QUALITY CONTROL(QC)
AND
QUALITY ASSURANCE(QA)
OF
X-RAY MACHINE
TOPIC TO BE DISCUSSED:
• QUALITY CONTRO AND QUALITY ASSURANCE
• TEST TO BE DONE TO CONTROL QUALITY
• LEVELS OF QUALITY CONTROL(FREQENCY)
• TOOLS REQUIRED
• PROBLEM BASED AREAS.
QUALITY CONTROL
• Series of tests to assure radiographic system
is functioning properly and radiographs are of
acceptable quality.
• Quality control tests should be conducted for
x ray machines.
• logs of tests performed dates results and
corrective action should be kept.
QUALITY ASSURANCE
• An organized effort by staff operating a facility
to ensure that the diagnostic image produced
are of a sufficiently high quality so that they
consistently provide adequate diagnostic
information at the lowest possible cost and
with the least possible exposure of the patient
to radiation
QUALITY ASSURANCE PROGRAM
• THIS PROGRAM HAS MANY FACETS
INCULDING
1) QUALITY COTROL TESTS
2 )ADMINSTRED PROCEDURE
3 )PREVENTIVE MAINTANCE PROCEDURE
4 )TRANING
QUALITY CONTROL TESTS
1 )CENTRAL BEAM ALIGNMENT
2)FOCAL SPOT SIZE
3)TUBE VOLTAGE
4)TIMER CHECK
5)TUBE HOUSING LEAKAGE
CENTRAL BEAM ALIGNMENT
• If the x ray beam is not perpendicular to the
image receptor ,the image may be distorted
• If grid is used the distortion will be magnified
resulting in total loss of minute details.
FOCAL SPOT SIZE
• The ability for resolving the smallest size of
the image (i.e. Details) in a radiograph
depends on the focal spot size.
• Since, the focal spot size may be altered as a
result of bombardment of electrons on the
target, it has to be checked periodically to
ensure that focal spot size is within acceptable
limits.
TUBE VOLTAGE(KVP)
• The applied kvp affects the quality and
quantity of x-rays reaching the image
receptors.
• If there is variation in kvp setting, it will affect
the image quality, hence it is necessary to
check the kvp settings periodically.
TIMER CHECKING
• If the exposure time set on the diagnostic x-ray
unit is not optimal, the radiograph can be under
exposed or over exposed.
• This may lead to repeat examinations. Hence,
there is a need to test the timer of the x-ray unit
periodically.
TUBE HOUSING LEAKAGE
• The leakage of radiation from r-ray tube
housing adds the radiation dose to the patient
and to the operator too.
• Tube housing leakage test must be periodically
done to insure the radiation safety.
TOOLS REQUIRED
BEAM ALIGNMENT TEST TOOL
• Beam alignment test tool is kept on the collimator
test tool, such way that the stainless steel ball of
lower side of tool is just above the centre of the
collimator test-tool. the film is exposed and
processed.
• If the beam alignment is perfect, the image of the
top ball will merge with the image of the ball at the
bottom. The deviation of beam from the
perpendicular is determined from the location of
the image of top ball in the circles in the
radiographs.
BEAM ALIGNMENT TEST TOOL
FOCAL SPOT TEST TOOL
• Test tool consist of hollow cylinder with an acrylic
circular disc fastened at one end which contains 12
groups of bar patterns ingraved on a tungsten plate.
• Each group consist of 6 lines-sub group of 3 parallel
lines,perpendicular to another sub group whose
size and spacing decreases by steps of 16%.
• The focal spot size value are computed for
magnification of 4/3 being ffd 60cm and OFD 45.as
shown in table:
• Test tool placed on a non screen cassette loaded
with film,(ffd 60 cm )constant such that vertical
pattern are within the anode to cathode direction.
• Film is exposed and processed and bar pattern on
radiograph is observed-smallest group on which all
6 bars are clearly resolved is identified and
corresponding focal spot size can be obtained from
the table.
FOCAL SPOT TEST TOOL
TIMER CHECKING
SPINNING TOP TEST TOOL
• Spinning top test tool is used to check the
exposure time. It consist of a rotating circular
brass plate with small rectangular portion
cut(hole) at its periphery.
• The tool is placed on the cassette, loaded with
film. For a set time, the unit is energized, while
the top is rotating.
• The pulses passing through the hole of the
circular plate, produces equally spaced
rectangular density patterns, on the film.
SPINNING TOP TEST TOOL
TUBE VOLTAGE
Kvp METER
• Kvp meter,employes two solid state detector
with different beam hardening filters.when
exposed to radiation,the ratio of the signals
produced by these detectors will be proportional
to the peak tube voltage.
• Corrections for beam filtration should be applied
if necessary .
Kvp METER
TUBE HOUSING LEAKAGE
RADIATION SURVEY METER
• The radiation leakage measurement is carried
with an ionization radiation survey meter. For
checking the leakage radiation,the collimator of
the tube housing is fully closed and the tube is
energized at maximum rated tube potential and
current at that kvp.
• The exposure rate at one meter from the target
is measured at different location from the tube
housing and collimator .
• The tolerance limit of leakage radiation at 1 m
from the focus is >115 mR in one hour
RADIATION SURVEY METER
FREQUENCY OF QUALITY CONTROL.
Test Frequency
kVp Once in three year
Timer Once in 3 months
Output,mR/ mAs Monthly
Inherent filtration Once in 3 months
Focal spot size Once in a year
Central beam alignment Once in 2 months
Congruence of radiation and
optical field
Once in 2 months
Grid alignment Annually and whenever film
density appears non uniform
QA COMMITTEE & THEIR ROLE
• Includes:
Medical physicist
Radiologist
Biomedical engineer
Imaging technologist (radiographer)
IT technician
MEDICAL PHYSICIST
• Conducting test to ensure the safety and proper
performance of imaging equipment used.
• Assignment of biomedical engineering service
staff for corrective maintenance or preventive
actions
• Training of personal utilized for quality control.
• Develop and implement a radiation protection
program.
RADIOLOGIST
• Determine the overall quality of the output image
• Select the technologist to be the primary QC
technologist, performing the prescribed QC test.
• Ensure that appropriate test equipment and
materials are available to perform the QC tests.
• Ensuring that medical physicist and radiographers
have adequate training and continuous education.
BIOMEDICAL ENGNEER
• Corrective and preventive maintenance.
• Fault reporting
IMAGING TECHNOLOGIST AND
RADIOGRAPHER
• Ensuring that the appropriate protocol and
technique factors are used for the requested
examination
• Ensuring that the QC tests are performed,
interpreted and recorded appropriately.
• Perform all the checks for the daily ,monthly
and quarterly QC testing of equipment.
• Report faults immediately, any deviation in
trend of equipment performance to QA
manager
• Undertaking additional continuous education
courses.
IT TECHNICIAN
• Maintenance of the integrity of
system database to ensure
continuous and accurate
operation of the information
systems

Contenu connexe

Tendances

Image quality
Image qualityImage quality
Image qualityairwave12
 
Soft tissue radiography.pptx
Soft tissue radiography.pptxSoft tissue radiography.pptx
Soft tissue radiography.pptxRitupanta1
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancementYashawant Yadav
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptxdypradio
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipmentSharmaRajan4
 
Fluoroscopy systems
Fluoroscopy systemsFluoroscopy systems
Fluoroscopy systemsRad Tech
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionNitish Virmani
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictorsvishwanath0908
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentationHuzaifa Oxford
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettesSudil Paudyal
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department AbubakarMustaphaAman
 
Radiographic Film and Its processing
Radiographic Film and Its processingRadiographic Film and Its processing
Radiographic Film and Its processingTarique Ajij
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxThejaTej6
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiationNitish Rajput
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 

Tendances (20)

Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Image quality
Image qualityImage quality
Image quality
 
Soft tissue radiography.pptx
Soft tissue radiography.pptxSoft tissue radiography.pptx
Soft tissue radiography.pptx
 
X ray film
X  ray filmX  ray film
X ray film
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancement
 
Emergency radiography
Emergency radiography Emergency radiography
Emergency radiography
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipment
 
Fluoroscopy systems
Fluoroscopy systemsFluoroscopy systems
Fluoroscopy systems
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentation
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettes
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department
 
Radiographic Film and Its processing
Radiographic Film and Its processingRadiographic Film and Its processing
Radiographic Film and Its processing
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiation
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 

Similaire à Qc in xray

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
Radiographic quality assurance & infection control
Radiographic quality assurance & infection controlRadiographic quality assurance & infection control
Radiographic quality assurance & infection controlBinaya Subedi
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxAALIA ABDULLAH
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentationMumba Chilimboyi
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instrumentsSolairajan A
 
Infrared spectroscopy calibration procedure
Infrared spectroscopy calibration procedureInfrared spectroscopy calibration procedure
Infrared spectroscopy calibration procedureVishnuyadav44
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjPratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxPratik434909
 
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxqualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxBhumiSuratiya
 

Similaire à Qc in xray (20)

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
Radiographic quality assurance & infection control
Radiographic quality assurance & infection controlRadiographic quality assurance & infection control
Radiographic quality assurance & infection control
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
QUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLCQUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLC
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
Infrared spectroscopy calibration procedure
Infrared spectroscopy calibration procedureInfrared spectroscopy calibration procedure
Infrared spectroscopy calibration procedure
 
Quality control
Quality  control Quality  control
Quality control
 
Intro to ndt ashish 1
Intro to ndt ashish 1Intro to ndt ashish 1
Intro to ndt ashish 1
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxqualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 

Plus de Anupam Niraula

Handling the emergencies in radiology and first aid in the x ray department
Handling the emergencies in radiology and first aid in the x ray departmentHandling the emergencies in radiology and first aid in the x ray department
Handling the emergencies in radiology and first aid in the x ray departmentAnupam Niraula
 
Infection control concepts
Infection control conceptsInfection control concepts
Infection control conceptsAnupam Niraula
 
Bedside radiography and patient care
Bedside radiography and patient careBedside radiography and patient care
Bedside radiography and patient careAnupam Niraula
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matterAnupam Niraula
 

Plus de Anupam Niraula (10)

Handling the emergencies in radiology and first aid in the x ray department
Handling the emergencies in radiology and first aid in the x ray departmentHandling the emergencies in radiology and first aid in the x ray department
Handling the emergencies in radiology and first aid in the x ray department
 
First aid .
First aid .First aid .
First aid .
 
Infection control concepts
Infection control conceptsInfection control concepts
Infection control concepts
 
Bedside radiography and patient care
Bedside radiography and patient careBedside radiography and patient care
Bedside radiography and patient care
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matter
 
Vital signs
Vital signsVital signs
Vital signs
 
First aid
First aidFirst aid
First aid
 
Mri safety
Mri safetyMri safety
Mri safety
 
Lead apron
Lead apronLead apron
Lead apron
 
Emergency radiology
Emergency  radiologyEmergency  radiology
Emergency radiology
 

Dernier

Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...fonyou31
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfchloefrazer622
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024Janet Corral
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
fourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingfourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingTeacherCyreneCayanan
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDThiyagu K
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 

Dernier (20)

Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdf
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
fourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingfourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writing
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 

Qc in xray

  • 2. TOPIC TO BE DISCUSSED: • QUALITY CONTRO AND QUALITY ASSURANCE • TEST TO BE DONE TO CONTROL QUALITY • LEVELS OF QUALITY CONTROL(FREQENCY) • TOOLS REQUIRED • PROBLEM BASED AREAS.
  • 3. QUALITY CONTROL • Series of tests to assure radiographic system is functioning properly and radiographs are of acceptable quality. • Quality control tests should be conducted for x ray machines. • logs of tests performed dates results and corrective action should be kept.
  • 4. QUALITY ASSURANCE • An organized effort by staff operating a facility to ensure that the diagnostic image produced are of a sufficiently high quality so that they consistently provide adequate diagnostic information at the lowest possible cost and with the least possible exposure of the patient to radiation
  • 5. QUALITY ASSURANCE PROGRAM • THIS PROGRAM HAS MANY FACETS INCULDING 1) QUALITY COTROL TESTS 2 )ADMINSTRED PROCEDURE 3 )PREVENTIVE MAINTANCE PROCEDURE 4 )TRANING
  • 6. QUALITY CONTROL TESTS 1 )CENTRAL BEAM ALIGNMENT 2)FOCAL SPOT SIZE 3)TUBE VOLTAGE 4)TIMER CHECK 5)TUBE HOUSING LEAKAGE
  • 7. CENTRAL BEAM ALIGNMENT • If the x ray beam is not perpendicular to the image receptor ,the image may be distorted • If grid is used the distortion will be magnified resulting in total loss of minute details.
  • 8. FOCAL SPOT SIZE • The ability for resolving the smallest size of the image (i.e. Details) in a radiograph depends on the focal spot size. • Since, the focal spot size may be altered as a result of bombardment of electrons on the target, it has to be checked periodically to ensure that focal spot size is within acceptable limits.
  • 9. TUBE VOLTAGE(KVP) • The applied kvp affects the quality and quantity of x-rays reaching the image receptors. • If there is variation in kvp setting, it will affect the image quality, hence it is necessary to check the kvp settings periodically.
  • 10. TIMER CHECKING • If the exposure time set on the diagnostic x-ray unit is not optimal, the radiograph can be under exposed or over exposed. • This may lead to repeat examinations. Hence, there is a need to test the timer of the x-ray unit periodically.
  • 11. TUBE HOUSING LEAKAGE • The leakage of radiation from r-ray tube housing adds the radiation dose to the patient and to the operator too. • Tube housing leakage test must be periodically done to insure the radiation safety.
  • 12. TOOLS REQUIRED BEAM ALIGNMENT TEST TOOL • Beam alignment test tool is kept on the collimator test tool, such way that the stainless steel ball of lower side of tool is just above the centre of the collimator test-tool. the film is exposed and processed. • If the beam alignment is perfect, the image of the top ball will merge with the image of the ball at the bottom. The deviation of beam from the perpendicular is determined from the location of the image of top ball in the circles in the radiographs.
  • 14. FOCAL SPOT TEST TOOL • Test tool consist of hollow cylinder with an acrylic circular disc fastened at one end which contains 12 groups of bar patterns ingraved on a tungsten plate. • Each group consist of 6 lines-sub group of 3 parallel lines,perpendicular to another sub group whose size and spacing decreases by steps of 16%. • The focal spot size value are computed for magnification of 4/3 being ffd 60cm and OFD 45.as shown in table:
  • 15.
  • 16. • Test tool placed on a non screen cassette loaded with film,(ffd 60 cm )constant such that vertical pattern are within the anode to cathode direction. • Film is exposed and processed and bar pattern on radiograph is observed-smallest group on which all 6 bars are clearly resolved is identified and corresponding focal spot size can be obtained from the table.
  • 18. TIMER CHECKING SPINNING TOP TEST TOOL • Spinning top test tool is used to check the exposure time. It consist of a rotating circular brass plate with small rectangular portion cut(hole) at its periphery. • The tool is placed on the cassette, loaded with film. For a set time, the unit is energized, while the top is rotating. • The pulses passing through the hole of the circular plate, produces equally spaced rectangular density patterns, on the film.
  • 20. TUBE VOLTAGE Kvp METER • Kvp meter,employes two solid state detector with different beam hardening filters.when exposed to radiation,the ratio of the signals produced by these detectors will be proportional to the peak tube voltage. • Corrections for beam filtration should be applied if necessary .
  • 22. TUBE HOUSING LEAKAGE RADIATION SURVEY METER • The radiation leakage measurement is carried with an ionization radiation survey meter. For checking the leakage radiation,the collimator of the tube housing is fully closed and the tube is energized at maximum rated tube potential and current at that kvp. • The exposure rate at one meter from the target is measured at different location from the tube housing and collimator . • The tolerance limit of leakage radiation at 1 m from the focus is >115 mR in one hour
  • 24. FREQUENCY OF QUALITY CONTROL. Test Frequency kVp Once in three year Timer Once in 3 months Output,mR/ mAs Monthly Inherent filtration Once in 3 months Focal spot size Once in a year Central beam alignment Once in 2 months Congruence of radiation and optical field Once in 2 months Grid alignment Annually and whenever film density appears non uniform
  • 25. QA COMMITTEE & THEIR ROLE • Includes: Medical physicist Radiologist Biomedical engineer Imaging technologist (radiographer) IT technician
  • 26. MEDICAL PHYSICIST • Conducting test to ensure the safety and proper performance of imaging equipment used. • Assignment of biomedical engineering service staff for corrective maintenance or preventive actions • Training of personal utilized for quality control. • Develop and implement a radiation protection program.
  • 27. RADIOLOGIST • Determine the overall quality of the output image • Select the technologist to be the primary QC technologist, performing the prescribed QC test. • Ensure that appropriate test equipment and materials are available to perform the QC tests. • Ensuring that medical physicist and radiographers have adequate training and continuous education.
  • 28. BIOMEDICAL ENGNEER • Corrective and preventive maintenance. • Fault reporting
  • 29. IMAGING TECHNOLOGIST AND RADIOGRAPHER • Ensuring that the appropriate protocol and technique factors are used for the requested examination • Ensuring that the QC tests are performed, interpreted and recorded appropriately. • Perform all the checks for the daily ,monthly and quarterly QC testing of equipment. • Report faults immediately, any deviation in trend of equipment performance to QA manager • Undertaking additional continuous education courses.
  • 30. IT TECHNICIAN • Maintenance of the integrity of system database to ensure continuous and accurate operation of the information systems