1. 07.02.2013
Sample & Assay Technologies
Setting up MiSeq to run GeneRead DNAseq Gene
Panel Libraries
02-07-2013
Sample & Assay Technologies
Holger Wedler
Check Software Version of the MiSeq Control Software
Get sure to have installed MiSeq Control Software 2.0.5.0 or later:
After booting of the MiSeq, usually the MiSeq Control Software is
launched automatically.
Otherwise double click symbol
on the desktop.
Click to get details
Title, Location, Date
2
1
2. 07.02.2013
Sample & Assay Technologies
Check Software Version of the MiSeq Control Software
Title, Location, Date
Sample & Assay Technologies
3
Check Software Version of the Experiment Manger
Get sure to have installed Experiment Manager Software 1.3 or
later:
Double click symbol
Title, Location, Date
on the desktop.
4
2
3. 07.02.2013
Sample & Assay Technologies
Check Software Version of the Sequencing Analysis Viewer
(SAV)
Get sure to have installed Sequencing Analysis Software 1.8.11 or
later:
Double click symbol
on the desktop.
Title, Location, Date
Sample & Assay Technologies
5
Check Software Version of the MiSeq Reporter
Get sure to have access to installed MiSeq Reporter Software
2.0.26 or later via Internet Browser:
Enter address „http://XX.XXX.XX.XX:XXXX/default.htm“ or
http://127.0.0.1:8042 in the Internet Explorer of the MiSeq.
Ask your IT specialists and/ Illumina support for details.
Title, Location, Date
6
3
4. 07.02.2013
How to Start a Run
Sample & Assay Technologies
Title, Location, Date
7
Create a Sample Sheet
Sample & Assay Technologies
Click „Folder Settings“
Double Click
.
Title, Location, Date
8
4
5. 07.02.2013
Sample & Assay Technologies
Create a Sample Sheet
Check Directories. Defaults are:
Title, Location, Date
Sample & Assay Technologies
9
Create a Sample Sheet
Click „Create Sample Sheet“
Title, Location, Date
10
5
6. 07.02.2013
Sample & Assay Technologies
Create a Sample Sheet
Click „MiSeq“
„Next“
Title, Location, Date
Sample & Assay Technologies
11
Create a Sample Sheet
Click „Other“
Title, Location, Date
„FASTQ Only“
„Next“
12
6
7. 07.02.2013
Sample & Assay Technologies
Create a Sample Sheet
Enter „Reagent Cartridge Barcode“ (required)
select „TruSeq
LT“
Select „Index Reads 1“
Select „Read Type Paired End“
Enter „Cycles Read 1: 151“, „Cycles Read 2: 151“
Check „Use
Adapter Trimming“
All other fields are optionally, but useful.
Title, Location, Date
Sample & Assay Technologies
Click „Next“
13
Create a Sample Sheet
Check „Maximize“
Title, Location, Date
14
7
8. 07.02.2013
Sample & Assay Technologies
Create a Sample Sheet
Press „Add Blank Row“ – one Click per Sample/Library – each
Click adds a row
Title, Location, Date
Sample & Assay Technologies
15
Create a Sample Sheet
In this example 5 rows were added. Entering „Sample ID“ and
selection of „Index 1 (I7) is mandatory, all other fields are optionally
Title, Location, Date
16
8
9. 07.02.2013
Sample & Assay Technologies
Create a Sample Sheet
Example:
This Warning
can be ignored
(it seems to be a bug)
Click „Finish“ to save the Sample Sheet in *csv format (s. Folder Settings).
Title, Location, Date
Sample & Assay Technologies
17
How to Start a Run
Follow Instructions in MiSeq™ Reagent Kit Reagent Preparation Guide
Title, Location, Date
18
9
10. 07.02.2013
Sample & Assay Technologies
How to Start a Run
Follow Instructions in MiSeq™ Reagent Kit Reagent Preparation Guide
Title, Location, Date
Sample & Assay Technologies
19
How to Start a Run
Follow Instructions in MiSeq® System User Guide
Title, Location, Date
20
10
11. 07.02.2013
Sample & Assay Technologies
How to Start a Run
Follow the Instructions of the Wizard in the MiSeq Control Software
Click „Sequence“
Title, Location, Date
Sample & Assay Technologies
21
How to Start a Run
Uncheck „Use Base Space....“
Title, Location, Date
Click „Next“
22
11
12. 07.02.2013
Sample & Assay Technologies
How to Start a Run
Follow the Instructions on the screen to load a flowcell, cartridge
and buffer...
Title, Location, Date
23
Title, Location, Date
24
Sample & Assay Technologies
12
13. 07.02.2013
Sample & Assay Technologies
Title, Location, Date
25
Title, Location, Date
26
Sample & Assay Technologies
13
14. 07.02.2013
Sample & Assay Technologies
Title, Location, Date
27
Title, Location, Date
28
Sample & Assay Technologies
14
15. 07.02.2013
Sample & Assay Technologies
Title, Location, Date
Sample & Assay Technologies
29
FASTQ Files
After the run, data analysis (generation of FASTQ files including
demultiplexing) is done automatically.
FASTQ files are zipped and stored in directory
D:/MiSeqAnalysis/<RunFolder>/Data/Intensities/BaseCalls
Title, Location, Date
30
15