SPECTROFLUOROMETER
Presented by: Nour Elrashed
Q.C master student
Damascus UniversityFaculty of Pharmacy
CONTENTS:
 Introduction
 the principle of spectrofluorometer
 What can specrofluorometer do?
 The components of specrofluorometer
 Fluorescence Spectra
 Fluorescence and chemical structure
 Applications of Spectrofluorometer
 New pharmaceutical studies conducted by
spectrofluorometer
 References
Emission
Molecular
Photoluminescence
Fluorescence
spectrofluorometer
Phosphorescence
Atomic
Absorption
HISTORY:
 The term fluorescence comes from the mineral
fluorspar (calcium fluoride) when Sir George G.
Stokes observed in 1852 that fluorspar would give
off visible light (fluoresce) when exposed to
electromagnetic radiation in the ultraviolet
wavelength.
THE PRINCIPLE OF SPECTROFLUOROMETER:
 It is an analytical device depends on the fluorescence
phenomenon which is a short-lived type of
photoluminescence created by electromagnetic excitation.
 That is, fluorescence is generated when a molecule
transmits from its ground state So to one of several
vibrational energy levels in the first excited electronic
state, S1, or the second electronic excited state, S2, both
of which are singlet states.
 Relaxation to the ground state from these excited states
occurs by emission of energy through heat and/or
photons.
 The difference between the excitation and emission
wavelengths is called the Stokes shift.
 Stokes’ studies of fluorescent substances led to the
formulation of Stokes’ Law, which states that the
wavelength of fluorescent light is always greater
than that of the exciting radiation. Thus, for any
fluorescent molecule, the wavelength of emission is
always longer than the wavelength of absorption.
WHAT IS THE FLUORESCENCE QUANTUM YIELD (ΦF)?
 It is the quantitative expression of the fluorescence
efficiency, which is the fraction of excited molecules
returning to the ground state by fluorescence.
 Quantum yields range from 1, when every molecule in
an excited state undergoes fluorescence, to 0 when
fluorescence does not occur.
 A molecule’s fluorescence quantum yield is
influenced by external Variables such as:
• temperature
• viscosity of solvent
• pH
 Increasing temperature generally decreases Φf
because more frequent collisions between the molecule
and the solvent increases external conversion.
 Decreasing the solvent’s viscosity decreases Φf for
similar reasons.
 For an analyte with acidic or basic functional groups, a
change in pH may change the analyte’s structure
and, therefore, its fluorescent properties.
 Emission of a photon
when the analyte
returns from a higher-
energy state to a lower-
energy state with the
opposite spin (triplet
excited state: An
excited state in which
unpaired electron spins
occur.).
 Average lifetime from
10-3 to 10+2 sec
 Emission of a photon
when the analyte
returns from a higher-
energy state to a lower-
energy state with the
same spin (singlet
excited state: An
excited state in which
all electron spins are
paired).
 Average lifetime from
10-9 to 10-7 sec
Phosphorescence Fluorescence
WHAT CAN SPECROFLUOROMETER DO?
 It has been used for the direct or indirect quantitative and
qualitative analysis by measuring the fluorescent
intensity F.
 It is relatively inexpensive and sensitive (the sensitivity
of fluorescence is approximately 1,000 times greater
than absorption spectrophotometric methods).
 fluorescent intensity F is dependent on both intrinsic
properties of the compound (fluorescence quantum yield
Φf), and on readily controlled experimental parameters
including:
• intensity of the absorbed light I0
• molar absorption coefficient Ɛ
• path length of the cell b
• concentration of the fluorophor in solution c
 At low concentrations of fluorophore, the fluorescence
intensity of a sample is essentially linearly proportional
to concentration.
 However, as the concentration increases, a point is
reached at which the intensity increase is progressively
less linear, and the intensity eventually decreases as
concentration increases further.
fluorescence intensity
concentrations
 The most common reason for this is Inner filter effect
that, as the absorbance of the sample increases, the
light intensity experienced by some of the fluorescent
molecules is lower than that experienced by others.
When excitation intensity decreases, so does
fluorescence emission intensity.
 It is generally necessary to use concentrations that
result in absorbance values of 0.1 or lower to observe
concentration dependent emission.
 As the concentration of molecules in a solution
increases, probability increases that excited molecules
will interact with each other and lose energy through
processes other than fluorescent emission. Any process
that reduces the probability of fluorescent emission is
known as quenching.
 Other parameters that can cause quenching include:
• presence of impurities
• increased temperature
• reduced viscosity of the solution media
THE COMPONENTS OF SPECROFLUOROMETER:
1. light source (75 to 450-W high-pressure xenon arc
lamp or Lasers)
2. excitation monochromator
3. sample holder (Quartz/Optical Glass/Plastic Cells)
4. emission monochromator
5. Detector (photomultiplier)
6. Most spectrofluorometers also have a reference
sample. The reference is generally a solution of a
strongly fluorescent molecule with a broad
absorbance spectrum such as rhodamine. The
reference is necessary to correct for lamp
output, especially when varying the excitation
wavelength, and to correct for differences in
detector sensitivity.
FLUORESCENCE SPECTRA:
 Photoluminescence spectra are recorded by measuring
the intensity of emitted radiation as a function of either
the excitation wavelength or the emission wavelength.
 The excitation spectra is determined by measuring the
emission intensity at a fixed wavelength , while varying
the excitation wavelength. It is useful for selecting the
best excitation wavelength for a quantitative or
qualitative analysis.
 The emission spectra is determined by measuring the
variation in emission intensity wavelength for a fixed
excitation wavelength.
FLUORESCENCE AND CHEMICAL STRUCTURE:
 Fluorescence is generally observed with molecules
where the lowest energy absorption is →*
transition, although some n→* transitions show weak
fluorescence.
 Most unsubstituted, nonheterocyclic aromatic
compounds show favorable fluorescence quantum
yields wich usually increases with the number of rings
and their degree of condensation. In
addition, substitution to the aromatic ring can have a
significant effect on Φf. For example, the presence of an
electron-withdrawing group, such as (NO2), decreases
Φf, whereas adding an electron-donating group, such as
(OH), increases Φf.
 Fluorescence also increases for aromatic ring
systems and for aromatic molecules with rigid planar
structures.
 The simple heterocyclics, such as
pyridine, furan, thiophene, and pyrrole do not exhibit
fluorescence; on the other hand, fused ring structures
ordinarily do.
SPECTROFLUOROMETER:APPLICATIONS OF
 Environmental Significance:
 To detect environmental pollutants such as polycyclic
aromatic hydrocarbons:
• pyrene
• benzopyrene
• organothiophosphorous pesticides
• carbamate insecticides
 Geology:
 Many types of calcite and amber will fluoresce under
shortwave UV. Rubies, emeralds, and the Hope
Diamond exhibit red fluorescence under short-wave UV
light; diamonds also emit light under X ray radiation.
SPECTROFLUOROMETER:APPLICATIONS OF
 Analytical chemistry
 to detect compounds from HPLC flow
 TLC plates can be visualized if the compounds or a
coloring reagent is fluorescent
 Biochemistry:
 used generally as a non-destructive way of tracking or
analysis of biological molecules (proteins)
 Possible direct or indirect analysis aromatic amino acids
(phenylalanine- tyrosine-tryptophan)
 Fingerprints can be visualized with fluorescent
compounds such as ninhydrin.
SPECTROFLUOROMETER:APPLICATIONS OF
 Medicine
 Blood and other substances are sometimes detected by
fluorescent reagents, particularly where their location was not
previously known.
 There has also been a report of its use in differentiating
malignant, bashful skin tumors from benign.
 Pharmacy:
 Possible direct or indirect analysis drugs such as:
• vitamins (vitamin A -vitamin B2 -vitamin B6 -vitamin B12 -
vitamin E -folic acid)
• catecholamines (dopamine-norepinephrine)
• Other drugs (quinine-salicylic acid–morphine-barbiturates –
LSD)
NEW PHARMACEUTICAL STUDIES CONDUCTED
BY SPECTROFLUOROMETER:
OPTIMIZATION AND VALIDATION OF SPECTROFLUORIMETRIC METHOD FOR
DETERMINATION OF CEFADROXILE AND CEFUROXIME SODIUM IN
PHARMACEUTICAL FORMULATIONS
.HYNEINE-BOULA,SMHMEDA,AALBASHIRE
CHEMISTRY DEPARTMENT, FACULTY OF SCIENCE, UNIVERSITY OF KHARTOUM, PO BOX 321, KHARTOUM, 11115, SUDAN.
 Abstract
A simple, accurate, precise and validated spectrofluorimetric method is proposed for
the determination of two cephalosporins, namely, cefadroxile (cefa) and cefuroxime
sodium (cefu) in pharmaceutical formulations. The method is based on a reaction
between cephalosporins with 1,2-naphthoquinone-4-sulfonate in alkaline medium, to
form fluorescent derivatives that are extracted with chloroform and subsequently
measured at 610 and 605 nm after excitation at 470 and 460 nm for cefa and cefu
respectively. The optimum experimental conditions have been studied. Beer's law is
obeyed over the concentrations of 20-70 ng/mL and 15-40 ng/mL for cefa and
cefu, respectively. The detection limits were 4.46 ng/mL and 3.02 ng/mL with a linear
regression correlation coefficient of 0.9984 and 0.998, and recoveries ranging 97.50-
109.96% and 95.73-98.89% for cefa and cefu, respectively. The effects of
pH, temperature, reaction time, 1,2-naphthoquinone-4-sulfonic concentration and
extraction solvent on the determination of cefa and cefu, have been examined. The
proposed method can be applied for the determination of cefa and cefu in
pharmaceutical formulations in quality control laboratories. Copyright © 2013 John
Wiley & Sons, Ltd.
ahead of print]Epub. [2481/bio.1002.10:doi.24Jan2013Luminescence.
cefadroxile
1,2-naphthoquinone-4-sulfonate
cefuroxime sodium
A HIGHLY SENSITIVE FLUORIMETRIC METHOD FOR DETERMINATION
VIACAPSULESANDFORMBULKITSINLENALIDOMIDEOF
DERIVATIZATION WITH FLUORESCAMINE
.NZLZOMANA,AHAKHEITB,NYHALILK,IAARWISHD
DEPARTMENT OF PHARMACEUTICAL CHEMISTRY, COLLEGE OF PHARMACY, KING SAUD UNIVERSITY,, P,O, BOX 2457, RIYADH, 11451, SAUDI
ARABIA. IDARWISH@KSU.EDU.SA.
ABSTRACT:
BACKGROUND: Lenalidomide (LND) is a potent novel thalidomide analog which demonstrated remarkable
clinical activity in treatment of multiple myeloma disease via a multiple-pathways mechanism. The strong
evidences-based clinical success of LND in patients has led to its recent approval by US-FDA under the trade
name of Revlimid® capsules by Celgene Corporation. Fluorimetry is a convenient technique for
pharmaceutical quality control, however there was a fluorimetric method for determination of LND in its bulk
and capsules.
RESULTS: A novel highly sensitive and simple fluorimetric method has been developed and validated for the
determination of lenalidmide (LND) in its bulk and dosage forms (capsules). The method was based on
nucleophilic substitution reaction of LND with fluorescamine (FLC) in aqueous medium to form a highly
fluorescent derivative that was measured at 494 nm after excitation at 381 nm. The factors affecting the
reaction were carefully studied and optimized. The kinetics of the reaction was investigated, and the reaction
mechanism was postulated. Under the optimized conditions, linear relationship with good correlation
coefficient (0.9999) was found between the fluorescence intensity and LND concentration in the range of 25-
300 ng/mL. The limits of detection and quantitation for the method were 2.9 and 8.7 ng/mL, respectively. The
precision of the method was satisfactory; the values of relative standard deviations did not exceed 1.4%. The
proposed method was successfully applied to the determination of LND in its bulk form and pharmaceutical
capsules with good accuracy; the recovery values were 97.8-101.4 ± 1.08-2.75%.
CONCLUSIONS: The proposed method is selective and involved simple procedures. In conclusion, the method is
practical and valuable for routine application in quality control laboratories for determination of LND.
.118-6-X153-1752/1186.10:doi.118):1(6;16Oct2012Cent J.Chem
lenalidomide
fluorescamine
INTIZANIDINEOFDETERMINATIONSPECTROFLUORIMETRICENSITIVES
PHARMACEUTICAL PREPARATIONS, HUMAN PLASMA AND URINE
THROUGH DERIVATIZATION WITH DANSYL CHLORIDE
.STLUU
DEPARTMENT OF ANALYTICAL CHEMISTRY, FACULTY OF PHARMACY, ISTANBUL UNIVERSITY, TURKEY. SEVGITATAR@YAHOO.COM
 Abstract
A sensitive spectrofluorimetric method was developed for the determination of
tizanidine in human plasma, urine and pharmaceutical preparations. The method
is based on reaction of tizanidine with 1-dimethylaminonaphthalene-5-sulphonyl
chloride (dansyl chloride) in an alkaline medium to form a highly fluorescent
derivative that was measured at 511 nm after excitation at 383 nm. The different
experimental parameters affecting the fluorescence intensity of tizanidine was
carefully studied and optimized. The fluorescence-concentration plots were
rectilinear over the ranges 50-500 and 20-300 ng/mL for plasma and
urine, respectively, detection limits of 1.81 and 0.54 ng/mL and quantification
limits of 5.43 and 1.62 ng/mL for plasma and urine, respectively. The method
presents good performance in terms of linearity, detection and quantification
limits, precision, accuracy and specificity. The proposed method was successfully
applied for the determination of tizanidine in pharmaceutical preparations. The
results obtained were compared with a reference method, using t- and F-tests.
.28Oct2011Epub.1367/bio.1002.10:doi.30-426):5(27Oct;-Sep2012Luminescence.
dansyl chloride
tizanidine
INHCLPAROXETINEOFDETERMINATIONPECTROFLUORIMETRICS
PHARMACEUTICALS VIA DERIVATIZATION WITH 4-
CHLORO-7- NITROBENZO-2-OXA-1,3-DIAZOLE (NBD-CL)
.HLMANSIE,NNANYE-LE,FELALB,MALSHW
DEPARTMENT OF ANALYTICAL CHEMISTRY, FACULTY OF PHARMACY, UNIVERSITY OF
MANSOURA, MANSOURA, EGYPT.
 Abstract:
A sensitive and simple spectrofluorimetric method has been developed and
validated for the determination of the antidepressant paroxetine HCl (PXT) in its
dosage forms. The method was based on coupling reaction of PXT with 4-
chloro-7-nitrobenzo-2- oxa-1,3-diazole (NBD-Cl) in an alkaline medium (pH 8)
to form a highly fluorescent derivative that was measured at 530 nm after
excitation at 460 nm. The factors affecting the formation and stability of the
reaction product were carefully studied and optimized. The fluorescence-
concentration plot is rectilinear over the range 0.2-6 μg/mL with LOD of 0.08
μg/mL and LOQ of 0.24 μg/mL respectively. The method was applied to the
analysis of commercial tablets and the results were in good agreement with
those obtained using the reference method. The mean percentage recoveries
for paxetin and xandol tablets were 101.27 ± 1.79 and 101.33 ± 1.19
respectively. A proposal of the reaction pathway was postulated.
J Fluoresc. 2011 Jan;21(1):105-12. doi: 10.1007/s10895-010-0693-2. Epub 2010 Jul 1.
paroxetine
4-chloro-7-nitrobenzo-2- oxa-1,3-diazole
HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC AND
SPECTROFLUOROMETRIC DETERMINATION OF ΑLPHA- TOCOPHEROL IN A
NATURAL PLANT: FERULA HERMONIS (ZALOOH ROOT)
CANNAHHALILK,BHIDIACCHIWAG,,BIMARAMILJ,AOUNAMILEE
A LEBANESE AGRICULTURAL RESEARCH INSTITUTE (LARI), FANAR, P.O. BOX 90 1965, JDEIDET EL-METN, LEBANON
B LABORATORY OF MOLECULAR CHEMISTRY, FACULTY OF SCIENCE II, THE LEBANESE UNIVERSITY, FANAR, P.O. BOX
26110217 FANAR-MATN, LEBANON
 Abstract:
A high-performance liquid chromatographic (HPLC) method for the determination of α-
tocopherol in a natural plant (Ferula hermonis–Zalooh roots) is reported. The method
includes saponification of samples and extraction of α-tocopherol with a mixture of
acetonitrile and methanol (1:1 v/v). However, the presence of α-tocopherol in Zalooh
is confirmed with HPLC-UV and fluorescence detection. A spectrofluorometric and the
internal addition standard methods are also used to quantify the α-tocopherol in the
plant. An internal standard method is based on a known concentration of α-tocopherol
that is added in every sample that is analyzed. Alpha-tocopherol levels as determined
in samples by HPLC with UV and fluorescence detection did not differ significantly
from the levels determined by Shimadzu spectrofluorometer . However, the amount of
tocopherol determined by both techniques in Zalooh roots was relatively very high.
Standards were checked for linearity giving correlation coefficients that were higher
than 0.99 in the concentration range of 1 and 6 μmol L−1.
615–607, Pages2005November,7, Issue18Volume
α-tocopherol
REFERENCES:
 www.pubmed.com
 www.sciencedirect.com
 Harvey.D, 2000, Modern Analytic Chemistry.
 Fluorescence and Fluorescence Applications-
Integrated DNA Technologies-2010
 Brandt.M, 2010, Spectrofluorometers and
Fluorescence Phenomena
Specrtofluorometer

Contenu connexe

Tendances

atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
sharmasuriti
 
Fluorescence spectroscopy
Fluorescence spectroscopyFluorescence spectroscopy
Fluorescence spectroscopy
Rahul Sharma
 

Tendances (20)

Fluorescence spectroscopy
Fluorescence spectroscopyFluorescence spectroscopy
Fluorescence spectroscopy
 
Spectroscopy - principle, procedure & application
Spectroscopy - principle, procedure & application Spectroscopy - principle, procedure & application
Spectroscopy - principle, procedure & application
 
Spectroscopy principles and applications
Spectroscopy  principles and applicationsSpectroscopy  principles and applications
Spectroscopy principles and applications
 
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
 
Flourescence
FlourescenceFlourescence
Flourescence
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
Infra Red Spectroscopy and Its Applications
Infra Red Spectroscopy and Its ApplicationsInfra Red Spectroscopy and Its Applications
Infra Red Spectroscopy and Its Applications
 
spectrofluorimetry ppt
spectrofluorimetry pptspectrofluorimetry ppt
spectrofluorimetry ppt
 
Beer lamberts law
Beer lamberts lawBeer lamberts law
Beer lamberts law
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
FTIR spectroscopy
FTIR spectroscopyFTIR spectroscopy
FTIR spectroscopy
 
Spectrofluorimetry
Spectrofluorimetry Spectrofluorimetry
Spectrofluorimetry
 
Fluorescence spectroscopy
Fluorescence spectroscopyFluorescence spectroscopy
Fluorescence spectroscopy
 
Fluorescence spectroscopy
Fluorescence spectroscopyFluorescence spectroscopy
Fluorescence spectroscopy
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRY
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPT
 
Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 

Similaire à Specrtofluorometer

fluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satarafluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satara
nikhil salunkhe
 

Similaire à Specrtofluorometer (20)

Spectrofluorimetry
SpectrofluorimetrySpectrofluorimetry
Spectrofluorimetry
 
Spectrofluorimetry
SpectrofluorimetrySpectrofluorimetry
Spectrofluorimetry
 
Fluorometry
FluorometryFluorometry
Fluorometry
 
Fluorometry & its application in lab.assay
Fluorometry & its application in lab.assayFluorometry & its application in lab.assay
Fluorometry & its application in lab.assay
 
Payalfluorimetryupload
PayalfluorimetryuploadPayalfluorimetryupload
Payalfluorimetryupload
 
Photoluminescence -chapter-6.pdf
Photoluminescence -chapter-6.pdfPhotoluminescence -chapter-6.pdf
Photoluminescence -chapter-6.pdf
 
Fluorometry PPT file.ppt
Fluorometry PPT file.pptFluorometry PPT file.ppt
Fluorometry PPT file.ppt
 
fluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satarafluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satara
 
2. Fluorimetry.pdf
2. Fluorimetry.pdf2. Fluorimetry.pdf
2. Fluorimetry.pdf
 
Flourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copyFlourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copy
 
Spectroflourimetry.pdf
Spectroflourimetry.pdfSpectroflourimetry.pdf
Spectroflourimetry.pdf
 
Flourimetry
FlourimetryFlourimetry
Flourimetry
 
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
 
Photoluminescence.pptx
Photoluminescence.pptxPhotoluminescence.pptx
Photoluminescence.pptx
 
spectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptxspectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptx
 
Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)
 
Phosphorimetry
PhosphorimetryPhosphorimetry
Phosphorimetry
 
Fluorimetry/ Fluoroscences
Fluorimetry/ FluoroscencesFluorimetry/ Fluoroscences
Fluorimetry/ Fluoroscences
 
Fluorometry
FluorometryFluorometry
Fluorometry
 
FLUOROSCENCE.pptx
FLUOROSCENCE.pptxFLUOROSCENCE.pptx
FLUOROSCENCE.pptx
 

Dernier

Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 

Dernier (20)

FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data DiscoveryTrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
 
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost SavingRepurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
Polkadot JAM Slides - Token2049 - By Dr. Gavin Wood
Polkadot JAM Slides - Token2049 - By Dr. Gavin WoodPolkadot JAM Slides - Token2049 - By Dr. Gavin Wood
Polkadot JAM Slides - Token2049 - By Dr. Gavin Wood
 
A Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source MilvusA Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source Milvus
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
Automating Google Workspace (GWS) & more with Apps Script
Automating Google Workspace (GWS) & more with Apps ScriptAutomating Google Workspace (GWS) & more with Apps Script
Automating Google Workspace (GWS) & more with Apps Script
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)Powerful Google developer tools for immediate impact! (2023-24 C)
Powerful Google developer tools for immediate impact! (2023-24 C)
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdf
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024
 

Specrtofluorometer

  • 1. SPECTROFLUOROMETER Presented by: Nour Elrashed Q.C master student Damascus UniversityFaculty of Pharmacy
  • 2. CONTENTS:  Introduction  the principle of spectrofluorometer  What can specrofluorometer do?  The components of specrofluorometer  Fluorescence Spectra  Fluorescence and chemical structure  Applications of Spectrofluorometer  New pharmaceutical studies conducted by spectrofluorometer  References
  • 4. HISTORY:  The term fluorescence comes from the mineral fluorspar (calcium fluoride) when Sir George G. Stokes observed in 1852 that fluorspar would give off visible light (fluoresce) when exposed to electromagnetic radiation in the ultraviolet wavelength.
  • 5. THE PRINCIPLE OF SPECTROFLUOROMETER:  It is an analytical device depends on the fluorescence phenomenon which is a short-lived type of photoluminescence created by electromagnetic excitation.  That is, fluorescence is generated when a molecule transmits from its ground state So to one of several vibrational energy levels in the first excited electronic state, S1, or the second electronic excited state, S2, both of which are singlet states.  Relaxation to the ground state from these excited states occurs by emission of energy through heat and/or photons.
  • 6.  The difference between the excitation and emission wavelengths is called the Stokes shift.  Stokes’ studies of fluorescent substances led to the formulation of Stokes’ Law, which states that the wavelength of fluorescent light is always greater than that of the exciting radiation. Thus, for any fluorescent molecule, the wavelength of emission is always longer than the wavelength of absorption.
  • 7.
  • 8. WHAT IS THE FLUORESCENCE QUANTUM YIELD (ΦF)?  It is the quantitative expression of the fluorescence efficiency, which is the fraction of excited molecules returning to the ground state by fluorescence.  Quantum yields range from 1, when every molecule in an excited state undergoes fluorescence, to 0 when fluorescence does not occur.
  • 9.  A molecule’s fluorescence quantum yield is influenced by external Variables such as: • temperature • viscosity of solvent • pH  Increasing temperature generally decreases Φf because more frequent collisions between the molecule and the solvent increases external conversion.  Decreasing the solvent’s viscosity decreases Φf for similar reasons.  For an analyte with acidic or basic functional groups, a change in pH may change the analyte’s structure and, therefore, its fluorescent properties.
  • 10.  Emission of a photon when the analyte returns from a higher- energy state to a lower- energy state with the opposite spin (triplet excited state: An excited state in which unpaired electron spins occur.).  Average lifetime from 10-3 to 10+2 sec  Emission of a photon when the analyte returns from a higher- energy state to a lower- energy state with the same spin (singlet excited state: An excited state in which all electron spins are paired).  Average lifetime from 10-9 to 10-7 sec Phosphorescence Fluorescence
  • 11.
  • 12.
  • 13. WHAT CAN SPECROFLUOROMETER DO?  It has been used for the direct or indirect quantitative and qualitative analysis by measuring the fluorescent intensity F.  It is relatively inexpensive and sensitive (the sensitivity of fluorescence is approximately 1,000 times greater than absorption spectrophotometric methods).
  • 14.  fluorescent intensity F is dependent on both intrinsic properties of the compound (fluorescence quantum yield Φf), and on readily controlled experimental parameters including: • intensity of the absorbed light I0 • molar absorption coefficient Ɛ • path length of the cell b • concentration of the fluorophor in solution c
  • 15.  At low concentrations of fluorophore, the fluorescence intensity of a sample is essentially linearly proportional to concentration.  However, as the concentration increases, a point is reached at which the intensity increase is progressively less linear, and the intensity eventually decreases as concentration increases further.
  • 17.  The most common reason for this is Inner filter effect that, as the absorbance of the sample increases, the light intensity experienced by some of the fluorescent molecules is lower than that experienced by others. When excitation intensity decreases, so does fluorescence emission intensity.  It is generally necessary to use concentrations that result in absorbance values of 0.1 or lower to observe concentration dependent emission.
  • 18.  As the concentration of molecules in a solution increases, probability increases that excited molecules will interact with each other and lose energy through processes other than fluorescent emission. Any process that reduces the probability of fluorescent emission is known as quenching.  Other parameters that can cause quenching include: • presence of impurities • increased temperature • reduced viscosity of the solution media
  • 19. THE COMPONENTS OF SPECROFLUOROMETER: 1. light source (75 to 450-W high-pressure xenon arc lamp or Lasers) 2. excitation monochromator 3. sample holder (Quartz/Optical Glass/Plastic Cells) 4. emission monochromator 5. Detector (photomultiplier) 6. Most spectrofluorometers also have a reference sample. The reference is generally a solution of a strongly fluorescent molecule with a broad absorbance spectrum such as rhodamine. The reference is necessary to correct for lamp output, especially when varying the excitation wavelength, and to correct for differences in detector sensitivity.
  • 20.
  • 21.
  • 22. FLUORESCENCE SPECTRA:  Photoluminescence spectra are recorded by measuring the intensity of emitted radiation as a function of either the excitation wavelength or the emission wavelength.  The excitation spectra is determined by measuring the emission intensity at a fixed wavelength , while varying the excitation wavelength. It is useful for selecting the best excitation wavelength for a quantitative or qualitative analysis.  The emission spectra is determined by measuring the variation in emission intensity wavelength for a fixed excitation wavelength.
  • 23.
  • 24. FLUORESCENCE AND CHEMICAL STRUCTURE:  Fluorescence is generally observed with molecules where the lowest energy absorption is →* transition, although some n→* transitions show weak fluorescence.  Most unsubstituted, nonheterocyclic aromatic compounds show favorable fluorescence quantum yields wich usually increases with the number of rings and their degree of condensation. In addition, substitution to the aromatic ring can have a significant effect on Φf. For example, the presence of an electron-withdrawing group, such as (NO2), decreases Φf, whereas adding an electron-donating group, such as (OH), increases Φf.
  • 25.  Fluorescence also increases for aromatic ring systems and for aromatic molecules with rigid planar structures.  The simple heterocyclics, such as pyridine, furan, thiophene, and pyrrole do not exhibit fluorescence; on the other hand, fused ring structures ordinarily do.
  • 26. SPECTROFLUOROMETER:APPLICATIONS OF  Environmental Significance:  To detect environmental pollutants such as polycyclic aromatic hydrocarbons: • pyrene • benzopyrene • organothiophosphorous pesticides • carbamate insecticides  Geology:  Many types of calcite and amber will fluoresce under shortwave UV. Rubies, emeralds, and the Hope Diamond exhibit red fluorescence under short-wave UV light; diamonds also emit light under X ray radiation.
  • 27. SPECTROFLUOROMETER:APPLICATIONS OF  Analytical chemistry  to detect compounds from HPLC flow  TLC plates can be visualized if the compounds or a coloring reagent is fluorescent  Biochemistry:  used generally as a non-destructive way of tracking or analysis of biological molecules (proteins)  Possible direct or indirect analysis aromatic amino acids (phenylalanine- tyrosine-tryptophan)  Fingerprints can be visualized with fluorescent compounds such as ninhydrin.
  • 28. SPECTROFLUOROMETER:APPLICATIONS OF  Medicine  Blood and other substances are sometimes detected by fluorescent reagents, particularly where their location was not previously known.  There has also been a report of its use in differentiating malignant, bashful skin tumors from benign.  Pharmacy:  Possible direct or indirect analysis drugs such as: • vitamins (vitamin A -vitamin B2 -vitamin B6 -vitamin B12 - vitamin E -folic acid) • catecholamines (dopamine-norepinephrine) • Other drugs (quinine-salicylic acid–morphine-barbiturates – LSD)
  • 29. NEW PHARMACEUTICAL STUDIES CONDUCTED BY SPECTROFLUOROMETER:
  • 30. OPTIMIZATION AND VALIDATION OF SPECTROFLUORIMETRIC METHOD FOR DETERMINATION OF CEFADROXILE AND CEFUROXIME SODIUM IN PHARMACEUTICAL FORMULATIONS .HYNEINE-BOULA,SMHMEDA,AALBASHIRE CHEMISTRY DEPARTMENT, FACULTY OF SCIENCE, UNIVERSITY OF KHARTOUM, PO BOX 321, KHARTOUM, 11115, SUDAN.  Abstract A simple, accurate, precise and validated spectrofluorimetric method is proposed for the determination of two cephalosporins, namely, cefadroxile (cefa) and cefuroxime sodium (cefu) in pharmaceutical formulations. The method is based on a reaction between cephalosporins with 1,2-naphthoquinone-4-sulfonate in alkaline medium, to form fluorescent derivatives that are extracted with chloroform and subsequently measured at 610 and 605 nm after excitation at 470 and 460 nm for cefa and cefu respectively. The optimum experimental conditions have been studied. Beer's law is obeyed over the concentrations of 20-70 ng/mL and 15-40 ng/mL for cefa and cefu, respectively. The detection limits were 4.46 ng/mL and 3.02 ng/mL with a linear regression correlation coefficient of 0.9984 and 0.998, and recoveries ranging 97.50- 109.96% and 95.73-98.89% for cefa and cefu, respectively. The effects of pH, temperature, reaction time, 1,2-naphthoquinone-4-sulfonic concentration and extraction solvent on the determination of cefa and cefu, have been examined. The proposed method can be applied for the determination of cefa and cefu in pharmaceutical formulations in quality control laboratories. Copyright © 2013 John Wiley & Sons, Ltd. ahead of print]Epub. [2481/bio.1002.10:doi.24Jan2013Luminescence.
  • 32. A HIGHLY SENSITIVE FLUORIMETRIC METHOD FOR DETERMINATION VIACAPSULESANDFORMBULKITSINLENALIDOMIDEOF DERIVATIZATION WITH FLUORESCAMINE .NZLZOMANA,AHAKHEITB,NYHALILK,IAARWISHD DEPARTMENT OF PHARMACEUTICAL CHEMISTRY, COLLEGE OF PHARMACY, KING SAUD UNIVERSITY,, P,O, BOX 2457, RIYADH, 11451, SAUDI ARABIA. IDARWISH@KSU.EDU.SA. ABSTRACT: BACKGROUND: Lenalidomide (LND) is a potent novel thalidomide analog which demonstrated remarkable clinical activity in treatment of multiple myeloma disease via a multiple-pathways mechanism. The strong evidences-based clinical success of LND in patients has led to its recent approval by US-FDA under the trade name of Revlimid® capsules by Celgene Corporation. Fluorimetry is a convenient technique for pharmaceutical quality control, however there was a fluorimetric method for determination of LND in its bulk and capsules. RESULTS: A novel highly sensitive and simple fluorimetric method has been developed and validated for the determination of lenalidmide (LND) in its bulk and dosage forms (capsules). The method was based on nucleophilic substitution reaction of LND with fluorescamine (FLC) in aqueous medium to form a highly fluorescent derivative that was measured at 494 nm after excitation at 381 nm. The factors affecting the reaction were carefully studied and optimized. The kinetics of the reaction was investigated, and the reaction mechanism was postulated. Under the optimized conditions, linear relationship with good correlation coefficient (0.9999) was found between the fluorescence intensity and LND concentration in the range of 25- 300 ng/mL. The limits of detection and quantitation for the method were 2.9 and 8.7 ng/mL, respectively. The precision of the method was satisfactory; the values of relative standard deviations did not exceed 1.4%. The proposed method was successfully applied to the determination of LND in its bulk form and pharmaceutical capsules with good accuracy; the recovery values were 97.8-101.4 ± 1.08-2.75%. CONCLUSIONS: The proposed method is selective and involved simple procedures. In conclusion, the method is practical and valuable for routine application in quality control laboratories for determination of LND. .118-6-X153-1752/1186.10:doi.118):1(6;16Oct2012Cent J.Chem
  • 34. INTIZANIDINEOFDETERMINATIONSPECTROFLUORIMETRICENSITIVES PHARMACEUTICAL PREPARATIONS, HUMAN PLASMA AND URINE THROUGH DERIVATIZATION WITH DANSYL CHLORIDE .STLUU DEPARTMENT OF ANALYTICAL CHEMISTRY, FACULTY OF PHARMACY, ISTANBUL UNIVERSITY, TURKEY. SEVGITATAR@YAHOO.COM  Abstract A sensitive spectrofluorimetric method was developed for the determination of tizanidine in human plasma, urine and pharmaceutical preparations. The method is based on reaction of tizanidine with 1-dimethylaminonaphthalene-5-sulphonyl chloride (dansyl chloride) in an alkaline medium to form a highly fluorescent derivative that was measured at 511 nm after excitation at 383 nm. The different experimental parameters affecting the fluorescence intensity of tizanidine was carefully studied and optimized. The fluorescence-concentration plots were rectilinear over the ranges 50-500 and 20-300 ng/mL for plasma and urine, respectively, detection limits of 1.81 and 0.54 ng/mL and quantification limits of 5.43 and 1.62 ng/mL for plasma and urine, respectively. The method presents good performance in terms of linearity, detection and quantification limits, precision, accuracy and specificity. The proposed method was successfully applied for the determination of tizanidine in pharmaceutical preparations. The results obtained were compared with a reference method, using t- and F-tests. .28Oct2011Epub.1367/bio.1002.10:doi.30-426):5(27Oct;-Sep2012Luminescence.
  • 36. INHCLPAROXETINEOFDETERMINATIONPECTROFLUORIMETRICS PHARMACEUTICALS VIA DERIVATIZATION WITH 4- CHLORO-7- NITROBENZO-2-OXA-1,3-DIAZOLE (NBD-CL) .HLMANSIE,NNANYE-LE,FELALB,MALSHW DEPARTMENT OF ANALYTICAL CHEMISTRY, FACULTY OF PHARMACY, UNIVERSITY OF MANSOURA, MANSOURA, EGYPT.  Abstract: A sensitive and simple spectrofluorimetric method has been developed and validated for the determination of the antidepressant paroxetine HCl (PXT) in its dosage forms. The method was based on coupling reaction of PXT with 4- chloro-7-nitrobenzo-2- oxa-1,3-diazole (NBD-Cl) in an alkaline medium (pH 8) to form a highly fluorescent derivative that was measured at 530 nm after excitation at 460 nm. The factors affecting the formation and stability of the reaction product were carefully studied and optimized. The fluorescence- concentration plot is rectilinear over the range 0.2-6 μg/mL with LOD of 0.08 μg/mL and LOQ of 0.24 μg/mL respectively. The method was applied to the analysis of commercial tablets and the results were in good agreement with those obtained using the reference method. The mean percentage recoveries for paxetin and xandol tablets were 101.27 ± 1.79 and 101.33 ± 1.19 respectively. A proposal of the reaction pathway was postulated. J Fluoresc. 2011 Jan;21(1):105-12. doi: 10.1007/s10895-010-0693-2. Epub 2010 Jul 1.
  • 38. HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC AND SPECTROFLUOROMETRIC DETERMINATION OF ΑLPHA- TOCOPHEROL IN A NATURAL PLANT: FERULA HERMONIS (ZALOOH ROOT) CANNAHHALILK,BHIDIACCHIWAG,,BIMARAMILJ,AOUNAMILEE A LEBANESE AGRICULTURAL RESEARCH INSTITUTE (LARI), FANAR, P.O. BOX 90 1965, JDEIDET EL-METN, LEBANON B LABORATORY OF MOLECULAR CHEMISTRY, FACULTY OF SCIENCE II, THE LEBANESE UNIVERSITY, FANAR, P.O. BOX 26110217 FANAR-MATN, LEBANON  Abstract: A high-performance liquid chromatographic (HPLC) method for the determination of α- tocopherol in a natural plant (Ferula hermonis–Zalooh roots) is reported. The method includes saponification of samples and extraction of α-tocopherol with a mixture of acetonitrile and methanol (1:1 v/v). However, the presence of α-tocopherol in Zalooh is confirmed with HPLC-UV and fluorescence detection. A spectrofluorometric and the internal addition standard methods are also used to quantify the α-tocopherol in the plant. An internal standard method is based on a known concentration of α-tocopherol that is added in every sample that is analyzed. Alpha-tocopherol levels as determined in samples by HPLC with UV and fluorescence detection did not differ significantly from the levels determined by Shimadzu spectrofluorometer . However, the amount of tocopherol determined by both techniques in Zalooh roots was relatively very high. Standards were checked for linearity giving correlation coefficients that were higher than 0.99 in the concentration range of 1 and 6 μmol L−1. 615–607, Pages2005November,7, Issue18Volume
  • 40. REFERENCES:  www.pubmed.com  www.sciencedirect.com  Harvey.D, 2000, Modern Analytic Chemistry.  Fluorescence and Fluorescence Applications- Integrated DNA Technologies-2010  Brandt.M, 2010, Spectrofluorometers and Fluorescence Phenomena