X30 SERIES FUME HOOD MONITORS
75 Discovery Way • Acton, Massachusetts 01720 • Telephone (978) 795-1285 • Fax (978) 795-1111
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—1 OF 16
Phoenix Controls X30 Series Fume Hood Monitors (FHMs) are
used on fume hoods with Phoenix Controls valves for airflow
control. Airflow control on these fume hoods is achieved with the
use of constant volume valves (CVV), two-position valves (PEV or
BEV) or variable air volume valves (VAV). Each FHM provides
two primary functions: indication of hood exhaust operating
condition and alarming. In VAV systems, each FHM also provides
face velocity control and optional energy-saving features.
SPECIFICATIONS
Enclosure
• Dimensions:
2.5" W x 6" H x 1.5" D
(64 mm x 152 mm x 38 mm)
• Color: White
• IP44 compliant
Operating Range
32-122 °F (0-50 °C) ambient
10-90% RH, non-condensing
8202’ (2500 m) altitude
Power Requirements for Each
Unit
24 Vac, ±10%, 50-60 Hz, 10 VA
±15 Vdc, ±5%, 220 mA
Inputs and Outputs
See table in "Features" section on page
2 for model-specific inputs and outputs.
Input to Optional Use LED
• Yellow LED indicates unique
customer conditions
• Wired directly from customer’s
device to two dedicated conductors
within the eight-wire signal cable
• Limited to ≤ 12 Vdc with maximum
current draw of 0.012 amps.
Customer must install a 1 K Ω
resistor in series with input signal.
• Visual indication only, no audible
Monitoring Points
• Airflow exhaust device command
• Actual exhaust airflow
• 0-10 Vdc alarm:
• 0 Vdc indicates normal operation
• 5 Vdc indicates incorrect airflow
• 10 Vdc indicates low differential
static pressure
• Sash position: 0-10 Vdc
• User status
• 10 K Ω minimum input impedence
for monitoring system
Backward Compatibility
X30 Series Fume Hood Monitors are
backward compatible with earlier
standard models of Phoenix fume hood
monitors.
Readout (FHM631 only)
The display shows one of the following
measurements:
• Cubic feet per minute (CFM)
• Meters cubed per hour (m3
/h)
• Liters per second (l/s)
• Feet per minute (fpm)
• Meters per second (m/s)
FCC Compliance
This equipment has been tested and
found to comply with the limits for a
Class B digital device, pursuant to Part
15 of the FCC Rules.
Power Loss Alarm Option
(±15 Vdc powered monitor only)
Indicates loss of power to the
fume hood system.
• During power failure, a red LED
flashes once every 4 sec
• Accompanied by short audible
alarm “chirp”
• Alarm continues for at least 64
hours or until power is restored
Dimensions
4.75" W x 4" H x 4.75" D
(121 mm x 102 mm x 121 mm)
Color
• Black enclosure
• Yellow switch
Operating Range
32-122 °F (0-50 °C) ambient
Power Requirements
±15 Vdc @ 0.015 amp
More details on this option are in the
"Features" section on page 2.
If the X30 Series Fume Hood Monitor
is used in a manner not specified by
the manufacturer, the protection
provided by the equipment may be
impaired.
! See wiring diagrams
ORDERING GUIDE
X30 Fume Hood Monitor (631 model shown).
Standard Operation
Flow Alarm
Emergency
Exhaust Mute
Standby Operation
BASE NUMBER
TYPE OF FUME HOOD MONITOR
430 = Base monitor, EXV applications
530 = CVV, PEV or BEV applications
631 = Face velocity monitor, EXV applications
FACEPLATE LANGUAGE
DAN = Danish GER = German
ENG = English JPN = Japanese
FRC = French Canadian NOR = Norwegian
FRN = French SWE = Swedish
OPTIONS
DFT = Drive failure text (FHM430 only)
DIV = Diversity LED to be driven by external contact closure from
a Celeris system or the BMS.
PLA = Power Loss Alarm (±15 volt only; see note below)
PRI = Primary (Available only on FHM430. Must be ordered with
at least one secondary.)
RET = Retrofit kit
RMT= Recess mount kit
SEC = Secondary (Available only on FHM430. Must be ordered
with primary.)
NOTE: By ordering the add-on option “PLA,” you will receive a
separate Power Loss Alarm board and box that also
mounts on top of the hood. This board interfaces to an
X30 Series Fume Hood Monitor that includes an
additional red LED with "Power Failure Alarm" text. This
unit can be installed on DC-powered monitors only.
FHM – –
TABLE OF CONTENTS
Specifications and Ordering Guide.............................1
Features......................................................................2
Applications: Hood Control.......................................4
Applications: Energy-saving........................................5
Installation .................................................................6
Points and Wiring ......................................................9
Maintenance and Troubleshooting ...........................14
2 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
FEATURES
Standard Operation
Flow Alarm
Emergency
Exhaust Mute
Standby Operation
Power Failure Alarm
Diversity Alarm
Face Velocity or Flow Display—Indicates the face velocity or flow of the hood. (Available
only on the FHM631.)
Standard Operation LED—When this LED is on, it indicates the hood is operating at the
standard face velocity or flow.
Standby Operation LED—When this LED is on, it indicates the hood is operating at a lower
face velocity or flow (active only with a ZPS).
Flow Alarm/Drive Failure LED—Indicates unsafe airflow condition when it is on.
Mute button—Push the button to silence an alarm. The Flow Alarm LED will remain on.
Mute mode is reset when the alarm conditions clear.
Emergency Exhaust button with LED—Push this button to activate the emergency
exhaust mode. The alarm will sound and the LED at the left of the Emergency Exhaust
button will blink. In this mode, exhaust air is at its maximum flow. (NOTE: On constant
volume hoods, this button is used only to test the alarm circuit, not modulate the exhaust
valve.) Push the button again to turn off the emergency exhaust/test mode.
Light sensor—Detects the light level in the room. Triggers the energy waste alert, if
configured. (Available only on the FHM631.)
OPTIONAL FEATURES
Diversity Alarm LED—Alerts lab users to reduce the total flow by closing their sashes. This visual alarm is
triggered when the flow demand exceeds the flow limit and a diversity alarm is generated by the Celeris system or
the BMS.
Power Failure Alarm LED—Activates when there is a loss of power. Used in conjunction with the Power Loss
Alarm (PLA). With this option, a separate PLA module is provided with the monitor.
The optional PLA circuit:
• Has a sealed lead acid battery that uses +15 Vdc to recharge the battery while the system is powered. The
battery recharges enough in 8 hours to power the alarm circuit for 24 hours. The battery’s expected service life
is five years.
• Detects the loss of ±15 Vdc system power. Not available for 24 Vac powered monitors.
• Trips a solid-state relay that causes the battery to provide power to the monitor’s alarm circuit.
• Drives the audible alarm and LED on the monitor to indicate loss of power.
The DIN standard requires PLAs on fume hood monitors, as well as periodic functional testing. Therefore, we
provide a test button on all PLA option enclosures. Units can be tested by pushing and holding the test button for
at least four seconds. This cuts the system’s power, trips the relay and tests the battery and alarm circuitry. No
other maintenance is required.
NOTE: The PLA must be powered by the same ±15 Vdc that powers the monitor.
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—3 OF 16
FEATURES (CONTINUED)
FEATURE/OPTION DESCRIPTION
CV
Two-
position
VAV
FHM530
(CVV)
FHM530
(PEV or
BEV)
FHM430
(EXV)
FHM631
(EXV)
FACEPLATE*
Face velocity display
Displays the current face velocity setting in either English or
metric units.
Operating mode LED Specifies use of hood.
Emergency exhaust LED Indicates that the emergency exhaust has been activated. **
Caution flow alarm Alarm due to either pressure alarm, valve jam alarm failure.
Spare or Diversity LED
Customer specified LED (spare) or diversity alarm condition
has been initiated.
†‡ †‡ †‡ †‡
Emergency exhaust button LED Emergency exhaust button has been pushed.
Setback LED Setback mode has been activated.
Power loss LED Power loss mode has been activated. † † † †
Emergency exhaust override button Provides a means of activating the emergency exhaust.
Mute button Provides a means of muting the audible alarms.
Light sensor Activates energy waste alert.
CONTROL
N.O. microswitch input Activates the two-state command to the valves.
Sash position input
Activates the two-state or variable air volume (VAV)
command to the valves.
Sash opening alarm setting Sets the sash height alarm set point.
Two-position switch point setting Sets the trip point for the two-state command operation.
Standby mode input (e.g., ZPS) Activates standby mode.
Switch input Activates decommission mode.
Emergency exhaust (locally or remotely) Ability to activate emergency exhaust with digital input. ***
VAV hood exhaust command output Command output to the exhaust source (valve).
VAV drive command output Command output to a fan drive.
24 Vdc relay output two-position mode
24 Vdc relay output to command 24 Vdc solenoids for two-
position application.
Primary-secondary option (e.g., teaching
hood)
Configures monitor for either primary or secondary control of
a teaching hood.
†
Standby velocity setting Sets the standby mode velocity.
Auto alarm mute
Audible alarms will mute automatically after 20 seconds when
this function is enabled.
Mute duration setting
Allows for resounding audible alarms after the Mute function
has been activated. Adjustable from 1-10, 15, or 20 minutes.
Sound volume setting
Adjust the sound volume. Two settings are available: high and
low.
Power loss alarm
Allows for power loss module that will display power loss
alarm when the monitor has lost main power (±15 Vdc only).
† † † †
Spare LED control
User-defined LED that can be engaged externally by means
of a digital input.
† † † †
Broken sash alarm
Alarm is generated when the sash sensor’s signal cable has
been broken or when a vertical sash sensor (VSS110)
retracting cable has detatched.
±15 Vdc or 24 Vac power May be powered by either AC or DC voltage.
MONITORING
Hood exhaust command/emergency
exhaust
0-10 Vdc signal represent command or > 10 Vdc represents
activation of emergency override.
Hood exhaust feedback 0-10 Vdc scaled feedback signal.
Alarm signal
0, 5, or 10 Vdc alarm signal generated at the valve can be
monitored at the fume hood monitor.
Normally open (NO) alarm relay NO contact output that de-energizes during alarm condition.
Sash position
0-10 Vdc signal proportionally represents sash position from
fully closed to fully open.
Hood decommissioned Monitors hood decommission command.
User status Monitors Zone Presence Sensor’s (ZPS’) activity.
* Faceplates are available in English, Danish, French Canadian, French, German, Japanese, Norwegian and Swedish.
** Does not modulate exhaust CV valve, but allows the operator to test the alarm circuit.
*** Cannot use this external input with decommission mode via the external switch or BMS command.
† Options
‡ Contact your representative to communicate details.
4 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
APPLICATIONS: HOOD CONTROL
Variable Air Volume (VAV) Fume Hoods (FHM430 and 631)
Fume hood containment is accomplished by maintaining proper face velocity through the variable sash opening.
Phoenix Controls fume hood monitors can be used on manifolded exhaust systems (with Phoenix Controls valves) and
on individual exhaust systems (with a variable speed drive by others).
Functions
• Constant face velocity control—The goal is to maintain a
constant face velocity (FV) as the sash opening varies.
Since the FV set point is known, a change in sash area
causes a linear change in exhaust flow
(FV x Area = Flow command).
Example: 5 ft2
x 100 ft/min = 500 ft3
/min
(or 0.5 m2 x 0.5 m/s x 3600 s/hr = 900 m3/hr)
• Setback of face velocity—Under many conditions, the
face velocity can be set back to provide safe containment
when the hood area is vacated. Setback face velocity is
adjustable to field conditions—typically between 60-100
fpm (or 0.3-0.5 m/s). (See the Zone Presence Sensor Product
Data Sheet for more details.)
• Alarms—The FHM430 and 631 monitors provide
indication of a fume hood’s operation. Alarms include:
• Insufficient differential static pressure as detected by
the valve’s pressure switch
• Incorrect airflow alarm (sash command ≠ closed-loop
feedback)
Hood
exhaust valve
200-1000 CFM
Sash sensor
To manifold
Variable
speed drive
200-1000 CFM
Sash sensor
Individual exhaust
Fume hood containment—valve system.
Standard Operation
Flow Alarm
Emergency
Exhaust Mute
Standby Operation
Variable Air Volume (VAV)
Standard Operation
Flow Alarm
Emergency
Exhaust Mute
Standby Operation
Fume hood containment—individual exhaust systems.
Variable Speed Drive
FV set
point
Hi
select
Exhaust flow
command
Sash Sensor
(position)
Area
Zone Presence Sensor
(operator status)
Reset
FV
Flow
command
FV x Area = Flow command
Emergency
exhaust override
Alarm
circuit
Alarm status
from valve or drive
Fume Hood Monitor
VAV fume hood monitor signal flow diagram.
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—5 OF 16
APPLICATIONS: HOOD CONTROL (CONTINUED)
Constant Volume (CV) and Two-state Fume Hoods (FHM530)
A Phoenix Controls (constant volume) CVV series airflow control valve provides the fume hood with a steady constant
exhaust flow, independent of duct pressure changes. The two-state airflow control valve (PEV or BEV series) provides
two-position exhaust control based on an operator’s presence at the fume hood.
Functions
• Alarm—The FHM530, together
with a differential pressure switch
mounted on a Phoenix Controls
airflow valve, indicates a fume
hood’s operation. An optional
sash opening alarm may be used
with a sash sensor or sash switch.
• Two-position control—The
FHM530, together with a Zone
Presence Sensor and a Phoenix
Controls PEV or BEV series
airflow control valve with an on-
board solenoid, provides two-
position exhaust control based on
an operator’s presence at the fume
hood. Other switching
mechanisms, such as a sash sensor
or sash switch, may be applied.
APPLICATIONS: ENERGY-SAVING
Face Velocity Setback (FHM430, 530, and 631)
Under many conditions, the face velocity can be set back to provide safe containment when the hood area is vacated.
Setback face velocity is adjustable to field conditions—typically between 60-100 fpm (or 0.3-0.5 m/s). (See the Zone
Presence Sensor Product Data Sheet for more details.)
Energy Waste Alert (FHM631)
The FHM631 is equipped with an energy waste alert, indicating the sash is open and the room is dark (adjustable light
intensity level). The display shows "ENRG," and an audible alarm sounds until the sash is closed or the lights are
illuminated.
Fume Hood Decommissioning Mode (FHM631)*
The fume hood decommissioning mode on the FHM631 allows a fume hood to be decommissioned when it is not in
use and the sash is fully closed. The exhaust flow is reduced below the fume hood’s minimum to the valve’s minimum
flow (e.g., 90 CFM for a 12-inch valve), and the display shows "OFF." This mode can be initiated in one of three ways:
1. Through pushbuttons on the monitor’s faceplate. Press and hold the Emergency Exhaust and Mute buttons for
three seconds. The display will flash "OFF?" Confirm by pressing the Mute button.
2. Through the external momentary switch (by others)
3. Through the BMS network command
The mode is exited automatically when the sash is opened.
* NOTES:
• Decommission mode cannot be used in drive applications.
• Proper standard operating procedures (SOPs) must be in place to remove all chemicals from the fume hood before it is decommissioned.
Fume hood
exhaust valve
600/1000 CFM
Zone Presence
Sensor
Fume hood
exhaust valve
1000 CFM
FHM530
(typical)
To manifold To manifold
Pressure
switch
Fume hood alarming and control—constant volume and two-state.
Standard Operation
Flow Alarm
Emergency
Exhaust Mute
Standby Operation
Constant Volume (CV) Two-state
6 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
INSTALLATION
The X30 Fume Hood Monitor Series is available with two mounting options: surface and flush.
Surface Mount Option
Required materials:
• Fume hood monitor, X30 series
• Phoenix Controls room wiring diagrams
• Two zinc-plated hex washer head slotted sheet metal
screws, Type A
Installation steps:
4. Use the mounting template to mark the points for
attaching the fume hood monitor to the fume hood (see
Figure 1).
5. Unsnap the front cover from the monitor:
• Loosen the bottom of the monitor housing and rotate
it slightly upward.
• Gently slide the enclosure up to release the two tabs
inside the top of the monitor housing from the two
grooves on the monitor base.
6. Set the monitor housing aside.
7. Using the enclosure base as a template, mark the mounting holes at both
ends. Be certain the fume hood monitor assembly is oriented as indicated in
the figure on the right. Set the enclosure base aside.
8. Drill pilot holes at the marked mounting holes.
9. Mount the monitor base to the fume hood using two sheet metal screws (see
Figure 2).
10. Run the cable through the mounting plate and strip back the conductors.
11. Remove the connectors from the monitor housing.
12. Complete the terminations to the connectors (refer to the wiring diagrams
provided).
13. Reinstall the connectors in the monitor housing.
14. Reattach the monitor housing to the monitor base (see Figure 3):
• Insert the two tabs at the top of the monitor housing in the grooves at
the top of the monitor base.
• Rotate the monitor housing downward and snap the bottom of the
cover onto the tabs in the monitor base to secure it.
Figure 1. Fume Hood Monitor mounting template (surface mount).
5"
(12.7 cm)
1.2"
(3.05 cm)
Completed enclosure
base with labels applied
Pilot hole (typical of 2)
(.2" dia. each/.5 cm)
.75" dia.
(1.9 cm)
Extents of monitor
6" x 2.5"
(15.24 x 6.35 cm)
5"
(12.7 cm)
Date
Init'l
Face Vel
Sash Hght
Min Flow
Max Flow
Standby Flow
Flow S-F
601-000-030
LBLLabel/SerialW/CE-USA
.625"
(1.59 cm)
Figure 2. Attaching the Fume Hood
Monitor to the fume hood (surface mount).
Slide enclosure cover onto
top of enclosure base.Then
rotate cover onto lower tabs
for a secure fit.Date
In
it'l
Fac
eVel
Sas
hHght
M
in
Flo
w
M
ax
Flo
w
StandbyFlow
Flo
w
S-F
601-000-030
LB
L
Label/Seria
lW
/C
E-U
SA
Sta
ndard
Operation
FlowAla
rm
Em
erg
enc
yExhau
st
Mute
Figure 3. Reattaching the Fume Hood Monitor
monitor housing to the base.
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—7 OF 16
INSTALLATION (CONTINUED)
Recess Mount Option
Required materials:
• Fume hood monitor, X30-RMT
series (includes recess mounting
bracket and screws)
• Phoenix Controls room wiring
diagrams
• Cable, based on wiring diagram
(not provided by Phoenix
Controls)
Installation steps:
1. Use the recess mounting template
to mark the area for cutout and
mounting holes for attaching the
fume hood monitor to the fume
hood (see Figure 4).
2. Cut a square hole, as necessary, for
the recess mounting kit.
3. Mark the holes for mounting the bracket:
• Using the recess mounting bracket as a template, place the bracket in the
cutout section of the fume hood and mark locations of holes required for
attaching the bracket to the fume hood.
• Set the bracket aside.
4. Drill .125" (.32 cm) pilot holes for mounting screws.
5. Unsnap the front cover from the monitor base:
• Loosen the bottom of the monitor housing and rotate it slightly upward.
• Gently slide the enclosure cover up to release the two tabs inside the top of
the monitor housing from the two grooves on the monitor base.
6. Set the enclosure cover aside.
7. Run the cable through the back of the monitor base.
8. Remove the connectors from the monitor housing.
9. Complete the terminations to the connectors (refer to provided wiring diagrams).
10. Reinstall the connectors in the monitor housing.
11. Attach the enclosure cover to the monitor base:
• Insert the two tabs at the top of the enclosure cover in the grooves at the top
of the monitor base.
• Rotate the enclosure cover downward and snap the bottom of the cover onto the tabs in the monitor base to
secure it.
12. Attach the monitor to the recess mounting bracket (see Figure 6 on the next page):
• Slide the monitor assembly into the mounting bracket from the back.
• Insert the L-strap into the top slot on the recess mounting bracket as shown.
• Rotate the L-strap and slide it downward through the bottom slot in the recess mounting bracket.
2.6"
(6.6 cm)
6.25"
(15.88 cm)
6.8"
(17.27 cm)
Pilot hole (typical of 2)
(.125" dia. each/.32 cm)
3.25"
(8.26 cm)
7.5"
(19.05 cm)
1.9"
(4.83 cm)
Figure 4. Fume Hood Monitor mounting template (recess mount).
Figure 5. Attaching the Fume Hood
Monitor to the fume hood (recess
mount).
8 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
INSTALLATION (CONTINUED)
Recess Mount Option (continued)
13. Mount the monitor and bracket assembly to the
fume hood (see Figure 5 on the previous page):
• Slide the assembly into the cutout area.
• Using the mounting screws provided, attach
the bracket to the fume hood.
Retrofit Jobs
Required materials:
• Fume hood monitor, X30 series (includes recess
mounting bracket assembly and screws)
• Phoenix Controls room wiring diagrams
• Cable, based on wiring diagram (not provided
by Phoenix Controls)
Installation steps:
1. Use the existing monitor to cut out hole and
screw pilot holes from the previous FHM X10
Series monitor installation. Ensure that the
dimensions of the cutout and mounting holes
match those given here (see Figure 7).
2. Follow steps 5-12 from the standard Recess
Mount installation procedure above.
3. Mount the monitor and bracket assembly to
the fume hood (see Figure 8):
• Slide the U-nut (included) over the
existing mounting holes with the spring
side inward.
• Slide the assembly into the cutout area.
• Using the mounting screws provided, attach the bracket to the fume hood.
Phoenix Controls Wiring Recommendations
• The fume hood monitor and exhaust valve should be powered from the same source.
• If control over the fume hoods is important during a power failure or if airflow
conservation is required, fume hood monitors should be placed on backup power;
otherwise, the fume hood valve defaults to its fail-safe state.
• Use cables recommended by Phoenix Controls.
• Stranded wire is strongly recommended for ease of installation.
• Follow good wiring practices:
• Do not run the signal cable in the same conduit or wireway as the power cables.
• If the signal cable must cross power cables, it is best to do so at a 90-degree angle.
• Shield or drain wires, if required, should be wrapped with insulating tape to prevent
contact with exposed conductors or contacts.
• Maintain a consistent color code or polarity all the way through the wiring system.
• All connections to the Celeris valve controller (LVC) must meet the requirements of an
NEC Class 2 circuit.
• Local and national electrical codes take precedence.
L-strap
Figure 6. Attaching the Fume Hood Monitor to the recess mounting bracke
8.13"
(20.65 cm)
9.38"
(23.83 cm)
2.5"
(6.35 cm)
3.25"
(8.26 cm)
10.5"
(26.67 cm)
9.38"
(23.83 cm)
Figure 7. Fume Hood Monitor mounting template (retrofit jobs).
U-nut
Figure 8. Attaching the Fume Hood
Monitor to the fume hood (retrofit
jobs).
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—9 OF 16
POINTS AND WIRING (see submittal wiring diagram for project-specific details)
J2
TB1
TB2
TB4
1
1
1 TB3
1
FHM430/631
To fume hood
exhaust valve
(See note 1)
VSS110
Zone Presence
Sensor (ZPS)
200 Series
1
2
3
4
5
6
7
8
TB1
TB2
1
2
3
4
5
6
7
7
6
5
4
3
2
1
TB1
To TB1-2
To TB1-7
Reel
Sensor
Red
Black
ISO Ground
Not Used
User Status
Secondary In
Not Used
Ground
+15 Vdc/24 Vac
3 2 1
TB4
From AC
Power
Supply
24VacH
24VacN
8c
(See note 2)
Notes:
1. Eight-conductor wire is Belden 9421 (22 AWG) or equivalent. (Tape back unused conductors.)
2. Sash sensor is provided with two-conductor cable. See combination sash sensors for exception.
FHM430 and 631
Typical Wiring Diagram
Monitor Termination Points Variable Air Volume (VAV) Application
TERMINAL BLOCK POINTS
1 2 3 4 5 6 7 8
TB1 +12 Vdc Analog Ground External
Emergency/
Decommission
Input
Command Feedback Tri-level Alarm User Status Sash Signal
TB2 Sash Sensor Analog Ground — Decommission
Output
Power Loss
Input
Spare LED (+) Spare LED (-) —
TB4 24 Vac H/
+15 Vdc
Functional
Ground
24 Vac N/
-15 Vdc
— — — — —
!
See pages 12-13 for DC wiring details, as well as the wiring of
options and ancillary equipment.
10 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details)
J2
TB1
TB2
TB3
TB4
J1
1
1
1
1
Note:
Requirements for transformers:
• Secondary power shall be fused externally to current limit of 4 amps, in accordance with NEC CL2 power requirements. Thermal interrupts
are not recommended.
• Do not earth ground secondary transformers.
• Maximum cable length for 96 VA load is 110’.
• Each pressurization zone shall have a dedicated single-phase primary circuit or a secondary circuit disconnect.
FHM530
Monitor Termination Points Constant Volume Application
COMPONENTS
TERMINAL BLOCK POINTS
Device Function
J1
IN for DC solenoid
OUT for AC solenoid
1 2 3 4 5 6 7 8
TB1 +12 Vdc Analog Ground External
Emergency
NOV Command NCV Command Pressure
Switch
User Status Sash Signal
TB2 Sash Sensor Analog Ground — — Power Loss
Input
Spare LED (+) Spare LED (-) —
TB3 Solenoid
Command
Common
Alarm NO Alarm Common — — — — —
TB4 24 Vac H/
+15 Vdc
NOV/NCV
Common
Functional
Ground
24 Vac N
/-15 Vdc
— — — — —
Phoenix Constant Volume
Venturi Valve
FHM530
From AC
Power Supply
24VacH
24VacN
TB1
1
2
3
4
5
6
7
8
3 2 1
TB4
Pressure Switch
NO
COM P
!
See page 12 for DC wiring details, as well as
the wiring of options and ancillary equipment.
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—11 OF 16
POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details)
AC Solenoid Valve
Pressure Switch
Phoenix Two-state
Venturi Valve
To TB1-4 (-)
To TB4-1 (+)
1
2
3
To TB1-2 (NO)
To TB1-6 (COM) P
I
3
2
1
TB3
FHM530
From AC
Power Supply
(See note 2)
24VacH
24VacN
TB1
1
2
3
4
5
6
7
8
3 2 1
TB4
Zone Presence
Sensor (ZPS)
200 Series
7
6
5
4
3
2
1
TB1 To TB1-2
To TB1-7
ISO Ground
Not Used
User Status
Secondary In
Not Used
Ground
+15 Vdc/24 Vac
J1 OUT
See note 1
DC Solenoid Valve
Pressure Switch
Phoenix Two-state
Venturi Valve
To TB1-4 (-)
To TB1-1 (+)
1
2
3
To TB1-2 (NO)
To TB1-6 (COM) P
H
3
2
1
TB3
FHM530
From DC
Power Supply
(See note 2)
+15Vdc
GROUND
TB1
1
2
3
4
5
6
7
8
3 2 1
TB4
Zone Presence
Sensor (ZPS)
200 Series
7
6
5
4
3
2
1
TB1 To TB1-2
To TB1-7
ISO Ground
Not Used
User Status
Secondary In
Not Used
Ground
+15 Vdc/24 Vac
J1 IN
-15Vdc
Notes:
1. For AC-powered solenoid, two-position applications, you must install a field jumper between TB3-1 and TB4-3 (not required for DC-
powered solenoids).
2. AC solenoids require the fume hood monitor to have AC power. DC solenoids require the fume hood monitor to have DC power.
FHM530 (continued)
Two-state Application: AC Solenoid
Two-state Application: DC Solenoid
!
See page 12 for the wiring of options and ancillary equipment.
See page 12 for the wiring of options and ancillary equipment.
12 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details)
Wiring of Options and Ancillary Equipment for X30 Fume Hood Monitor Series
FHM430
Primary/Secondary Wiring
(for AC-powered units)
FHM430
Primary/Secondary Wiring
(for DC-powered units)
FHM430/631/530 External Emergency
Override, Method 1
FHM430/631/530 External Emergency
Override, Method 2
FHM430/631/530 Power Loss Alarm FHM430/631/530 Optional/Diversity LED
FHM530 Alarm Relay Zone Presence Sensor (ZPS100 Series) X30 Series: DC Wiring Details
Primary Unit
3 2 1
TB4
From
Power
Supply
24VacH
24VacN
TB3
1
2
3
3 2 1
TB4
24VacH
24VacN
TB3
Secondary Unit
2c
1
2
3
Primary Unit
From
Power
Supply
Secondary Unit
TB2
1
2
3
4
5
6
7
3 2 1
TB4
+15
GROUND
-15
TB3
1
2
3
TB3
2c
1
2
3
TB2
1
2
3
4
5
6
7
3 2 1
TB4
+15
GROUND
-15
FHMX30
Contact Closure
(provided by others)
1
2
3
4
5
6
7
8
TB1
2c
FHMX30
1
2
3
4
5
6
TB1
10-volt command
from BMS to initiate
emergency override
2c
Power Loss
Alarm Unit
1
2
3
4
TB1
To FHMX30
Monitor
To TB4-1
To TB2-2
To TB4-3
To TB2-5
4c
+15
Ground
-15
Output
NOTE: The Power Loss Alarm unit will work
only if powered by the same DC power
source serving the monitor. AC-powered
monitors cannot have a Power Loss Alarm.
FHMX30
1
2
3
4
5
6
7
8
TB1
TB2
1
2
3
4
5
6
7
2c
Contact Closure
(provided by others)
1 K Ohm Resistor
(provided by others)
FHM530
1
2
3
TB3
To BMS for
monitoring
purposes
NO
COM
ZPS110
7
6
5
4
3
2
1
TB1
To FHM X30
Monitor
To TB4-3
To TB1-2
To TB4-1
4c
Sync Out
Sync In
User Status
Secondary
-15
Ground
+15
To TB1-7
NOTE: ZPS100 series units can be powered
by only DC voltage. If the FHM X30 is
powered by AC, then separate power must
be brought to the ZPS100 unit.
TB2
1
2
3
4
5
6
7
3 2 1
TB4
From DC
Power
Supply
+15
GROUND
-15
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—13 OF 16
POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details)
Decommission Mode Wiring for FHM631 (in addition to the wiring shown on page 9)
Decommission Mode through the Fume Hood Monitor’s Pushbutton
Decommission Mode through a Momentary External Switch
Decommission Mode through a BMS Command*
TB2
1
2
3
4
5
6
7
GROUND
Celeris Valve
Controller (LVC)
1
2
3
4
5
6
7
8
TB1
FHM631
DIGITAL OUT1
DIGITAL IN+
DIGITAL IN-
8c
GROUND
Celeris Valve
Controller (LVC)
1
2
3
4
5
6
7
8
TB1
FHM631
DIGITAL OUT1
DIGITAL IN+
DIGITAL IN-
EXTERNAL
SWITCH
8c
2c
1
2
3
4
5
6
7
8
TB1
TB2
1
2
3
4
5
6
7
+12 V
+12 V
Celeris Valve
Controller (LVC)
1
2
3
4
5
6
7
8
TB1
FHM631
DIGITAL OUT1
DIGITAL IN+
DIGITAL IN-
TB2
1
2
3
4
5
6
7
NO
COM
2c
1
2
3
4
5
6
7
8
TB1
TB2
1
2
3
4
5
6
7
GROUND
8c
*Note: Do not use this configuration if the LVC and fume hood monitor are using separate power sources.
14 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
MAINTENANCE
Phoenix Controls fume hood monitors require no ongoing preventative maintenance. Once the field engineer has
completed the field setup, the monitors will provide years of continuous operation.
TROUBLESHOOTING
Phoenix Controls fume hood monitors alert the operator of alarm conditions. Generally, this alarm is caused by a
problem condition in the exhaust duct (e.g., fan failure). A trained facilities person may troubleshoot the system from
the monitor with a digital voltmeter and perhaps a magnehelic gauge.
Alarms
The types of alarms for the X30 Series Fume Hood Monitor, the audible and visual cues for each alarm, and the purpose
for each alarm are listed in the table below.
Replacement Part Part Number
FHM631 board 860-200-108
FHM630 board 860-200-102
FHM530 board 860-200-109
FHM430 board 860-200-111
FHM430 primary board 860-200-110
FHM430 secondary board 860-200-112
X30 recess mount retrofit kit (replaces X10 models with X30) 260-270-004
X30 recess mount kit 260-270-005
Type of Alarm
Alarm Indicators
Reason for AlarmVisual Alarm
(LED blink rate*)
Audible Alarm
(beep rate**)
Signals (as seen
in analog valve
systems)
Not commissioned
Standard Operation—Fast
Flow Alarm—Fast
631 Display—Er-c
None N/A The unit requires commissioning.
Low differential static pressure
Flow Alarm—Slow
FHM631 display—Blank
Fast TB1, 6 = 10 Vdc
The hood’s pressure switch is open
(FHM631 and 430 only).
Flow Alarm—Slow Fast N/A
The hood’s pressure switch is open
(FHM530 only).
Incorrect airflow (i.e., valve jam) Flow Alarm—Fast Slow TB1, 6 = 5 Vdc
The valve’s position is jammed
(FHM631 and 430 only).
Sash opening alarm Flow Alarm—Slow Slow TB1, 6 = 0 Vdc***
The sash opening exceeds the calibrated alarm
set point.
Broken sash sensor alarm Flow Alarm—Fast Fast
TB1, 6 = 0 Vdc***
TB1, 8 > 10.4 Vdc
The sash input is greater than the maximum
resistive input for the application, or the sash
sensor signal cable is broken or loose.
Emergency exhaust override
alarm (local)
Emerg Exh—Slow Slow
TB1, 6 = 0 Vdc
TB1, 4 > 10.4 Vdc
The emergency exhaust button was pushed.
Emergency exhaust override
alarm (BMS external)
Emerg Exh—Fast Slow
TB1, 6 = 0 Vdc***
TB1, 4 > 10.4 Vdc
The externally controlled emergency input is in
emergency state.
Energy waste alert FHM631 display—ENRG Slow N/A
The sash is open and room lights are off
(FHM631 only).
Power fail alarm
Power loss LED on every 4
seconds
3 beeps every 10
seconds
N/A The fume hood monitor lost ±15 Vdc power.
Optional alarm
Optional alarm LED driven
by customer
None TB1, 6 = 0 Vdc Customer specific alarm or diversity alarm
Loss of alarm line
Flow Alarm—Slow
631 Display—Blank
Fast TB1, 6 = 12 Vdc The alarm signal cable is broken.
Loss of feedback signal
Standard Operation—Solid
Flow Alarm—Slow
631 Display—Low value
Slow TB1, 6 = 0 Vdc Feedback signal cable is broken.
Notes:
* A fast blink rate resembles a flashing light, while a slow rate resembles a fluttering light.
** A fast beep rate is 10 beeps per second, while a slow rate is 2 beeps per second.
*** 5 Vdc will be detected when these alarms are generated by a Celeris system. If the fume hood monitor generates this condition first, the Celeris signal will
supercede after it is generated.
©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—15 OF 16
TROUBLESHOOTING (CONTINUED)
Troubleshooting Guide
Use the information in the tables below to troubleshoot the X30 Series Fume Hood Monitor.
Variable Air Volume (VAV) Valve Systems (FHM430 and 631 only)
Constant Volume (CV) and Two-state Valve Systems (FHM530)
Problem at Fume
Hood Monitor
Voltage at TB1-6
in Monitor
Possible Cause/Solution
1. The monitor is in flow
alarm.
> 10 V
5 V
Voltages may vary.
A. Low static pressure across valve
• Loss of airflow—Check fan operation and duct blockage.
• Too many sashes open at one time—Close sashes.
• Valve failed open—Check pneumatic and mechanical connections.
B. Incorrect valve position
• Valve failed open.
• Sash open beyond the maximum allowable position—Lower sash.
• A broken sash cable—Check all sash sensor connections.
C. Equipment/connection problems
• Blocked or kinked pressure switch tubing—Correct tubing.
• Wiring terminations between monitor and control device—Correct terminations.
• Malfunctioning alarm circuits—Verify proper static.*
• Monitor miscalibration—Recalibrate.*
2. The monitor indicates
normal operation, but
the actual face velocity
has been measured
high or low.
0 V A. Low static pressure
• Differential pressure between the pressure switch set point and the low end of the
static pressure operating range [i.e., 0.3" and 0.6" wc (75-150 Pa) for medium
pressure valves] will not trip the alarm circuit.
• Measure differential pressure. If it is low, see Possible Cause/Solution 1A above.
B. Monitor miscalibration—Recalibrate.*
3. The monitor has
malfunctioned.
• No display.
• Cannot mute alarm.
Voltages may vary. A. Loss of power
• Check power at monitor TB4-1 (24 Vac H/+15 Vdc) and TB4-3 (24 Vac N/-15 Vdc).
• Check wiring connections at monitor, valve, and power supply.
• Verify power supply has input voltage (120 Vac or 240 Vac).
B. Defective monitor—Replace board on monitor.*
4. The flow remains
constant through sash
travel.
A. Broken sash sensor cable—Replace sensor.
B. Monitor miscalibration—Recalibrate.*
* Contact Phoenix Controls Product Support Center for assistance.
Problem Possible Cause/Solution
1. The monitor is in flow alarm. Low static pressure across valve.
• Loss of airflow. Check fan operation and duct blockage.
• Blocked or kinked pressure switch tubing. Correct the tubing.
2. The airflow was measured
improperly in a:
A. Constant volume system.
B. Two-position system.
A. Incorrect valve position. Realign the pivot arm.
B. Incorrect valve position. Check pneumatic tubing and pressure. Also check solenoid valve operation.
16 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07
TROUBLESHOOTING (CONTINUED)
Variable Air Volume (VAV) Drive Systems (FHM430 and 631 only)
Energy-saving Options (FHM631 only)
Problem Possible Cause/Solution
1. The monitor is in flow alarm. A. Loss of airflow
• Check fan operation and duct blockage.
• Check the drive for proper operation.
B. Monitor/connection problems
• The wiring terminations between monitor and control device are incorrect. Correct the
terminations.
• The monitor was miscalibrated. Recalibrate the monitor as needed.*
• The monitor is defective. Replace the board on the monitor.*
2. The monitor indicates normal
operation, but actual face velocity
has been measured high or low.
The monitor was miscalibrated. Recalibrate the monitor as needed.*
3. The monitor is malfunctioning.
• No display
• Cannot mute alarm
A. Loss of power
• Check the wiring connections at the monitor, drive and power supply.
• Verify that the power supply has input voltage.
• Check power at the monitor.
B. The monitor is defective. Replace the board on the monitor. Recalibrate the monitor as needed.*
* Contact Phoenix Controls Product Support Center for assistance.
Problem Possible Cause/Solution
1. The monitor will not enter energy
waste alert.
A. The low-light level setting is too high. Recalibrate the monitor.
B. The energy waste alert is not enabled. Recalibrate the monitor.
2. The monitor will not exit energy
waste alert.
The high-light level setting is too high. Recalibrate the monitor.
3. The monitor will not enter
decommission mode.
A. The sash is not fully closed. Close the sash.
B. Decommission mode is not enabled. Recalibrate the monitor.
C. Check the wiring to the valve.
4. The monitor will not exit
decommission mode.
A. The digital input (DI) on the valve is not configured properly. Reconfigure this input.
B. The monitor is defective. Replace and recalibrate the monitor.

Contenu connexe

Tendances

RA TechED 2019 - CL01 - Integrated Architecture System Hardware - what's new
RA TechED 2019 -  CL01 - Integrated Architecture System Hardware - what's newRA TechED 2019 -  CL01 - Integrated Architecture System Hardware - what's new
RA TechED 2019 - CL01 - Integrated Architecture System Hardware - what's newRockwell Automation
 
Distributed Control System (Presentation)
Distributed Control System (Presentation)Distributed Control System (Presentation)
Distributed Control System (Presentation)Thunder Bolt
 
CENTUM VP - 2014 version
CENTUM VP - 2014 versionCENTUM VP - 2014 version
CENTUM VP - 2014 versionYokogawa
 
When is a SIL Rating of a Valve Required?
When is a SIL Rating of a Valve Required?When is a SIL Rating of a Valve Required?
When is a SIL Rating of a Valve Required?ISA Interchange
 
Honeywell Experion HS
Honeywell Experion HSHoneywell Experion HS
Honeywell Experion HSShivam Singh
 
Client oriented approach for physical examination of animals
Client oriented approach for physical examination of animalsClient oriented approach for physical examination of animals
Client oriented approach for physical examination of animalsHimanshu Pandey
 
PLC Introduction Details
PLC Introduction DetailsPLC Introduction Details
PLC Introduction Detailssuhaskhadake
 
Foundation fieldbus technology
Foundation fieldbus technologyFoundation fieldbus technology
Foundation fieldbus technologyMohamed A Hakim
 
Fieldbus Tutorial Part 1 - Fieldbus Overview
Fieldbus Tutorial Part 1 - Fieldbus OverviewFieldbus Tutorial Part 1 - Fieldbus Overview
Fieldbus Tutorial Part 1 - Fieldbus OverviewEmerson Exchange
 
Internal and external parasites prof.dr Hamed Attia
Internal and external parasites prof.dr Hamed AttiaInternal and external parasites prof.dr Hamed Attia
Internal and external parasites prof.dr Hamed Attiahamed attia
 

Tendances (20)

Cours_pneumatique.ppt
Cours_pneumatique.pptCours_pneumatique.ppt
Cours_pneumatique.ppt
 
RA TechED 2019 - CL01 - Integrated Architecture System Hardware - what's new
RA TechED 2019 -  CL01 - Integrated Architecture System Hardware - what's newRA TechED 2019 -  CL01 - Integrated Architecture System Hardware - what's new
RA TechED 2019 - CL01 - Integrated Architecture System Hardware - what's new
 
CENTUM CS300 R3
CENTUM CS300 R3CENTUM CS300 R3
CENTUM CS300 R3
 
Distributed Control System (Presentation)
Distributed Control System (Presentation)Distributed Control System (Presentation)
Distributed Control System (Presentation)
 
CENTUM VP - 2014 version
CENTUM VP - 2014 versionCENTUM VP - 2014 version
CENTUM VP - 2014 version
 
PlantPAx Process
PlantPAx ProcessPlantPAx Process
PlantPAx Process
 
When is a SIL Rating of a Valve Required?
When is a SIL Rating of a Valve Required?When is a SIL Rating of a Valve Required?
When is a SIL Rating of a Valve Required?
 
Dcs
DcsDcs
Dcs
 
Honeywell Experion HS
Honeywell Experion HSHoneywell Experion HS
Honeywell Experion HS
 
NFS-320
NFS-320NFS-320
NFS-320
 
Intro to IO-Link
Intro to IO-LinkIntro to IO-Link
Intro to IO-Link
 
Client oriented approach for physical examination of animals
Client oriented approach for physical examination of animalsClient oriented approach for physical examination of animals
Client oriented approach for physical examination of animals
 
Plc 101
Plc 101Plc 101
Plc 101
 
Mammary tumors
Mammary tumorsMammary tumors
Mammary tumors
 
PLC Introduction Details
PLC Introduction DetailsPLC Introduction Details
PLC Introduction Details
 
Foundation fieldbus technology
Foundation fieldbus technologyFoundation fieldbus technology
Foundation fieldbus technology
 
Fieldbus Tutorial Part 1 - Fieldbus Overview
Fieldbus Tutorial Part 1 - Fieldbus OverviewFieldbus Tutorial Part 1 - Fieldbus Overview
Fieldbus Tutorial Part 1 - Fieldbus Overview
 
Presentaton on Plc & Scada
Presentaton on Plc & ScadaPresentaton on Plc & Scada
Presentaton on Plc & Scada
 
Foundation fieldbus
Foundation fieldbusFoundation fieldbus
Foundation fieldbus
 
Internal and external parasites prof.dr Hamed Attia
Internal and external parasites prof.dr Hamed AttiaInternal and external parasites prof.dr Hamed Attia
Internal and external parasites prof.dr Hamed Attia
 

Similaire à Phoenix controls

4603-0001 - LCD Annunciator - 4603-9101.PDF
4603-0001 - LCD Annunciator - 4603-9101.PDF4603-0001 - LCD Annunciator - 4603-9101.PDF
4603-0001 - LCD Annunciator - 4603-9101.PDFSourovSamadder
 
Machine monitoring system
Machine monitoring systemMachine monitoring system
Machine monitoring systemPrem Baboo
 
Datasheet foxboro eckardt srd991 18463765
Datasheet foxboro eckardt srd991 18463765Datasheet foxboro eckardt srd991 18463765
Datasheet foxboro eckardt srd991 18463765IkhuoshoAbode
 
4903-0010 - Conventional Strobe with Horn.pdf
4903-0010 - Conventional Strobe with Horn.pdf4903-0010 - Conventional Strobe with Horn.pdf
4903-0010 - Conventional Strobe with Horn.pdfSourovSamadder
 
Siemens Terminal Box Controller (APOGEE Automation System )
Siemens Terminal Box Controller (APOGEE Automation System )Siemens Terminal Box Controller (APOGEE Automation System )
Siemens Terminal Box Controller (APOGEE Automation System )CONTROLS & SYSTEMS
 
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.com
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.comCatalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.com
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.comBeeteco
 
K85000-0391 -- Releasing Module Flyer.pdf
K85000-0391 -- Releasing Module Flyer.pdfK85000-0391 -- Releasing Module Flyer.pdf
K85000-0391 -- Releasing Module Flyer.pdfSecorpSAS
 
Elc Inverter Introduction
Elc Inverter IntroductionElc Inverter Introduction
Elc Inverter Introductionscottburge
 
Innovations - Ability 11.02.14
Innovations - Ability 11.02.14Innovations - Ability 11.02.14
Innovations - Ability 11.02.14Overbury
 
Horn with Strobe 49AV-WRF.pdf
Horn with Strobe 49AV-WRF.pdfHorn with Strobe 49AV-WRF.pdf
Horn with Strobe 49AV-WRF.pdfSourovSamadder
 
LCD Annunciators
LCD AnnunciatorsLCD Annunciators
LCD Annunciatorsaescotom
 
Controlador ORP Milton Roy
Controlador ORP Milton RoyControlador ORP Milton Roy
Controlador ORP Milton RoyVibropac
 
Functional Safety Frequency to DC Transmitter
Functional Safety Frequency to DC TransmitterFunctional Safety Frequency to DC Transmitter
Functional Safety Frequency to DC TransmitterArjay Automation
 
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec Sheet
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec SheetHEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec Sheet
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec SheetThorne & Derrick UK
 

Similaire à Phoenix controls (20)

4603-0001 - LCD Annunciator - 4603-9101.PDF
4603-0001 - LCD Annunciator - 4603-9101.PDF4603-0001 - LCD Annunciator - 4603-9101.PDF
4603-0001 - LCD Annunciator - 4603-9101.PDF
 
Machine monitoring system
Machine monitoring systemMachine monitoring system
Machine monitoring system
 
Datasheet foxboro eckardt srd991 18463765
Datasheet foxboro eckardt srd991 18463765Datasheet foxboro eckardt srd991 18463765
Datasheet foxboro eckardt srd991 18463765
 
4903-0010 - Conventional Strobe with Horn.pdf
4903-0010 - Conventional Strobe with Horn.pdf4903-0010 - Conventional Strobe with Horn.pdf
4903-0010 - Conventional Strobe with Horn.pdf
 
13.Safety & navigation
13.Safety & navigation13.Safety & navigation
13.Safety & navigation
 
Siemens Terminal Box Controller (APOGEE Automation System )
Siemens Terminal Box Controller (APOGEE Automation System )Siemens Terminal Box Controller (APOGEE Automation System )
Siemens Terminal Box Controller (APOGEE Automation System )
 
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.com
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.comCatalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.com
Catalog Biến tần Frenic ACE Mini (Fuji Electric) - Beeteco.com
 
K85000-0391 -- Releasing Module Flyer.pdf
K85000-0391 -- Releasing Module Flyer.pdfK85000-0391 -- Releasing Module Flyer.pdf
K85000-0391 -- Releasing Module Flyer.pdf
 
Elc Inverter Introduction
Elc Inverter IntroductionElc Inverter Introduction
Elc Inverter Introduction
 
Dn 60779 pdf
Dn 60779 pdfDn 60779 pdf
Dn 60779 pdf
 
Innovations - Ability 11.02.14
Innovations - Ability 11.02.14Innovations - Ability 11.02.14
Innovations - Ability 11.02.14
 
EL-FLOW Prestige Mass Flow Controller and Flow Meter
EL-FLOW Prestige Mass Flow Controller and Flow MeterEL-FLOW Prestige Mass Flow Controller and Flow Meter
EL-FLOW Prestige Mass Flow Controller and Flow Meter
 
Fadal Diagnostics Maintenance Manual
Fadal Diagnostics Maintenance ManualFadal Diagnostics Maintenance Manual
Fadal Diagnostics Maintenance Manual
 
CATV Optical Transmitter
CATV Optical TransmitterCATV Optical Transmitter
CATV Optical Transmitter
 
Horn with Strobe 49AV-WRF.pdf
Horn with Strobe 49AV-WRF.pdfHorn with Strobe 49AV-WRF.pdf
Horn with Strobe 49AV-WRF.pdf
 
LCD Annunciators
LCD AnnunciatorsLCD Annunciators
LCD Annunciators
 
Controlador ORP Milton Roy
Controlador ORP Milton RoyControlador ORP Milton Roy
Controlador ORP Milton Roy
 
Functional Safety Frequency to DC Transmitter
Functional Safety Frequency to DC TransmitterFunctional Safety Frequency to DC Transmitter
Functional Safety Frequency to DC Transmitter
 
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec Sheet
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec SheetHEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec Sheet
HEATFAST HF9101-v1 Ambient Sensing Heat Tracing Thermostat - Spec Sheet
 
Bien tan-mini
Bien tan-miniBien tan-mini
Bien tan-mini
 

Dernier

Air Compressor reciprocating single stage
Air Compressor reciprocating single stageAir Compressor reciprocating single stage
Air Compressor reciprocating single stageAbc194748
 
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best ServiceTamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Servicemeghakumariji156
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...HenryBriggs2
 
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKARHAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKARKOUSTAV SARKAR
 
Hostel management system project report..pdf
Hostel management system project report..pdfHostel management system project report..pdf
Hostel management system project report..pdfKamal Acharya
 
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...soginsider
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startQuintin Balsdon
 
Unleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapUnleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapRishantSharmaFr
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfJiananWang21
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdfKamal Acharya
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdfKamal Acharya
 
Computer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to ComputersComputer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to ComputersMairaAshraf6
 
Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdfKamal Acharya
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptDineshKumar4165
 
Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086anil_gaur
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadhamedmustafa094
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwaitjaanualu31
 
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Arindam Chakraborty, Ph.D., P.E. (CA, TX)
 

Dernier (20)

Air Compressor reciprocating single stage
Air Compressor reciprocating single stageAir Compressor reciprocating single stage
Air Compressor reciprocating single stage
 
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best ServiceTamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
 
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKARHAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
 
Hostel management system project report..pdf
Hostel management system project report..pdfHostel management system project report..pdf
Hostel management system project report..pdf
 
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Unleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapUnleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leap
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdf
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdf
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced LoadsFEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
 
Computer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to ComputersComputer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to Computers
 
Integrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - NeometrixIntegrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - Neometrix
 
Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdf
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.ppt
 
Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal load
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
 
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
 

Phoenix controls

  • 1. X30 SERIES FUME HOOD MONITORS 75 Discovery Way • Acton, Massachusetts 01720 • Telephone (978) 795-1285 • Fax (978) 795-1111 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—1 OF 16 Phoenix Controls X30 Series Fume Hood Monitors (FHMs) are used on fume hoods with Phoenix Controls valves for airflow control. Airflow control on these fume hoods is achieved with the use of constant volume valves (CVV), two-position valves (PEV or BEV) or variable air volume valves (VAV). Each FHM provides two primary functions: indication of hood exhaust operating condition and alarming. In VAV systems, each FHM also provides face velocity control and optional energy-saving features. SPECIFICATIONS Enclosure • Dimensions: 2.5" W x 6" H x 1.5" D (64 mm x 152 mm x 38 mm) • Color: White • IP44 compliant Operating Range 32-122 °F (0-50 °C) ambient 10-90% RH, non-condensing 8202’ (2500 m) altitude Power Requirements for Each Unit 24 Vac, ±10%, 50-60 Hz, 10 VA ±15 Vdc, ±5%, 220 mA Inputs and Outputs See table in "Features" section on page 2 for model-specific inputs and outputs. Input to Optional Use LED • Yellow LED indicates unique customer conditions • Wired directly from customer’s device to two dedicated conductors within the eight-wire signal cable • Limited to ≤ 12 Vdc with maximum current draw of 0.012 amps. Customer must install a 1 K Ω resistor in series with input signal. • Visual indication only, no audible Monitoring Points • Airflow exhaust device command • Actual exhaust airflow • 0-10 Vdc alarm: • 0 Vdc indicates normal operation • 5 Vdc indicates incorrect airflow • 10 Vdc indicates low differential static pressure • Sash position: 0-10 Vdc • User status • 10 K Ω minimum input impedence for monitoring system Backward Compatibility X30 Series Fume Hood Monitors are backward compatible with earlier standard models of Phoenix fume hood monitors. Readout (FHM631 only) The display shows one of the following measurements: • Cubic feet per minute (CFM) • Meters cubed per hour (m3 /h) • Liters per second (l/s) • Feet per minute (fpm) • Meters per second (m/s) FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. Power Loss Alarm Option (±15 Vdc powered monitor only) Indicates loss of power to the fume hood system. • During power failure, a red LED flashes once every 4 sec • Accompanied by short audible alarm “chirp” • Alarm continues for at least 64 hours or until power is restored Dimensions 4.75" W x 4" H x 4.75" D (121 mm x 102 mm x 121 mm) Color • Black enclosure • Yellow switch Operating Range 32-122 °F (0-50 °C) ambient Power Requirements ±15 Vdc @ 0.015 amp More details on this option are in the "Features" section on page 2. If the X30 Series Fume Hood Monitor is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. ! See wiring diagrams ORDERING GUIDE X30 Fume Hood Monitor (631 model shown). Standard Operation Flow Alarm Emergency Exhaust Mute Standby Operation BASE NUMBER TYPE OF FUME HOOD MONITOR 430 = Base monitor, EXV applications 530 = CVV, PEV or BEV applications 631 = Face velocity monitor, EXV applications FACEPLATE LANGUAGE DAN = Danish GER = German ENG = English JPN = Japanese FRC = French Canadian NOR = Norwegian FRN = French SWE = Swedish OPTIONS DFT = Drive failure text (FHM430 only) DIV = Diversity LED to be driven by external contact closure from a Celeris system or the BMS. PLA = Power Loss Alarm (±15 volt only; see note below) PRI = Primary (Available only on FHM430. Must be ordered with at least one secondary.) RET = Retrofit kit RMT= Recess mount kit SEC = Secondary (Available only on FHM430. Must be ordered with primary.) NOTE: By ordering the add-on option “PLA,” you will receive a separate Power Loss Alarm board and box that also mounts on top of the hood. This board interfaces to an X30 Series Fume Hood Monitor that includes an additional red LED with "Power Failure Alarm" text. This unit can be installed on DC-powered monitors only. FHM – – TABLE OF CONTENTS Specifications and Ordering Guide.............................1 Features......................................................................2 Applications: Hood Control.......................................4 Applications: Energy-saving........................................5 Installation .................................................................6 Points and Wiring ......................................................9 Maintenance and Troubleshooting ...........................14
  • 2. 2 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 FEATURES Standard Operation Flow Alarm Emergency Exhaust Mute Standby Operation Power Failure Alarm Diversity Alarm Face Velocity or Flow Display—Indicates the face velocity or flow of the hood. (Available only on the FHM631.) Standard Operation LED—When this LED is on, it indicates the hood is operating at the standard face velocity or flow. Standby Operation LED—When this LED is on, it indicates the hood is operating at a lower face velocity or flow (active only with a ZPS). Flow Alarm/Drive Failure LED—Indicates unsafe airflow condition when it is on. Mute button—Push the button to silence an alarm. The Flow Alarm LED will remain on. Mute mode is reset when the alarm conditions clear. Emergency Exhaust button with LED—Push this button to activate the emergency exhaust mode. The alarm will sound and the LED at the left of the Emergency Exhaust button will blink. In this mode, exhaust air is at its maximum flow. (NOTE: On constant volume hoods, this button is used only to test the alarm circuit, not modulate the exhaust valve.) Push the button again to turn off the emergency exhaust/test mode. Light sensor—Detects the light level in the room. Triggers the energy waste alert, if configured. (Available only on the FHM631.) OPTIONAL FEATURES Diversity Alarm LED—Alerts lab users to reduce the total flow by closing their sashes. This visual alarm is triggered when the flow demand exceeds the flow limit and a diversity alarm is generated by the Celeris system or the BMS. Power Failure Alarm LED—Activates when there is a loss of power. Used in conjunction with the Power Loss Alarm (PLA). With this option, a separate PLA module is provided with the monitor. The optional PLA circuit: • Has a sealed lead acid battery that uses +15 Vdc to recharge the battery while the system is powered. The battery recharges enough in 8 hours to power the alarm circuit for 24 hours. The battery’s expected service life is five years. • Detects the loss of ±15 Vdc system power. Not available for 24 Vac powered monitors. • Trips a solid-state relay that causes the battery to provide power to the monitor’s alarm circuit. • Drives the audible alarm and LED on the monitor to indicate loss of power. The DIN standard requires PLAs on fume hood monitors, as well as periodic functional testing. Therefore, we provide a test button on all PLA option enclosures. Units can be tested by pushing and holding the test button for at least four seconds. This cuts the system’s power, trips the relay and tests the battery and alarm circuitry. No other maintenance is required. NOTE: The PLA must be powered by the same ±15 Vdc that powers the monitor.
  • 3. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—3 OF 16 FEATURES (CONTINUED) FEATURE/OPTION DESCRIPTION CV Two- position VAV FHM530 (CVV) FHM530 (PEV or BEV) FHM430 (EXV) FHM631 (EXV) FACEPLATE* Face velocity display Displays the current face velocity setting in either English or metric units. Operating mode LED Specifies use of hood. Emergency exhaust LED Indicates that the emergency exhaust has been activated. ** Caution flow alarm Alarm due to either pressure alarm, valve jam alarm failure. Spare or Diversity LED Customer specified LED (spare) or diversity alarm condition has been initiated. †‡ †‡ †‡ †‡ Emergency exhaust button LED Emergency exhaust button has been pushed. Setback LED Setback mode has been activated. Power loss LED Power loss mode has been activated. † † † † Emergency exhaust override button Provides a means of activating the emergency exhaust. Mute button Provides a means of muting the audible alarms. Light sensor Activates energy waste alert. CONTROL N.O. microswitch input Activates the two-state command to the valves. Sash position input Activates the two-state or variable air volume (VAV) command to the valves. Sash opening alarm setting Sets the sash height alarm set point. Two-position switch point setting Sets the trip point for the two-state command operation. Standby mode input (e.g., ZPS) Activates standby mode. Switch input Activates decommission mode. Emergency exhaust (locally or remotely) Ability to activate emergency exhaust with digital input. *** VAV hood exhaust command output Command output to the exhaust source (valve). VAV drive command output Command output to a fan drive. 24 Vdc relay output two-position mode 24 Vdc relay output to command 24 Vdc solenoids for two- position application. Primary-secondary option (e.g., teaching hood) Configures monitor for either primary or secondary control of a teaching hood. † Standby velocity setting Sets the standby mode velocity. Auto alarm mute Audible alarms will mute automatically after 20 seconds when this function is enabled. Mute duration setting Allows for resounding audible alarms after the Mute function has been activated. Adjustable from 1-10, 15, or 20 minutes. Sound volume setting Adjust the sound volume. Two settings are available: high and low. Power loss alarm Allows for power loss module that will display power loss alarm when the monitor has lost main power (±15 Vdc only). † † † † Spare LED control User-defined LED that can be engaged externally by means of a digital input. † † † † Broken sash alarm Alarm is generated when the sash sensor’s signal cable has been broken or when a vertical sash sensor (VSS110) retracting cable has detatched. ±15 Vdc or 24 Vac power May be powered by either AC or DC voltage. MONITORING Hood exhaust command/emergency exhaust 0-10 Vdc signal represent command or > 10 Vdc represents activation of emergency override. Hood exhaust feedback 0-10 Vdc scaled feedback signal. Alarm signal 0, 5, or 10 Vdc alarm signal generated at the valve can be monitored at the fume hood monitor. Normally open (NO) alarm relay NO contact output that de-energizes during alarm condition. Sash position 0-10 Vdc signal proportionally represents sash position from fully closed to fully open. Hood decommissioned Monitors hood decommission command. User status Monitors Zone Presence Sensor’s (ZPS’) activity. * Faceplates are available in English, Danish, French Canadian, French, German, Japanese, Norwegian and Swedish. ** Does not modulate exhaust CV valve, but allows the operator to test the alarm circuit. *** Cannot use this external input with decommission mode via the external switch or BMS command. † Options ‡ Contact your representative to communicate details.
  • 4. 4 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 APPLICATIONS: HOOD CONTROL Variable Air Volume (VAV) Fume Hoods (FHM430 and 631) Fume hood containment is accomplished by maintaining proper face velocity through the variable sash opening. Phoenix Controls fume hood monitors can be used on manifolded exhaust systems (with Phoenix Controls valves) and on individual exhaust systems (with a variable speed drive by others). Functions • Constant face velocity control—The goal is to maintain a constant face velocity (FV) as the sash opening varies. Since the FV set point is known, a change in sash area causes a linear change in exhaust flow (FV x Area = Flow command). Example: 5 ft2 x 100 ft/min = 500 ft3 /min (or 0.5 m2 x 0.5 m/s x 3600 s/hr = 900 m3/hr) • Setback of face velocity—Under many conditions, the face velocity can be set back to provide safe containment when the hood area is vacated. Setback face velocity is adjustable to field conditions—typically between 60-100 fpm (or 0.3-0.5 m/s). (See the Zone Presence Sensor Product Data Sheet for more details.) • Alarms—The FHM430 and 631 monitors provide indication of a fume hood’s operation. Alarms include: • Insufficient differential static pressure as detected by the valve’s pressure switch • Incorrect airflow alarm (sash command ≠ closed-loop feedback) Hood exhaust valve 200-1000 CFM Sash sensor To manifold Variable speed drive 200-1000 CFM Sash sensor Individual exhaust Fume hood containment—valve system. Standard Operation Flow Alarm Emergency Exhaust Mute Standby Operation Variable Air Volume (VAV) Standard Operation Flow Alarm Emergency Exhaust Mute Standby Operation Fume hood containment—individual exhaust systems. Variable Speed Drive FV set point Hi select Exhaust flow command Sash Sensor (position) Area Zone Presence Sensor (operator status) Reset FV Flow command FV x Area = Flow command Emergency exhaust override Alarm circuit Alarm status from valve or drive Fume Hood Monitor VAV fume hood monitor signal flow diagram.
  • 5. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—5 OF 16 APPLICATIONS: HOOD CONTROL (CONTINUED) Constant Volume (CV) and Two-state Fume Hoods (FHM530) A Phoenix Controls (constant volume) CVV series airflow control valve provides the fume hood with a steady constant exhaust flow, independent of duct pressure changes. The two-state airflow control valve (PEV or BEV series) provides two-position exhaust control based on an operator’s presence at the fume hood. Functions • Alarm—The FHM530, together with a differential pressure switch mounted on a Phoenix Controls airflow valve, indicates a fume hood’s operation. An optional sash opening alarm may be used with a sash sensor or sash switch. • Two-position control—The FHM530, together with a Zone Presence Sensor and a Phoenix Controls PEV or BEV series airflow control valve with an on- board solenoid, provides two- position exhaust control based on an operator’s presence at the fume hood. Other switching mechanisms, such as a sash sensor or sash switch, may be applied. APPLICATIONS: ENERGY-SAVING Face Velocity Setback (FHM430, 530, and 631) Under many conditions, the face velocity can be set back to provide safe containment when the hood area is vacated. Setback face velocity is adjustable to field conditions—typically between 60-100 fpm (or 0.3-0.5 m/s). (See the Zone Presence Sensor Product Data Sheet for more details.) Energy Waste Alert (FHM631) The FHM631 is equipped with an energy waste alert, indicating the sash is open and the room is dark (adjustable light intensity level). The display shows "ENRG," and an audible alarm sounds until the sash is closed or the lights are illuminated. Fume Hood Decommissioning Mode (FHM631)* The fume hood decommissioning mode on the FHM631 allows a fume hood to be decommissioned when it is not in use and the sash is fully closed. The exhaust flow is reduced below the fume hood’s minimum to the valve’s minimum flow (e.g., 90 CFM for a 12-inch valve), and the display shows "OFF." This mode can be initiated in one of three ways: 1. Through pushbuttons on the monitor’s faceplate. Press and hold the Emergency Exhaust and Mute buttons for three seconds. The display will flash "OFF?" Confirm by pressing the Mute button. 2. Through the external momentary switch (by others) 3. Through the BMS network command The mode is exited automatically when the sash is opened. * NOTES: • Decommission mode cannot be used in drive applications. • Proper standard operating procedures (SOPs) must be in place to remove all chemicals from the fume hood before it is decommissioned. Fume hood exhaust valve 600/1000 CFM Zone Presence Sensor Fume hood exhaust valve 1000 CFM FHM530 (typical) To manifold To manifold Pressure switch Fume hood alarming and control—constant volume and two-state. Standard Operation Flow Alarm Emergency Exhaust Mute Standby Operation Constant Volume (CV) Two-state
  • 6. 6 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 INSTALLATION The X30 Fume Hood Monitor Series is available with two mounting options: surface and flush. Surface Mount Option Required materials: • Fume hood monitor, X30 series • Phoenix Controls room wiring diagrams • Two zinc-plated hex washer head slotted sheet metal screws, Type A Installation steps: 4. Use the mounting template to mark the points for attaching the fume hood monitor to the fume hood (see Figure 1). 5. Unsnap the front cover from the monitor: • Loosen the bottom of the monitor housing and rotate it slightly upward. • Gently slide the enclosure up to release the two tabs inside the top of the monitor housing from the two grooves on the monitor base. 6. Set the monitor housing aside. 7. Using the enclosure base as a template, mark the mounting holes at both ends. Be certain the fume hood monitor assembly is oriented as indicated in the figure on the right. Set the enclosure base aside. 8. Drill pilot holes at the marked mounting holes. 9. Mount the monitor base to the fume hood using two sheet metal screws (see Figure 2). 10. Run the cable through the mounting plate and strip back the conductors. 11. Remove the connectors from the monitor housing. 12. Complete the terminations to the connectors (refer to the wiring diagrams provided). 13. Reinstall the connectors in the monitor housing. 14. Reattach the monitor housing to the monitor base (see Figure 3): • Insert the two tabs at the top of the monitor housing in the grooves at the top of the monitor base. • Rotate the monitor housing downward and snap the bottom of the cover onto the tabs in the monitor base to secure it. Figure 1. Fume Hood Monitor mounting template (surface mount). 5" (12.7 cm) 1.2" (3.05 cm) Completed enclosure base with labels applied Pilot hole (typical of 2) (.2" dia. each/.5 cm) .75" dia. (1.9 cm) Extents of monitor 6" x 2.5" (15.24 x 6.35 cm) 5" (12.7 cm) Date Init'l Face Vel Sash Hght Min Flow Max Flow Standby Flow Flow S-F 601-000-030 LBLLabel/SerialW/CE-USA .625" (1.59 cm) Figure 2. Attaching the Fume Hood Monitor to the fume hood (surface mount). Slide enclosure cover onto top of enclosure base.Then rotate cover onto lower tabs for a secure fit.Date In it'l Fac eVel Sas hHght M in Flo w M ax Flo w StandbyFlow Flo w S-F 601-000-030 LB L Label/Seria lW /C E-U SA Sta ndard Operation FlowAla rm Em erg enc yExhau st Mute Figure 3. Reattaching the Fume Hood Monitor monitor housing to the base.
  • 7. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—7 OF 16 INSTALLATION (CONTINUED) Recess Mount Option Required materials: • Fume hood monitor, X30-RMT series (includes recess mounting bracket and screws) • Phoenix Controls room wiring diagrams • Cable, based on wiring diagram (not provided by Phoenix Controls) Installation steps: 1. Use the recess mounting template to mark the area for cutout and mounting holes for attaching the fume hood monitor to the fume hood (see Figure 4). 2. Cut a square hole, as necessary, for the recess mounting kit. 3. Mark the holes for mounting the bracket: • Using the recess mounting bracket as a template, place the bracket in the cutout section of the fume hood and mark locations of holes required for attaching the bracket to the fume hood. • Set the bracket aside. 4. Drill .125" (.32 cm) pilot holes for mounting screws. 5. Unsnap the front cover from the monitor base: • Loosen the bottom of the monitor housing and rotate it slightly upward. • Gently slide the enclosure cover up to release the two tabs inside the top of the monitor housing from the two grooves on the monitor base. 6. Set the enclosure cover aside. 7. Run the cable through the back of the monitor base. 8. Remove the connectors from the monitor housing. 9. Complete the terminations to the connectors (refer to provided wiring diagrams). 10. Reinstall the connectors in the monitor housing. 11. Attach the enclosure cover to the monitor base: • Insert the two tabs at the top of the enclosure cover in the grooves at the top of the monitor base. • Rotate the enclosure cover downward and snap the bottom of the cover onto the tabs in the monitor base to secure it. 12. Attach the monitor to the recess mounting bracket (see Figure 6 on the next page): • Slide the monitor assembly into the mounting bracket from the back. • Insert the L-strap into the top slot on the recess mounting bracket as shown. • Rotate the L-strap and slide it downward through the bottom slot in the recess mounting bracket. 2.6" (6.6 cm) 6.25" (15.88 cm) 6.8" (17.27 cm) Pilot hole (typical of 2) (.125" dia. each/.32 cm) 3.25" (8.26 cm) 7.5" (19.05 cm) 1.9" (4.83 cm) Figure 4. Fume Hood Monitor mounting template (recess mount). Figure 5. Attaching the Fume Hood Monitor to the fume hood (recess mount).
  • 8. 8 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 INSTALLATION (CONTINUED) Recess Mount Option (continued) 13. Mount the monitor and bracket assembly to the fume hood (see Figure 5 on the previous page): • Slide the assembly into the cutout area. • Using the mounting screws provided, attach the bracket to the fume hood. Retrofit Jobs Required materials: • Fume hood monitor, X30 series (includes recess mounting bracket assembly and screws) • Phoenix Controls room wiring diagrams • Cable, based on wiring diagram (not provided by Phoenix Controls) Installation steps: 1. Use the existing monitor to cut out hole and screw pilot holes from the previous FHM X10 Series monitor installation. Ensure that the dimensions of the cutout and mounting holes match those given here (see Figure 7). 2. Follow steps 5-12 from the standard Recess Mount installation procedure above. 3. Mount the monitor and bracket assembly to the fume hood (see Figure 8): • Slide the U-nut (included) over the existing mounting holes with the spring side inward. • Slide the assembly into the cutout area. • Using the mounting screws provided, attach the bracket to the fume hood. Phoenix Controls Wiring Recommendations • The fume hood monitor and exhaust valve should be powered from the same source. • If control over the fume hoods is important during a power failure or if airflow conservation is required, fume hood monitors should be placed on backup power; otherwise, the fume hood valve defaults to its fail-safe state. • Use cables recommended by Phoenix Controls. • Stranded wire is strongly recommended for ease of installation. • Follow good wiring practices: • Do not run the signal cable in the same conduit or wireway as the power cables. • If the signal cable must cross power cables, it is best to do so at a 90-degree angle. • Shield or drain wires, if required, should be wrapped with insulating tape to prevent contact with exposed conductors or contacts. • Maintain a consistent color code or polarity all the way through the wiring system. • All connections to the Celeris valve controller (LVC) must meet the requirements of an NEC Class 2 circuit. • Local and national electrical codes take precedence. L-strap Figure 6. Attaching the Fume Hood Monitor to the recess mounting bracke 8.13" (20.65 cm) 9.38" (23.83 cm) 2.5" (6.35 cm) 3.25" (8.26 cm) 10.5" (26.67 cm) 9.38" (23.83 cm) Figure 7. Fume Hood Monitor mounting template (retrofit jobs). U-nut Figure 8. Attaching the Fume Hood Monitor to the fume hood (retrofit jobs).
  • 9. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—9 OF 16 POINTS AND WIRING (see submittal wiring diagram for project-specific details) J2 TB1 TB2 TB4 1 1 1 TB3 1 FHM430/631 To fume hood exhaust valve (See note 1) VSS110 Zone Presence Sensor (ZPS) 200 Series 1 2 3 4 5 6 7 8 TB1 TB2 1 2 3 4 5 6 7 7 6 5 4 3 2 1 TB1 To TB1-2 To TB1-7 Reel Sensor Red Black ISO Ground Not Used User Status Secondary In Not Used Ground +15 Vdc/24 Vac 3 2 1 TB4 From AC Power Supply 24VacH 24VacN 8c (See note 2) Notes: 1. Eight-conductor wire is Belden 9421 (22 AWG) or equivalent. (Tape back unused conductors.) 2. Sash sensor is provided with two-conductor cable. See combination sash sensors for exception. FHM430 and 631 Typical Wiring Diagram Monitor Termination Points Variable Air Volume (VAV) Application TERMINAL BLOCK POINTS 1 2 3 4 5 6 7 8 TB1 +12 Vdc Analog Ground External Emergency/ Decommission Input Command Feedback Tri-level Alarm User Status Sash Signal TB2 Sash Sensor Analog Ground — Decommission Output Power Loss Input Spare LED (+) Spare LED (-) — TB4 24 Vac H/ +15 Vdc Functional Ground 24 Vac N/ -15 Vdc — — — — — ! See pages 12-13 for DC wiring details, as well as the wiring of options and ancillary equipment.
  • 10. 10 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details) J2 TB1 TB2 TB3 TB4 J1 1 1 1 1 Note: Requirements for transformers: • Secondary power shall be fused externally to current limit of 4 amps, in accordance with NEC CL2 power requirements. Thermal interrupts are not recommended. • Do not earth ground secondary transformers. • Maximum cable length for 96 VA load is 110’. • Each pressurization zone shall have a dedicated single-phase primary circuit or a secondary circuit disconnect. FHM530 Monitor Termination Points Constant Volume Application COMPONENTS TERMINAL BLOCK POINTS Device Function J1 IN for DC solenoid OUT for AC solenoid 1 2 3 4 5 6 7 8 TB1 +12 Vdc Analog Ground External Emergency NOV Command NCV Command Pressure Switch User Status Sash Signal TB2 Sash Sensor Analog Ground — — Power Loss Input Spare LED (+) Spare LED (-) — TB3 Solenoid Command Common Alarm NO Alarm Common — — — — — TB4 24 Vac H/ +15 Vdc NOV/NCV Common Functional Ground 24 Vac N /-15 Vdc — — — — — Phoenix Constant Volume Venturi Valve FHM530 From AC Power Supply 24VacH 24VacN TB1 1 2 3 4 5 6 7 8 3 2 1 TB4 Pressure Switch NO COM P ! See page 12 for DC wiring details, as well as the wiring of options and ancillary equipment.
  • 11. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—11 OF 16 POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details) AC Solenoid Valve Pressure Switch Phoenix Two-state Venturi Valve To TB1-4 (-) To TB4-1 (+) 1 2 3 To TB1-2 (NO) To TB1-6 (COM) P I 3 2 1 TB3 FHM530 From AC Power Supply (See note 2) 24VacH 24VacN TB1 1 2 3 4 5 6 7 8 3 2 1 TB4 Zone Presence Sensor (ZPS) 200 Series 7 6 5 4 3 2 1 TB1 To TB1-2 To TB1-7 ISO Ground Not Used User Status Secondary In Not Used Ground +15 Vdc/24 Vac J1 OUT See note 1 DC Solenoid Valve Pressure Switch Phoenix Two-state Venturi Valve To TB1-4 (-) To TB1-1 (+) 1 2 3 To TB1-2 (NO) To TB1-6 (COM) P H 3 2 1 TB3 FHM530 From DC Power Supply (See note 2) +15Vdc GROUND TB1 1 2 3 4 5 6 7 8 3 2 1 TB4 Zone Presence Sensor (ZPS) 200 Series 7 6 5 4 3 2 1 TB1 To TB1-2 To TB1-7 ISO Ground Not Used User Status Secondary In Not Used Ground +15 Vdc/24 Vac J1 IN -15Vdc Notes: 1. For AC-powered solenoid, two-position applications, you must install a field jumper between TB3-1 and TB4-3 (not required for DC- powered solenoids). 2. AC solenoids require the fume hood monitor to have AC power. DC solenoids require the fume hood monitor to have DC power. FHM530 (continued) Two-state Application: AC Solenoid Two-state Application: DC Solenoid ! See page 12 for the wiring of options and ancillary equipment. See page 12 for the wiring of options and ancillary equipment.
  • 12. 12 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details) Wiring of Options and Ancillary Equipment for X30 Fume Hood Monitor Series FHM430 Primary/Secondary Wiring (for AC-powered units) FHM430 Primary/Secondary Wiring (for DC-powered units) FHM430/631/530 External Emergency Override, Method 1 FHM430/631/530 External Emergency Override, Method 2 FHM430/631/530 Power Loss Alarm FHM430/631/530 Optional/Diversity LED FHM530 Alarm Relay Zone Presence Sensor (ZPS100 Series) X30 Series: DC Wiring Details Primary Unit 3 2 1 TB4 From Power Supply 24VacH 24VacN TB3 1 2 3 3 2 1 TB4 24VacH 24VacN TB3 Secondary Unit 2c 1 2 3 Primary Unit From Power Supply Secondary Unit TB2 1 2 3 4 5 6 7 3 2 1 TB4 +15 GROUND -15 TB3 1 2 3 TB3 2c 1 2 3 TB2 1 2 3 4 5 6 7 3 2 1 TB4 +15 GROUND -15 FHMX30 Contact Closure (provided by others) 1 2 3 4 5 6 7 8 TB1 2c FHMX30 1 2 3 4 5 6 TB1 10-volt command from BMS to initiate emergency override 2c Power Loss Alarm Unit 1 2 3 4 TB1 To FHMX30 Monitor To TB4-1 To TB2-2 To TB4-3 To TB2-5 4c +15 Ground -15 Output NOTE: The Power Loss Alarm unit will work only if powered by the same DC power source serving the monitor. AC-powered monitors cannot have a Power Loss Alarm. FHMX30 1 2 3 4 5 6 7 8 TB1 TB2 1 2 3 4 5 6 7 2c Contact Closure (provided by others) 1 K Ohm Resistor (provided by others) FHM530 1 2 3 TB3 To BMS for monitoring purposes NO COM ZPS110 7 6 5 4 3 2 1 TB1 To FHM X30 Monitor To TB4-3 To TB1-2 To TB4-1 4c Sync Out Sync In User Status Secondary -15 Ground +15 To TB1-7 NOTE: ZPS100 series units can be powered by only DC voltage. If the FHM X30 is powered by AC, then separate power must be brought to the ZPS100 unit. TB2 1 2 3 4 5 6 7 3 2 1 TB4 From DC Power Supply +15 GROUND -15
  • 13. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—13 OF 16 POINTS AND WIRING (CONTINUED) (see submittal wiring diagram for project-specific details) Decommission Mode Wiring for FHM631 (in addition to the wiring shown on page 9) Decommission Mode through the Fume Hood Monitor’s Pushbutton Decommission Mode through a Momentary External Switch Decommission Mode through a BMS Command* TB2 1 2 3 4 5 6 7 GROUND Celeris Valve Controller (LVC) 1 2 3 4 5 6 7 8 TB1 FHM631 DIGITAL OUT1 DIGITAL IN+ DIGITAL IN- 8c GROUND Celeris Valve Controller (LVC) 1 2 3 4 5 6 7 8 TB1 FHM631 DIGITAL OUT1 DIGITAL IN+ DIGITAL IN- EXTERNAL SWITCH 8c 2c 1 2 3 4 5 6 7 8 TB1 TB2 1 2 3 4 5 6 7 +12 V +12 V Celeris Valve Controller (LVC) 1 2 3 4 5 6 7 8 TB1 FHM631 DIGITAL OUT1 DIGITAL IN+ DIGITAL IN- TB2 1 2 3 4 5 6 7 NO COM 2c 1 2 3 4 5 6 7 8 TB1 TB2 1 2 3 4 5 6 7 GROUND 8c *Note: Do not use this configuration if the LVC and fume hood monitor are using separate power sources.
  • 14. 14 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MAINTENANCE Phoenix Controls fume hood monitors require no ongoing preventative maintenance. Once the field engineer has completed the field setup, the monitors will provide years of continuous operation. TROUBLESHOOTING Phoenix Controls fume hood monitors alert the operator of alarm conditions. Generally, this alarm is caused by a problem condition in the exhaust duct (e.g., fan failure). A trained facilities person may troubleshoot the system from the monitor with a digital voltmeter and perhaps a magnehelic gauge. Alarms The types of alarms for the X30 Series Fume Hood Monitor, the audible and visual cues for each alarm, and the purpose for each alarm are listed in the table below. Replacement Part Part Number FHM631 board 860-200-108 FHM630 board 860-200-102 FHM530 board 860-200-109 FHM430 board 860-200-111 FHM430 primary board 860-200-110 FHM430 secondary board 860-200-112 X30 recess mount retrofit kit (replaces X10 models with X30) 260-270-004 X30 recess mount kit 260-270-005 Type of Alarm Alarm Indicators Reason for AlarmVisual Alarm (LED blink rate*) Audible Alarm (beep rate**) Signals (as seen in analog valve systems) Not commissioned Standard Operation—Fast Flow Alarm—Fast 631 Display—Er-c None N/A The unit requires commissioning. Low differential static pressure Flow Alarm—Slow FHM631 display—Blank Fast TB1, 6 = 10 Vdc The hood’s pressure switch is open (FHM631 and 430 only). Flow Alarm—Slow Fast N/A The hood’s pressure switch is open (FHM530 only). Incorrect airflow (i.e., valve jam) Flow Alarm—Fast Slow TB1, 6 = 5 Vdc The valve’s position is jammed (FHM631 and 430 only). Sash opening alarm Flow Alarm—Slow Slow TB1, 6 = 0 Vdc*** The sash opening exceeds the calibrated alarm set point. Broken sash sensor alarm Flow Alarm—Fast Fast TB1, 6 = 0 Vdc*** TB1, 8 > 10.4 Vdc The sash input is greater than the maximum resistive input for the application, or the sash sensor signal cable is broken or loose. Emergency exhaust override alarm (local) Emerg Exh—Slow Slow TB1, 6 = 0 Vdc TB1, 4 > 10.4 Vdc The emergency exhaust button was pushed. Emergency exhaust override alarm (BMS external) Emerg Exh—Fast Slow TB1, 6 = 0 Vdc*** TB1, 4 > 10.4 Vdc The externally controlled emergency input is in emergency state. Energy waste alert FHM631 display—ENRG Slow N/A The sash is open and room lights are off (FHM631 only). Power fail alarm Power loss LED on every 4 seconds 3 beeps every 10 seconds N/A The fume hood monitor lost ±15 Vdc power. Optional alarm Optional alarm LED driven by customer None TB1, 6 = 0 Vdc Customer specific alarm or diversity alarm Loss of alarm line Flow Alarm—Slow 631 Display—Blank Fast TB1, 6 = 12 Vdc The alarm signal cable is broken. Loss of feedback signal Standard Operation—Solid Flow Alarm—Slow 631 Display—Low value Slow TB1, 6 = 0 Vdc Feedback signal cable is broken. Notes: * A fast blink rate resembles a flashing light, while a slow rate resembles a fluttering light. ** A fast beep rate is 10 beeps per second, while a slow rate is 2 beeps per second. *** 5 Vdc will be detected when these alarms are generated by a Celeris system. If the fume hood monitor generates this condition first, the Celeris signal will supercede after it is generated.
  • 15. ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 MKT-0044 MPC-1093 X30 SERIES FUME HOOD MONITORS—15 OF 16 TROUBLESHOOTING (CONTINUED) Troubleshooting Guide Use the information in the tables below to troubleshoot the X30 Series Fume Hood Monitor. Variable Air Volume (VAV) Valve Systems (FHM430 and 631 only) Constant Volume (CV) and Two-state Valve Systems (FHM530) Problem at Fume Hood Monitor Voltage at TB1-6 in Monitor Possible Cause/Solution 1. The monitor is in flow alarm. > 10 V 5 V Voltages may vary. A. Low static pressure across valve • Loss of airflow—Check fan operation and duct blockage. • Too many sashes open at one time—Close sashes. • Valve failed open—Check pneumatic and mechanical connections. B. Incorrect valve position • Valve failed open. • Sash open beyond the maximum allowable position—Lower sash. • A broken sash cable—Check all sash sensor connections. C. Equipment/connection problems • Blocked or kinked pressure switch tubing—Correct tubing. • Wiring terminations between monitor and control device—Correct terminations. • Malfunctioning alarm circuits—Verify proper static.* • Monitor miscalibration—Recalibrate.* 2. The monitor indicates normal operation, but the actual face velocity has been measured high or low. 0 V A. Low static pressure • Differential pressure between the pressure switch set point and the low end of the static pressure operating range [i.e., 0.3" and 0.6" wc (75-150 Pa) for medium pressure valves] will not trip the alarm circuit. • Measure differential pressure. If it is low, see Possible Cause/Solution 1A above. B. Monitor miscalibration—Recalibrate.* 3. The monitor has malfunctioned. • No display. • Cannot mute alarm. Voltages may vary. A. Loss of power • Check power at monitor TB4-1 (24 Vac H/+15 Vdc) and TB4-3 (24 Vac N/-15 Vdc). • Check wiring connections at monitor, valve, and power supply. • Verify power supply has input voltage (120 Vac or 240 Vac). B. Defective monitor—Replace board on monitor.* 4. The flow remains constant through sash travel. A. Broken sash sensor cable—Replace sensor. B. Monitor miscalibration—Recalibrate.* * Contact Phoenix Controls Product Support Center for assistance. Problem Possible Cause/Solution 1. The monitor is in flow alarm. Low static pressure across valve. • Loss of airflow. Check fan operation and duct blockage. • Blocked or kinked pressure switch tubing. Correct the tubing. 2. The airflow was measured improperly in a: A. Constant volume system. B. Two-position system. A. Incorrect valve position. Realign the pivot arm. B. Incorrect valve position. Check pneumatic tubing and pressure. Also check solenoid valve operation.
  • 16. 16 OF 16—X30 SERIES FUME HOOD MONITORS MKT-0044 MPC-1093 ©2005 Phoenix Controls Corporation. Specifications subject to change without notice. Rev. 9/07 TROUBLESHOOTING (CONTINUED) Variable Air Volume (VAV) Drive Systems (FHM430 and 631 only) Energy-saving Options (FHM631 only) Problem Possible Cause/Solution 1. The monitor is in flow alarm. A. Loss of airflow • Check fan operation and duct blockage. • Check the drive for proper operation. B. Monitor/connection problems • The wiring terminations between monitor and control device are incorrect. Correct the terminations. • The monitor was miscalibrated. Recalibrate the monitor as needed.* • The monitor is defective. Replace the board on the monitor.* 2. The monitor indicates normal operation, but actual face velocity has been measured high or low. The monitor was miscalibrated. Recalibrate the monitor as needed.* 3. The monitor is malfunctioning. • No display • Cannot mute alarm A. Loss of power • Check the wiring connections at the monitor, drive and power supply. • Verify that the power supply has input voltage. • Check power at the monitor. B. The monitor is defective. Replace the board on the monitor. Recalibrate the monitor as needed.* * Contact Phoenix Controls Product Support Center for assistance. Problem Possible Cause/Solution 1. The monitor will not enter energy waste alert. A. The low-light level setting is too high. Recalibrate the monitor. B. The energy waste alert is not enabled. Recalibrate the monitor. 2. The monitor will not exit energy waste alert. The high-light level setting is too high. Recalibrate the monitor. 3. The monitor will not enter decommission mode. A. The sash is not fully closed. Close the sash. B. Decommission mode is not enabled. Recalibrate the monitor. C. Check the wiring to the valve. 4. The monitor will not exit decommission mode. A. The digital input (DI) on the valve is not configured properly. Reconfigure this input. B. The monitor is defective. Replace and recalibrate the monitor.