Ed Israelski PhD,
Director of Human Factors
Abbott
Abbott Park, IL, USA
Overview of International
Medical Device Human
Factors Standards
Outline
 History of Medical Device Human Factors Standards
 Summary of major current international HF Standard
History of Medical Device HF Standards Development
Standard Title Main Purpose
ANSI/AAMI HE48: 1993
Human Factors Engineering
Guidelines and Preferred Practices
For the Design of Medical Devices
General design practices for user
interfaces (from MIL-STD 1472)
Superseded by HE-75
ANSI/AAMI HE74: 2001
Human Factors Design Process
For Medical Devices
HFE Process related to Design Controls
in US 21CFR 820.30
IEC 60601-1-6: 2006.
Now replaced by 2010
edition
Medical Electrical Equipment -
General requirements for basic
safety and essential performance -
Collateral standard: Usability
2006 edition: HFE process applied to
medical electrical equipment.
2010 edition points to IEC 62366:2007
IEC 62366: 2007
Medical Devices – Application of
Usability Engineering To Medical
Devices
 HFE process extended to all
medical devices with emphasis on
Risk Management and lifecyle
 Harmonized by European Union as
EN 62366:2008
AAMI HE75: 2009
Human Factors Engineering -
Design of Medical Devices
Interface design practices; special
medical application issues; test &
evaluation methods
Standard Title Main Purpose
IEC 60601-1:2005
ANSI/AAMI
ES60601-1:2011,
3rd ed.
General safety & essential
performance standard for
medical electrical equipment
Introduction to broad standard with
numerous subparts for a wide
variety of electrical medical
devices. (multiple subparts for type
acceptance)
IEC 60601-1-8:
2006
Collateral to IEC 60601-
1:2005 on general
requirements, tests and
guidance for alarm systems
Recommends visual and auditory
alarm design parameters, e.g.
color, frequency and cadence.
IEC TR
60878:2003
Graphical symbols for
electrical equipment in
medical practice
Collects existing symbols
applicable to medical devices and
presents them in 15 medical device
categories
ISO 15223-1:2007
ISO 15223-2:2010
Medical devices – Symbols
to be used with medical
device labels, labeling and
information to be supplied
Part 1– Lists MDD + IVD Symbols
Part 2 – Symbol development,
selection and validation
Other Medical Device Standards
Standard Title Main Purpose
ISO 14971:2007
Medical devices -- Application
of risk management to
medical devices
The definitive standard on principles
of risk management, e.g. FTA,
FMEA to medical devices
IEC 60601-1-11:
2011
- Collateral Standard:
Requirements for medical
electrical equipment and
medical electrical systems
used in the home healthcare
environment
Describes particular requirements
for home healthcare medical
devices
ISO 80369 –
1:2010
Small-bore connectors for
liquids and gases in
healthcare applications - Part
1: General Requirements
Parts 2 to 7 for particular
devices
Describes standard connectors that
are usable and impossible to
misconnect across medical device
categories
IECEE – TRF’s TRF – Test Report Forms
Used by Notified Bodies in EU and
elsewhere to gauge compliance with
IEC/ISO standards
Other Medical Device Standards
 Usability Engineering process document like HE74 & IEC 60601-1-6
 IEC 60601-1-6:2010 essentially is a reference/pointer to IEC 62366:2007
– i.e. links IEC 62366:2007 into the IEC 60601-series for medical electrical
equipment
 Similar content to previous version (2007) of IEC 60601-1-6 with some
changes
– “Medical Electrical equipment” changed to all medical devices
– “Operator” changed to “user”
– Includes Risk Management language (ISO 14971:2007)
– Has many more Informative Annexes
 Contains HE-74 as Annex D
 Although uses term “usability”, also stresses risk assessment and control
 IEC (EN) 62366:2008 harmonized under the European MDD directive as
of November 27, 2008
IEC 62366: Medical Devices –
Application of Usability Engineering to Medical Devices
IEC 62366: Medical Devices: General Requirements for
Safety and Essential Performance-Usability
1: Scope
2: Normative references
3: Terms and definitions
4: Principles
– 4.1 General Requirements
• 4.1.1 Usability Engineering
Process
• 4.1.2 Residual Risk
• 4.1.3 Information for Safety
– 4.2 Usability Engineering File
– 4.3 Scaling the Process
The Manufacturer shall establish, document and
maintain a Usability Eng Process to provide safety
for the Patient, User and others... The process shall
address User interactions with the Medical Device…
If the Usability Eng Process detailed in this ..
standard has been complied with and the
acceptance criteria documented in the Usability
Validation Plan have been met, .. for the purposes
of ISO 14971, the residual risks associated with
usability of the device shall be .. acceptable, unless
there is objective evidence to the contrary.
..Usab. Eng Process may be scaled .. based on nature of
device…extent of modifications determined by risk analysis…
If Information for Safety is a risk control, it shall be subjected
to the Usability Eng Process..
IEC 62366: Medical Devices: General Requirements for Safety
and Essential Performance-Usability
 5: Usability Engineering Process
– 5.1: Application Specification
– 5.2: Frequently Used Functions
– 5.3 Identification. Hazardous Situations
• 5.3.1 Identify characteristics related to
safety
• 5.3.2 Identify known or foreseeable hazards
– 5.4: Primary Operating Functions
– 5.5: Usability Specification
– 5.6 Usability Validation Plan
– 5.7 User Interface Implementation
– 5.8 Usability Verification
– 5.9 Usability Validation
 6: Accompanying Document
 7: Training/Training Materials
..shall provide testable requirements for
Usability Verification; and .. for Usability of the
Primary Operating Functions including criteria
for determining the adequacy of Risk Control
achieved by the Usab. Eng.Process.
.. criteria for determining successful Validation of
the usability of the Primary Oper. Functions based
on the Usability Specification; and involvement of
representative users…May be quantitative or
qualitative. Sim. Use environment..
Relationship of FDA HF Guidance to IEC 62366
 5: Usability Engineering Process
– 5.1: Application Specification
– 5.2: Frequently Used Functions
– 5.3 Ident. Hazardous Situations
• 5.3.1 Identify characteristics related to
safety
• 5.3.2 Identify known or foreseeable hazards
– 5.4: Primary Operating Functions
– 5.5: Usability Specification
– 5.6 Usability Validation Plan
– 5.7 User Interface Implementation
– 5.8 Usability Verification
– 5.9 Usability Validation
 6: Accompanying Document
 7: Training/Training Materials
FDA’s
Preliminary
Evaluation
FDA’s Formative
Evaluation & Design
Mods
FDA’s
Validation
Testing
FDA’s
Intended Use
ANSI/AAMI HE74:2001: Human Factors Design
Process For Medical Devices
– Main Features
• Shows relationship of all HFE processes to design
cycle (Design Controls, etc.)
• Provides detail on recommended analysis and
testing methods with examples
– HE-74 has been incorporated into IEC 62366 as an
annex. It is no longer a stand-alone standard.
– Revisions to the HE-74 HFE process will be
incorporated in future revisions to IEC 62366
Future Revisions to IEC 62366:2007
• Feedback is that IEC 62366 could be more usable
• Fast Track Revision
– Add annex that will address process for:
• Legacy Devices,
• Changes to Legacy Devices
• Additions of accessories not subject to IEC 62366 (Usability of Unknown Provenance)
• Standard will be rewritten and split
– IEC 62366-1 (Shorter Normative Standard)
– IEC TR 62366-2 (Informative Tutorial Tech Report on Usability
Engineering for Medical Devices)
– Will be Harmonized with new FDA HF Guidance (draft June, 2011)
– Target is for 2014 Publication
Company Confidential
© 2009 Abbott
12
Use Error Chart from IEC 62366:2007
Normal Use
Company Confidential
© 2009 Abbott
14
Example of a Fault Tree (FTA)
Reviewed/
Confirmed
incorrect Rx
Incorrect Rx
programmed
Initial Program Change to Rx
PCA Dose
AND
Misprogrammed
Rx
Initial Loading
Dose
Delivery Mode Lockout Rate Dose Limit Concentration Units
PCA Dose
Additional
Loading Dose
Delivery Mode Lockout Rate Dose Limit
Press Start
Key
OR
OR
OR
Company Confidential
© 2009 Abbott
15
Example of a Fault Tree
Gate1
Free T4 drift resulting in
falsely elevated FT4 result
Q:1.00004e-006
Gate5
Undetected drift-- due to
microparticle deposition
Q:3.75615e-011
Gate3
Microparticle deposition in
bottle
Q:1.87808e-005
Gate7
Insufficient agitation
Q:0.000751982
Gate8
Manual swirl procedure failed to
remove microparticle deposition
Q:0.0312
Event7
Swirl- insufficient
magnitude
Q:0.0001
Event8
Swirl done infrequently,
intermittent deposition
Q:0.0001
Gate12
Users failed to perform
manual swirl
Q:0.031
Event18
Users did not receive
notification.
Q:0.001
Event19
User failed to read instructions
or forgot to manually swirl
Q:0.03
Gate9
Automatic swirl procedure failed
to remove microparticles
Q:0.024102
Event11
Electrical or mechanical
failure- agitation failure
Q:1e-006
Event12
Daily maintenance repeatedly interrupted
before swirl completion
Q:1e-006
Gate10
Automatic swirl (within daily
maintenance) not performed.
Q:0.0011
Event13
Automatic swirl software
not installed.
Q:0.001
Event14
Operator failed to perform
daily maintenance
Q:0.0001
Event16
Automatic swirl failed to
remove deposition
Q:0.023
Event5
Reagent(s) left on-board
Q:0.999
Event10
*Usage rate < 100 tests per
month or <500 test per month
Q:0.05
Event15
Instrument left in
continuous run mode
Q:0.5
Gate6
Controls failed to detect
drift
Q:2e-006
Event3
Users/ lab consistently
failed to run controls
Q:1e-006
Event4
Controls run-- failed to
detect drift
Q:1e-006
Event1
Other causes for drift- see
FMEA
Q:1e-006
Company Confidential
© 2009 Abbott
16
Example of Use Error FMEA
Task
# TASK Hazard Faults Prob Crit RI Method of Control
Effectiveness
of Control
Risk
Acceptability Reference Category Notes
Set-up
Delay in
therapy
User fails to seat
vial fully into
bottom bracket Occ Mod Med
Message
displayed/alarm/will
not function if not
properly seated
Reduces to
Low Acceptable
Damage
to vial or
device
User fails to
squeeze cradle
release while
pushing down on
top bracket to
seat vial. Rare Mod Low
Message
Displayed/alarm/will
not function if not
properly seated
Will not
change Acceptable
Patient
overdose
Use non-Abbott
tubing, causing
free flow Rare Maj Med
Labeling on pump,
manual and set
requires Abbott sets
Reduces to
Low Acceptable
Data
Retention
Purging
the
System
Infusion of
drug
remaining
in line
User fails to
disconnect
patient and does
not purge line
before infusing
another drug Occ Mod Low
Message on pump
reminds user to
disconnect set and
must press and hold
purge key to purge.
Will not
change Acceptable
Delay in
therapy
PCA line is
connected to
patient -
established as
secondary
instead of
primary line. Rare Mod Low
Labeling on pump,
manual and set
requires Abbott sets
Will not
change Acceptable
Summary
 Human Factors Standards have been expanded, revised,
replaced
• Process Standards: (IEC 62366: 2007, HE-74)
• 62366 is now the usability engineering standard as 60601-
1-6:2010 (3rd ed.) is a “pointer” to it
• HE-74 has been phased out, incorporated into IEC 62366
as annex
• Design Principles: HE-75 replaces HE-48
• There are individual standards for Alarms, Symbols, Home
Devices and Risk Management that relate to HF
• More Standards being written (HF Complaints, Contextual
Inquiry, Healthcare IT)
• IEC 62366 will be split and updated by 2014

Contenu connexe

Tendances

ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness TrainingISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
Operational Excellence Consulting
 
Design and development of medical device
Design and development of medical deviceDesign and development of medical device
Design and development of medical device
Malesh M
 
Medical Device Development Lifecycle
Medical Device Development LifecycleMedical Device Development Lifecycle
Medical Device Development Lifecycle
Tim Blair
 
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdfbsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
Nash229987
 

Tendances (20)

ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness TrainingISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
 
Design and development of medical device
Design and development of medical deviceDesign and development of medical device
Design and development of medical device
 
Understanding IEC 62304
Understanding IEC 62304Understanding IEC 62304
Understanding IEC 62304
 
15 Steps to get Approval to IEC 60601-1
15 Steps to get Approval to IEC 60601-115 Steps to get Approval to IEC 60601-1
15 Steps to get Approval to IEC 60601-1
 
Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices
 
medical device regulatory approval in USA
medical device regulatory approval in USAmedical device regulatory approval in USA
medical device regulatory approval in USA
 
Medical Device Development Lifecycle
Medical Device Development LifecycleMedical Device Development Lifecycle
Medical Device Development Lifecycle
 
Medical device design and development | Combination Product
Medical device design and development | Combination ProductMedical device design and development | Combination Product
Medical device design and development | Combination Product
 
ISO: 14971 Quality risk management of medical devices
ISO: 14971 Quality risk management  of medical devicesISO: 14971 Quality risk management  of medical devices
ISO: 14971 Quality risk management of medical devices
 
Iso 14971 2019
Iso 14971 2019Iso 14971 2019
Iso 14971 2019
 
Medical devices introduction
Medical devices introductionMedical devices introduction
Medical devices introduction
 
Risk Management in Medical Device Development
Risk Management in Medical Device DevelopmentRisk Management in Medical Device Development
Risk Management in Medical Device Development
 
The EU’s Medical Device Regulation
The EU’s Medical Device RegulationThe EU’s Medical Device Regulation
The EU’s Medical Device Regulation
 
Medical Device Regulation
Medical Device RegulationMedical Device Regulation
Medical Device Regulation
 
The European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final textThe European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final text
 
Risk management in-60601-1
Risk management in-60601-1Risk management in-60601-1
Risk management in-60601-1
 
EU MDR
EU MDR EU MDR
EU MDR
 
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdfbsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
bsi-md-symbols-and-information-to-be-provided-by-the-manufacturer-webinar.pdf
 
FDA and Medical Device Reporting
FDA and Medical Device ReportingFDA and Medical Device Reporting
FDA and Medical Device Reporting
 
What You Need to Know About Medical Electrical Standards Updates (and how the...
What You Need to Know About Medical Electrical Standards Updates (and how the...What You Need to Know About Medical Electrical Standards Updates (and how the...
What You Need to Know About Medical Electrical Standards Updates (and how the...
 

En vedette

Usability Testing Medical Devices
Usability Testing Medical DevicesUsability Testing Medical Devices
Usability Testing Medical Devices
Dan Berlin
 
ZMPCZM016000.13.03 Certificate of compliance
ZMPCZM016000.13.03 Certificate of complianceZMPCZM016000.13.03 Certificate of compliance
ZMPCZM016000.13.03 Certificate of compliance
painezeeman
 
사용성 평가의 계보
사용성 평가의 계보사용성 평가의 계보
사용성 평가의 계보
Slava Han
 

En vedette (14)

Usability Testing Medical Devices
Usability Testing Medical DevicesUsability Testing Medical Devices
Usability Testing Medical Devices
 
User Interface Design for Medical Devices - The Relationship Between Usabilit...
User Interface Design for Medical Devices - The Relationship Between Usabilit...User Interface Design for Medical Devices - The Relationship Between Usabilit...
User Interface Design for Medical Devices - The Relationship Between Usabilit...
 
7 Components to Medical Device Usability Testing Success
7 Components to Medical Device Usability Testing Success7 Components to Medical Device Usability Testing Success
7 Components to Medical Device Usability Testing Success
 
Medical Device Usability: Polly Shelton presents at UK UPA (Usability Profess...
Medical Device Usability: Polly Shelton presents at UK UPA (Usability Profess...Medical Device Usability: Polly Shelton presents at UK UPA (Usability Profess...
Medical Device Usability: Polly Shelton presents at UK UPA (Usability Profess...
 
Human Factors – Device Integration
Human Factors – Device IntegrationHuman Factors – Device Integration
Human Factors – Device Integration
 
Cathal Brennan , Medical Device Assessor- Human Products Authorisation and Re...
Cathal Brennan , Medical Device Assessor- Human Products Authorisation and Re...Cathal Brennan , Medical Device Assessor- Human Products Authorisation and Re...
Cathal Brennan , Medical Device Assessor- Human Products Authorisation and Re...
 
Product Safety Testing Reduces the Risk of Shock, Fire, Explosions
Product Safety Testing Reduces the Risk of Shock, Fire, ExplosionsProduct Safety Testing Reduces the Risk of Shock, Fire, Explosions
Product Safety Testing Reduces the Risk of Shock, Fire, Explosions
 
Applying IEC 62304 Risk Management in Aligned Elements - the medical device ALM
Applying IEC 62304 Risk Management in Aligned Elements - the medical device ALMApplying IEC 62304 Risk Management in Aligned Elements - the medical device ALM
Applying IEC 62304 Risk Management in Aligned Elements - the medical device ALM
 
Company x anti infective orthopedic implant drug device preliminary final rep...
Company x anti infective orthopedic implant drug device preliminary final rep...Company x anti infective orthopedic implant drug device preliminary final rep...
Company x anti infective orthopedic implant drug device preliminary final rep...
 
ZMPCZM016000.13.03 Certificate of compliance
ZMPCZM016000.13.03 Certificate of complianceZMPCZM016000.13.03 Certificate of compliance
ZMPCZM016000.13.03 Certificate of compliance
 
UX 관점에서 사용성 평가하기
UX 관점에서 사용성 평가하기UX 관점에서 사용성 평가하기
UX 관점에서 사용성 평가하기
 
사용성 평가의 계보
사용성 평가의 계보사용성 평가의 계보
사용성 평가의 계보
 
Home Healthcare, IEC 60601-1-11
Home Healthcare, IEC 60601-1-11Home Healthcare, IEC 60601-1-11
Home Healthcare, IEC 60601-1-11
 
Usability Engineering in Medizintechnik-Projekten
Usability Engineering in Medizintechnik-ProjektenUsability Engineering in Medizintechnik-Projekten
Usability Engineering in Medizintechnik-Projekten
 

Similaire à Abbott overview medical device human factors standards

Massmedic mar2004g
Massmedic mar2004gMassmedic mar2004g
Massmedic mar2004g
mitmohan
 
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
ulmedical
 
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
DeepaPansare
 
Usability in healthcare, general overview on new standards and metrics (Inter...
Usability in healthcare, general overview on new standards and metrics (Inter...Usability in healthcare, general overview on new standards and metrics (Inter...
Usability in healthcare, general overview on new standards and metrics (Inter...
Stella Tsank
 
presentation-ema-experience-mdr-implementation_en.pdf
presentation-ema-experience-mdr-implementation_en.pdfpresentation-ema-experience-mdr-implementation_en.pdf
presentation-ema-experience-mdr-implementation_en.pdf
Balaji Kuzhandhaivelu
 

Similaire à Abbott overview medical device human factors standards (20)

TÜV SÜD Design Dossiers_Med info
 TÜV SÜD Design Dossiers_Med info TÜV SÜD Design Dossiers_Med info
TÜV SÜD Design Dossiers_Med info
 
Massmedic mar2004g
Massmedic mar2004gMassmedic mar2004g
Massmedic mar2004g
 
20210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr202120210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr2021
 
ISO/IEC80001 - Do we need another standard?
ISO/IEC80001 - Do we need another standard?ISO/IEC80001 - Do we need another standard?
ISO/IEC80001 - Do we need another standard?
 
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
“CFDA Registration – Market Access Before Investment” delivered by Tim Lin, T...
 
Critical Steps in Software Development: Enhance Your Chances for a Successful...
Critical Steps in Software Development: Enhance Your Chances for a Successful...Critical Steps in Software Development: Enhance Your Chances for a Successful...
Critical Steps in Software Development: Enhance Your Chances for a Successful...
 
Compliance with medical standards iec 62304, iso 14971, iec 60601, fda title ...
Compliance with medical standards iec 62304, iso 14971, iec 60601, fda title ...Compliance with medical standards iec 62304, iso 14971, iec 60601, fda title ...
Compliance with medical standards iec 62304, iso 14971, iec 60601, fda title ...
 
China Compulsory Certification Requirement And Procedures Of Medical Device
China Compulsory Certification Requirement And Procedures Of Medical DeviceChina Compulsory Certification Requirement And Procedures Of Medical Device
China Compulsory Certification Requirement And Procedures Of Medical Device
 
China Compulsory Certification Requirement And Procedures Of Medical Device
China Compulsory Certification Requirement And Procedures Of Medical DeviceChina Compulsory Certification Requirement And Procedures Of Medical Device
China Compulsory Certification Requirement And Procedures Of Medical Device
 
Pa pdf-output-specs-section-c9-biomedical-engineering-final-no
Pa pdf-output-specs-section-c9-biomedical-engineering-final-noPa pdf-output-specs-section-c9-biomedical-engineering-final-no
Pa pdf-output-specs-section-c9-biomedical-engineering-final-no
 
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
13 Medical Device Single Audit Program (MDSAP) Working Group.pdf
 
Usability in healthcare, general overview on new standards and metrics (Inter...
Usability in healthcare, general overview on new standards and metrics (Inter...Usability in healthcare, general overview on new standards and metrics (Inter...
Usability in healthcare, general overview on new standards and metrics (Inter...
 
Health apps regulation and quality control case studies and session 2 present...
Health apps regulation and quality control case studies and session 2 present...Health apps regulation and quality control case studies and session 2 present...
Health apps regulation and quality control case studies and session 2 present...
 
Health apps regulation and quality control case studies and session 2 present...
Health apps regulation and quality control case studies and session 2 present...Health apps regulation and quality control case studies and session 2 present...
Health apps regulation and quality control case studies and session 2 present...
 
Clinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical DevicesClinical Evaluation Report for Medical Devices
Clinical Evaluation Report for Medical Devices
 
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca MosesmHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
 
presentation-ema-experience-mdr-implementation_en.pdf
presentation-ema-experience-mdr-implementation_en.pdfpresentation-ema-experience-mdr-implementation_en.pdf
presentation-ema-experience-mdr-implementation_en.pdf
 
TwinSPIN_Lecture.ppt
TwinSPIN_Lecture.pptTwinSPIN_Lecture.ppt
TwinSPIN_Lecture.ppt
 
Software design for_medical_devices_europe_conferent_19012011[1]
Software design for_medical_devices_europe_conferent_19012011[1]Software design for_medical_devices_europe_conferent_19012011[1]
Software design for_medical_devices_europe_conferent_19012011[1]
 
Presentation: Proposed Reforms to the Regulation of Software, Including Softw...
Presentation: Proposed Reforms to the Regulation of Software, Including Softw...Presentation: Proposed Reforms to the Regulation of Software, Including Softw...
Presentation: Proposed Reforms to the Regulation of Software, Including Softw...
 

Dernier

Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and Myths
Joaquim Jorge
 

Dernier (20)

Boost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdfBoost Fertility New Invention Ups Success Rates.pdf
Boost Fertility New Invention Ups Success Rates.pdf
 
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
 
Artificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and MythsArtificial Intelligence: Facts and Myths
Artificial Intelligence: Facts and Myths
 
A Year of the Servo Reboot: Where Are We Now?
A Year of the Servo Reboot: Where Are We Now?A Year of the Servo Reboot: Where Are We Now?
A Year of the Servo Reboot: Where Are We Now?
 
Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024Axa Assurance Maroc - Insurer Innovation Award 2024
Axa Assurance Maroc - Insurer Innovation Award 2024
 
The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024The 7 Things I Know About Cyber Security After 25 Years | April 2024
The 7 Things I Know About Cyber Security After 25 Years | April 2024
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Real Time Object Detection Using Open CV
Real Time Object Detection Using Open CVReal Time Object Detection Using Open CV
Real Time Object Detection Using Open CV
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data DiscoveryTrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
TrustArc Webinar - Unlock the Power of AI-Driven Data Discovery
 
GenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdfGenAI Risks & Security Meetup 01052024.pdf
GenAI Risks & Security Meetup 01052024.pdf
 
Manulife - Insurer Innovation Award 2024
Manulife - Insurer Innovation Award 2024Manulife - Insurer Innovation Award 2024
Manulife - Insurer Innovation Award 2024
 
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
Bajaj Allianz Life Insurance Company - Insurer Innovation Award 2024
 
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
Mastering MySQL Database Architecture: Deep Dive into MySQL Shell and MySQL R...
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 

Abbott overview medical device human factors standards

  • 1. Ed Israelski PhD, Director of Human Factors Abbott Abbott Park, IL, USA Overview of International Medical Device Human Factors Standards
  • 2. Outline  History of Medical Device Human Factors Standards  Summary of major current international HF Standard
  • 3. History of Medical Device HF Standards Development Standard Title Main Purpose ANSI/AAMI HE48: 1993 Human Factors Engineering Guidelines and Preferred Practices For the Design of Medical Devices General design practices for user interfaces (from MIL-STD 1472) Superseded by HE-75 ANSI/AAMI HE74: 2001 Human Factors Design Process For Medical Devices HFE Process related to Design Controls in US 21CFR 820.30 IEC 60601-1-6: 2006. Now replaced by 2010 edition Medical Electrical Equipment - General requirements for basic safety and essential performance - Collateral standard: Usability 2006 edition: HFE process applied to medical electrical equipment. 2010 edition points to IEC 62366:2007 IEC 62366: 2007 Medical Devices – Application of Usability Engineering To Medical Devices  HFE process extended to all medical devices with emphasis on Risk Management and lifecyle  Harmonized by European Union as EN 62366:2008 AAMI HE75: 2009 Human Factors Engineering - Design of Medical Devices Interface design practices; special medical application issues; test & evaluation methods
  • 4. Standard Title Main Purpose IEC 60601-1:2005 ANSI/AAMI ES60601-1:2011, 3rd ed. General safety & essential performance standard for medical electrical equipment Introduction to broad standard with numerous subparts for a wide variety of electrical medical devices. (multiple subparts for type acceptance) IEC 60601-1-8: 2006 Collateral to IEC 60601- 1:2005 on general requirements, tests and guidance for alarm systems Recommends visual and auditory alarm design parameters, e.g. color, frequency and cadence. IEC TR 60878:2003 Graphical symbols for electrical equipment in medical practice Collects existing symbols applicable to medical devices and presents them in 15 medical device categories ISO 15223-1:2007 ISO 15223-2:2010 Medical devices – Symbols to be used with medical device labels, labeling and information to be supplied Part 1– Lists MDD + IVD Symbols Part 2 – Symbol development, selection and validation Other Medical Device Standards
  • 5. Standard Title Main Purpose ISO 14971:2007 Medical devices -- Application of risk management to medical devices The definitive standard on principles of risk management, e.g. FTA, FMEA to medical devices IEC 60601-1-11: 2011 - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment Describes particular requirements for home healthcare medical devices ISO 80369 – 1:2010 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General Requirements Parts 2 to 7 for particular devices Describes standard connectors that are usable and impossible to misconnect across medical device categories IECEE – TRF’s TRF – Test Report Forms Used by Notified Bodies in EU and elsewhere to gauge compliance with IEC/ISO standards Other Medical Device Standards
  • 6.  Usability Engineering process document like HE74 & IEC 60601-1-6  IEC 60601-1-6:2010 essentially is a reference/pointer to IEC 62366:2007 – i.e. links IEC 62366:2007 into the IEC 60601-series for medical electrical equipment  Similar content to previous version (2007) of IEC 60601-1-6 with some changes – “Medical Electrical equipment” changed to all medical devices – “Operator” changed to “user” – Includes Risk Management language (ISO 14971:2007) – Has many more Informative Annexes  Contains HE-74 as Annex D  Although uses term “usability”, also stresses risk assessment and control  IEC (EN) 62366:2008 harmonized under the European MDD directive as of November 27, 2008 IEC 62366: Medical Devices – Application of Usability Engineering to Medical Devices
  • 7. IEC 62366: Medical Devices: General Requirements for Safety and Essential Performance-Usability 1: Scope 2: Normative references 3: Terms and definitions 4: Principles – 4.1 General Requirements • 4.1.1 Usability Engineering Process • 4.1.2 Residual Risk • 4.1.3 Information for Safety – 4.2 Usability Engineering File – 4.3 Scaling the Process The Manufacturer shall establish, document and maintain a Usability Eng Process to provide safety for the Patient, User and others... The process shall address User interactions with the Medical Device… If the Usability Eng Process detailed in this .. standard has been complied with and the acceptance criteria documented in the Usability Validation Plan have been met, .. for the purposes of ISO 14971, the residual risks associated with usability of the device shall be .. acceptable, unless there is objective evidence to the contrary. ..Usab. Eng Process may be scaled .. based on nature of device…extent of modifications determined by risk analysis… If Information for Safety is a risk control, it shall be subjected to the Usability Eng Process..
  • 8. IEC 62366: Medical Devices: General Requirements for Safety and Essential Performance-Usability  5: Usability Engineering Process – 5.1: Application Specification – 5.2: Frequently Used Functions – 5.3 Identification. Hazardous Situations • 5.3.1 Identify characteristics related to safety • 5.3.2 Identify known or foreseeable hazards – 5.4: Primary Operating Functions – 5.5: Usability Specification – 5.6 Usability Validation Plan – 5.7 User Interface Implementation – 5.8 Usability Verification – 5.9 Usability Validation  6: Accompanying Document  7: Training/Training Materials ..shall provide testable requirements for Usability Verification; and .. for Usability of the Primary Operating Functions including criteria for determining the adequacy of Risk Control achieved by the Usab. Eng.Process. .. criteria for determining successful Validation of the usability of the Primary Oper. Functions based on the Usability Specification; and involvement of representative users…May be quantitative or qualitative. Sim. Use environment..
  • 9. Relationship of FDA HF Guidance to IEC 62366  5: Usability Engineering Process – 5.1: Application Specification – 5.2: Frequently Used Functions – 5.3 Ident. Hazardous Situations • 5.3.1 Identify characteristics related to safety • 5.3.2 Identify known or foreseeable hazards – 5.4: Primary Operating Functions – 5.5: Usability Specification – 5.6 Usability Validation Plan – 5.7 User Interface Implementation – 5.8 Usability Verification – 5.9 Usability Validation  6: Accompanying Document  7: Training/Training Materials FDA’s Preliminary Evaluation FDA’s Formative Evaluation & Design Mods FDA’s Validation Testing FDA’s Intended Use
  • 10. ANSI/AAMI HE74:2001: Human Factors Design Process For Medical Devices – Main Features • Shows relationship of all HFE processes to design cycle (Design Controls, etc.) • Provides detail on recommended analysis and testing methods with examples – HE-74 has been incorporated into IEC 62366 as an annex. It is no longer a stand-alone standard. – Revisions to the HE-74 HFE process will be incorporated in future revisions to IEC 62366
  • 11. Future Revisions to IEC 62366:2007 • Feedback is that IEC 62366 could be more usable • Fast Track Revision – Add annex that will address process for: • Legacy Devices, • Changes to Legacy Devices • Additions of accessories not subject to IEC 62366 (Usability of Unknown Provenance) • Standard will be rewritten and split – IEC 62366-1 (Shorter Normative Standard) – IEC TR 62366-2 (Informative Tutorial Tech Report on Usability Engineering for Medical Devices) – Will be Harmonized with new FDA HF Guidance (draft June, 2011) – Target is for 2014 Publication
  • 12. Company Confidential © 2009 Abbott 12 Use Error Chart from IEC 62366:2007 Normal Use
  • 13.
  • 14. Company Confidential © 2009 Abbott 14 Example of a Fault Tree (FTA) Reviewed/ Confirmed incorrect Rx Incorrect Rx programmed Initial Program Change to Rx PCA Dose AND Misprogrammed Rx Initial Loading Dose Delivery Mode Lockout Rate Dose Limit Concentration Units PCA Dose Additional Loading Dose Delivery Mode Lockout Rate Dose Limit Press Start Key OR OR OR
  • 15. Company Confidential © 2009 Abbott 15 Example of a Fault Tree Gate1 Free T4 drift resulting in falsely elevated FT4 result Q:1.00004e-006 Gate5 Undetected drift-- due to microparticle deposition Q:3.75615e-011 Gate3 Microparticle deposition in bottle Q:1.87808e-005 Gate7 Insufficient agitation Q:0.000751982 Gate8 Manual swirl procedure failed to remove microparticle deposition Q:0.0312 Event7 Swirl- insufficient magnitude Q:0.0001 Event8 Swirl done infrequently, intermittent deposition Q:0.0001 Gate12 Users failed to perform manual swirl Q:0.031 Event18 Users did not receive notification. Q:0.001 Event19 User failed to read instructions or forgot to manually swirl Q:0.03 Gate9 Automatic swirl procedure failed to remove microparticles Q:0.024102 Event11 Electrical or mechanical failure- agitation failure Q:1e-006 Event12 Daily maintenance repeatedly interrupted before swirl completion Q:1e-006 Gate10 Automatic swirl (within daily maintenance) not performed. Q:0.0011 Event13 Automatic swirl software not installed. Q:0.001 Event14 Operator failed to perform daily maintenance Q:0.0001 Event16 Automatic swirl failed to remove deposition Q:0.023 Event5 Reagent(s) left on-board Q:0.999 Event10 *Usage rate < 100 tests per month or <500 test per month Q:0.05 Event15 Instrument left in continuous run mode Q:0.5 Gate6 Controls failed to detect drift Q:2e-006 Event3 Users/ lab consistently failed to run controls Q:1e-006 Event4 Controls run-- failed to detect drift Q:1e-006 Event1 Other causes for drift- see FMEA Q:1e-006
  • 16. Company Confidential © 2009 Abbott 16 Example of Use Error FMEA Task # TASK Hazard Faults Prob Crit RI Method of Control Effectiveness of Control Risk Acceptability Reference Category Notes Set-up Delay in therapy User fails to seat vial fully into bottom bracket Occ Mod Med Message displayed/alarm/will not function if not properly seated Reduces to Low Acceptable Damage to vial or device User fails to squeeze cradle release while pushing down on top bracket to seat vial. Rare Mod Low Message Displayed/alarm/will not function if not properly seated Will not change Acceptable Patient overdose Use non-Abbott tubing, causing free flow Rare Maj Med Labeling on pump, manual and set requires Abbott sets Reduces to Low Acceptable Data Retention Purging the System Infusion of drug remaining in line User fails to disconnect patient and does not purge line before infusing another drug Occ Mod Low Message on pump reminds user to disconnect set and must press and hold purge key to purge. Will not change Acceptable Delay in therapy PCA line is connected to patient - established as secondary instead of primary line. Rare Mod Low Labeling on pump, manual and set requires Abbott sets Will not change Acceptable
  • 17. Summary  Human Factors Standards have been expanded, revised, replaced • Process Standards: (IEC 62366: 2007, HE-74) • 62366 is now the usability engineering standard as 60601- 1-6:2010 (3rd ed.) is a “pointer” to it • HE-74 has been phased out, incorporated into IEC 62366 as annex • Design Principles: HE-75 replaces HE-48 • There are individual standards for Alarms, Symbols, Home Devices and Risk Management that relate to HF • More Standards being written (HF Complaints, Contextual Inquiry, Healthcare IT) • IEC 62366 will be split and updated by 2014