SlideShare une entreprise Scribd logo
1  sur  23
Gas chromatography
• Gas chromatography is a separation
technique based on partitioning
analytes between two immiscible
phases: gaseous mobile phase (Carrier
gas) and a stationary solid or
immobilized liquid phase (packed or
hollow capillary column).
Gas chromatography
• The sample is first vaporized by a heated inlet system to be passed through a gaseous
carrier into GC column. The passed analytes adsorbed on the stationary phase of the
GC column, the adsorbed analytes eluted by applying heating program (the stronger
adsorption of an analyte, the higher heat required for its elution). Therefore, GC is
suitable for analysis of volatile (or semi-volatile) and a thermally stable analyte.
Stauffer E, Dolan JA, and Newman R (2008) Fire Debris
Analysis, p. 246. Burlington, MA: Elsevier Academic Press
Carrier gas
• A carrier gas is responsible for carrying the sample molecules into the
column and finally to the detector.
• Carrier gas must be: inert with the stationary phase, of high purity 99.999
%. (for residue analysis 99.9999 % is preferred).
Agilent, fundamental gas chromatography, G1176-90000
It is better to be vertically standing
GC Injection Techniques
• Split/split-less
• Programmed temperature vaporization (PTV)
injection.
• On-column injection
Split injector
• In the split injection mode,
sample enters the hot liner and
volatilized rapidly.
• Vaporized sample is mixed with
a carrier gas (diluted).
• Finally, a large part of the
diluted vaporized sample is split
away from the Colum, while a
small part will enter the column.
• This mode of injection is used
for analysis of samples of high
analyte concentrations.
Agilent, fundamental gas chromatography, G1176-90000
Split-less injector
• Is the split-less injection mode,
sample enter the hot liner and
volatilized rapidly.
• Vaporized sample is mixed with
a carrier gas (diluted).
• Finally, all the diluted vaporized
sample enter the column.
• After that, the split valve is
opened to remove residual
vapors.
• This mode of injection is used
for analysis of samples of trace
analyte concentrations (residue
analysis). Agilent, fundamental gas chromatography, G1176-90000
• In the program temperature vaporized injector
the sample is injected in a cooled liner (enable
sample and analyte adsorption). While, the
solvent will be split away. Therefore, injection
in PTV should be carried out slowly.
• After that, split valve closed, and beginning
ramping temperature increase (ramp PTV) to
vaporize adsorbed sample matrix and analytes
to be carried into the column
Baffled
Liner
Vent
Syringe
needle
Column
Sample
Analyte
Solvent
PTV injector
• sample aliquots are directly introduced onto the
analytical column ( 0.2-0.5 ) at low temperatures (60–
80°C).
• On column injection is favored for an analyte that can
be thermally degradated at the elevated heated split or
split-less mode (around 200 C).
• These injection mode require careful awareness to attain
a good reproducibility.
• A liner of the wider volume is favorable for this
injection.
On-column injection
https://www.sigmaaldrich.com/analytical-
chromatography/analytical-
products.html?TablePage=110598941
• The used solvent (containing the analytes) shall have
an expansion volume that will not exceed 75 % of the
liner volume (especially for split-less mode of
injection).
• Inner-surface of the used liner shall be highly inert,
especially for residue analysis.
• If the sample is not so cleaned up, liners containing
glass wool is more favorable.
• The temperature of the liner should be enough to
evaporate the desired analytes and avoid its
degradation.
• A bottom tap liner is more favorable for splitless
injection.
Choosing the right liner
http://www.restek.com/Supplies-Accessories/GC-
Accessories/Inlet-Liners-Liner-Supplies
https://www.chromservis.eu/c/gc-liners
GC-COLUMN
Capillary columns
Open tubular (OT)
K. Maštovska, S.J. Lehotay, J. Chromatogr. A, 1000 (2003) 153
Packed columns
Agilent, fundamental gas chromatography, G1176-90000Agilent, fundamental gas chromatography, G1176-90000
• The commonly used stationary phase for capillary GC column is
poly dimethyl siloxane [PDMS]. The separation, in this case, will
depend upon dispersive interactions (van der Waals) with nonpolar
part of the analytes.
• The PDMS stationary phases can be more polar by adding phenyl,
cyano, or trifluoro functional groups each of these stationary phases
leads to different separation results.
• In case of using PDMS with cyano, trifluoro, or (especially)
hydroxyl functional groups, analyte separation can occur based on
hydrogen bonding interactions (the strongest intermolecular forces
in capillary gas chromatography (GC).
GC-COLUMN
GC-COLUMN
Jana Hajšlová, Tomášajka, Gas chromatography–mass spectrometry (GC–
MS), Food Toxicants Analysis, 2007
• A longer column will increase the resolution (selectivity), but it will
also increase analysis time, and cost.
• Reduced column internal diameter double the efficiency and leads
to better selectivity. This will increase retention time when using
isothermal separations. such columns easily contaminated and suffer
from peak broadening after routine work.
• Changes in film thickness effects retention of analyte species,
interaction with the silica tubing (increased with increasing film
thickness). Usually, thin films (0.10-0.25 µm)are used for trace
analysis.
GC-COLUMN
A common GC-COLUMN/ HP 5 MS
Internal diameter 0.25 mm
Length 30 m
Film 0.50 µm
Temperature Limits -60 - 325/350 °C
• Stationary phase coating is : 5 %-Phenyl-methyl poly siloxane
Column Temperature
• Every capillary column has its working temperature range that maintains the
stateof the used coated stationary phase.
• Working on GC column
should be below its maximum
temperature.
Column starting temperature
• The starting temperature for the column (oven temperature) should be much
lower than the lowest boiling points of tested analytes.
• For analytes of low boiling points, the adsorption of both solvent and these
analytes is required at the front column inlet (starting column temperature of 20 C
lower than the used solvent) which will subsequently enhance its peak shape.
• While, for highly boiling points analytes (late eluting peaks), adsorption of the
used solvent isn’t required (it will cost a longer run time) only start with a
temperature below the boiling points of these analytes but that give enough
adsorption at the front column inlet (no peak tailing)
Column starting temperature
GC Inlets An Introduction, Agilent. 5958-9468
Journal of Chromatographic Science, Vol. 34, December 1996
GC-MS, GC-MS/MS
Retention time, related terms and definitions
• Chromatogram : A plot of the detector response related to the effluent time.
• Mass spectrum: A plot of the intensities versus m/z of specific peak.
• Retention time tR : Time taken by the carrier gas from the analyte injection to its
completely detection.
• Dead time (holdup time) tM : Is the time taken by the carrier gas from the point
of injection to the detector.
• Adjusted retention time : = tR - tM
• Distribution constant K: The ratio of analyte concentration in the stationary
phase to its concentration in the mobile phase 𝐾 =
𝐶𝑠
𝐶𝑚
Column Resolution, related terms and definitions
• Separation factor (selectivity factor): is the relative retention times of two
adjacent eluted peaks, α=
tR1
tR2
.
• Peak resolution based on their peak widths Rs= 2(tR1− tR2)
Wb1
+
Wb2
.
• Peak resolution based on their peak widths Rh= h1−hw
h1
• The number of theoretical plates, N = 16 (
tR
W
) 2
• Efficiency (N) ,for thin film column, is related to column
length (L) and internal diameter (dc), N= 𝐿/𝑑𝑐
https://doi.org/10.1016/B978-0-12-409547-2.14348-3
References, and useful links
• Yehua Han, Yanfen Zhang, Huwei Liu https://doi.org/10.1016/B978-0-12-409547-2.14348-3
• A.I. Ruiz-Matute, S. Rodrı´guez-Sa´nchez, M.L. Sanz and A.C. Soria, https://doi.org/10.1016/B978-0-12-814264-
6.00012-8
• Alejandra Garcia Piantanida , Andrew R. Barron, Principles of Gas Chromatography, 2014.
http://cnx.org/content/m50228/1.2/
• Fundamentals of Gas Chromatography, Agilent Technologies, Inc., 2002.
G1176-90000
• https://www.restek.com/pdfs/GNBR1724-UNV.pdf
• http://www.chromatographyonline.com/pragmatic-rules-gc-column-selection
• https://chem.libretexts.org/Textbook_Maps/Analytical_Chemistry/Supplemental_Modules_(Analytical_Chemistry)/I
nstrumental_Analysis/Chromatography/Gas_Chromatography.
Thank you
Dr. Sherif M. Taha,
sherif2taha@gmail.com

Contenu connexe

Tendances

Hplc presentation for class
Hplc presentation for classHplc presentation for class
Hplc presentation for class
Dr. Ravi Sankar
 

Tendances (20)

Gas chromatography-GC
Gas chromatography-GCGas chromatography-GC
Gas chromatography-GC
 
gas chromatography
gas chromatographygas chromatography
gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Hplc by naveed 27 nov 2018
Hplc by naveed   27 nov 2018Hplc by naveed   27 nov 2018
Hplc by naveed 27 nov 2018
 
Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Use of gc and gcms in api analysis
Use of gc and gcms in api analysis
 
Gas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfGas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdf
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
HPLC Presentation by moun
HPLC Presentation by mounHPLC Presentation by moun
HPLC Presentation by moun
 
GC - MS
GC - MSGC - MS
GC - MS
 
Supercritical fluid chromatography
Supercritical fluid chromatographySupercritical fluid chromatography
Supercritical fluid chromatography
 
Hplc presentation for class
Hplc presentation for classHplc presentation for class
Hplc presentation for class
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography [gc]
Gas chromatography [gc]Gas chromatography [gc]
Gas chromatography [gc]
 
Applications of Chiral Supercritical Fluid Chromatography
Applications of Chiral Supercritical Fluid ChromatographyApplications of Chiral Supercritical Fluid Chromatography
Applications of Chiral Supercritical Fluid Chromatography
 
Gc troubleshooting
Gc troubleshootingGc troubleshooting
Gc troubleshooting
 
Gas chromatography
Gas chromatography Gas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 

Similaire à Gas chromatography

GC PPT...pptx
GC PPT...pptxGC PPT...pptx
GC PPT...pptx
Dipu77
 

Similaire à Gas chromatography (20)

GAS CHROMATOGRAPHY
GAS  CHROMATOGRAPHYGAS  CHROMATOGRAPHY
GAS CHROMATOGRAPHY
 
Chapter 5 (GC)
Chapter 5 (GC)Chapter 5 (GC)
Chapter 5 (GC)
 
GC.ppt
GC.pptGC.ppt
GC.ppt
 
Lec#4_Column Efficiency_GC.ppt
Lec#4_Column Efficiency_GC.pptLec#4_Column Efficiency_GC.ppt
Lec#4_Column Efficiency_GC.ppt
 
GC.ppt
GC.pptGC.ppt
GC.ppt
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
 
GAS CHROMATOGRAPHY
GAS CHROMATOGRAPHYGAS CHROMATOGRAPHY
GAS CHROMATOGRAPHY
 
lecture-8-gc c.pptx
lecture-8-gc c.pptxlecture-8-gc c.pptx
lecture-8-gc c.pptx
 
Rajeshri patil
Rajeshri patilRajeshri patil
Rajeshri patil
 
GC PPT...pptx
GC PPT...pptxGC PPT...pptx
GC PPT...pptx
 
Gas chromatograph analyzer
Gas chromatograph analyzerGas chromatograph analyzer
Gas chromatograph analyzer
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Chromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.pptChromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.ppt
 
Chromatography
ChromatographyChromatography
Chromatography
 
GC.pptx
GC.pptxGC.pptx
GC.pptx
 
Gc
GcGc
Gc
 
Gas Chromatography
Gas Chromatography Gas Chromatography
Gas Chromatography
 
GC and GC-MS slide.ppt
GC and GC-MS slide.pptGC and GC-MS slide.ppt
GC and GC-MS slide.ppt
 
Gas chromatography instrumentation
Gas chromatography instrumentationGas chromatography instrumentation
Gas chromatography instrumentation
 

Plus de sherif Taha

Plus de sherif Taha (7)

A practical guide for hplc beginner users
A practical guide for hplc beginner usersA practical guide for hplc beginner users
A practical guide for hplc beginner users
 
Common multi pesticide residue extraction methods
Common multi pesticide residue  extraction methodsCommon multi pesticide residue  extraction methods
Common multi pesticide residue extraction methods
 
Pesticides classification and maximum residue limits in food
Pesticides classification and maximum residue limits in foodPesticides classification and maximum residue limits in food
Pesticides classification and maximum residue limits in food
 
mass spectrometry for pesticides residue analysis- L4
mass spectrometry for pesticides residue analysis- L4mass spectrometry for pesticides residue analysis- L4
mass spectrometry for pesticides residue analysis- L4
 
mass spectrometry for pesticides residue analysis- L3
mass spectrometry for pesticides residue analysis- L3mass spectrometry for pesticides residue analysis- L3
mass spectrometry for pesticides residue analysis- L3
 
mass spectrometry for pesticides residue analysis- L2
mass spectrometry for pesticides residue analysis- L2mass spectrometry for pesticides residue analysis- L2
mass spectrometry for pesticides residue analysis- L2
 
mass spectrometry for pesticides residue analysis- L1
mass spectrometry for pesticides residue analysis- L1mass spectrometry for pesticides residue analysis- L1
mass spectrometry for pesticides residue analysis- L1
 

Dernier

Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
KarakKing
 

Dernier (20)

Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
Interdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptxInterdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptx
 
Graduate Outcomes Presentation Slides - English
Graduate Outcomes Presentation Slides - EnglishGraduate Outcomes Presentation Slides - English
Graduate Outcomes Presentation Slides - English
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptxHMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docx
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
Spatium Project Simulation student brief
Spatium Project Simulation student briefSpatium Project Simulation student brief
Spatium Project Simulation student brief
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 

Gas chromatography

  • 2. • Gas chromatography is a separation technique based on partitioning analytes between two immiscible phases: gaseous mobile phase (Carrier gas) and a stationary solid or immobilized liquid phase (packed or hollow capillary column). Gas chromatography • The sample is first vaporized by a heated inlet system to be passed through a gaseous carrier into GC column. The passed analytes adsorbed on the stationary phase of the GC column, the adsorbed analytes eluted by applying heating program (the stronger adsorption of an analyte, the higher heat required for its elution). Therefore, GC is suitable for analysis of volatile (or semi-volatile) and a thermally stable analyte. Stauffer E, Dolan JA, and Newman R (2008) Fire Debris Analysis, p. 246. Burlington, MA: Elsevier Academic Press
  • 3. Carrier gas • A carrier gas is responsible for carrying the sample molecules into the column and finally to the detector. • Carrier gas must be: inert with the stationary phase, of high purity 99.999 %. (for residue analysis 99.9999 % is preferred). Agilent, fundamental gas chromatography, G1176-90000 It is better to be vertically standing
  • 4. GC Injection Techniques • Split/split-less • Programmed temperature vaporization (PTV) injection. • On-column injection
  • 5. Split injector • In the split injection mode, sample enters the hot liner and volatilized rapidly. • Vaporized sample is mixed with a carrier gas (diluted). • Finally, a large part of the diluted vaporized sample is split away from the Colum, while a small part will enter the column. • This mode of injection is used for analysis of samples of high analyte concentrations. Agilent, fundamental gas chromatography, G1176-90000
  • 6. Split-less injector • Is the split-less injection mode, sample enter the hot liner and volatilized rapidly. • Vaporized sample is mixed with a carrier gas (diluted). • Finally, all the diluted vaporized sample enter the column. • After that, the split valve is opened to remove residual vapors. • This mode of injection is used for analysis of samples of trace analyte concentrations (residue analysis). Agilent, fundamental gas chromatography, G1176-90000
  • 7. • In the program temperature vaporized injector the sample is injected in a cooled liner (enable sample and analyte adsorption). While, the solvent will be split away. Therefore, injection in PTV should be carried out slowly. • After that, split valve closed, and beginning ramping temperature increase (ramp PTV) to vaporize adsorbed sample matrix and analytes to be carried into the column Baffled Liner Vent Syringe needle Column Sample Analyte Solvent PTV injector
  • 8. • sample aliquots are directly introduced onto the analytical column ( 0.2-0.5 ) at low temperatures (60– 80°C). • On column injection is favored for an analyte that can be thermally degradated at the elevated heated split or split-less mode (around 200 C). • These injection mode require careful awareness to attain a good reproducibility. • A liner of the wider volume is favorable for this injection. On-column injection https://www.sigmaaldrich.com/analytical- chromatography/analytical- products.html?TablePage=110598941
  • 9. • The used solvent (containing the analytes) shall have an expansion volume that will not exceed 75 % of the liner volume (especially for split-less mode of injection). • Inner-surface of the used liner shall be highly inert, especially for residue analysis. • If the sample is not so cleaned up, liners containing glass wool is more favorable. • The temperature of the liner should be enough to evaporate the desired analytes and avoid its degradation. • A bottom tap liner is more favorable for splitless injection. Choosing the right liner http://www.restek.com/Supplies-Accessories/GC- Accessories/Inlet-Liners-Liner-Supplies https://www.chromservis.eu/c/gc-liners
  • 10. GC-COLUMN Capillary columns Open tubular (OT) K. Maštovska, S.J. Lehotay, J. Chromatogr. A, 1000 (2003) 153 Packed columns Agilent, fundamental gas chromatography, G1176-90000Agilent, fundamental gas chromatography, G1176-90000
  • 11. • The commonly used stationary phase for capillary GC column is poly dimethyl siloxane [PDMS]. The separation, in this case, will depend upon dispersive interactions (van der Waals) with nonpolar part of the analytes. • The PDMS stationary phases can be more polar by adding phenyl, cyano, or trifluoro functional groups each of these stationary phases leads to different separation results. • In case of using PDMS with cyano, trifluoro, or (especially) hydroxyl functional groups, analyte separation can occur based on hydrogen bonding interactions (the strongest intermolecular forces in capillary gas chromatography (GC). GC-COLUMN
  • 12. GC-COLUMN Jana Hajšlová, Tomášajka, Gas chromatography–mass spectrometry (GC– MS), Food Toxicants Analysis, 2007
  • 13. • A longer column will increase the resolution (selectivity), but it will also increase analysis time, and cost. • Reduced column internal diameter double the efficiency and leads to better selectivity. This will increase retention time when using isothermal separations. such columns easily contaminated and suffer from peak broadening after routine work. • Changes in film thickness effects retention of analyte species, interaction with the silica tubing (increased with increasing film thickness). Usually, thin films (0.10-0.25 µm)are used for trace analysis. GC-COLUMN
  • 14. A common GC-COLUMN/ HP 5 MS Internal diameter 0.25 mm Length 30 m Film 0.50 µm Temperature Limits -60 - 325/350 °C • Stationary phase coating is : 5 %-Phenyl-methyl poly siloxane
  • 15. Column Temperature • Every capillary column has its working temperature range that maintains the stateof the used coated stationary phase. • Working on GC column should be below its maximum temperature.
  • 16. Column starting temperature • The starting temperature for the column (oven temperature) should be much lower than the lowest boiling points of tested analytes. • For analytes of low boiling points, the adsorption of both solvent and these analytes is required at the front column inlet (starting column temperature of 20 C lower than the used solvent) which will subsequently enhance its peak shape. • While, for highly boiling points analytes (late eluting peaks), adsorption of the used solvent isn’t required (it will cost a longer run time) only start with a temperature below the boiling points of these analytes but that give enough adsorption at the front column inlet (no peak tailing)
  • 17. Column starting temperature GC Inlets An Introduction, Agilent. 5958-9468
  • 18. Journal of Chromatographic Science, Vol. 34, December 1996
  • 20. Retention time, related terms and definitions • Chromatogram : A plot of the detector response related to the effluent time. • Mass spectrum: A plot of the intensities versus m/z of specific peak. • Retention time tR : Time taken by the carrier gas from the analyte injection to its completely detection. • Dead time (holdup time) tM : Is the time taken by the carrier gas from the point of injection to the detector. • Adjusted retention time : = tR - tM • Distribution constant K: The ratio of analyte concentration in the stationary phase to its concentration in the mobile phase 𝐾 = 𝐶𝑠 𝐶𝑚
  • 21. Column Resolution, related terms and definitions • Separation factor (selectivity factor): is the relative retention times of two adjacent eluted peaks, α= tR1 tR2 . • Peak resolution based on their peak widths Rs= 2(tR1− tR2) Wb1 + Wb2 . • Peak resolution based on their peak widths Rh= h1−hw h1 • The number of theoretical plates, N = 16 ( tR W ) 2 • Efficiency (N) ,for thin film column, is related to column length (L) and internal diameter (dc), N= 𝐿/𝑑𝑐 https://doi.org/10.1016/B978-0-12-409547-2.14348-3
  • 22. References, and useful links • Yehua Han, Yanfen Zhang, Huwei Liu https://doi.org/10.1016/B978-0-12-409547-2.14348-3 • A.I. Ruiz-Matute, S. Rodrı´guez-Sa´nchez, M.L. Sanz and A.C. Soria, https://doi.org/10.1016/B978-0-12-814264- 6.00012-8 • Alejandra Garcia Piantanida , Andrew R. Barron, Principles of Gas Chromatography, 2014. http://cnx.org/content/m50228/1.2/ • Fundamentals of Gas Chromatography, Agilent Technologies, Inc., 2002. G1176-90000 • https://www.restek.com/pdfs/GNBR1724-UNV.pdf • http://www.chromatographyonline.com/pragmatic-rules-gc-column-selection • https://chem.libretexts.org/Textbook_Maps/Analytical_Chemistry/Supplemental_Modules_(Analytical_Chemistry)/I nstrumental_Analysis/Chromatography/Gas_Chromatography.
  • 23. Thank you Dr. Sherif M. Taha, sherif2taha@gmail.com