_issue_final-00308068
Magnetom flash 51_scmr_issue_final-00308068Jhon Arriaga Cordova
 
Event recognition image & video segmentation
Event recognition image & video segmentationEvent recognition image & video segmentation
Event recognition image & video segmentationeSAT Journals
 
Bone Suppression for Chest Radiographic Images
Bone Suppression for Chest Radiographic ImagesBone Suppression for Chest Radiographic Images
Bone Suppression for Chest Radiographic ImagesCarestream
 
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...Reconstruction of the upper human femur from microCT images and FEM(Post-grad...
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...Katerina Stamou
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareCarestream
 
4 D Adaptive Radiotherapy & Tomotherapy
4 D Adaptive Radiotherapy & Tomotherapy4 D Adaptive Radiotherapy & Tomotherapy
4 D Adaptive Radiotherapy & Tomotherapyfondas vakalis
 
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Images
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR ImagesFuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Images
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Imagesijfls
 
Detection of exudates draft
Detection of exudates draftDetection of exudates draft
Detection of exudates draftnitheshksuvarna
 
Aneurysm Enhancement Poster AMIA 2009
Aneurysm Enhancement Poster AMIA 2009Aneurysm Enhancement Poster AMIA 2009
Aneurysm Enhancement Poster AMIA 2009ktdiedrich
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTMelissa McClement
 
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...sipij
 
Paulin hansen.2011.gaze interaction from bed
Paulin hansen.2011.gaze interaction from bedPaulin hansen.2011.gaze interaction from bed
Paulin hansen.2011.gaze interaction from bedmrgazer
 
Detection of exudates draft
Detection of exudates draftDetection of exudates draft
Detection of exudates draftnitheshksuvarna
 

Tendances (20)

DIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareDIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus Software
 
Neurorobotics and Advances in rehabilitation engineering
Neurorobotics and Advances in rehabilitation engineeringNeurorobotics and Advances in rehabilitation engineering
Neurorobotics and Advances in rehabilitation engineering
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
 
NIH TGM Guo 38 Real-Time 3D
NIH TGM Guo 38 Real-Time 3DNIH TGM Guo 38 Real-Time 3D
NIH TGM Guo 38 Real-Time 3D
 
Tomo Qa
Tomo QaTomo Qa
Tomo Qa
 
Altaire Performance Paper
Altaire Performance PaperAltaire Performance Paper
Altaire Performance Paper
 
Magnetom flash 51_scmr_issue_final-00308068
Magnetom flash 51_scmr_issue_final-00308068Magnetom flash 51_scmr_issue_final-00308068
Magnetom flash 51_scmr_issue_final-00308068
 
Event recognition image & video segmentation
Event recognition image & video segmentationEvent recognition image & video segmentation
Event recognition image & video segmentation
 
Bone Suppression for Chest Radiographic Images
Bone Suppression for Chest Radiographic ImagesBone Suppression for Chest Radiographic Images
Bone Suppression for Chest Radiographic Images
 
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...Reconstruction of the upper human femur from microCT images and FEM(Post-grad...
Reconstruction of the upper human femur from microCT images and FEM(Post-grad...
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization Software
 
4 D Adaptive Radiotherapy & Tomotherapy
4 D Adaptive Radiotherapy & Tomotherapy4 D Adaptive Radiotherapy & Tomotherapy
4 D Adaptive Radiotherapy & Tomotherapy
 
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Images
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR ImagesFuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Images
Fuzzy Clustering Based Segmentation of Vertebrae in T1-Weighted Spinal MR Images
 
Detection of exudates draft
Detection of exudates draftDetection of exudates draft
Detection of exudates draft
 
Aneurysm Enhancement Poster AMIA 2009
Aneurysm Enhancement Poster AMIA 2009Aneurysm Enhancement Poster AMIA 2009
Aneurysm Enhancement Poster AMIA 2009
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRT
 
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...
INHIBITION AND SET-SHIFTING TASKS IN CENTRAL EXECUTIVE FUNCTION OF WORKING ME...
 
Paulin hansen.2011.gaze interaction from bed
Paulin hansen.2011.gaze interaction from bedPaulin hansen.2011.gaze interaction from bed
Paulin hansen.2011.gaze interaction from bed
 
Detection of exudates draft
Detection of exudates draftDetection of exudates draft
Detection of exudates draft
 
Medphysics Planning
Medphysics PlanningMedphysics Planning
Medphysics Planning
 

En vedette

Advances in neuroimaging
Advances in neuroimagingAdvances in neuroimaging
Advances in neuroimagingfahad shafi
 
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...Dr. Muhammad Bin Zulfiqar
 
Radiological approach to Congenital skeletal dysplasia
Radiological approach to Congenital skeletal dysplasiaRadiological approach to Congenital skeletal dysplasia
Radiological approach to Congenital skeletal dysplasiaSachin Balutkar
 
CT ANATOMY OF PARA NASAL SINUSES
CT ANATOMY OF PARA NASAL SINUSESCT ANATOMY OF PARA NASAL SINUSES
CT ANATOMY OF PARA NASAL SINUSESPrasanna Kumaravel
 
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...drmcbansal
 

En vedette (7)

Advances in neuroimaging
Advances in neuroimagingAdvances in neuroimaging
Advances in neuroimaging
 
Mri Pulse Seqs
Mri Pulse SeqsMri Pulse Seqs
Mri Pulse Seqs
 
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...
Role of medical imaging in management of arteriovenous fistula Dr. Muhammad B...
 
Radiological approach to Congenital skeletal dysplasia
Radiological approach to Congenital skeletal dysplasiaRadiological approach to Congenital skeletal dysplasia
Radiological approach to Congenital skeletal dysplasia
 
CT ANATOMY OF PARA NASAL SINUSES
CT ANATOMY OF PARA NASAL SINUSESCT ANATOMY OF PARA NASAL SINUSES
CT ANATOMY OF PARA NASAL SINUSES
 
RADIOLOGY PRESENTATION
RADIOLOGY PRESENTATIONRADIOLOGY PRESENTATION
RADIOLOGY PRESENTATION
 
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...
Imaging in obstetrics & gynaecology (part 1- Gynaecological scans in benign c...
 

Similaire à Motion hot topics_broc

A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysis
A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysisA (quick) introduction to Magnetic Resonance Imagery preprocessing and analysis
A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysisStephen Larroque
 
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...INFOGAIN PUBLICATION
 
IRJET-Motion Segmentation
IRJET-Motion SegmentationIRJET-Motion Segmentation
IRJET-Motion SegmentationIRJET Journal
 
iaetsd Image fusion of brain images using discrete wavelet transform
iaetsd Image fusion of brain images using discrete wavelet transformiaetsd Image fusion of brain images using discrete wavelet transform
iaetsd Image fusion of brain images using discrete wavelet transformIaetsd Iaetsd
 
Analysis of Efficient Wavelet Based Volumetric Image Compression
Analysis of Efficient Wavelet Based Volumetric Image CompressionAnalysis of Efficient Wavelet Based Volumetric Image Compression
Analysis of Efficient Wavelet Based Volumetric Image CompressionCSCJournals
 
Poster RITS motion_correction
Poster RITS motion_correctionPoster RITS motion_correction
Poster RITS motion_correctionRania BERRADA
 
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...cscpconf
 
Final Progress Report- Farzana
Final Progress Report- FarzanaFinal Progress Report- Farzana
Final Progress Report- FarzanaFarzana Ali
 
IEEE Medical image Title and Abstract 2016
IEEE Medical image Title and Abstract 2016 IEEE Medical image Title and Abstract 2016
IEEE Medical image Title and Abstract 2016 tsysglobalsolutions
 
19 440 Publication in NAUN journal
19 440 Publication in NAUN journal 19 440 Publication in NAUN journal
19 440 Publication in NAUN journal Pawitra Masa-ah
 
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative OrganIOSR Journals
 
IEEE Bio medical engineering 2016 Title and Abstract
IEEE Bio medical engineering  2016 Title and Abstract IEEE Bio medical engineering  2016 Title and Abstract
IEEE Bio medical engineering 2016 Title and Abstract tsysglobalsolutions
 
Image restoration model with wavelet based fusion
Image restoration model with wavelet based fusionImage restoration model with wavelet based fusion
Image restoration model with wavelet based fusionAlexander Decker
 
Image processing techniques in nm 08,09
Image processing techniques in nm 08,09Image processing techniques in nm 08,09
Image processing techniques in nm 08,09Rutuja Solkar
 
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...IJECEIAES
 
Neurography Supplement to Issue 50
Neurography Supplement to Issue 50Neurography Supplement to Issue 50
Neurography Supplement to Issue 50Jhon Arriaga Cordova
 
Pattern Recognition final project
Pattern Recognition final projectPattern Recognition final project
Pattern Recognition final projectMaher Nadar
 

Similaire à Motion hot topics_broc (20)

A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysis
A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysisA (quick) introduction to Magnetic Resonance Imagery preprocessing and analysis
A (quick) introduction to Magnetic Resonance Imagery preprocessing and analysis
 
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...
Brain Image Fusion using DWT and Laplacian Pyramid Approach and Tumor Detecti...
 
IRJET-Motion Segmentation
IRJET-Motion SegmentationIRJET-Motion Segmentation
IRJET-Motion Segmentation
 
iaetsd Image fusion of brain images using discrete wavelet transform
iaetsd Image fusion of brain images using discrete wavelet transformiaetsd Image fusion of brain images using discrete wavelet transform
iaetsd Image fusion of brain images using discrete wavelet transform
 
Analysis of Efficient Wavelet Based Volumetric Image Compression
Analysis of Efficient Wavelet Based Volumetric Image CompressionAnalysis of Efficient Wavelet Based Volumetric Image Compression
Analysis of Efficient Wavelet Based Volumetric Image Compression
 
Poster RITS motion_correction
Poster RITS motion_correctionPoster RITS motion_correction
Poster RITS motion_correction
 
Whole brain optical imaging
Whole brain optical imagingWhole brain optical imaging
Whole brain optical imaging
 
Confer
ConferConfer
Confer
 
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...
ADAPTIVE, SCALABLE, TRANSFORMDOMAIN GLOBAL MOTION ESTIMATION FOR VIDEO STABIL...
 
Final Progress Report- Farzana
Final Progress Report- FarzanaFinal Progress Report- Farzana
Final Progress Report- Farzana
 
IEEE Medical image Title and Abstract 2016
IEEE Medical image Title and Abstract 2016 IEEE Medical image Title and Abstract 2016
IEEE Medical image Title and Abstract 2016
 
19 440 Publication in NAUN journal
19 440 Publication in NAUN journal 19 440 Publication in NAUN journal
19 440 Publication in NAUN journal
 
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
 
2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE
 
IEEE Bio medical engineering 2016 Title and Abstract
IEEE Bio medical engineering  2016 Title and Abstract IEEE Bio medical engineering  2016 Title and Abstract
IEEE Bio medical engineering 2016 Title and Abstract
 
Image restoration model with wavelet based fusion
Image restoration model with wavelet based fusionImage restoration model with wavelet based fusion
Image restoration model with wavelet based fusion
 
Image processing techniques in nm 08,09
Image processing techniques in nm 08,09Image processing techniques in nm 08,09
Image processing techniques in nm 08,09
 
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...
Neuroendoscopy Adapter Module Development for Better Brain Tumor Image Visual...
 
Neurography Supplement to Issue 50
Neurography Supplement to Issue 50Neurography Supplement to Issue 50
Neurography Supplement to Issue 50
 
Pattern Recognition final project
Pattern Recognition final projectPattern Recognition final project
Pattern Recognition final project
 

Plus de Jhon Arriaga Cordova (20)

Resonancia no ecoplanar en colesteatoma.pptx
Resonancia no ecoplanar en colesteatoma.pptxResonancia no ecoplanar en colesteatoma.pptx
Resonancia no ecoplanar en colesteatoma.pptx
 
Oído.pptx
Oído.pptxOído.pptx
Oído.pptx
 
Toshiba Visions 23
Toshiba Visions 23Toshiba Visions 23
Toshiba Visions 23
 
FieldStrength 49
FieldStrength 49FieldStrength 49
FieldStrength 49
 
FieldStrength 50 advanced neuro
FieldStrength 50 advanced neuroFieldStrength 50 advanced neuro
FieldStrength 50 advanced neuro
 
Ct somatom sessions-33
Ct somatom sessions-33Ct somatom sessions-33
Ct somatom sessions-33
 
Somatom sessions 32
Somatom sessions 32Somatom sessions 32
Somatom sessions 32
 
Somatom sessions 31
Somatom sessions 31Somatom sessions 31
Somatom sessions 31
 
Somatom sessions 30
Somatom sessions 30Somatom sessions 30
Somatom sessions 30
 
Somatom sessions 26
Somatom sessions 26Somatom sessions 26
Somatom sessions 26
 
Somatom sessions 25
Somatom sessions 25Somatom sessions 25
Somatom sessions 25
 
Somatom sessions 24
Somatom sessions 24Somatom sessions 24
Somatom sessions 24
 
Somatom session 29
Somatom session 29Somatom session 29
Somatom session 29
 
Somatom session 29
Somatom session 29Somatom session 29
Somatom session 29
 
Somatom session 28
Somatom session 28Somatom session 28
Somatom session 28
 
Somatom session 27
Somatom session 27Somatom session 27
Somatom session 27
 
Somatom session 27
Somatom session 27Somatom session 27
Somatom session 27
 
ISMRM Edition - Issue 49
ISMRM Edition - Issue 49ISMRM Edition - Issue 49
ISMRM Edition - Issue 49
 
Pediatric Imaging - Issue 47
Pediatric Imaging - Issue 47Pediatric Imaging - Issue 47
Pediatric Imaging - Issue 47
 
ISMRM Edition – Issue 58
ISMRM Edition – Issue 58ISMRM Edition – Issue 58
ISMRM Edition – Issue 58
 

Motion hot topics_broc

  • 1. MRI Hot Topics Motion Correction for MR Imaging s medical
  • 2. Motion Correction for MR Imaging Kyle A. Salem, PhD Motion Artifacts in a Clinical Setting or be scanned with another, less ideal, modality. Patient motion is probably the most common cause Integrated Parallel Acquisition Techniques (iPAT) of artifacts in today’s MR imaging. This can range significantly reduce scan times and are useful, but in from physiological motion, such as respiratory or a few cases strategies for motion control are needed. cardiac movement, to physical movement by the In the setting of the MR department, fast, effective patient. Physical movement may be due to voluntary motion control provides better care, serves more movements or involuntary movements as in patients (which increases revenue), and makes Parkinsonism. Physiologic motion can be controlled scanning more efficient by eliminating repeat by gating or in the sequence design, so it is really scanning (which could potentially increase profit). the physical patient movement that creates the Current Strategies for Motion Control largest concern and will be addressed here. Today, the most common strategy for handling The reality is that patients are often in pain and may motion artifacts is to use retrospective motion or may not be cooperative. Many times, the patient correction. These post-processing approaches use for whom the results are most needed is also the a variety of algorithmic, iterative approaches applied patient who can be the least cooperative. Beyond in the image domain as well as in k-space.1 While that, the longer a patient spends in the scanner the they can be effective and are certainly applicable more likely they are to become restless, agitated, or to any acquired image, once the data is corrupted nervous, which again leads to motion in the images. by motion, you cannot recover ideal data. A better strategy, as seen in some of the newer techniques Patient Motion and the Clinical Exam and research, is to perform the motion control Since most exams in the chest and abdomen are during the acquisition, or “inline”, so that the data dominated by physiologic motion, the goal of most is never corrupted. motion correction algorithms is to produce ideal images in the head or extremities. In the clinic a Navigators typical head examination consists of T1, T2 Turbo The fastest “inline” method of controlling motion Spin Echo (TSE), Dark-Fluid (FLAIR) inversion is the “navigator” technique, 1D-PACE (Prospective recovery (IR), and possibly a diffusion scan (EPI or Acquisition Correction) to Siemens.2 This technique TSE). Practical imaging times for these exams at only adds about 30 ms to your scan and is typically 1.5T are about 3 minutes, 2 minutes, 3 minutes, used for controlling physiologic motion, such as and 1 minute, respectively. Notice that the sequences respiratory motion in cardiac or abdominal exams. with the longest scan times are also those where It works by acquiring a single line of data from the most motion artifact can occur, simply because a pencil-shaped volume that crosses the boundary the patient has more time to move. So the diffusion of the diaphragm. This single line of data allows scans, and to a lesser extent the T2 TSE sequences, the scanner, in real time, to know the exact position are not as susceptible to patient motion since they of the diaphragm and trigger a scan only when are also the fastest. the diaphragm is at the appropriate position. So But motion is still a very real problem that requires misalignment is eliminated by only acquiring data an efficient solution. Images that are corrupted by when the anatomy is in a specific location. motion artifacts can be rendered unreadable. Many Siemens provides an enhancement to the traditional sick patients are unable to cooperate long enough navigator technique that makes it more robust and to be imaged without artifacts. Some are simply adds quality and reliability. The 2D-PACE technique unable to complete an exam and must have the scan acquires a low resolution gradient echo image in repeated, be called back for another appointment, approximately 100 ms. A low flip angle is used to 1
  • 3. Figure 2: Finger tapping fMRI: trained stimulus-correlated 1.5° rotation head motion. t-test: 3D spatial filter & t-test: Motion Correction: None ART 3D PACE ensure that the magnetization is only minimally An alternative strategy is to use each acquired volume effected maintaining final image homogeneity. to estimate any motion that may have occurred Instead of a pencil beam, a box is positioned by using a similar rigid body technique. Each volume the user and a 2D evaluation of the diaphragm is compared with the previous volume to calculate position is used. The 2D-image provides much motion in six dimensions (3 translations and 3 more information making the technique broadly rotational axes). Then, in real time, the system uses applicable. This lead to the emergence of excellent this information to adjust the acquisition parameters quality “free breathing” or “multi-breath-hold” studies (the imaged volume) so that the brain stays in the like 3D TSE MRCP. (See Figure 1.) same position in the image, no matter where the head has moved. Siemens is the only vendor to It’s not hard to see, have implemented this strategy. It is called 3D-PACE however, that navigator and is used with functional MRI exams.3 3D-PACE techniques provide little prospectively adjusts gradient commands according help in the brain where bulk to 6-dimensional head motion providing accurate motion of the entire imaged fMRI results. (See Figure 2.) volume is the problem. Additionally, since this Emerging Strategies motion is not periodic and In recent years there have been a number of new Figure 1: T2 TSE 3D Restore; there is no guarantee that methods proposed for motion control, though few PAT 2; 2D PACE; 1 mm x 1 mm the head will ever move have caught on. Two methods of note are PROPELLER x 1.5 mm; Body Matrix, Spine Matrix. back to the original position, and octant or cloverleaf navigators. the user is left with only post-processing techniques. A technique termed “periodically rotated overlapping parallel lines with enhanced reconstruction” 3D Strategies (PROPELLER) was introduced by Pipe in 1999.4 Other methods have been developed specifically to Originally developed but not implemented on a deal with this unpredictable bulk motion that is seen Siemens MR scanner, over the last 5 years this in head examinations. Often rigid body registration method has received quite a bit of attention and techniques are used as a post-processing strategy. has proven itself to be useful for motion correction While from an ideal perspective, these should be in some cases. The PROPELLER technique collects more than sufficient, practical issues such as partial data in concentric rectangular strips rotated about volume effects and through-plane signal changes the k-space origin. (See Figure 3.) The central region can create tremendous problems if all of the motion of k-space is sampled for every strip, which control is left until after the acquisition is complete. (a) allows one to correct spatial inconsistencies 2
  • 4. The results, however, have been positive. In a number of studies, radiologists have preferred to view PROPELLER images over conventional T2-weighted TSE or EPI-DWI scans.6-8 One drawback, recently noted, was Blade that EPI-DWI is superior to PROPELLER for quantitative width analysis (i.e. FA mapping) due to artifactual high-signal bands in the PROPELLER images.9 X A concern for PROPELLER is total time required. In today’s clinical workflow, the goal is to decrease the Y amount of time needed to perform an examination, from patient set-up to scanning, etc. PROPELLER, Figure 3: PROPELLER. however, increases the time it takes to perform a scan. Specifically, it adds a factor of pi/2 to the in position, rotation, and phase between strips, acquisition due to the oversampling of the center (b) allows one to reject data based on a correlation of k-space.4 To its credit, this does increase SNR due measure indicating through-plane motion, and to increased k-space sampling. Additionally, the series (c) further decreases motion artifacts through of corrections that are applied to the 2D slices in an averaging effect for low spatial frequencies. PROPELLER to account for translational, rotation, However, while PROPELLER does acquire data and phase changes that may occur during acquisition differently, all of the motion correction occurs each add to the reconstruction time. The intense as a post-processing technique that attempts to post-processing of the PROPELLER technique adds correct corrupted data instead of inline correction about 2 to 3 minutes of additional reconstruction to start with the best data. time per scan.6 In PROPELLER, you can acquire any 2D T2-weighted In all, while PROPELLER can reduce apparent motion turbo spin echo (TSE) scan4 or a diffusion weighted artifacts that appear in T2 and diffusion weighted multi-shot TSE scan.5 (See Figures 4A–4B.) These are TSE exams, it fails to seamlessly fit into today’s clinical workflow and is outperformed by traditional EPI techniques for quantitative diffusion analysis. Another new method for controlling motion is the octant or cloverleaf navigator approach. Developed by Dale and Van Der Kouwe in 2004, cloverleaf navigators are an improved k-space trajectory and associated mapping procedure that allows rapid, inline correction or rotations and translations with A B minimal additional acquisition time.10, 11 Figures 4A–4B: Head with axial rotation in plane. Figure 4A head without The cloverleaf trajectory covers each of the principal propeller; Standard TSE. Figure 4B head with propeller; motion corrected. axes and three arcs, which connect each of the three important images to be able to acquire without axis pairs. A complete navigator set takes only 2.2 ms image artifacts, however, this falls far short of being inside of each TR to be played out. The inherent 3D able to provide a full clinical exam, with both T1 nature of the navigator makes it appropriate for both and IR, on an uncooperative patient. Additionally, 2D and 3D acquisitions and it could be added to the PROPELLER technique does not provide through- almost any imaging sequence. This provides distinct plane motion correction.4,6 So while a patient can advantages since it can be used for any contrast, rotate their head “in-plane”, through-plane motion any acquisition method, and could be applied to is accounted for by eliminating data. It is rather an entire study for an uncooperative patient. unrealistic to think that a real patient would only The navigator correction system works in concert move their head in the imaging plane. with a map of k-space in the vicinity of the navigator 3
  • 5. acquired in a short preliminary mapping sequence of 1. Medley M and Yan H. Motion artifact suppression: 15 s. By playing out the navigator in each TR, a linear A review of post-processing techniques. Magnetic Resonance Imaging. 10(4): 627-635, 1992. mapping between the acquired navigator and the preliminary map allows the angle of rotation to be 2. Ehman RL and Felmlee JP. Adaptive technique for high- definition MR imaging of moving structures. Radiology. determined with a simple matrix multiplication. 173(1): 255-263, 1989. The changes due to translational and rotational 3. Thesen S, Heid O, Mueller E and Schad LR. Prospective motion are then fed back to the MR system that acquisition corrections for head motion with image- applies gradient corrections during the acquisition. based tracking for real-time fMRI. Magnetic Resonance This technique carries with it the advantage of in Medicine. 44(3): 457-465, 2000. finding and correcting motion during the acquisition, 4. Pipe JG. Motion correction with PROPELLER MRI: as opposed to using post-processing. application to head motion and free-breathing cardiac imaging. Magnetic Resonance in Medicine. A very new and promising technique, cloverleaf 42(5): 963-969, 1999. navigators have shown great promise for inline 5. Pipe JG, Farthing VG and Forbes KP. Multi-shot diffusion control of motion without the drawback of additional weighted FSE using PROPELLER MRI. Magnetic scan and reconstruction time. It is hoped that this Resonance in Medicine. 47(1): 42-52, 2002. will provide a robust method for motion control, 6. Forbes KP, Pipe JG, Bird CR and Heiserman JE. PROPELLER applicable to today’s demanding clinical environment, MRI: clinical testing of a novel technique for but to date this technique is unproven clinically. quantification and compensation of head motion. Journal of Magnetic Resonance Imaging. 14(3): 215- Conclusion 222, 2001. A series of motion correction techniques have 7. Forbes KP, Pipe JG, Karis JP, Farthing V and Heiserman JE. Brain imaging in the usedated pediatric patient: been proposed over the years. There is common comparison of periodically rotated overlapping parallel understanding that the underlying goal is to provide lines with enhanced reconstruction and single-shot fast MRI to everyone who needs a scan. While many spin-echo sequences. AJNR American Journal of believe this is best achieved through motion control, Neuroradiology. 24(5): 794-798, 2003. there are still a number of retrospective correction 8. Forbes KP, Pipe JG, Karis JP and Heiserman JE. Improved algorithms available. The new techniques, while image quality and detection of acute cerebral infarction promising, have their drawbacks, as well. The hope with PROPELLER diffusion-weighted imaging. Radiology. 225(2): 551-555, 2002. is that through further refinement and innovation, 9. Abe O, Mori H, Aoki S, Kunimatsu A, Hayashi N, a robust method (or combination of methods) will Masumoto T, Yamada H, Masutani Y, Kabasawa H and provide perfect motion control for all patients. Ohtomo K. Periodically rotated overlapping parallel lines with enhanced reconstruction-based diffusion tensor imaging: comparison with echo planar imaging-based diffusion tensor imaging. Journal of Computer Assisted Tomography. 28(5): 654-660, 2004. 10. van der Kouwe AJ and Dale AM. Rapid real-time prospective rigid body motion correction during imaging using clover-leaf navigators. Proceedings of the International Society of Magnetic Resonance in Medicine. 11: 95, 2004. 11. van der Kouwe AJ and Dale AM. Real-time motion correction using octant navigators. NeuroImage. 13(6): S48, 2001. 4
  • 6. On account of certain regional limitations of sales rights and service availability, we cannot guarantee that all products included in this brochure are available through the Siemens sales organization worldwide. Availability and packaging may vary by country and is subject to change without prior notice. Some/All of the features and products described herein may not be available in the United States. The information in this document contains general technical descriptions of specifications and options as well as standard and optional features which do not always have to be present in individual cases. Siemens reserves the right to modify the design, packaging, specifications and options described herein without prior notice. Please contact your local Siemens sales representative for the most current information. Note: Any technical data contained in this document may vary within defined tolerances. Original images always lose a certain amount of detail when reproduced. Siemens Medical Solutions USA, Inc. Siemens Medical Magnetic Resonance Division Solutions that help 51 Valley Stream Parkway Malvern, PA 19355-1406 USA Telephone: 1-888-826-9702 www.usa.siemens.com/medical © 2005 Siemens Medical Solutions Order No. A91004-M2220-M100-12-4A00 Printed in USA